COBAS B 123 POC SYSTEM, COBAS B 123 AUTOQC PACK TRI-LEVEL, COBAS B 123 AUTOCVC PACK, ROCHE COMBITROL PLUS B
Applicant
Roche Diagnostics
Product Code
CHL · Clinical Chemistry
Decision Date
May 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1120
Device Class
Class 2
Intended Use
The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Na⁺, K⁺, iCa²⁺ (ISE), hematocrit (Hct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O₂Hb, HHb, COHb, MetHb), and oxygen saturation (SO₂). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option. pH, pO₂ and pCO₂: pH, pO₂ and pCO₂ measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances. Sodium (Na⁺): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance. Potassium (K⁺): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Calcium (Ca²⁺): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). Hematocrit (Hct): hematocrit measurements are used to distinguish normal from abnormal states of whole blood, such as anemia and erythrocytosis (an increase in the number of red cells). Glucose (Glu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Lactate (Lac): Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.) Total Hemoglobin (tHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia. O₂Hb: oxyhemoglobin as a fraction of total hemoglobin. COHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning. MetHb: methemoglobin as a fraction of total hemoglobin. HHb: reduced hemoglobin as a fraction of total hemoglobin. SO₂: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus reduced hemoglobin.
Device Story
The cobas b 123 POC system is a modular, automated analyzer for whole blood analysis. It utilizes thick film electrochemical sensor technology (potentiometric for pH, pCO2, Na+, K+, iCa2+; amperometric for pO2, glucose, lactate) and conductivity for hematocrit. An optional oximeter module uses spectrophotometry and an ultrasonic hemolyzer to measure hemoglobin derivatives (tHb, O2Hb, HHb, COHb, MetHb, SO2). The system features a Linux-based touch screen interface, a self-contained disposable sample/reagent delivery system, and an optional AutoQC module for automated quality control. Smart memory chips in sensors and reagent packs track lot numbers, expiration, and usage. Operated by clinicians in POC or laboratory settings, the device provides rapid results to inform clinical decision-making for critical care patients. It benefits patients by enabling immediate assessment of acid-base status, electrolyte balance, and oxygenation, facilitating timely therapeutic interventions.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP5-A2) conducted with N=80 replicates per level. Linearity studies (CLSI EP6) confirmed measuring ranges. Interference studies (CLSI EP7-A2) evaluated potential substances. Method comparison study across 6 sites (2 lab, 4 POC) compared results against predicate devices, showing high correlation (r > 0.97 for all analytes).
Technological Characteristics
Modular analyzer with thick-film microelectrode array technology. Potentiometric sensors (pH, pCO2, Ca2+, K+, Na+), amperometric sensors (pO2, Glucose, Lactate), and spectrophotometric oximeter module. Conductivity-based hematocrit measurement. Disposable self-contained reagent/sample delivery system with smart memory chip tracking. Linux-based software. Operating temperature 15-32°C.
Indications for Use
Indicated for use in Point-of-Care and laboratory environments for the quantitative measurement of pH, blood gases, electrolytes, hematocrit, metabolites, and hemoglobin derivatives in whole blood to aid in the diagnosis and treatment of acid-base disturbances, electrolyte imbalances, anemia, carbohydrate metabolism disorders, and carbon monoxide poisoning.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
K092686 — ABL90 FLEX SERIES · Radiometer Medical Aps · Aug 6, 2010
K110648 — NOVA STAT PROFILE PHOX ULTRA ANALYZER SYSTEM · Nova Biomedical Corporation · Aug 29, 2011
K061974 — GEM PREMIER 4000 WITH IQM, MODEL 4000; GEM CVP 1 AND 2 WITH CO-OX; GEM CVP 3 AND 4 HEMATOCRIT · Instrumentation Laboratory CO · Sep 15, 2006
K203790 — GEM Premier 5000 · Instrumentation Laboratory CO · Aug 20, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
cobas® b 123 POC System Section III: 510(k) Summary
## Section III: 510(k) Summary
1111 88
### 510(k) Summary
MAY 1 4 2012
Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
and the contraction of the comments of the consisted on the contribution of the contribution of the contribution of the contribution of the contribution of the contribution o
Submitter name, address, contact
Roche Diagnostics Corporation 9115 Hague Road Indianapolis, IN 46250
Contact Person: Sarah Baumann Phone: 317-521-3952 Fax: 317-521-2324 Email: sarah.baumann@roche.com
Secondary Contact: Stephanie Greeman Phone: 317-521-2458 Fax: 317-521-2324 Email: stephanie.greeman@roche.com
Date Prepared: May 14, 2012
