← Product Code [DSI](/productcode/DSI) · K111160

# INTRICON DATRIX CENTAURI MT AMBULATORY PATIENT ECG TELEMETRY SYSTEM (APETS) (K111160)

_Intricon Datrix Corporation · DSI · Aug 1, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111160

## Device Facts

- **Applicant:** Intricon Datrix Corporation
- **Product Code:** [DSI](/productcode/DSI.md)
- **Decision Date:** Aug 1, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The IntriCon Datrix Centauri MT Ambulatory Patient ECG Telemetry System (APETS) is intended for diagnostic evaluation of patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors, automatically generates an event triggered by an arrhythmia detection algorithm or manually by the patient, and automatically transmits the recorded cardiac activity associated with these symptoms for review by a licensed physician.

## Device Story

Centauri MT APETS is an ambulatory ECG telemetry system for up to 30-day cardiac monitoring. Device acquires ECG signals via silver-chloride electrodes; supports 1, 2, or 3-channel recording. System continuously monitors cardiac activity; utilizes an internal arrhythmia detection algorithm to trigger event recording; allows patient-initiated manual event triggering. Recorded data transmitted via phone service to a receiving station for physician review. Used in outpatient settings; operated by patients; reviewed by licensed physicians. Output enables diagnostic evaluation of transient symptoms; assists in clinical decision-making regarding cardiac arrhythmia management.

## Clinical Evidence

Bench testing only. Performance, functionality, and reliability characteristics were verified following established test procedures in a quality system. Compliance with IEC 60601-1:1998 (2nd ed), IEC 60601-1-2:2001, and AAMI EC 38-1998 standards confirmed.

## Technological Characteristics

Ambulatory ECG telemetry recorder; 1, 2, or 3-channel capability; silver-chloride electrode interface; battery-powered; up to 30-day recording duration. Connectivity via phone service to receiving station. Standards: IEC 60601-1, IEC 60601-1-2, AAMI EC 38.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- Braemar, Inc. Fusion Wireless Recorder ([K081444](/device/K081444.md))
- Datrix E-Tac EX-1000 Electrocardiographic Event Recorder ([K042022](/device/K042022.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K111160
p1/2

## 510(k) SUMMARY (as required by 807.92(c))

.

AUG - 1 2011

ﯿﺴﻪ

ェ

.

| Regulatory Correspondent: | AJW Technology Consultants Inc<br>962 Allegro Lane<br>Apollo Beach, FL 33572<br>Tanya O'Brien<br>tmobrien@ajwtech.com<br>813-645-2855<br>813-677-4787                                                                                                                                                                                                                                                                                                                                                          |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of 510(k):      | IntriCon Datrix<br>340 State Place<br>Escondido, CA 92029<br>Terrie Heidemann<br>theidemann@ecgrecorder.com                                                                                                                                                                                                                                                                                                                                                                                                    |
| Date of Summary:          | 20 April 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Trade/Proprietary Name:   | IntriCon Datrix Centauri MT Ambulatory Patient<br>ECG Telemetry System (APETS)                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name:      | Arrhythmia Detector and Alarm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code:             | DSI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended Use:             | The IntriCon Datrix Centauri MT Ambulatory<br>Patient ECG Telemetry System (APETS) is<br>intended for diagnostic evaluation of patients who<br>experience transient symptoms that may suggest<br>cardiac arrhythmia. The device continuously<br>monitors, automatically generates an event triggered<br>by an arrhythmia detection algorithm or manually<br>by the patient, and automatically transmits the<br>recorded cardiac activity associated with these<br>symptoms for review by a licensed physician. |
| Device Description:       | The IntriCon Datrix Centauri MT APETS is a<br>device used for recording patient ECG data for as<br>long as 30 days.  Selected data are sent for review<br>from the recorder, through a phone service to a<br>receiving station where they are available for<br>review by a physician or other qualified medical<br>professional. The Centauri MT APETS is capable of                                                                                                                                           |

.

{1}------------------------------------------------

..............................................................................................................................................................................

|                          | 1, 2, and 3-channel recording, depending on the<br>type of cable being used. The recorder can run for<br>up to 30 days. Battery life depends on the mode and<br>battery option chosen by the user. The patient's<br>ECG signal is acquired via commercially available<br>silver-chloride electrodes appropriate for long-term<br>recording. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device:        | Braemar, Inc. Fusion Wireless Recorder (K081444),<br>Datrix E-Tac EX-1000 Electrocardiographic Event<br>Recorder (K042022)                                                                                                                                                                                                                  |
| Substantial Equivalence: | Comparison to the predicate devices listed shows<br>nearly identical technical data, same indications for<br>use, same safety standards tested to, and raises no<br>new questions of safety or efficacy.                                                                                                                                    |
| Performance Testing:     | Testing to applicable standards: IEC60601-1:1998,<br>2nd edition, IEC 60601-1-2: 2001, AAMI EC 38-<br>1998.<br><br>Testing for the performance, functionality, and<br>reliability characteristics of the device followed<br>established test procedures in a quality system.                                                                |
| Conclusion               | Comparison to the predicate devices listed shows<br>nearly identical technical data, same indications for<br>use, same safety standards tested to, and raises no<br>new questions of safety or efficacy. Therefore, the<br>IntriCon Datrix Centauri MT APETS supports a<br>claim of substantial equivalence.                                |

and the contraction of the comments of the consistence the consistence and any and

·

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

. . . . . . . . . .

and the contraction of the states of the subscribed on

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

IntriCon Datrix Corporation c/o Ms. Tanya O'Brien Clinical Affairs Specialist AJW Technology Consultants, Inc. 962 Allegro Lane Pollo Beach, FL 33572

1 2011 ઠ્ઠાંડિ

K11160 Re:

KTTT100
Trade/Device Name: IntriCon Datrix Centrauri Ambulatory Patient ECG Telemetry System (APETS) Regulatory Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: II (two) Product Code: DSI Dated: July 11, 2011 Received: July 14, 2011

Dear Ms. O'Brien:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 commored print to thirs 20, 377 7, 101 11:51
devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de rices that have been require approval of a premarket approval application (PMA). and Cosmetic Act (71ct) that to not require subject to the general controls provisions of the Act. The rou may, merelove, manel of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is elassified (600 a007 0) . Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Tanya O'Brien

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a decemmations administered by other Federal agencies. You must of any I odolar statuates and regarations, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire specific da vies to: your your centers Offices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulation antined, "incertains of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known):

Device Name: IntriCon Datrix Centauri Ambulatory Patient ECG Telemetry System (APETS)

IntriCon Datrix Centauri MT Model:

Indications for Use: The IntriCon Datrix Centauri MT Ambulatory Patient ECG Telemetry System (APETS) is intended for diagnostic evaluation of patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously montons, automatically generates an event triggered by an arrhythmia detection algorithm or manuoly by the patient, and automatically transmits the recorded cardiac activity associated with these symptoms for review by a licensed physician.

Contraindications:

- 1. Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
- Patients who the attending physician thinks should be hospitalized. 2.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Intricon Centauri MT 510(k)

---

**Source:** [https://fda.innolitics.com/device/K111160](https://fda.innolitics.com/device/K111160)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
