The Amsco Chimeron Small Steam Sterilizer models 16, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: Prevacuum (Models 16, 16C and 20, 20C) - are equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) - are equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
Device Story
Small steam sterilizer for healthcare facilities; uses steam as sterilizing agent to process heat/moisture-stable materials. Models feature varying chamber sizes (16"x16"x26" or 20"x20"x38"). Configurations include Prevacuum or Steam Flush Pressure-Pulse (SFPP) cycles, plus Gravity, Liquid, Leak Test, and DART cycles. Device uses embedded controller with touch screen interface; monitors chamber conditions via transducers and dual-element drain sensors. Safety features include pressure relief valve, float switch, and emergency stop. Healthcare personnel operate device to sterilize instruments/packs; output is a sterilized load verified by biological/chemical indicators. Benefits include effective pathogen reduction and cross-contamination prevention.
Clinical Evidence
Bench testing only. Validation performed per ANSI/AAMI ST8:2008. Studies included empty chamber testing, fabric test packs, and full-load instrument trays. All cycles achieved SAL of at least 10^-6. Liquid cycles validated for water loss and sealing. DART cycles validated for air removal. No clinical data.
Indicated for sterilization of heat and moisture-stable materials in healthcare facilities. Applicable to porous and non-porous loads, including fabric packs and wrapped/unwrapped instrument trays.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
K 111223
ﺎ
#### STERIS ABBREVIATED 510(k) PREMARKET NOTIFICATION SUMMARY Amsco Chimeron Small Steam Sterilizer
**Device Name**
SEP 20 2011
| Trade Name: | Amsco Chimeron Small Steam Sterilizer |
|----------------------|--------------------------------------------------------|
| Models: | 16, 16C, 16CS, 16S, 20, 20C, 20CS, and 20S |
| Common/Usual Name: | Steam Sterilizer |
| Classification Name: | Steam Sterilizer (21 CFR 880.6880)<br>Product Code FLE |
#### 2. Predicate Device
1.
K030789, Amsco Century Small Steam Sterilizer, product code [FLE] cleared March 26th, 2003.
#### 3. Description of Device
The Amsco Chimeron Small Steam Sterilizer models 16, 16CS. 16S. 20, 20C. 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:
- Prevacuum (Models 16, 16C and 20, 20C) are equipped with Prevacuum. . Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
- . Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20CS) – are equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
The chamber sizes of the proposed Amsco Chimeron Small Steam Sterilizer Models are as follows:
- · 16" x 16" x 26" (Models 16, 16C, 16CS, and 16S)
- · 20" x 20" x 38" (Models 20, 20C, 20CS, and 20S)
#### Intended Use 4.
The Amsco Chimeron Small Steam Sterilizer Models 16, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configurations:
- Models 16, 16C, 20, 20C: . Prevacuum - equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
12
{1}------------------------------------------------
- . Models 16CS, 16S, 20CS, and 20S: . Steam Flush Pressure-Pulse (SFPP) – equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
The Amsco Chimeron Small Prevacuum Steam Sterilizers (Models 16, 16C, 20, and 20C) are equipped with the following factory-validated sterilization cycles (Table 5-1):
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Factory<br>Default | Recommended Load |
|------------------------------|--------------------------|-------------------|------------|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F<br>(132°C) | 4 minutes | 20 minutes | Yes | Fabric Packs. Refer to<br>Table 5-3 for<br>recommended<br>quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 30 minutes | Yes | Double wrapped<br>instrument trays,<br>maximum weight 25<br>lbs (11.3 kg) each.<br>Refer to Table 5-3 for<br>recommended<br>quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 5 minutes | No | Single Fabric Pack |
