CR MAMMOGRAPHY SYSTEM WITH DX-M DIGITIZER

K111324 · Agfa Healthcare N.V. · MUE · Dec 22, 2011 · Radiology

Device Facts

Record IDK111324
Device NameCR MAMMOGRAPHY SYSTEM WITH DX-M DIGITIZER
ApplicantAgfa Healthcare N.V.
Product CodeMUE · Radiology
Decision DateDec 22, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1715
Device ClassClass 2

Indications for Use

Agfa's CR Mammography System with DX-M Digitizer is indicated for use in diagnostic and screening mammography. It is intended to be used wherever conventional film/ mammography screen systems are used.

Device Story

System comprises DX-M digitizer, HM5.0 image detectors, and NX user workstation. Detectors capture X-ray images; digitizer scans detectors to create electronic images. NX workstation enables patient identification, image preview, and adjustment. Images transmitted to PACS or printer for physician review. Used in clinical settings by radiology staff. MUSICA² software provides image processing, including Micro Calcification Enhancement (MCE). System facilitates diagnostic and screening mammography, aiding clinical decision-making through digital image acquisition and enhancement.

Clinical Evidence

Bench testing only. Evaluated sensitometric response, spatial resolution, noise, dynamic range, erasure, fading, repeat exposures, and AEC performance per IEC 62220-1-2. Clinical image quality assessed by MQSA-qualified radiologists; reviewers evaluated breast positioning, exposure, contrast, sharpness, and tissue visibility. All studies deemed acceptable for clinical use. No patient exposures or treatments involved in device development.

Technological Characteristics

Detector: CsBr:Eu2+ phosphor-coated image plates. Digitizer: Scans detectors. Workstation: NX for image management. Connectivity: DICOM. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 62220-1-2, ISO 13485, ISO 14971. Image processing: MUSICA² with Micro Calcification Enhancement (MCE). Acquisition: 16-bit square root compression.

Indications for Use

Indicated for diagnostic and screening mammography in patients requiring breast imaging, intended for use in clinical settings where conventional film/screen mammography is utilized.

