SEGMENTATION ASSISTANT FOR PROSTATE-AUTO CONTOURING
K111311 · Omisa Inc. (Segasist Technologies) · MUJ · Oct 17, 2011 · Radiology
Device Facts
Record ID
K111311
Device Name
SEGMENTATION ASSISTANT FOR PROSTATE-AUTO CONTOURING
Applicant
Omisa Inc. (Segasist Technologies)
Product Code
MUJ · Radiology
Decision Date
Oct 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Prostate gland contouring
Atlas-based segmentation software
—
—
—
—
—
—
Indications for Use
Segasist P-AC contouring software is a standalone software application for Windows platforms that assists clinicians in generating estimates of the anatomy boundary contours of the prostate gland in Computed Tomography (CT) scans, Magnetic Resonance (MR) images and ultrasound (sonography) scans to aid in patient diagnosis, treatment-planning and post-treatment monitoring. The software is intended to be used to provide clinicians with tools to efficiently contour/delineate the prostate gland in volume data and save the results in DICOM and BMP format. The clinician has the ability to use the saved contours directly or import them in other software tools to perform the task at hand. The clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions based on their standard practices and visual comparison of the individual images. The Segasist P-AC software tool is a compliment to manual contouring techniques.
Device Story
Segasist P-AC is a standalone atlas-based segmentation software for Windows; processes CT, MR, and ultrasound images to delineate prostate gland boundaries. Input: DICOM images; Output: prostate contours saved as DICOM or BMP files. Operation: clinician-driven; requires user input (clicks/ROIs) for semi-automated or automated segmentation. Features include edge enhancement, volume calculation, and manual editing tools. Used in clinical settings by radiologists/oncologists to improve contouring efficiency. Clinician retains final authority to accept, reject, or modify software-generated contours; software does not provide automated diagnosis or auto-saving. Benefits: reduces manual contouring time; provides standardized, editable anatomical boundaries for treatment planning.
Clinical Evidence
Bench testing only. No clinical trials conducted. Validation performed using imported prostate image datasets from CT, MR, and ultrasound modalities. Testing confirmed software functionality, accuracy of contouring, and adherence to software requirements. Results deemed acceptable for clinical use by experienced medical professionals.
Technological Characteristics
Standalone Windows-based software. Atlas-based segmentation algorithm. Supports DICOM import/export and BMP output. Features include edge enhancement, manual/semi-automated/automated contouring, and volume measurement. No patient-contacting materials. Non-real-time processing.
Indications for Use
Indicated for clinicians (radiologists, oncologists, medical physicists) to assist in contouring/delineating the prostate gland in CT, MR, and ultrasound images for diagnosis, treatment planning, and post-treatment monitoring.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
#### CONFIDENTIAL
OCT 1 7 2011
1
# 510(k) Summary
| Date Summary<br>Prepared: | May 4, 2011 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Segasist Prostate Auto-Contouring Software |
| Common Name: | Segasist P-AC/Medical Image Processing Software |
| Classification: | Class II device |
| Classification Name: | System, Image Processing, Radiological |
| CFR Classification: | 21 CFR 892.2050 |
| Product Code : | LLZ |
| Manufacturer: | Segasist Technologies<br>MaRS Centre, South Tower,<br>101 College St., Ste. 200,<br>Toronto, Ontario, Canada, M5G 1L7<br>(647) 260-7834 |
| Contact Name: | Allison Manners<br>Regulatory Advisor<br>Tel: (416) 567-5176<br>email: Allison.manners@segasist.com |
### Legally Marketed Predicate Devices:
The Segasist P-AC contouring software utilizes the same technological characteristics, has the same intended use and is substantially equivalent to at least four predicate devices previously cleared for commercial distribution. The enclosed information for four of these predicates demonstrates substantial equivalence: These commercial products include:
- 1. Medviso AB's Segment software cleared for market via K090833 Trade Name: Segment / Image Processing System Product Code: LLZ Classification: 21 CFR 892.2050 Class: Il
- 2. Computerized Medical Systems, Inc.'s Atlas-Based Autosegmentation Software, cleared for market via K080799 Trade Name: Atlas-Based Autosegmentation Product Code: LHN and MUJ Classification: 21 CFR 892.5050 Class: Il
- 3. IKOEtech, LLC's IKOEngelo™ software cleared for market via K083591 Trade Name: IKOEngelo ™ Product Code: KPQ Classification: 21 CFR 892.5840 Class: Il
- 4. Siemens AG, Contouring Software Package, cleared for market via K071310 Trade Name: syngo CT Oncology Product Code: JAK
{2}------------------------------------------------
## Classification: 21 CFR 892.1750 Class: Il
In each of these software devices, including the Segasist P-AC software, the analysis software is used to receive images from various sources, which can then be stored, communicated, manipulated, annotated, measured, compared and displayed at workstations with the appropriate configurations.
