MICROSCAN (R) MICROSTREP PLUS (R) PANELS

K111205 · Siemens Healthcare Diagnostics, Inc. · LRG · Jul 20, 2011 · Microbiology

Device Facts

Record IDK111205
Device NameMICROSCAN (R) MICROSTREP PLUS (R) PANELS
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeLRG · Microbiology
Decision DateJul 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The MicroScan® MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually. Alternatively, the panels can be incubated in and read by the MicroScan® WalkAway System, according to the Package Insert. This particular submission is for the addition of the antimicrobial Moxifloxacin at concentrations of 0.03 to 8 ug/ml to the test panel. The organisms which may be used for Moxifloxacin susceptibility testing in this panel are: Streptococcus pneumoniae (including penicillin resistant strains) Streptococcus pyogenes Streptoccocus agalactiae Streptococcus constellatus Streptococcus anginosus Viridans group streptococci

Device Story

MicroScan® MICroSTREP plus Panels are miniaturized broth dilution susceptibility tests; dehydrated antimicrobial agents in wells. User inoculates panels with standardized organism suspension in Mueller-Hinton broth with 3% lysed horse blood. Panels incubated 20-24 hours at 35°C. Minimum Inhibitory Concentration (MIC) determined by manual visual observation of growth inhibition or via automated MicroScan® WalkAway System. Output is MIC value and categorical interpretation (SIR). Used in clinical microbiology laboratories to guide antimicrobial therapy for streptococcal infections.

Clinical Evidence

Bench testing only. Performance evaluated using fresh and stock efficacy isolates and stock challenge strains. Compared against CLSI frozen reference panels. Overall Essential Agreement for Moxifloxacin was 97.3%. Reproducibility and precision testing confirmed acceptable performance across both manual and MicroScan WalkAway instrument read methods. Quality control testing met acceptance criteria.

Technological Characteristics

Miniaturized broth microdilution panels; dehydrated antimicrobial agent. Rehydrated with Mueller-Hinton broth + 3% lysed horse blood. Growth-based detection principle. Manual visual read or automated optical detection via MicroScan® WalkAway System. Sterilization method not specified.

