VISION ONE LASER SYSTEM

K111213 · Lumenis, Inc. · GEX · Jul 25, 2011 · General, Plastic Surgery

Device Facts

Record IDK111213
Device NameVISION ONE LASER SYSTEM
ApplicantLumenis, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 25, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Vision One Laser System is intended for use in the treatment of ocular pathology. The Vision One Laser System green, yellow, and red wavelengths are indicated for use in photocoagulation of both anterior and posterior segments including: - Retinal photocoagulation, panretinal photocoagulation, and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroid including: - Proliferative and nonproliferative diabetic retinopathy - Choroidal neovascularization - Branch retinal vein occlusion - Retinal tears and detachments - Retinopathy of prematurity - Iridotomy, iridectomy, and trabeculoplasty in angle closure glaucoma and open angle glaucoma

Device Story

Vision One Laser System is a Class IV ophthalmic laser console; utilizes Optically Pumped Semiconductor Laser (OPSL) and Nd:YAG (subsets of Diode Pumped Solid State Laser) to generate green, yellow, and red wavelengths. System includes console, control/display panel, remote, door interlock, and footswitch. Compatible with various delivery devices including integrated slit lamps, LaserLinks, and Laser Indirect Ophthalmoscopes (LIO). Operated by clinicians in clinical settings for photocoagulation of ocular pathology. Device delivers laser energy to target retinal/choroidal vascular/structural abnormalities or for glaucoma procedures. Output allows precise tissue coagulation; aids in managing diabetic retinopathy, neovascularization, and retinal detachments. System configurable for up to three treatment lasers.

Clinical Evidence

Bench testing only; safety, performance, and functional testing conducted to demonstrate substantial equivalence.

Technological Characteristics

Class IV laser; utilizes Optically Pumped Semiconductor Laser (OPSL) and Nd:YAG (Diode Pumped Solid State Laser) technology. Configurable for up to three treatment lasers. Includes console, control/display panel, remote, interlock, and footswitch. Compatible with various slit lamp delivery systems and LIOs.