{1}------------------------------------------------
| Device name | Proprietary Name: | (1) cobas® b 123 POC System<br>(2) cobas® b 123 AutoQC Pack Tri-Level<br>(3) cobas® b 123 AutoCVC Pack<br>(4) Roche COMBITROL PLUS B |
|-------------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| | Common Name: | (1) Blood gas analyzer<br>(2) AutoQC Pack Tri-Level<br>(3) AutoCVC Pack<br>(4) COMBITROL PLUS B |
. '
| Analyte | Classification Name | Regulation Section | Product<br>Code |
|--------------------------------------|-----------------------------------------------------|--------------------|-----------------|
| PCO2, PO2, pH | Blood gases (PCO2, PO2) and blood<br>pH test system | 21 CFR § 862.1120 | CHL |
| Na+ (sodium) | Sodium test system | 21 CFR § 862.1665 | JGS |
| K+ (potassium) | Potassium test system | 21 CFR § 862.1600 | CEM |
| Ca2+ (Calcium) | Calcium test system | 21 CFR § 862.1145 | JFP |
| Hct (Hematocrit) | Automated hematocrit instrument | 21 CFR § 864.5600 | GKF |
| Glucose | Glucose test system | 21 CFR § 862.1345 | CGA |
| Lactate | Lactic acid test system | 21 CFR § 862.1450 | KHP |
| tHb | Whole blood hemoglobin assays<br>(tHb) | 21 CFR § 864.7500 | KHG |
| O2Hb | Whole blood hemoglobin assays<br>(Oxyhemoglobin) | 21 CFR § 864.7500 | GGZ |
| HHb | Deoxyhemoglobin | 21 CFR § 864.7500 | GKA |
| MetHb | Methemoglobin | 21 CFR § 864.5620 | KHG |
| SO2 | Oxygen Saturation | 870.1230 Oximeter | GLY |
| COHb (carboxyhemoglobin) | Carboxyhemoglobin assay | 21 CFR § 864.7425 | GHS |
| cobas b 123 AutoQC Pack<br>Tri-Level | Quality control material<br>(assayed and unassayed) | 21 CFR § 862.1660 | JJY |
| cobas b 123 AutoCVC Pack | | | |
| Roche COMBITROL PLUS B | | | |
Continued on next page
Confidential
{2}------------------------------------------------
.
:
·
## 510(k) Summary, Continued
| Device description | (1) cobas b 123 POC System: |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The cobas b 123 POC system consists of a modular analyzer incorporating a Linux-based graphical user interface with a large color touch screen interfacing the analyzer electronic, sensor, fluidic and AutoQC modules. The user interface module contains the analyzer CPU and all required electronic interfaces for external communication, data storage and data transfer. |
| | A single electrochemical sensor system independent of the reagent delivery system that utilizes Roche thick film technology consolidates the following analytes: |
| | <ul><li>pCO2, pH, calcium, potassium and sodium (potentiometric measurement)</li><li>pO2 (amperometric measurement)</li><li>Hct (conductivity measurement)</li><li>Glucose and Lactate enzyme reaction (amperometric measurement)</li></ul> |
| | An optionally integrated oximeter module consisting of a spectrometer, measurement and calibration light source, respectively, an ultrasonic hemolyzer and thermostatic components measure SO2, tHb, O2Hb, HHb, COHb, and MetHb. |
| | A disposable, self-contained sample and reagent delivery system contains: |
| | <ul><li>Liquid reagents, calibrators and waste container, stable for 42 days on-board</li><li>Built-in safety shielded sample port</li><li>Built-in oximeter cuvette</li><li>Two peristaltic pump fluidics system</li><li>Built-in air filter</li></ul> |
| | The system also includes an optional integrated AutoQC module which utilizes a disposable cassette containing three levels of quality control material. |
| | A smart memory chip is incorporated into each biosensor, reagent pack (sample and reagent delivery system) and AutoQC cassette providing the lot number, expiration date and value assignments (for QC and CVC materials). The chip also tracks and monitors sensor, reagent pack, AutoQC and AutoCVC cassette usage. |
| | Continued on next page |
| Confidential | Page 3 of 21 |
{3}------------------------------------------------
(assayed and unassayed) (continued)
#### (2) cobas b 123 AutoQC Pack Tri-Level:
The cobas b 123 AutoQC pack Tri-Level contains 24 single glass ampoules of multi-analyte controls in 3 levels:
Level 1: representing concentrations of a patient's respiratory acidosis as given by low pH and high PCO2 values, low PO2, low values for Na", K*, and tHb, normal values for glucose and high values for Hct,
iCa** and lactate.
Level 2: representing concentrations of normal values for all parameters except glucose, which is low.
Level 3: representing concentrations of a patient's respiratory alkalosis as given by high pH and low PCO2 values, high PO2, high values for Na , K , tHb, and glucose, and low values for Hct, iCa + and lactate.
### (3) cobas® b 123 AutoCVC Pack:
The cobas b 123 AutoCVC pack contains 24 single glass ampoules of multi-analyte controls in 6-levels covering the range of instrument performance. The CVC materials are an aqueous-based solution of salts, organic and carbonate buffers and metabolites equilibrated with predetermined levels of oxygen, carbon dioxide, nitrogen and dyes.