| Prevac | 275°F<br>(135°C) | 3 minutes | 30 minutes | Yes | Double wrapped<br>instrument trays,<br>maximum weight 25<br>lbs (11.3 kg) each.<br>Refer to Table 5-3 for<br>recommended<br>quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F<br>(132°C) | 4 minutes | 1 minutes | Yes | Unwrapped instrument<br>tray with a single<br>instrument or multiple<br>unwrapped instrument<br>trays, maximum<br>weight 25 lbs (11.3 kg)<br>each. Refer to Table<br>5-3 for recommended<br>quantities. |
| Gravity | 250°F<br>(121°C) | 30 minutes | 15 minutes | No | Fabric Packs. Refer to<br>Table 5-3 for<br>recommended<br>quantities. |
| Gravity | 270°F<br>(132°C) | 15 minutes | 30 minutes | No | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30 minutes | 30 minutes | No | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Gravity | 270°F<br>(132°C) | 25 minutes | 15 minutes | No | Fabric Packs. Refer to Table 5-3 for recommended quantities. |
| Gravity<br>(Immediate Use) | 270°F<br>(132°C) | 3 minutes | 1 minutes | Yes | Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Gravity<br>(Immediate Use) | 270°F<br>(132°C) | 10 minutes | 1 minutes | Yes | Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Liquid | 250°F<br>(121°C) | 45 minutes | N/A | No | Validated with 3 1000ml bottles. Refer to Table 5-4 for recommended full load quantities. |
| DART Warm-Up | 270°F<br>(132°C) | 3 minutes | 1 minute | Yes | N/A |
| DART | 270°F<br>(132°C) | 3 ½<br>minutes | 1 minute | Yes | Bowie-Dick Test Pack,<br>DART Test Pack |
| Leak Test | N/A | N/A | N/A | Yes | N/A |
# Table 5-1
{2}------------------------------------------------
. . .
1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
{3}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The Amsco Chimeron Small Steam Flush Pressure-Pulse (SFPP) Sterilizers (Models 16CS, 16S, 20CS, and 20S) are equipped with the following factoryvalidated sterilization cycles (Table 5-2):
# Table 5-2
.
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Factory Default | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SFPP | 270°F<br>(132°C) | 4 minutes | 30 minutes | Yes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| SFPP | 270°F<br>(132°C) | 4 minutes | 20 minutes | Yes | Fabric Packs.<br>Refer to Table 5-3 for recommended quantities. |
| SFPP | 275°F<br>(135°C) | 3 minutes | 30 minutes | Yes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| SFPP | 270°F<br>(132°C) | 4 minutes | 5 minutes | No | Single Fabric Pack |
| Prevac | 270°F<br>(132°C) | 4 minutes | 20 minutes | No | Fabric Packs.<br>Refer to Table 5-3 for recommended quantities |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Factory Default | Recommended Load |
| Prevac | 270°F<br>(132°C) | 4 minutes | 30 minutes | No | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 5 minutes | No | Single Fabric Pack |
| Prevac | 275°F<br>(135°C) | 3 minutes | 30 minutes | No | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Prevac<br>(Immediate Use) | 270°F<br>(132°C) | 4 minutes | 1 minutes | No | Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30 minutes | 15 minutes | No | Fabric Packs. Refer to Table 5-3 for recommended quantities. |
| Gravity | 270°F<br>(132°C) | 15 minutes | 30 minutes | No | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. |
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Factory<br>Default | Recommended<br>Load |
| Gravity | 250°F<br>(121°C) | 30 minutes | 30 minutes | No | Double wrapped<br>instrument trays,<br>maximum weight<br>25 lbs (11.3 kg).<br>each. Refer to<br>Table 5-3 for<br>recommended<br>quantities. |
| Gravity | 270°F<br>(132°C) | 25 minutes | 15 minutes | No | Fabric Packs.<br>Refer to Table 5-3<br>for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 3 minutes | 1 minutes | Yes | Unwrapped<br>instrument tray<br>with a single<br>instrument or<br>multiple<br>unwrapped<br>instrument trays,<br>non-porous load,<br>maximum weight<br>25 lbs (11.3 kg)<br>each. Refer to<br>Table 5-3 for<br>recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 10 minutes | 1 minutes | Yes | Unwrapped<br>instrument tray<br>with a single<br>instrument or<br>multiple<br>unwrapped<br>instrument trays,<br>non-porous load,<br>maximum weight<br>25 lbs (11.3 kg)<br>each. Refer to<br>Table 5-3 for<br>recommended<br>quantities |