Detailed information regarding each of these predicates, including their 510(k) Summaries and Indications for Use can be found in Section 23 - Other: Predicates, of this Traditional Premarket Notification 510(k) submission for ease of reference.
#### Device Description:
Segasist P-AC (Segmentation Assistant for Prostate – Auto-Contouring) is a standalone atlas-based segmentation software tool for auto-contouring of the prostate gland from different input image modalities (Computed Tomography (CT) scans, Magnetic Resonance (MR) images, ultrasound scans). The software can read, write and display DICOM images from/to local directories, and offers the possibility of defining regions of interest (ROls) around the prostate gland in order to delineate the prostate for contouring, visual assessment, and size and volume calculation purposes, either manually, or via semi-automated or automated processes.
The Segasist P-AC software is a tool that has been designed and developed to assist clinicians (radiologists, oncologists, medical physicists etc.) in performing contouring/delineation of the prostate gland in images in multiple modalities more efficiently. The software is capable of segmenting the prostate gland in individual slices, in a choice of different modalities, and for any given view (axial, sagittal, or coronal). This is done by requiring some user input (clicks or drawing ROIs).
For volume prostate data. Segasist P-AC calculates the prostate volume in cubic centimeters and displays the contours on each slice. The results (contours) can be saved as DICOM or binary images (BMP), which can be edited/modified at any time, completely dismissed or accepted and saved by the end user.
The efficiency of contouring performed by the Segasist P-AC software may be improved by generating/using an advanced atlas using gold standard images created by the experienced clinician(s). This requires the software to be trained (atlas creation) before being used. The software can be delivered pre-trained with the comprehensive atlas or the end user can generate their own atlas; a well-established practice for atlas-based segmentation software products.
Segasist P-AC also offers a built-in editor, enabling the user to edit, modify, or change the extracted prostate boundaries to their desired configuration based on their medical and clinical knowledge and experience. The results provided by the Segasist P-AC software needs to be approved by the experienced clinician and can always be modified or corrected by him/her. In addition, the end user can delineate the prostate gland manually using the P-AC software, if necessary or desired. As a result, when Segasist P-AC generates a result, the expert user always has the final decision to override the software result, if deemed appropriate in his/her clinical judgment. It is up to the expert user to accept the result without any change, reject it completely and delineate manually, or modify the Segasist P-AC result and then save it. The software does not provide any auto-detection or auto-saving functionalities. Regardless of the accuracy of Segasist P-AC result. it is always the experienced clinician that remains the decision maker regarding the acceptability of the computed segmentation. Therefore, the final decision
{3}------------------------------------------------
on diagnosis, treatment and overall management of the patient is not based on the software result.
Segasist P-AC software does not alter the original input images of the prostate gland. nor does it change the final results obtained once approved by the clinical expert.
Segasist P-AC offers several features and functionalities such as, but not limited to:
- . Import/Export DICOM images
- . Saving Contours to DICOM or BMB format
- . Semi-automated Segmentation
- . Auto-Segmentation (fast slice-to-slice auto-segmentation with minimal user interactions)
- . Volume Segmentation and measurement
- . Edge Enhancement (contour enhancement by user controlled edge snapping).
- . Standard Functionalities for Image Visualization (windowing, contrast, brightness, zoom, panning etc)
- . Advanced Functionalities for Contour Editing For Manual Segmentation (drawing, inflating, deflating, shifting, cut & add etc)
- . User access to modify the resulting contours at any time
#### Proposed Intended Use / Indications for Use:
Segasist P-AC contouring software is a standalone software application for Windows platforms that assists clinicians in generating estimates of the anatomy boundary contours of the prostate gland in Computed Tomography (CT) scans, Magnetic Resonance (MR) images and ultrasound (sonography) scans to aid in patient diagnosis, treatment-planning and post-treatment monitoring. The software is intended to be used to provide clinicians with tools to efficiently contour/delineate the prostate gland in volume data and save the results in DICOM and BMP format. The clinician has the ability to use the saved contours directly or import them in other software tools to perform the task at hand.
The clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions based on their standard practices and visual comparison of the individual images. The Segasist P-AC software tool is a compliment to manual contouring techniques.
#### Testing:
The Segasist P-AC software was designed and developed with the input and collaboration of experienced and trained professionals, such as radiologists, oncologists or other highly qualified medical clinicians that are proficient in reading, evaluating and interpreting images of the prostate produced by MR, CT or ultrasound devices. The input was captured in a written and approved Software Requirement Specifications Document. The Segasist P-AC software has been developed in a manner consistent with accepted standards for software development, including both unit and system integration testing. Testing was conducted both internally at Segasist Technologies and independently by experienced and trained medical professionals who are representative of the commercial end users of the software using clinical test cases. No clinical testing was conducted to demonstrate safety or effectiveness as device bench testing was completed using imported images from the various imaging modalities of the prostate in sufficient numbers to support the intended use of the device. This bench testing deemed that the
{4}------------------------------------------------
Segasist P-AC software works as intended, was acceptable for clinical use, and did not introduce any new concerns of safety or effectiveness compared to predicate products or manual contouring of the prostate gland.
As noted, imported prostate image datasets from the various imaging modalities were used as input for testing of the software functionalities in accordance with the software validation/verification plans. A full description of the software functionality, device hazard analysis, software requirements, and verification and validation testing can be found in Section 18 - Software, of this submission.
The device has no patient contacting materials and is utilized only by trained professionals. The output of the device is evaluated by professionals skilled in reading, evaluating and interpreting images of the prostate gland produced by CT, MR or ultrasound, which allows sufficient review for identification and intervention in the event of a malfunction. Device output and analysis is not real-time with respect to patient involvement or presence, and is only used to aid the end user in the further evaluation of the available prostate imaging data. Additional clinically relevant parameters / tests, including data from alternative imaging modalities, must be used in addition to the output of the Segasist P-AC software, in the diagnosis, treatment planning and on-going management of the patient. The device does not impact the quality or status of the original acquired imaging data.
#### Conclusions:
The Segasist P-AC contouring software has the same intended use and similar technological characteristics as the previously noted predicate software devices. As demonstrated within this Traditional Premarket Notification, there are no substantial differences between the Segasist P-AC contouring software and the stated predicate devices, and therefore, the Segasist P-AC software does not introduce any new concerns related to safety or effectiveness. The data contained within this Premarket Notification 510(k) submission is sufficient to deem the Segasist P-AC software substantially equivalent to those predicates described within this application.
4
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUM." are arranged in a semi-circle around the left side of the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
OMISA Inc. (Segasist Technologies) % Ms. Allison Manners Manners Regulatory Management Inc. 1666 12th Line, RR#3 LAKEFIELD ONTARIO K0L 2H0 CANADA
OCT 1 7 2011
Re: K111311
Trade/Device Name: Segasist Prostate Auto-Contouring (Segasist P-AC) Software Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ, LLZ Dated: September 6, 2011 Received: September 7, 2011
Dear Ms. Manners:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{6}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/RevortaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Segasist Prostate Auto-Contouring (Segasist P-AC) Software
Indications for Use:
Segasist P-AC contouring software is a standalone software application for Windows platforms that assists clinicians in generating estimates of the anatomy boundary contours of the prostate gland in Computed Tomography (CT) scans, Magnetic Resonance (MR) images and ultrasound (sonography) scans to aid in patient diagnosis, treatment planning and post-treatment monitoring. The software is intended to be used to provide clinicians with tools to efficiently contour/delineate the prostate gland in volume data and save the results in DICOM and BMP format. The clinician has the ability to use the saved contours directly or import them in other software tools to perform the task at hand.
The clinician retains the ultimate responsibility for making the pertinent diagnosis and patient management decisions based on their standard practices and visual comparison of the individual images. The Segasist P-AC software tool is a compliment to manual contouring techniques.
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|---|
|----------------------------------------------|---|
AND/OR
Over-The (21 CFRe-Counter Use र 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary Pastel
(Division Sign-Off)
Division of Radiclogical Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.