#### (4) Roche COMBITROL PLUS B:
COMBITROL PLUS B multi-analyte controls are available in 3 levels:
Level 1: representing concentrations of a patient's respiratory acidosis as given by low pH and high PCO2 values, low PO2, low values for Na , K , and tHb, normal values for glucose and high values for Hct,
iCa2+, urea/BUN and lactate.
Level 2: representing concentrations of normal values for all parameters except glucose, which is low.
Level 3: representing concentrations of a patient's respiratory alkalosis as given by high pH and low PCO2 values, high PO2, high values for Na *, K , tHb, and glucose, and low values for Hct, iCa -, lactate and urea/BUN.
{4}------------------------------------------------
#### (1) cobas b 123 POC System: Intended use
The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Na *, K *, iCa2+ (ISE), hematocrit (Hct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters.
It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option.
Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC materials (Table 1):
| Instrument Version | Measured Parameters | Optional Modules |
|------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------|
| cobas b 123 <1> System | pH, BG (pO2, pCO2), ISE (Na+, K+,<br>Ca2+), Hct, Glu, Lac | N/A |
| cobas b 123 <2> System | pH, BG (pO2, pCO2), ISE (Na+, K+,<br>Ca2+), Hct, Glu, Lac | AutoQC Module |
| cobas b 123 <3> System | pH, BG (pO2, pCO2), ISE (Na+, K+,<br>Ca2+), Hct, Glu, Lac, tHb, O2Hb, HHb,<br>COHb, MetHb, SO2 | Oximeter Module |
| cobas b 123 <4> System | pH, BG (pO2, pCO2), ISE (Na+, K+,<br>Ca2+), Hct, Glu, Lac, tHb, O2Hb, HHb,<br>COHb, MetHb, SO2 | AutoQC Module<br>and<br>Oximeter Module |
#### Table 1. Versions of the cobas b 123 POC System
#### (2) cobas b 123 AutoQC Pack Tri-Level:
The cobas b 123 AutoQC pack Tri-Level is a multi-analyte control intended for use as control material to monitor the measurement of pH, PCO2, PO2, SO2, Na*, K*, iCa2+, Hct, tHb and Hb derivatives as well as glucose and lactate on cobas b 123 systems with an AutoQC module.
{5}------------------------------------------------
Intended use (continued)
(3) cobas® b 123 AutoCVC Pack:
The cobas b 123 AutoCVC pack is a multi-analyte control, intended for use in calibration verification of the measuring range established by the cobas b 123 POC system for pH, PCO2, PO2, SO2, Na+, K+, iCa2+, Hct, tHb and Hb derivatives as well as glucose and lactate on cobas b 123 systems with an AutoQC module.
(4) Roche COMBITROL PLUS B:
COMBITROL PLUS B is a multi-analyte control intended for use as control material to monitor the measurement of pH, PCO2, PO2, SO2, Nat, K+, Cl , iCa2+, Hct, tHb and Hb derivatives as well as glucose, lactate, urea/BUN and bilirubin on Roche OMNI S or cobas b 221 analyzers with an oximeter module, and cobas b 123 analyzers (except urea/BUN, chloride, and bilirubin). COMBITROL PLUS B control material is not intended for use with analyzers from other manufacturers.
{6}------------------------------------------------
devices
### 510(k) Summary, Continued
The cobas b 123 POC System is substantially equivalent to other Predicate products in commercial distribution intended for similar use. We claim equivalency to the following currently marketed products:
- cobas b 221 (OMNI S) Blood Gas System-K032311 .
- Hitachi 902 Analyzer (Glucose and Lactate)—K921661 .
- cobas c 501 Analyzer (Glucose and Lactate)-K060373/A001 .
- COMBITROL PLUS B / AUTOTROL PLUS B-K032453 .
{7}------------------------------------------------
Na*
.
.
Substantial The cobas b 123 POC System is substantially equivalent to other equivalence devices legally marketed in the United States. The cobas b 123 POC System is substantially equivalent to the cobas b 221 (OMNI S Analyzer), cleared on K032311, for the following parameters: K+ O2Hb pH . . · SO2 . iCa2+ pO2 . . HHP ● Hct' pCO2 СОНЫ . . .
tHb
The cobas b 123 POC system is substantially equivalent to the Roche Gluco-Quant Glucose/HK assay on the Hitachi 902 (K921661) and the Roche Glucose HK Assay on the cobas c 501 analyzer (K060373/A001) for the measurement of glucose.
MetHb
.
The cobas b 123 POC system is substantially equivalent to the Roche Lactate assay on the Hitachi 902 analyzer (K921661) and the Lactate Generation 2 (Gen.2) assay on the cobas c 501 analyzer (K060373/A001) for the measurement of lactate. Per 21 CFR 862.1450, a Lactic Acid Test System is exempt to measure lactic acid in whole blood and plasma.