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Factory<br>Default | Recommended<br>Load |
| Liquid | 250°F<br>(121°C) | 45<br>minutes | N/A | No | Validated with 3<br>1000ml bottles.<br>Refer to Table 5-4<br>for recommended<br>full load<br>quantities. |
| DART<br>Warm-Up | 270°F<br>(132°C) | 3 minutes | 1 minute | Yes | N/A |
| DART | 270°F<br>(132°C) | 3 ½<br>minutes | 1 minute | Yes | Bowie-Dick Test<br>Pack, DART Test<br>Pack |
| Leak Test | N/A | N/A | N/A | Yes | N/A |
{4}------------------------------------------------
:
{5}------------------------------------------------
人
{6}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
# Table 5-3 Amsco Chimeron Small Steam Sterilizers recommended loads per sterilizer size
| Sterilizer Size | Wrapped Instrument<br>Trays | Fabric Packs |
|-----------------|-----------------------------|--------------|
| 16" x 16" x 26" | 2 | 4 |
| 20" x 20" x 38" | 3 | 6 |
The following table is a guideline for liquid cycle processing:
# Table 5-4 Amsco Chimeron Small Steam Sterilizers Liquid Cycle Guideline
| Sterilizer<br>Size | Number of<br>Containers<br>for full<br>load | Number of<br>Containers<br>for validation | Volume of Liquid In<br>One Container | Minimum Recommended<br>Sterilize, Time at 250°F<br>(121°C) |
|--------------------|---------------------------------------------|-------------------------------------------|--------------------------------------|------------------------------------------------------------|
| 16" x 16"<br>x 26" | 15 | 3 | 1000 ml | 45 minutes |
| 20" x 20"<br>x 38" | 32 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
of 12
{7}------------------------------------------------
#### 5. Description of Safety and Substantial Equivalence
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A summary of the technological characteristics of the device subject of this premarket notification in comparison to those of the predicate devices is included in Table 5-1.
| Table 5-1 |
|-----------------------------------------------|
| Summary of the Proposed and predicate Devices |
| Technological Characteristics |
| General Sterilizer<br>Features | Amsco Chimeron Small<br>Steam Sterilizer<br>(All Models Included unless Specified) | Amsco Century Small<br>Steam Sterilizer<br>(K030789) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | A steam sterilizer intended for<br>sterilization of non-porous and porous,<br>heat- and moisture-stable materials in<br>healthcare facilities. | A steam sterilizer intended for<br>sterilization of non-porous and porous,<br>heat- and moisture-stable materials in<br>healthcare facilities. |
| Operating Principle | Steam is the sterilizing agent. | Steam is the sterilizing agent. |
| Sterilization Cycles<br>Offered | Prevac<br>Gravity<br>SFPP (Models 16CS, 16S, 20CS, and 20S<br>Only)<br>Liquid | Prevac<br>Gravity |
| Chamber Sizes | 16" x 16" x 26"<br>16" x 16" x 38" | 16" x 16" x 26"<br>16" x 16" x 38" |
| Chamber Door | Type 316L stainless steel<br>Vertical Sliding (26"x26") | Type 316L stainless steel<br>Vertical Sliding (26"x26") |
| Shell Assembly | Type 316L stainless steel<br>ASME certified | Type 316L stainless steel<br>ASME certified |
| Control Technology | Embedded Controller<br>Touch Screen<br>320 x 240 Pixel Display<br>Ink on Paper Printer | Embedded Controller<br>Touch Screen<br>320 x 240 Pixel Display<br>Ink on Paper Printer |
| Process Monitors | Chamber Transducer<br>Dual Element Chamber Drain Sensor | Chamber Transducer<br>Single Element Chamber Drain Sensor |
| Safety Devices | Pressure Relief Valve<br>Chamber Float Switch<br>Emergency Stop Switch | Pressure Relief Valve<br>Chamber Float Switch<br>Emergency Stop Switch |
{8}------------------------------------------------
# Effectiveness
Effectiveness of sterilizer function and exposure time recommendations was demonstrated by complete kill of biological indicators and by verifying an appropriate safety factor or sterility assurance level (SAL) of at least 10° probability of survival. STERIS validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The Amsco Chimeron Steam Sterilizers have been validated to meet the requirements of ANSVAAMI ST8:2008, December 2008.
