K111278 · Otto Bock Healthcare LP · ITI · Sep 16, 2011 · Physical Medicine
Device Facts
Record ID
K111278
Device Name
Z10 WHEELCHAIR PUSH AND BRAKE ASSIST DEVICE
Applicant
Otto Bock Healthcare LP
Product Code
ITI · Physical Medicine
Decision Date
Sep 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
In combination with a manual wheelchair, the z10 power add-on drive is intended exclusively as a push and brake assist for the transportation by an assistant of persons who are unable to walk or who have walking disabilities. In combination with a manual wheelchair, the z10-ce power add-on drive is intended exclusively as a power add-on drive with tiller control for individual use by persons who are unable to walk or who have a walking disability, in order to transport themselves. The wheelchair/z10 combination may only be used with the components/options listed in the instructions for use. The z10 power add-on drive may only be used by persons who have received training. Training in use is a prerequisite in order to protect persons from danger and to operate the wheelchair/z10 combination safely and without mistakes.
Device Story
The z10 is a powered add-on drive for manual wheelchairs; it functions as a push/brake assist (operated by an assistant) or a tiller-controlled drive (operated by the user). The device attaches to the wheelchair via an adapter plate. It utilizes a 24V DC worm gear motor powered by a Lithium-Ion battery to provide traction and propulsion. The user controls speed (up to 6 km/h) via a control panel. The device is designed for indoor and outdoor use on slopes up to 20% and uneven surfaces. It enhances patient mobility by providing powered assistance for those unable to walk or with walking disabilities. The system includes safety features like anti-tippers (in the z10-ce configuration) and requires user training for safe operation.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance were validated through testing to ISO 7176-14:2008, ISO 7176-21:2009, ANSI/RESNA WC/Vol. 2:1998, and ISO 10993-1:2003 (cytotoxicity).
Technological Characteristics
Powered by 24V DC worm gear motor; 24V/18.4 Ah Lithium-ion battery; 10" airless drive wheels; 8" dual steering casters (z10-ce); enAble 40 control unit; max speed 6 km/h; max load 160 kg; attachment via bolt/locking pin system; ISO 7176-4 range compliance; optional Hella HL 2000 light module.
Indications for Use
Indicated for persons physically challenged and limited to sitting positions due to palsies/paralyses, loss of limbs, defective/deformed limbs, joint contractures, joint defects, or other diseases requiring mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Otto Bock. The logo is in black and white and features the company name in a stylized font. Below the name is the phrase "QUALITY FOR LIFE" in a simple, sans-serif font.
SEP 1 6 2011
K111278
### 510(k) SUMMARY of SAFETY and EFFECTIVENESS
## A. General Information
| 1. Submitter's Name.: | OTTO BOCK Health Care, LP |
|-------------------------|-------------------------------------------------------------------------------|
| 2. Address: | Two Carlson Parkway North, Suite 100<br>Minneapolis, Minnesota USA 55447-4467 |
| 3. Telephone: | 763-253-5610 |
| 4. Contact Person: | William Kabitz, Quality Assurance Manager |
| 5. Date Prepared: | May 1st, 2011 |
| 6. Registration Number: | 2182293 |
#### B. Device
,
ﻨﺘ
| 1. Name: | z10 Push and Brake Assist |
|-------------------------|---------------------------|
| 2. Trade Name: | z10 Push and Brake Assist |
| 3. Common Name: | Powered Wheelchair Mover |
| 4. Classification Name: | Wheelchair, Powered |
| 5. Product Code: | ITI |
| 6. Class: | II |
| 7. Regulation Number: | 21 CFR 890.3860 |
## C. Identification of Legally Marketed Predicate Device
| 1. Name: | Dane Technologies Wheelchair Mover |
|------------------|------------------------------------|
| 2. Manufacture: | Dane Technologies |
| 3. K Number: | K073701 |
| 4. Date Cleared: | January 8, 2008 |
#### D. Description of the Device
The z10 is a powerful push and brake assist for manual wheelchairs. It can also be converted into the z10-ce add-on drive with tiller control quickly and easily. In this product version, the user drives and controls the wheelchair independently.
{1}------------------------------------------------
Otto Bock®
JALITY FOR LI
#### E. Features
The special features of the z10 push and brake assist for wheelchairs includes:
- Very good operating performance, including uneven surfaces, thanks . to the motor unit, suspension, and generous ground clearance of 8 cm.
- . Optimum traction, even when going uphill or downhill, thanks to the rear-mounted drive.
- Range of up to 25 km* for pushing and driving, no creeping loss of . battery capacity thanks to the latest Lithium-lon battery technology.
- . Speed control up to 6 km/h in 5 levels at the push of a button.
- Hills and slopes up to 20% and obstacle clearance up to 5 cm . possible.
- Excellent maneuverability, including indoors. .
- . Very quiet operation due to the worm gear motor.
#### F. Intended Use Statement
In combination with a manual wheelchair, the z10 power add-on drive is intended exclusively as a push and brake assist for the transportation by an assistant of persons who are unable to walk or who have walking disabilities.
In combination with a manual wheelchair, the z10-ce power add-on drive is intended exclusively as a power add-on drive with tiller control for individual use by persons who are unable to walk or who have a walking disability, in order to transport themselves.