The cobas b 123 AutoOC Pack Tri-Level, cobas b 123 AutoCVC Pack, and COMBITROL PLUS B materials are substantially equivalent to the Roche AUTOTROL PLUS B and Roche COMBITROL PLUS B materials cleared on K032453.
<sup>1</sup> Hematocrit measurements on the cobas b 123 POC system were also compared to the hemofuge, which represents the gold standard (reference) method for measurement of hematocrit in blood. This analysis was performed as part of the non-clinical (internal) method comparison study and the results may be located in Section IV (Volume 2) of this Premarket Notification 510(k) submission. Het measurements on the cobas b 123 POC System were compared against the predicate device, cobas b 221, as part of the clinical (external) study; the results may be located in Section V (Volume 3).
{8}------------------------------------------------
Substantial
equivalence
comparison—
similarities
Table 2 provides the similarities between the **cobas b** 123 POC System (K111188) and the **cobas b** 221 predicate device (K032311).
### Table 2. Comparison of Candidate (cobas b 123) and Predicate (cobas b 221) Devices—Similarities
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>cobas b 221<br>(K032311) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Intended Use | The cobas b 123 POC system is a fully automated<br>POC system for whole blood in vitro measurement<br>of pH, blood gases (BG), electrolytes Na+, K+, iCa2+<br>(ISE), hematocrit (Hct), metabolites (Glu, Lac), total<br>hemoglobin (tHb), hemoglobin derivatives (O2Hb,<br>HHb, COHb, MetHb), and oxygen saturation (SO2).<br>In addition, the cobas b 123 POC system calculates<br>derived parameters.<br>It is dedicated for use in a Point-of-Care<br>environment and laboratory. The integrated AutoQC<br>module and the oximeter module are available as an<br>option.<br>Depending on the equipment configuration of the<br>instrument, the Sensor Cartridge and the Fluid Pack<br>used, the following parameters are measured in<br>human whole blood and QC materials. See Table 1<br>(page 5). | Same |
| Blood Gas<br>Measurement | pH and pCO2 by potentiometry<br>pO2 by amperometry | Same |
| Electrolyte<br>Measurement | K+, Na+, iCa2+ by potentiometry | Same |
| Metabolite<br>Measurement | Glucose and Lactate by amperometry | Same |
Continued on next page
Confidential
{9}------------------------------------------------
Table 2 provides the similarities between the cobas b 123 POC Substantial equivalence System (K111188) and the cobas b 221 predicate device (K032311). comparisonsimilarities (continued)
#### Table 2. Comparison of Candidate (cobas b 123) and Predicate (cobas b 221) Devices-Similarities, cont.
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K11188) | Predicate Device<br>cobas b 221<br>(K032311) |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Hemoglobin<br>Measurement | tHb, SO2, O2Hb, HHb, COHb, MetHb:<br>Spectrophotometric | Same |
| Hematocrit<br>Measurement | Conductivity | Same |
| Sensor Technology | Amperometric and potentiometric thick film<br>microelectrode array technology for<br>Glucose and Lactate | Same |
| Sample<br>Introduction | Syringe and capillary aspiration | Same |
| On-Board<br>(In-Use) Reagent<br>Stability | Up to 42 days | Same |
| Reagent Tracking | Memory chip technology for identification,<br>lot specifications, usage tracking and<br>traceability allowing pack to be moved from<br>one system to another | Same |
| QC Material | COMBITROL PLUS B (manual QC) and<br>AUTOTROL PLUS B (automated QC) | Same |
| Calibration | Two-point liquid calibration | Same |
| Graphical User<br>Interface | Menu-driven touch screen | Same |
| Additional<br>Analyzer Hardware | Hard drive and printer | Same |
| Operating System<br>Software | Linux-based | Same |
| System Operating<br>Temperature | 15-32°C | Same |
{10}------------------------------------------------
| Substantial<br>equivalence<br>comparison- | Table 3 provides the similarities between the cobas b 123 POC<br>System (K111188) and the following predicate devices: | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------|--|
| similarities<br>(continued) | • Roche Glucose HK Assays (K921661 and K060703/A001)<br>• Roche Lactate Assays (K921661 and K060703/A001) | |
#### Table 3. Comparison of Candidate and Predicate Devices for Glucose and Lactate Measurement—Similarities
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>Roche Hitachi Glucose HK Assays:<br>Hitachi 902 (K921661) and<br>cobas c 501 (K060373/A001) |
|----------------|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Test Principle | Enzymatic | Same |
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>Roche Hitachi Lactate Assays:<br>Hitachi 902 (K921661) and<br>cobas c 501 (K060373/A001) |
| Sample Type | Whole blood | Same |
| Test Principle | Enzymatic | Same |
{11}------------------------------------------------
| Substantial<br>equivalence<br>comparison—<br>similarities<br>(continued) | Table 4 provides the similarities between the cobas b 123 AutoQC<br>Pack Tri-Level quality control materials (K111188) and the predicate<br>device, Roche AUTOTROL PLUS B (K032453). |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
·
### Table 4. Comparison of Candidate and Predicate Devices (AutoQC Materials)-Similarities
| Characteristic | Candidate Device<br>cobas b 123 AutoQC Pack Tri-Level<br>(K111188) | Predicate Device<br>Roche AUTOTROL PLUS B<br>(K032453) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Number of Levels | 3 | Same |
| Matrix | Buffered, aqueous electrolyte solution<br>equilibrated with carbon dioxide and oxygen<br>gas mixture | Same |
| Technological Characteristics | The material consists of buffered aqueous<br>electrolyte solutions with clinically relevant<br>concentrations of the targeted analytes,<br>tonometered with precision gas mixtures of<br>carbon dioxide and oxygen to achieve pH<br>and blood gas values that span the range of<br>values typical for such products with the<br>same intended use. A mixture of dyes is used<br>to simulate absorbance of hemoglobin<br>derivatives. Hematocrit is simulated by<br>conductivity. | Same |
.