The results of the Amsco Chimeron Steam Sterilizer verification studies demonstrate that the sterilizers perform as intended. The results are summarized as follows:
- Empty chamber testing was performed as described in Section 5.4.2.5 of . ANSVAAMI-ST8. for the Prevac. Gravity. Liquid and SFPP cycles. These cycles demonstrated that the sterilizer is capable of providing steady state thermal conditions within the chamber that are consistent with the predicted sterility assurance level (SAL) in the load. The sterilizer meets the requirements of Sections 4.4.2 of ANSVAAMI-ST8.
- AII SFPP cycles were validated using the fabric test pack, described in Section . 5.5.2.1 of ANSI/AAMI-ST8, and were qualified according to Section 5.5.2 ANSVAAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10th through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention of less than 3% increase in presterilization test pack weight, and no evidence of wet spots.
- All SFPP cycles were validated using full load instrument trays, described in . 5.5.4.1 of ANSI/AAMI-ST8, and were qualified according to Section 5.5.4 ANSVAAMI-ST8. These cycles demonstrated a sterility assurance level of at least 106 using half-cycle cycle analysis, moisture retention of less than 20% increase in presterilization weight of the towel, and no evidence of wet spots on the outer wrapper.
- All GRAVITY cycles were validated using the fabric test pack, described in . Section 5.5.2.1 of ANSUAAMI-ST8, and were qualified according to Section 5.5.2 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10° through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention of less than 3% increase in presterilization test pack weight, and no evidence of wet spots.
- All GRAVITY cycles were validated using full load instrument trays, described . in 5.5.4.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.4
{9}------------------------------------------------
ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 100 using half-cycle cycle analysis, moisture retention of less than 20% increase in presterilization weight of the towel, and no evidence of wet spots on the outer wrapper.
- All Immediate Use (Short or Abbreviated) GRAVITY cycles were validated . using both a single, nonporous instrument weighing no more than 100g in an otherwise empty, unwrapped instrument tray and a full, non-porous load of unwrapped instrument trays as described in 5.5.5.1.2 of ANSVAAMI-ST8. and were qualified according to Section 5.5.5 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10th using half-cycle cvcle analysis.
- All PREVAC cycles were validated using the fabric test pack, described in . Section 5.5.2.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.2 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 100 through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, moisture retention of less than 3% increase in presterilization test pack weight, and exhibited no wet spots.
- All PREVAC cycles were validated using full load instrument trays, described . in 5.5.4.1 of ANSVAAMI-ST8, and were qualified according to Section 5.5.4 ANSVAAMI-ST8. These cycles demonstrated a sterility assurance level of at least 100 using half-cycle analysis, moisture retention of less than 20% increase in presterilization weight of the towel, and no evidence of wet spots on the outer wrapper.
- All Immediate Use (Short or Abbreviated) PREVAC cycles were validated . using both a single, nonporous instrument weighing no more than 100g in an otherwise empty, unwrapped instrument tray and a full, non-porous load of unwrapped instrument trays, described in 5.5.5.1.2 of ANSVAAMI-ST8, and were qualified according to Section 5.5.5 ANSVAAMI-ST8. These cycles demonstrated a sterility assurance level of at least 100 using half-cycle cycle analysis.
- All LIQUID cycles were validated using three 1,000 ml flasks, described in . Section 5.5.3.1 of the ANSVAAMI-ST8, and were qualified according to Section 5.5.3 of ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10th through achievement of a time-at-temperature sufficient to produce an Fo value of at least 12, water loss not exceeding 50 ml, and automatic sealing of the flask closure.