Regulatory Classification

Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K11324 DEC 2 2 2011 # 510(K) SUMMARY: AGFA DX-M Common/Classification Name: 21CFR 892.1715 and 21CFR892.1650 Proprietary Name: CR Mammography System with DX-M Digitizer Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Phil Cuscuna, Prepared: May 6, 2011 Telephone: (416) 240-7317 Facsimile: (416) 240-7359 ### LEGALLY MARKETED PREDICATE DEVICES A. This is a 510(K) for Agfa's CR Mammography System with DX-M Digitizer, a detector-only FFDM system. It is substantially equivalent to Fujifilm Medical Systems' Fuji Computed Radiography Mammography Suite (FCRm). #### DEVICE DESCRIPTION B. The new device is defined as: - . Agfa's DX-M digitizer. - HM5.0 image detectors designed for mammography, and . - . The NX user workstation with a mammography license. The device uses detectors which are exposed to x-rays. The detectors are then scanned by the digitizer to create an electronic images are sent to a user workstation for patient identification, image previewing and adjustment, before sending them to a viewing workstation (PACS system) or printer for use by the physician. Principles of operation and technological characteristics of the new and predicate devices are the same. With optional general radiography screens and cassettes, the DX-M digitizer has the same capabilities as Agfa's DX-G. #### C. INTENDED USE Agfa's CR Mammography System with DX-M Digitizer is indicated for use in diagnostic and screening mammography. It is intended to be used wherever conventional film/ mammography screen systems are used. #### SUBSTANTIAL EQUIVALENCE SUMMARY D. Agfa's CR Mammography System with DX-M Digitizer has an Indications For Use statement nearly . identical to the statement for the predicate device (Fuji FCRm). Intended uses are the same. The devices have the same technological characteristics. Descriptive characteristics and performance data are adequate to ensure equivalence. Differences in devices do not alter the intended therapeutic/diagnostic effect. {2}------------------------------------------------ | PRODUCT COMPARISON TABLE | | | |---------------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------| | | NEW DEVICE<br>Agfa CR Mammography System<br>with DX-M Digitizer | Predicate - PA05014<br>Fuji Computed Radiography for<br>Mammography, FCRm | | Indications | Same as the predicate | Mammography | | Communications | Same as the predicate | DICOM | | Detector Material | CsBr:Eu2+ | BaFBr₁₋ₓIₓ:Eu²⁺ | | Detector Sizes | Same as the predicate | 18x24 cm<br>24x30 cm | | Pixel Matrix | HM5.0:<br>18x24 cm – 3508x4644<br>24x30 cm – 4708x5844 | HR-BD:<br>18x24 cm – 3450x4740<br>24x30 cm – 4728x5928 | | Scanning resolutions | 50μ | 50μ | | Acquisition Gray<br>Scale (Bit Depth) | 16 bit sq. root compression | 12 bit logarithmic compression | | Displayed Image File<br>Gray Scale | 12 bit | 12 bit | | Maximum Image<br>Acquisitions/hr. | 41 (mammo, 24x30) | 65/hr - Profect CS (24x30)<br>40/hr - Profect One (24x30) | | Dimensions and<br>Weight | 43.5x20x48.4 in.<br>397 lb. | Profect CS<br>• 26x29x58 in.<br>• 628 lb.<br>Profect One<br>• 26x29x52 in.<br>• 529 lb. | | Image processing | MUSICA² for Mammography with<br>Micro Calcification Enhancement<br>(MCE) | Dynamic Range Control/Multi-<br>Frequency Processing (DRC/MFP)<br>Pattern Enhancement for<br>Mammography (PEM) | | X-ray System | Same as the predicate | Legally on the market,<br>Designed specifically for<br>mammography | | Softcopy Display | Same as the predicate | Legally on the market for<br>mammography | | Hardcopy Output | Same as the predicate | Legally on the market for<br>mammography | | Electrical Safety | Same as the predicate | IEC-60601-1 | : · {3}------------------------------------------------ ### E. TECHNOLOGICAL CHARACTERISTICS Agfa's CR Mammography System with DX-M Digitizer consists of: - . Mammography detectors containing phosphor coated image plates that capture the patient image. - . A digitizer which scans the detector and extract the image. - A NX workstation to identify the patient, preview and adjust the image, then direct the . image to where it is needed. ### F. TESTING The new device includes proven technology from the other Agfa computed radiography systems which has been tested to demonstrate its suitability for digital mammography. The device has been shown to conform to the electronic medical product safety, radiology, and medical imaging standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 6220-1-2, IEC 61223-2-3 and ACR/NEMA PS3.1-3.18 (DICOM). The following management standards have also been applied: ISO 13485, ISO 14971. Laboratory imaging tests of the new device consistent with IEC 62220-1-2 has been completed. Laboratory tests considered sensitometric response, spatial resolution, noise analysis, dynamic range, erasure, fading, repeat exposures, AEC performance and an evaluation of images from mammography phantoms. Comparisons to the predicate device followed FDA guidance. Test results demonstrated performance equivalent or better than the predicate device. {4}------------------------------------------------ Image /page/3/Figure/1 description: This figure shows a plot of DQE [u] vs spatial frequency [mm^-1]. The plot shows two curves, one for 'Slow scan' and one for 'fast scan'. The DQE values range from 0.0 to 1.0, and the spatial frequency values range from 0 to 10. Both curves decrease as spatial frequency increases. Figure 1 DOE data at ca. 100 µGy exposure level. An evaluation of clinical images by expert, MQSA qualified radiologists was conducted. Reviewers evaluated breast positioning, exposure, image contrast, sharpness, tissue visibility at the skin line noise and overall clinical image quality. Anonymized images were evaluated to determine if they were of acceptable quality for mammographic use. Reviewers found all studies to be of acceptable overall clinical image quality. Development of the new device did not involve patient exposures or treatment. ### ে CONCLUSIONS This 510(K) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image contains a circular logo for the Department of Health & Human Services - USA. The logo features the department's name encircling a stylized symbol. The symbol consists of three abstract, curved shapes that resemble a person embracing another person. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 DEC 2 2 2011 Agfa HealthCare N.V. % Mr. Phil Cuscuna Regional Regulatory Affairs & QS Manager Agfa HealthCare Corp. 10 South Academy Street GREENVILLE SC 29601 Re: K111324 Trade/Device Name: Agfa's CR Mammography System with DX-M digitizer Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: October 17, 2011 Received: October 19, 2011 Dear Mr. Cuscuna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {6}------------------------------------------------ # Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(K) Number (if known): Device Name: CR Mammography System with DX-M Digitizer ### Indications for Use: Agfa's CR Mammography System with DX-M Digitizer is indicated for use in diagnostic and screening mammography. It is intended to be used wherever conventional film/screen mammography systems are used. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Richard D'Olier Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(K) K111324 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...