Continued on next page
·
{12}------------------------------------------------
Substantial equivalence comparisonsimilarities (continued)
Table 5 provides the similarities between the modified Roche COMBITROL PLUS B quality control material (K111188) and the predicate device, Roche COMBITROL PLUS B (K032453).
#### Table 5. Comparison of Candidate and Predicate Devices (COMBITROL PLUS B)—Similarities
| Characteristic | Candidate Device<br>Modified COMBITROL PLUS B<br>(K111188) | Predicate Device<br>Roche COMBITROL PLUS B<br>(K032453) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Number of Levels | 3 | Same |
| Fill Volume | 1.7 mL | Same |
| Matrix | Buffered, aqueous electrolyte solution<br>equilibrated with carbon dioxide and oxygen<br>gas mixture | Same |
| Technological Characteristics | The material consists of buffered aqueous<br>electrolyte solutions with clinically relevant<br>concentrations of the targeted analytes,<br>tonometered with precision gas mixtures of<br>carbon dioxide and oxygen to achieve pH<br>and blood gas values that span the range of<br>values typical for such products with the<br>same intended use. A mixture of dyes is used<br>to simulate absorbance of hemoglobin<br>derivatives. Hematocrit is simulated by<br>conductivity. | Same |
{13}------------------------------------------------
Substantial Table 6 provides the similarities between the cobas b 123 AutoCVC equivalence Pack (K111188) and the predicate device, Roche AUTOTROL comparison-PLUS B (K032453). similarities (continued)
#### Table 6. Comparison of Candidate and Predicate Devices (CVC Materials)-Similarities
| Characteristic | Candidate Device<br>cobas b 123 AutoCVC Pack<br>(K111188) | Predicate Device<br>Roche AUTOTROL PLUS B<br>(K032453) |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Matrix | Buffered, aqueous electrolyte solution<br>equilibrated with carbon dioxide and oxygen<br>gas mixture | Same. |
| Technological<br>Characteristics | The material consists of buffered aqueous<br>electrolyte solutions with clinically relevant<br>concentrations of the targeted analytes,<br>tonometered with precision gas mixtures of<br>carbon dioxide and oxygen to achieve pH<br>and blood gas values that span the range of<br>values typical for such products with the<br>same intended use. A mixture of dyes is used<br>to simulate absorbance of hemoglobin<br>derivatives. Hematocrit is simulated by<br>conductivity. | Same. |
{14}------------------------------------------------
.