- The DART cycle was validated using the Bowie-Dick Test Pack, as described in . 5.6.1.1 of the ANSI/AAMI-ST8, was qualified according to Section 5.6.1 of the ANSVAAMI-ST8, and demonstrated a uniform color change throughout the test sheet. The DART cycle uses the same dynamic air removal routines as the 270°F Prevac cycle.
- The software validation for the cycle operation was performed according to t FDA's moderate level of concern recommendations provided in the document "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (05/11/05). "
{10}------------------------------------------------
# Safety :
STERIS sterilizers including the Amsco Chimeron Steam Sterilizers have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The Amsco Chimeron Steam Sterilizers comply with the following requirements:
- ANSI/UL 61010-1 2nd Edition (7/22/2005) "Safety Requirements for . Electrical Equipment for Measurement, Control and Laboratory Use - Part 1 General Requirements.
- CAN/CSA C22.2 No. 61010-1 2nd Edition (7/12/2004) "Safety . Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1 General Requirements".
- UL 61010A-2-041 1st Edition (3/27/2002) "Particular Requirements for . Autoclaves Using Steam for Treatment of Medical Materials and for · Laboratory Processes"
- CAN/CSA C22.2 No 1010.2.041 (R2004) "Particular Requirements for ● Autoclaves Using Steam for Treatment of Medical Materials and for Laboratory Processes".
- ASME Boiler Pressure Vessel Code, Section VIII (Division 1) (7/1/2007) ●
# Hazards - Failure of Performance
Failure of the sterilization process can lead to incidence of cross contamination, the transmission of potentially infectious organisms from one infected person to another who was not otherwise infected prior to the incident.
To avoid failure, the user must ensure that the materials, instruments and devices to be sterilized are thoroughly cleaned, the manufacturer's instructions for use are followed, the cycle to be used for each type of sterilizer load has been validated, the sterilizer has been maintained in accordance with the sterilizer manufacturer's recommended maintenance schedule and is operating properly, and each sterilizer load is monitored with available and validated biological and chemical sterilization process indicators.
The technology designed into STERIS steam sterilizers including the Amsco Chimeron Steam Sterilizers provide control safeguards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.
{11}------------------------------------------------
# Conclusion
The Amsco Chimeron Small Steam Sterilizer has been validated to meet the requirements of ANSVAAMI ST8:2008, December 4, 2008.
The results of the Amsco Chimeron Small Steam Sterilizer validation studies demonstrate that the sterilizer performs as intended.
Based on the information provided in this premarket notification, it can be concluded that the Amsco Chimeron Small Steam Sterilizer is substantially equivalent to the predicate device and is safe and effective when used as intended.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Robert F. Sullivan Senior Director FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060
NOV 1 7 2011
Re: K111223
Trade/Device Name: Amsco Chimeron Small Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: August 24, 2011 Received: August 25, 2011
Dear Mr. Sullivan:
This letter corrects our substantially equivalent letter of September 20, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{13}------------------------------------------------
Page 2- Mr. Sullivan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{14}------------------------------------------------
# Indications for Use
K 111223 510(k) Number (if known):
Device Name:
Amsco Chimeron Small Steam Sterilizer
Indications For Use:
The Amsco Chimeron Small Steam Sterilizer models 16, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:
- . Prevacuum (Models 16, 16C and 20, 20C) - are equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
- Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) are ● equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
The Amsco Chimeron Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-1):
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
| Prevac<br>(Immediate<br>Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray<br>with a single instrument or<br>multiple unwrapped<br>instrument trays, maximum<br>weight 25 lbs (11.3 kg)<br>each. Refer to Table 4-3<br>for recommended<br>quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to<br>Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 4-3 for recommended<br>quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to<br>Table 4-3 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray<br>with a single instrument or<br>multiple unwrapped<br>instrument trays, non-<br>porous load, maximum<br>weight 25 lbs (11.3 kg)<br>each. Refer to Table 4-3<br>for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10 minutes | 1 minutes | Unwrapped instrument tray<br>with a single instrument or<br>multiple unwrapped<br>instrument trays, non-<br>porous load, maximum<br>weight 25 lbs (11.3 kg)<br>each. Refer to Table 4-3<br>for recommended<br>quantities. |
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Timer | Recommended Load |
| Liquid | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml<br>bottles. Refer to Table 4-4<br>for recommended full load<br>quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack,<br>DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