Indications for Use

Indicated for patients requiring photocoagulation of anterior and posterior ocular segments, including treatment for proliferative/nonproliferative diabetic retinopathy, choroidal neovascularization, branch retinal vein occlusion, retinal tears/detachments, retinopathy of prematurity, and glaucoma (iridotomy, iridectomy, trabeculoplasty).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the Lumenis logo. The logo consists of a stylized letter "L" inside a black circle on the left, followed by the word "LUMENIS" in a bold, sans-serif font. Below the word "LUMENIS" is the tagline "Enhancing Life. Advancing Technology." K11/215 Page ① of ③ # 5. 510(k) SUMMARY # 510(k) Summary JUL 25 2011 # Submitter General Information: . | Device Owner Name: | Lumenis, Inc. | |----------------------|--------------------------------------------------| | Address: | 3959 West 1820 South<br>Salt Lake City, UT 84104 | | Contact: | Jace R. McLane | | Phone: | 801-656-2328 | | Fax: | 801-656-2627 | | Date of Preparation: | April 29, 2011 | # II. Device Name: | Trade Name: | Vision One Laser System | |-------------------------|-----------------------------------------------------------------------------------------------------------| | Common/Usual Name(s): | Surgical Laser, Ophthalmic Laser | | Class: | Class II | | Classification Name(s): | 79 GEX, Laser Powered Surgical Instrument | | | 86 HQF, Laser, Ophthalmic | | CFR Reference: | 21 CFR § 878.4810, Laser surgical instrument for use in<br>general and plastic surgery and in dermatology | | | 21 CFR § 886.4390, Ophthalmic Laser | # III. Predicate Device Name(s): Novus Varia Laser System (K022181) Trade Name(s): # IV. Device Description: The Vision One Laser System is classified as a Class IV laser by the Center for Devices and Radiological Health of the Food and Drug Administration. The Vision One Laser System is intended for use in the treatment of ocular pathology. The system's green, yellow, and red wavelengths are indicated for use in photocoagulation of both anterior and posterior segments. The Vision One Laser System will incorporate the newest laser technology - Optically Pumped Semiconductor Laser (OPSL) in additions to Nd:YAG which are both subsets of DPSS (Diode Pumped Solid State Laser). The system shall be configurable so that any combination of up to three (3) treatment lasers can be installed and used. The Vision One Laser System is comprised of the laser console, control and display panel, remote control, an external door interlock plug, & footswitch. Lumenis, Inc. 3959 West 1820 South Salt Lake City, UT 84104, USA Image /page/0/Picture/18 description: The image shows three horizontal black bars stacked vertically on top of each other. The bars are all the same width and height. The background of the image is white. The bars are solid black, with no visible texture or detail. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Lumenis. The logo consists of a stylized "L" inside of a circle, followed by the word "LUMENIS" in all caps. Below the word "LUMENIS" is the tagline "Enhancing Life. Advancing Technology." K11/2/3 Page (2) of (3) The following accessories & delivery devices are compatible and available for the Vision One Laser System: · Lumenis 1000 Integrated Slit Lamp, Lumenis InSight Integrated Slit Lamp, LaserLinks: SL130 Zeiss Slit Lamp, 30SL Zeiss Slit Lamp, 20SL Zeiss Slit Lamp, 125SL Zeiss Slit Lamp, Haag Streit Slit Lamp, Lumenis LaserLink S, Eye Safety Filter for all LaserLinks listed, Single and Dual Endo Kits for Zeiss and Leica / Wild microscopes, Endo Kit for Leica M840, Keeler LIO (Laser Indirect Ophthalmoscope), & Heine LIO (Laser Indirect Ophthalmoscope). The Vision One Laser System uses the same types of delivery devices and accessories as the predicate device. #### V. Intended Use: The Vision One Laser System is intended for use in the treatment of ocular pathology. The Vision One Laser System green, yellow, and red wavelengths are indicated for use in photocoagulation of both anterior and posterior segments including: - Retinal photocoagulation, panretinal photocoagulation, and intravitreal . endophotocoagulation of vascular and structural abnormalities of the retina and choroid including: - Proliferative and nonproliferative diabetic retinopathy . - Choroidal neovascularization . - Branch retinal vein occlusion . - Retinal tears and detachments . - . Retinopathy of prematurity - Iridotomy, iridectomy, and trabeculoplasty in angle closure . glaucoma and open angle glaucoma # VI. Substantial Equivalence Summary: The Vision One Laser System has the same intended use as the predicate device -Novus Varia Ophthalmic Laser (K022181). It uses the same fundamental scientific technology as the predicate device. In addition, Vision One Laser System and the predicate device have the same patient population, principles of operation, method of energy delivery, critical materials, and basal design. # VII. Performance Data: The appropriate testing, including safety, performance and functional testing, to determine substantial equivalence has been conducted. Lumenis, Inc. 3959 West 1820 South Salt Lake City, UT 84104, USA Image /page/1/Picture/21 description: The image shows three horizontal bars stacked vertically. The bars are black with some white speckling. The speckling is more prominent in the bottom bar, with the top bar having the least amount of speckling. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Lumenis. The logo consists of a stylized "L" inside of a circle, followed by the word "LUMENIS" in all caps. Below the word "LUMENIS" is the tagline "Enhancing Life. Advancing Technology." K11/13 page 3 of (3) # VIII. Conclusion: The subject device, the Vision One Laser System, has the same intended use, general design, and fundamental scientific technology as the predicate device (K022181). The Vision One Laser System uses technology substantially equivalent to the Novus Varia Ophthalmic Laser (K022181). There are no new hazards introduced by the Vision One Laser System as compared with the predicate device. Image /page/2/Picture/5 description: The image shows the text "Page 11 of 25". This indicates that it is a page number within a document. The document has at least 25 pages. The page number is in a simple, sans-serif font. Lumenis, Inc. 3959 West 1820 South Salt Lake City, UT 84104, USA Image /page/2/Picture/8 description: The image shows three horizontal bars. The bars are black with white speckles. The speckles are randomly distributed across the bars. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms or flowing lines, positioned within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration , 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Lumenis, Inc. % Ms. Jace R. McLane 3959 West 1820 South Regulatory Affairs Specialist Salt Lake City, Utah 84104 JUL 2 5 2011 Re: K111213 Trade/Device Name: Vision One Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 15, 2011 Received: July 18, 2011 Dear Ms. McLane: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Jace R. McLane comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcestorYou/Industry/default.htm. Sincerely yours, Eunel Keith for Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Lumensis logo. The logo consists of a circle with a white "L" inside, followed by the word "LUMENIS" in all caps. Below the word "LUMENIS" is the tagline "Enhancing Life. Advancing Technology." # 4. INDICATIONS FOR USE STATEMENT Indications for Use KILLE 510(k) Number (if known): Device Name: _Vision One Laser System_ Indications for Use: The Vision One Laser System is intended for use in the treatment of ocular pathology. The Vision One Laser System green, vellow, and red wavelengths are indicated for use in photocoagulation of both anterior and posterior segments including: - . Retinal photocoagulation, panretinal photocoagulation, and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroid including: - Proliferative and nonproliferative diabetic retinopathy . - Choroidal neovascularization . - Branch retinal vein occlusion . - . Retinal tears and detachments - Retinopathy of prematurity . - Iridotomy, iridectomy, and trabeculoplasty in angle closure glaucoma and . open angle glaucoma Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page U of O Neil R.P. Oden for man (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K111213 Page 8 of 25 Lumenis, Inc. 3959 West 1820 South .Salt Lake City, UT 84104, USA Image /page/5/Picture/24 description: The image shows three horizontal bars stacked on top of each other. The top bar is a solid dark color, while the middle and bottom bars have a speckled or textured appearance. The bottom bar also has a white line running horizontally through the middle of it.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...