The wheelchair/z10 combination may only be used with the components/options listed in the instructions for use.
The z10 power add-on drive may only be used by persons who have received training. Training in use is a prerequisite in order to protect persons from danger and to operate the wheelchair/z10 combination safely and without mistakes.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Otto Bock. The logo is in black and white and features a stylized, cursive font. A registered trademark symbol is present to the right of the text.
QUALITY FOR LIFE
# G. Technological Characteristics Summary
| | z10 | z10-ce |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| General | | |
| Housing Metal: | Anthracite metallic | |
| Drive wheel | 10", airless, grey | |
| Steering caster | N/A | 8" dual, airless, grey |
| Anti-tipper | | Swinging anti-tipper |
| Driving Data | | |
| Speed | Max. 6 km/h, 5 levels, continuous fine adjustment within each speed level | |
| Range | 25 km according to ISO 7176-4 | |
| Max. climbing ability | Up to 20% | |
| Maximum obstacle height | 6 cm | 4 cm |
| Turning radius | Depending on the radius of the manual wheelchair | |
| Weights | | |
| Drive unit (without battery) | 21 kg | |
| Tiller control with control panel | 1.5 kg | |
| Battery pack | 3.8 kg | |
| Wheelchair adapter plate | 1.0 kg | |
| z10-ce steering caster unit with adapter | N/A | 8.0 kg |
| Total weight, z10 without adapter plate | 26.3 kg | 34.3 kg |
| Maximum load capacity (including wheelchair) | 160 kg | |
| Minimum load capacity | 20 kg | |
| Dimensions (W x H x D) | | |
| Drive unit including battery | 335 x 580 x 450 mm | |
| Tiller control with control panel | 550 x 520 x 180 mm | |
| Wheelchair with z10 | Depends on the manual wheelchair, z10 does not change dimensions | |
| Electrical Installation | | |
| Lithium-ion battery | 24V / 18.4 Ah | |
| Charging time (battery drained) | Approx. 4 h | |
| Battery lifespan (complete charging cycles) | 700 cycles (subsequently 80% charging capacity) | |
| Control unit | enAble 40 | |
| Operating voltage | 24 V DC | |
| Motor power | 100 W | |
| Self-contacting | Yes | |
| Light module | (front) Optional: Hella HL 2000 | |
| Wheelchair Adaptation | | |
| Minimum seat height | 38 cm | |
| Minimum seat width | 38 cm | |
| Adaptation system | Attachment bolt with locking pin | |
| Compatible Otto Bock wheelchairs | Start M1/M2/M3/M4/M5, Motus, Avantgarde T/VR, Centro A3 | |
| Compatible third-party wheelchairs | Adaptable to most wheelchair models commonly available on the<br>market, limitations depending on the product design of the<br>wheelchair, please request a list from Otto Bock | |
| Charger | | |
| Type | Mentzer | |
| Mains plug | Euro-plug, 2-pin | |
| Power requirements | 110 V | |
| Mains frequency | 50/60 Hz | |
| Dimensions (W x H x D) | 170 x 140 x 85 mm | |
| Weight | | |
{3}------------------------------------------------
Otto Bock®
QUALITY FOR LI
## H. Safety Testing
The safety of the z10 Push and Brake Assist has been confirmed by CE certification, and was tested by Mikes Testing Partners in Strasskarchen Germany and TÜV SÜD Product Services GmbH, Hanover, Germany to the following standards:
ISO 7176-14:2008 ISO 7176-21:2009 ANSI/RESNA WC/Vol. 2 : 1998 ISO 10993-1:2003 Cytotoxicity ISO 9999:2007 Assistive products for persons with disability
Refer to Appendic C for test reports.
Signatures:
Quality Assurance Manager_
President & CEO US Healthcare _
lthcare
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a circular seal with text around the perimeter and a stylized bird-like symbol in the center. The text around the perimeter appears to be the name of a government agency or organization, possibly related to health and human services. The bird-like symbol in the center is composed of three curved lines, creating a sense of movement and flight.
Food and Drug Adminstration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 16 2511
Otto Bock Healthcare, LP % Mr. William Kabitz Quality Assurance Manager Two Carlson Parkway North, Suite 100 Minneapolis, Minnesota 55447-4467
Re: K111278
Trade/Device Name: 210 and z10ce Push and Brake Assist Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 30, 2011 Received: September 02, 2011
Dear Mr. Kabitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 – Mr. William Kabitz
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Pari 803). please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Sincerely yours,
FOR Pete Vimmerstedt
Mark N. Melkerson
Dic Linda
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Otto Bock. The logo is in black and white and features the company name in a stylized font. Below the name is the phrase "QUALITY FOR LIFE" in a simple, sans-serif font.
## APPENDIX J
#### Indications For Use Form
510(k) Number (if known): To be determined
Device Name: z10 and z10ce Push and Brake Assist
Indications for Use:
Provide mobility to persons physically challenged and limited to sitting positions due to:
- Palsies/Paralyses .
- Loss of limbs
- Defective and/or deformed limbs ●
- Joint contractures .
- Joint defects ●
- . Other diseases
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Signature
(Division Sign Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111278
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.