K111296 · Zoll Medical Corporation, World Wide Headquarters · MSX · Oct 27, 2011 · Cardiovascular
Device Facts
Record ID
K111296
Device Name
ZOLL RESCUENETLINK
Applicant
Zoll Medical Corporation, World Wide Headquarters
Product Code
MSX · Cardiovascular
Decision Date
Oct 27, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
RescueNet Link is intended for use in the collection and display of data that is entered by a user (caregiver), or captured from specified medical devices or from a vehicle navigation system. RescueNet Link is indicated for use by health care providers whenever there is a need for collection and display of patient, care reporting and transport data.
Device Story
RescueNet Link is a software-only EMS data system comprising MobileLink and FieldLink components. MobileLink resides in ambulances, collecting/integrating data from ZOLL patient monitors/defibrillators (E Series), electronic patient care reporting (ePCR) systems, and vehicle navigation systems via hardwired or wireless connections. It displays this data on a touch screen for caregivers. MobileLink transmits data via cellular/internet to a central FieldLink server. FieldLink provides web-based access for hospital emergency department staff to view real-time ambulance location, patient status, and charting information. The system is non-alarming. It facilitates clinical decision-making by providing remote visibility into patient status and transport progress, potentially improving care coordination between EMS and receiving facilities.
Clinical Evidence
Bench testing only. Performance testing demonstrated that the device meets all functional requirements and performance specifications, confirming substantial equivalence to the predicate device.
Technological Characteristics
Software-only system utilizing general-purpose, commercially available computer hardware, networking, and communications equipment. Operates via cellular/internet connectivity. Non-alarming. Designed for integration with ZOLL E Series monitor/defibrillators, RescueNet ePCR, and RescueNet Navigator systems.
Indications for Use
Indicated for use by health care providers for the collection and display of patient, care reporting, and transport data in pre-hospital and in-hospital settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 8, 2023
ZOLL Medical Corporation, World Wide Headquarters Charles W. Kolifrath Regulatory Affairs Manager 269 Mill Road Chelmsford, Massachusetts 01824-4105
Re: K111296
Trade/Device Name: ZOLL RescueNet Link Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MSX
Dear Charles W. Kolifrath:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 27, 2011. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.
Sincerely,
Aneesh S. Deoras -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
OCT 2 7 2011
ZOLL Medical Corporation c/o Mr. Charles W. Kolifrath Regulatory Affairs Manager 269 Mill Road Chelmsford, MA 01824
Re: K11296
Trade/Device Name: ZOLL RescueNet Link Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MSX, NSX Dated: August 29, 2011 Received: August 31, 2011
Dear Mr. Kolifrath:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{2}------------------------------------------------
## Page 2 - Mr. Kolifrath
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
KK
/ R / R Z
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Section 4 - Indications for Use
510(k) Number (if known): KII 2 96 RescueNet Link Device Name:
## Indications For Use:
RescueNet Link is intended for use in the collection and display of data that is entered by a user (caregiver), or captured from specified medical devices or from a vehicle navigation user (caregiver), or ouplarou from opson on opeon thealth care providers whenever there is a system. TNCBodonot Link lolay of patient, care reporting and transport data.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence/of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K111296
Page 1 of 1
{4}------------------------------------------------
111296 p1 1/3
OCT 27 2011
## 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Charles W. Kolifrath
(978) 421-9786
Date Summary Prepared:
August 29, 2011
Device:
RescueNet Link
Classification:
System, Network and Communication, Physiological Monitors. Product Code MSX. Device Class II.
Software, Transmission and Storage, Patient Data. Product code: NSX. Device Class: Not Classified.
#### Description:
The proposed RescueNet Link is a software-only product.
The ZOLL RescueNet Link system is an Emergency Medical Services (EMS) data system that electronically captures and displays patient monitoring, charting and ambulance location and navigation system data within the ambulance and communicates it for display to other locations throughout the pre-hospital and inhospital system of care. The system is comprised of two computer hardware/software based components, MobileLink and FieldLink. Both components use general purpose, commercially available computers, networking and communications hardware combined with software supplied by ZOLL.
The MobileLink component resides within each EMS ambulance where it collects, integrates and displays patient monitoring, care reporting and ambulance location and navigation data on a large, easy to read touch screen display. Patient, care and transport data collected by MobileLink are transmitted to a central FieldLink server via cellular/internet connections. This server
{5}------------------------------------------------
provides a distribution point from which internet accessible web pages displaying the information collected within each ambulance can be viewed. The FieldLink server also provides each receiving hospital with web pages showing information about the MobileLink-equipped ambulances currently on route to the medical facility and monitoring and charting information about the patients they are transporting.
RescueNet Link is non-alarming software.
### Intended Use:
The RescueNet Link system including its MobileLink and FieldLink subsystems is intended to be used for the following purposes:
### MobileLink
MobileLink is intended to be used for collecting and displaying manually entered information and electronic data output by a ZOLL patient monitor/defibrillator, and by ZOLL patient charting systems and vehicle navigation systems used in EMS ambulances. The system displays information collected within the ambulance and transmits selected portions of that information via cellular/internet connections for display at central EMS headquarters and hospital emergency departments.
MobileLink is intended to be used for the wireless collection of data from ZOLL's E Series monitor/defibrillator. It is also intended to be used for collecting data (either hardwired or wirelessly) from ZOLL RescueNet ePCR electronic patient care reporting system and ZOLL RescueNet Navigator vehicle location system.
The MobileLink sub-system is further intended to store and display (on demand) treatment protocols and other documents defined by the ambulance service's Medical Director.
### FieldLink
The FieldLink sub-system is intended to be used by a single EMS agency to collect patient and transmitted from each of its MobileLink equipped emergency vehicles and to provide internet connected web servers that allow hospital emergency department staff to view information related to the ambulances and patients destined for their facility.
#### Indications For Use:
RescueNet Link is intended for use in the collection and display of data that is entered by a user (caregiver), or captured from specified medical devices or from a vehicle navigation system. RescueNet Link is indicated for use by health care providers whenever there is a need for collection and display of patient, care reporting and transport data.
{6}------------------------------------------------
## Substantial Equivalence:
The features and functions of the proposed RescueNet Link are substantially The leatures and functions of the physic of the Physio-Control equivalent to the (K102757, cleared for use on 11/5/2010).
. K111296 pg 3/3
# Comparison of Technological Characteristics
RescueNet Link features and functions are similar to the corresponding features RescueNet Link loature and farrodicate device. Both RescueNet Link and the and functions of the indicated products intended for the collection Indicated predicate device are contracts on a user (caregiver) and data collected of data, including uuta that is only issues of safety or effectiveness are raised by this premarket notification.
## Performance Testing:
Extensive performance testing ensures that RescueNet Link performs as well as Extensive performanoo tooling ond meets all of its functional requirements and performance specifications.
#### Conclusion
Performance testing of RescueNet Link demonstrates that its features and f chomanos are substantially equivalent to the corresponding features and functions of the indicated commercially distributed predicate device with regard to performance, safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.