### 510(k) Summary, Continued
| Substantial equivalence comparison-differences | Table 7 provides the differences between the cobas b 123 POC System (K111188) and the cobas b 221 predicate device (K032311). |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
### Table 7. Comparison of Candidate (cobas b 123) and Predicate (cobas b 221) Devices—Differences
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>cobas b 221<br>(K032311) |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Sample Volume | 123 µl | 210 µl |
| Electrochemical<br>sensors | pH, PO2, PCO2, K+, Na+, iCa2+, Glucose,<br>Reference microelectrode array sensor<br>with up to 28 days in-use life<br><br>Note: Sensors with Lactate are stable up to<br>21 days. | pH, PO2, PCO2, K+, Na+, Cl-,<br>iCa2+ Reference electrodes with<br>6 to 16 months in-use life,<br>depending on the electrode type. |
| Electrochemical<br>sensor storage<br>stability | Sensor: 2°C – 8°C for 4 months | Sensor: 25°C for 18 or 24<br>months, depending on electrode<br>type |
| Auto QC<br>Module | AutoQC Module holds one AutoQC Pack,<br>which is comprised of 24 AUTOTROL<br>PLUS B tri-level ampoules. | QC module holds 120<br>AUTOTROL PLUS B tri-level<br>ampoules. |
{15}------------------------------------------------
| Substantial<br>equivalence<br>comparison—<br>differences<br>(continued) | Table 8 provides the differences between the cobas b 123 POC<br>System (K111188) and the Roche Glucose HK Assay predicate<br>devices (K921661 and K060703/A001). |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Table 8. Comparison of Candidate and Predicate Devices for Glucose Measurement—Differences
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>Roche Hitachi Glucose HK<br>Assays:<br>Hitachi 902 (K921661) and<br>cobas c 501 (K060373/A001) |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The cobas b 123 POC system is a fully<br>automated POC system for whole blood<br><i>in vitro</i> measurement of pH, blood gases<br>(BG), electrolytes Na+,<br>K+, iCa2+ (ISE),<br>hematocrit (Hct), metabolites (Glu, Lac),<br>total hemoglobin (tHb), hemoglobin<br>derivatives (O2Hb, HHb, COHb, MetHb),<br>and oxygen saturation (SO2).<br>In addition, the cobas b 123 POC system<br>calculates derived parameters.<br>It is dedicated for use in a Point-of-Care<br>environment and laboratory. The<br>integrated AutoQC module and the<br>oximeter module are available as an<br>option.<br>Depending on the equipment<br>configuration of the instrument, the<br>Sensor Cartridge and the Fluid Pack used,<br>the following parameters are measured in<br>human whole blood and QC materials.<br>See Table 1 (page 5). | Gluco-quant Glucose/HK Assay<br>(Hitachi 902): Enzymatic <i>in vitro</i><br>test for the quantitative<br>determination of glucose in human<br>serum, plasma, urine and CSF on<br>Roche automated clinical<br>chemistry analyzers.<br><br>Glucose HK Assay<br>(cobas c 501):<br><i>In vitro</i> test for the quantitative<br>determination of glucose in human<br>serum, plasma, urine, CSF and<br>hemolysate on Roche/Hitachi<br>cobas c systems. |
| Sample Type | Whole Blood | Serum, plasma, urine, CSF and<br>hemolysate |
{16}------------------------------------------------
Table 8 provides the differences between the cobas b 123 POC Substantial equivalence System (K11188) and the Roche Glucose HK Assay predicate comparisondevices (K921661 and K060703/A001). differences (continued)
### Table 8. Comparison of Candidate and Predicate Devices for Glucose Measurement-Differences, cont.
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>Roche Hitachi Glucose HK Assays:<br>Hitachi 902 (K921661) and<br>cobas c 501 (K060373/A001) |
|--------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Measurement<br>Principle | Amperometric measurement of the<br>detectable product (H2O2) under a<br>polarization voltage of 350 mV. | UV photometric measurement of the<br>detectable product (NADPH) at 340 nm. |
{17}------------------------------------------------
Substantial equivalence comparisondifferences (continued)
Table 9 provides the differences between the cobas b 123 POC System (K111188) and the Roche Lactate Assay predicate devices (K921661 and K060703/A001).
#### Table 9. Comparison of Candidate and Predicate Devices for Lactate Measurement—Differences
| Characteristic | Candidate Device<br>cobas b 123 POC System<br>(K111188) | Predicate Device<br>Roche Hitachi Lactate Assays:<br>Hitachi 902 (K921661) and<br>cobas c 501 (K060373/A001) |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The cobas b 123 POC system is a fully<br>automated POC system for whole blood<br>in vitro measurement of pH, blood<br>gases (BG), electrolytes Na+, K+, iCa2+<br>(ISE), hematocrit (Hct), metabolites<br>(Glu, Lac), total hemoglobin (tHb),<br>hemoglobin derivatives (O2Hb, HHb,<br>COHb, MetHb), and oxygen saturation<br>(SO2). In addition, the cobas b 123<br>POC system calculates derived<br>parameters.<br>It is dedicated for use in a Point-of-Care<br>environment and laboratory. The<br>integrated AutoQC module and the<br>oximeter module are available as an<br>option.<br>Depending on the equipment<br>configuration of the instrument, the<br>Sensor Cartridge and the Fluid Pack<br>used, the following parameters are<br>measured in human whole blood and<br>QC materials. See Table 1 (page 5). | Lactate Assay (Hitachi 902): For the<br>quantitative determination of<br>L-Lactate in plasma, cerebrospinal<br>fluid or whole blood on<br>Roche/Hitachi automated clinical<br>chemistry analyzers. L-Lactate<br>levels that evaluate the acid-base<br>status are used in the diagnosis and<br>treatment of lactic acidosis.<br>Lactate Gen. 2 Assay<br>(cobas c 501):<br>In vitro test for the quantitative<br>determination of lactate in human<br>plasma and CSF on Roche/Hitachi<br>cobas c systems. |