#### Table 4-1.
Amsco Chimeron Small Prevacuum Steam Sterilizer Models 16, 16, 20, 20C factoryvalidated sterilization cvcles and cvcle values
P. 1 46
{15}------------------------------------------------
.
.
·
{16}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The Amsco Chimeron Small Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-2):
| Table 4-2. |
|------------------------------------------------------------------------------|
| Amsco Chimeron Small Steam Flush Pressure-Pulse (SFPP)Sterilizer Models 16S, |
| 16CS, 20S, 20CS factory-validated sterilization cycles and cycle values |
| Cycles > >>Sterilize: >>>Sterilize: >>>Dry Time >> Recommended Load >> |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SFPP | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped<br>instrument trays,<br>maximum weight 25 lbs<br>(11.3 kg) each and Fabric<br>Packs. Refer to Table 4-3<br>for recommended<br>quantities. |
| SFPP | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to<br>Table 4-3 for<br>recommended quantities. |
| SFPP | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped<br>instrument trays,<br>maximum weight 25 lbs<br>(11.3 kg) each. Refer to<br>Table 4-3 for<br>recommended quantities. |
| SFPP | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to<br>Table 4-3 for<br>recommended quantities. |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. |
| Prevac (Immediate Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended quantities. |
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument<br>tray with a single<br>instrument or multiple<br>unwrapped instrument<br>trays, non-porous load,<br>maximum weight 25 lbs<br>(11.3 kg) each. Refer to<br>Table 4-3 for<br>recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10 minutes | 1 minutes | Unwrapped instrument<br>tray with a single<br>instrument or multiple<br>unwrapped instrument<br>trays, non-porous load,<br>maximum weight 25 lbs<br>(11.3 kg) each. Refer to<br>Table 4-3 for<br>recommended quantities. |
| Liquid | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml<br>bottles. Refer to Table 4-4<br>for recommended full load<br>quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minutė | N/A |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack,<br>DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
{17}------------------------------------------------
·
{18}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-----------------|--------------------------|--------------|
| 16" x 16" x 26" | 2 | 4 |
| 20" x 20" x 38" | 3 | 6 |
p. 50f 6
・
{19}------------------------------------------------
The following table is a guideline for liquid cycle processing:
## Table 4-4 Amsco Chimeron Small Steam Sterilizer Liquid Cycle Guideline
| Sterilizer<br>Size | Number of<br>Containers<br>for full<br>load | Number of<br>Containers<br>for validation | Volume of Liquid<br>In One Container | Minimum Recommended<br>Sterilize Time at 250°F<br>(121°C) |
|--------------------|---------------------------------------------|-------------------------------------------|--------------------------------------|-----------------------------------------------------------|
| 16" x 16"<br>x 26" | 15 | 3 | 1000 ml | 45 minutes |
| 20" x 20"<br>x 38" | 32 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The Amsco Chimeron Small Steam Sterilizer is offered in the following chamber sizes:
- 16" x 16" x 26" (Models 16, 16C, 16CS, and 16S) .
- 20" x 20" x 38" (Models 20, 20C, 20CS, and 20S) ●
Prescription Use _ (Part 21 CFR 801 Subpart D) . AND/OR
Over-The-Counter Use __ X (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eluhut S. Clavijo-Willian
Division Sign-Off) vision of Anesthesiology, General Hospital tection Control, Dental Devices
510(k) Number: K111233
p. 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.