Indications for Use

Indicated for determining quantitative and/or qualitative antimicrobial susceptibility of aerobic streptococci, including S. pneumoniae, S. pyogenes, S. agalactiae, S. constellatus, S. anginosus, and Viridans group streptococci. For prescription use only.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111205 B. Purpose for Submission: To obtain substantial equivalence determination for the addition of moxifloxacin at concentrations of 0.03-8 µg/mL to the Microscan® MICroSTREP plus Panel. C. Measurand: Moxifloxacin 0.03-8 µg/mL D. Type of Test: Quantitative growth-based detection algorithm using optics light detection E. Applicant: Siemens Healthcare Diagnostics, Inc. F. Proprietary and Established Names: MicroScan® MICroSTREP plus Panels G. Regulatory Information: 1. Regulation section: 866.1640 - Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: LRG- Instrument for Auto Reader &amp; Interpretation of Overnight Antimicrobial Susceptibility Systems JWY - Manual Antimicrobial Susceptibility Test Systems LTW – Susceptibility Test Cards, Antimicrobial {2} 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use: MicroScan® Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci including S. pneumoniae. 2. Indications for use: The MicroScan® MICroSTREP plus Panels are used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci including S. pneumoniae. After inoculation, panels are incubated for 20-24 hours at 35°C +/- 1°C in a non-CO₂ incubator, and read visually. Alternatively, the panels can be incubated in and read by the MicroScan® WalkAway System, according to the Package insert. This particular submission is for the addition of antimicrobial Moxifloxacin at concentrations 0.03 to 8μg/mL to the test panel. The organisms which may be used for susceptibility testing in this panel are: - Streptococcus pneumoniae (including penicillin resistant strains) - Streptococcus pyogenes - Streptococcus agalactiae - Streptococcus constellatus - Streptococcus anginosus - Viridans group streptococci 3. Special conditions for use statement(s): - For prescription use only - Turbidity inoculation method only 4. Special instrument requirements: MicroScan® WalkAway® System is the alternate read method for Moxifloxacin 2 {3} 3 I. Device Description: The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. The antimicrobial agent is diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115μl Mueller-Hinton broth supplemented with 3% lysed horse blood (LHB), after inoculation of the broth with a standardized suspension of the organism. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth. Alternatively, the panel can be incubated in and read by the MicroScan® WalkAway System. J. Substantial Equivalence Information: 1. Predicate device name(s): MicroScan MICroStrep Panels- Levofloxacin 2. Predicate 510(k) number(s): k020556 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies, grown on solid media of aerobic streptococci including S. pneumoniae. | Same | | Inoculum preparation | Inoculum prepared from isolated colonies using the Turbidity method | Same | | Technology | Growth based after 20-24 hours | Same | | Results | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Same | | | | | | Differences | | | | Item | Device | Predicate | | Antibiotic | Moxifloxacin at 0.03 - 8μg/mL μg/mL | Levofloxacin at 0.12- 16 μg/mL | | Instrument | WalkAway® or Manual visual read | Manual visual read | K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test {4} (AST) Systems; Guidance for Industry and FDA”; Clinical and Laboratory Standards Institute (CLSI) M07-A8 “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”; M100-S19 “Performance Standards for Antimicrobial Susceptibility Testing” ## L. Test Principle: The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. The antimicrobial agent is diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 µl Mueller-Hinton broth supplemented with 3% lysed horse blood (LHB), after inoculation of the broth with a standardized suspension of the organism. After incubation in a non-CO₂ incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth. Alternatively, the panel can be incubated in and read by the MicroScan® WalkAway System. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Reproducibility was demonstrated using 10 isolates tested at three sites on three separate days in triplicate. The study included the turbidity inoculum method with reading performed manually, by WalkAway instrument. Both within site and site to site results were &gt;95% reproducible. | Difference in the number of dilutions between the mode of the MicroScan® result and the actual result for between site reproducibility | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Inoculation Method | Read Method | ≥Minus 2 dilutions | Minus 1 dilution | Exact | Plus 1 Dilution | ≥Plus 2 dilutions | % Reproducible | | Turbidity | Manual | | 13 | 235 | 21 | | 99.6% | | Turbidity | WalkAway | 3 | 22 | 232 | 12 | | 98.9% | There were more results in the minus category (one dilution lower) by the WalkAway; however, there were more results in the plus category by the manual method. #### b. Linearity/assay reportable range: Not Applicable {5} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended QC isolate, S. pneumoniae ATCC 49619 was tested a sufficient number of times with acceptable results most of the time with the reference method. Quality control results demonstrated the ability of the different reading parameters (manual, and WalkAway) by Turbidity inoculation methods to produce acceptable results. The following table provides the frequency of the results in each concentration with the expected range stated. | Moxifloxacin | | | | | | --- | --- | --- | --- | --- | | Organism | μg/mL | | | | | S. pneumoniae ATCC 49619 | | Reference | Manual | WalkAway | | Expected | <=0.03 | | | | | range | 0.06 | | | | | 0.06- 0.25* | 0.12 | 101 | 87 | 104 | | μg/mL | 0.25 | 10 | 24 | 5 | *This quality control range is applicable only to broth microdilution susceptibility tests using cation-adjusted Mueller-Hinton broth with 2 - 5% lysed horse blood. Inoculum density control: A turbidity meter was used for the turbidity inoculation method. Inoculum density was controlled by monitoring the daily results of the MicroScan® Turbidity Meter to ensure the correct final concentration of the organisms for both the reference and test device. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Clinical efficacy testing was conducted at three external sites using fresh isolates supplemented with stock isolates. The study included a total of 443 Streptococci isolates, with a no growth rate of 1.1% (5/448). There were 319 (72%) fresh and 124 (28%) stock isolates. There were 50 challenge isolates {6} tested at one site and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. All isolates grew in the MICroStrep panels. Efficacy testing was performed using the turbidity inoculation method and read manually after incubation for 20- 24 hours at 35°C +/- 1°C in a non-CO₂ incubator. A comparison to the reference method was provided with the following agreement. Overall Performance Summary- Overnight Manual (Efficacy + Challenge) | | Total | EA | %EA | Total evaluable | EA of evaluable | %EA | CA | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Efficacy | 443 | 431 | 97.3 | 434 | 426 | 98.2 | 442 | 99.8 | 4 | 1 | 0 | 0 | | Challenge | 50 | 50 | 100 | 50 | 50 | 100 | 50 | 100 | 0 | 0 | 0 | 0 | | Combined | 493 | 481 | 97.6 | 484 | 476 | 98.3 | 492 | 99.8 | 4 | 1 | 0 | 0 | EA - Essential Agreement CA - Category Agreement R- Resistant maj - Major Discrepancies vmj -- Very Major Discrepancies min - Minor Discrepancies EA is when there is agreement between the reference method and the new method is within plus or minus one serial two-fold dilution of antibiotic. Category agreement (CA) is when the new method result interpretation agrees exactly with the reference panel result interpretation. Evaluable EA is when the MIC result is on scale for both the new method and the reference method and have on-scale EA. The performance of the alternate MicroScan® WalkAway System (incubation and read methods) was demonstrated in challenge, quality control, and reproducibility studies with acceptable results. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable {7} 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Streptococcus spp, including S. pneumoniae ≤1 (S), 2 (I), ≥4 (R) CLSI interpretive breakpoints have not been established for Streptococcus spp other than S. pneumoniae when Moxifloxacin was submitted for review. N. Proposed Labeling: The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...