| Measurement<br>Principle | Amperometric measurement of the<br>detectable product (H2O2) under a<br>polarization voltage of 350 mV. | UV photometric measurement of the<br>detectable product (chromogen) at<br>340 nm. |
{18}------------------------------------------------
| Substantial<br>equivalence<br>comparison—<br>differences<br>(continued) | Table 10 provides the differences between the cobas b 123 AutoQC<br>Pack Tri-Level quality control materials (K111188) and the predicate<br>device, Roche AUTOTROL PLUS B (K032453). |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### Table 10. Comparison of Candidate and Predicate Devices (AutoQC Materials)-Differences
| Characteristic | Candidate Device<br>cobas b 123 AutoQC Pack Tri-Level<br>(K111188) | Predicate Device<br>Roche AUTOTROL PLUS B<br>(K032453) |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The cobas b 123 AutoQC pack<br>Tri-Level is a multi-analyte control<br>intended for use as control material to<br>monitor the measurement of pH, pCO2,<br>pO2, SO2, Na+, K+, iCa2+, Hct, tHb and<br>Hb derivatives as well as glucose and<br>lactate on cobas b 123 systems with an<br>AutoQC module. | AUTOTROL PLUS B, a Multi-<br>analyte control, is intended for use<br>as a control material to monitor the<br>measurement of pH, pCO2, pO2,<br>SO2, Na+, K+, Cl-, iCa2+, Hct, tHb<br>and Hb derivatives as well as<br>Glucose, Lactate, Urea/BUN and<br>Bilirubin on Roche OMNI S and<br>cobas b 221 analyzers with an<br>oximeter module. This control<br>material is packaged for use with the<br>AutoQC module, an option to the<br>Roche OMNI S and cobas b 221<br>analyzers. AUTO-TROL PLUS B<br>control material is not intended to be<br>used with analyzers from other<br>manufacturers. |
| Auto QC Pack | The AutoQC Pack is comprised of 24<br>AUTOTROL PLUS B tri-level<br>ampoules.<br><br>The AutoQC Pack also contains a smart<br>memory chip for tracking and<br>traceability, allowing the pack to be<br>moved from one system to another. | The QC Pack holds 120<br>AUTOTROL PLUS B tri-level<br>ampoules.<br><br>The QC Pack does not contain a<br>smart memory chip. |
| Fill Volume | 1.0 mL | 1.7 mL |
{19}------------------------------------------------
Table 11 provides the differences between the modified Roche Substantial COMBITROL PLUS B quality control material (K111188) and the equivalence comparisonpredicate device, Roche COMBITROL PLUS B (K032453). differences (continued)
### Table 11. Comparison of Candidate and Predicate Devices (COMBITROL PLUS B)-Differences
| Characteristic | Candidate Device<br>Modified COMBITROL PLUS B<br>(K111188) | Predicate Device<br>Roche COMBITROL PLUS B<br>(K032453) |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | COMBITROL PLUS B is a multi-<br>analyte control intended for use as<br>control material to monitor the<br>measurement of pH, PCO2, PO2, SO2,<br>Na+, K+, Cl-, iCa2+, Hct, tHb and Hb<br>derivatives as well as glucose, lactate,<br>urea/BUN and bilirubin on Roche<br>OMNI S or cobas b 221 analyzers with<br>an oximeter module, and cobas b 123<br>analyzers (except urea/BUN, chloride,<br>and bilirubin).<br>COMBITROL PLUS B control<br>material is not intended for use with<br>analyzers from other manufacturers. | COMBITROL PLUS B, a Multi-<br>analyte control, is intended for use<br>as control material to monitor the<br>measurement of pH, PCO2, PO2,<br>SO2, Na+, K+, Cl-, iCa2+, Hct, tHb<br>and Hb derivatives as well as<br>glucose, lactate, urea/BUN and<br>bilirubin on Roche OMNI S or<br>cobas b 221 analyzers with an<br>oximeter module..<br>COMBITROL PLUS B control<br>material is not intended for use with<br>analyzers from other manufacturers. |
{20}------------------------------------------------
Substantial Table 12 provides the differences between the cobas b 123 AutoCVC equivalence Pack (K111188) and the predicate device, Roche AUTOTROL comparison-PLUS B (K032453). differences (continued)
#### Table 12. Comparison of Candidate and Predicate Devices (CVC Materials)-Differences
| Characteristic | Candidate Device<br>cobas b 123 AutoCVC Pack<br>(K111188) | Predicate Device<br>Roche AUTOTROL PLUS B<br>(K032453) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The cobas b 123 AutoCVC pack is a multi-analyte control, intended for use in calibration verification of the measuring range established by the cobas b 123 POC system for pH, PCO2, PO2, SO2, Na+, K+, iCa2+, Hct, tHb and Hb derivatives as well as glucose and lactate on cobas b 123 systems with an AutoQC module. | AUTOTROL PLUS B, a Multi-analyte control, is intended for use as a control material to monitor the measurement of pH, pCO2, pO2, SO2, Na+, K+, Cl-, iCa2+, Hct, tHb and Hb derivatives as well as Glucose, Lactate, Urea/BUN and Bilirubin on Roche OMNI S and cobas b 221 analyzers with an oximeter module. This control material is packaged for use with the AutoQC module, an option to the Roche OMNI S and cobas b 221 analyzers. AUTO-TROL PLUS B control material is not intended to be used with analyzers from other manufacturers. |
| Indications for Use | The cobas b 123 AutoCVC Pack is an assayed control for use in calibration verification of the measuring range established by the cobas b 123 POC system for analytes listed in the package insert. | COMBITROL PLUS B / AUTOTROL PLUS B assayed controls are intended to be used to monitor and evaluate the analytical performance of the Roche OMNI S for analytes listed in the package insert. |
| Fill Volume | 1.0 mL | 1.7 mL |
#### The information provided in this Premarket Notification [510(k)] will Conclusion support a determination of substantial equivalence for the cobas b 123 POC System.
End of document
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration
Roche Diagnostics Corporation c/o Sarah Baumann Regulatory Affairs Consultant 9115 Hague Road Indianapolis, IN 46250
10903 New Hampshire Avenue Silver Spring, MD 20993
MAY 14 2012
Re: k111188
Trade Name: Cobas® b 123 POC System, Cobas® b 123 AutoQC Pack Tri-Level Cobas® b 123 AutoCVC Pack, Roche COMBITROL PLUS B Regulation Number:21 CFR §862.1120 Regulation Name: Blood gases (PC02, PO2) and blood pH test system Regulatory Class: II Product Code: CHL, JGS, CEM, JFP, GKF, CGA, KHP, KHG, GGZ, GKA, GLY, GHS, JJY Dated May 9, 2012 Received: May 10, 2012
Dear Ms. Baumann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{22}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
S. H. Liu, Ph.D.
Countney H. Lias, Ph.D. Director
Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety
Center for Devices and Radiological Health
Enclosure
{23}------------------------------------------------
510(k) Number (if known): K111188
#### Device Name: cobas® b 123 POC System
Indications for Use:
The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Nat, K+, iCa27 (ISE), hematocrit (Hct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters.
It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option.
pH, pO2 and pCO2 : pH, pO2 and pCO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Sodium (Na*): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
Potassium (K+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
Calcium (Ca2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Prescription Use __ X . AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rute Cene
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K ( 1 ) 8 8
{24}------------------------------------------------
510(k) Number (if known): K111188
### Device Name: cobas® b 123 POC System
Indications for Use, continued:
Hematocrit (Hct): hematocrit measurements are used to distinguish normal from abnormal states of whole blood, such as anemia and erythrocytosis (an increase in the number of red cells).
Glucose (Glu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic. islet cell carcinoma.
Lactate (Lac): Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.)
Total Hemoglobin (tHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.
O2Hb: oxyhemoglobin as a fraction of total hemoglobin.
COHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
MetHb: methemoglobin as a fraction of total hemoglobin.
HHb: reduced hemoglobin as a fraction of total hemoglobin.
SO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus reduced hemoglobin.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Ruti (hm
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) ¥ 1 1 1 1 8 8
Page 2 of 4
{25}------------------------------------------------
510(k) Number (if known): K111188
### Device Name: cobas® b 123 AutoQC Pack Tri-Level
Indications for Use:
The cobas b 123 AutoQC pack Tri-Level is a multi-analyte control intended for use as control material to monitor the measurement of pH, PCO2, PO2, Na , K , iCa *, Hct, tHb and Hb derivatives as well as glucose and lactate on cobas b 123 systems with an AutoQC module.
#### Device Name: cobas® b 123 AutoCVC Pack
Indications for Use:
The cobas b 123 AutoCVC Pack is a multi-analyte control, intended for use in calibration verification of the measuring range established by the cobas b 123 POC system for pH, PCO2, PO2, SO2, Na , K , iCa , Hct, tHb and Hb derivatives as well as glucose and lactate on cobas b 123 systems with an AutoQC module.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rute Cima
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k 111 88.
Page 3 of 4
{26}------------------------------------------------
510(k) Number (if known): K111188
#### Device Name: Roche COMBITROL PLUS B
Indications for Use:
COMBITROL PLUS B is a multi-analyte control intended for use as control material to monitor the measurement of pH, PCO2, PO2, SO2, Na , K , CI , iCa , Hct, tHb and Ho derivatives as well as glucose, lactate, urea/BUN and bilirubin on Roche OMNI S or cobas b 221 analyzers with an oximeter module, and cobas b 123 analyzers (except urea/BUN, chloride, and bilirubin). COMBITROL PLUS B control material is not intended for use with analyzers from other manufacturers.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR ·
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rose Cieri
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) 88
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.