K111078 · Aseptico, Inc. · EKX · Jun 14, 2011 · Dental
Device Facts
Record ID
K111078
Device Name
AEU-26L ELCTRONIC ENDODONTIC SYSTEM
Applicant
Aseptico, Inc.
Product Code
EKX · Dental
Decision Date
Jun 14, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
For use in a wide range of dental procedures including; endodontic surgeries, such as drilling in to the tooth canal, and general dentistry, such as removing carious material from the dentin.
Device Story
AEU-26L is a software-controlled dental handpiece system comprising a controller console, low-voltage electric motor (AE-250L-30), foot pedal, and motor holder. Used in dental clinics by practitioners for endodontic and general procedures. System inputs include user-defined speed (300-30,000 RPM), torque settings (gram/cm), gear ratios, and drive modes (forward, reverse, reciprocating) via console interface. Foot pedal provides operational control. Motor supports geared handpieces, including those with LED optics for illumination. Device transforms electrical input into mechanical rotation for handpiece operation. Output is controlled mechanical motion for drilling or material removal. Clinical benefit includes precise control over speed/torque and illumination for improved visibility during procedures. System allows user-defined presets for workflow efficiency.
Clinical Evidence
Bench testing only. Performance assessed via comparative design verification against predicate devices. Safety and electromagnetic compatibility verified per IEC 60601-1 and IEC 60601-1-2. Motor performance verified per ISO 11498.
Technological Characteristics
Software-controlled dental handpiece system. Features: 300-30,000 RPM motor speed, adjustable torque (gram/cm), gear ratio selection, forward/reverse/reciprocating drive modes, and LED illumination support. Connectivity: Electronic foot pedal control. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 11498 (low-voltage motors).
Indications for Use
Indicated for patients undergoing dental procedures, including endodontic surgeries (e.g., tooth canal drilling) and general dentistry (e.g., carious material removal).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
K111078
Image /page/0/Picture/1 description: The image shows the logo for ASEPTICO. The logo is in a stylized font with a swirl design on the left side. The word "ASEPTICO" is written in all caps, with a registered trademark symbol to the right of the "O". The logo has a horizontal line extending from the "O" to the right.
SHIPPING: 8333 - 216th Street SE. Woodin
(425) 487-3157 WORLDWIDE
(800) 426-5913 TOLL FREE USA & CANADA
FAX: (360) 668-8722
www.aseptico.com
Email: info@aseptico.com
**DEPT OF QUALITY AND REGULATORY AFFAIRS**
Image /page/0/Picture/6 description: The image shows a series of national flags. The flags include the flags of Australia, Canada, the United States, Japan, and New Zealand. The flags are all displayed in a row, and they are all of similar size and shape. The flags are all easily recognizable.
# 510(k) Summary
#### Prepared by: Grant Ramaley
March 16, 2011
#### Device Description
Model Number: AEU-26L
Classification Name: 872.4200 Dental handpiece and accessories
Product Code: EKX handpiece, direct drive, ac-powered
Primary accessories exclusive to the AEU-26L system:
- a) AEU-26L Consol (Controller)
- b) AE-250L-30 Low Voltage Electric Motor
- c) On/Off Foot Pedal
- d) Motor holder attaching bracket
#### Intended Use
For use in a wide range of dental procedures including; endodontic surgeries, such as drilling in to the tooth canal, and general dentistry, such as removing carious material from the dentin.
#### Summary of Technological Characteristics
#### General operation and use
The AEU-26L is software controlled and has the following features:
- a) Allows adjustment of the motor speed from 300 to 30,000 RPM
- b) Allows setting the torque applied to the motor in gram/cm
- c) Allows selection of gear ratios for different geared E-Type handpieces
- d) Allows selection of forward or reverse drive rotation
- e) Allows reciprocating drive (forward/reverse cycling)
- Allows use of a foot pedal control to operate the attached handpiece motor ರಿ)
- Allows the user to define their own presets for speed (RPM) and torque (gram g)
- h) Allows for attaching geared handpieces that include optics for a light.
See Table 1 of comparisons provided at the end of this summary
{1}------------------------------------------------
#### Substantial Equivalence
#### Feature equivalency to Aseptico Model AEU-925:
All but two of the features listed above "Technological Characteristics" for the AEU-26L are already part of Aseptico's model AEU-925, which was cleared under 510(k) number K030163. Both models can be used for endodontic surgery.
#### **Allows attachment of lighted handpieces
The AEU-26L allows use of contra angle handpieces that include optics for a light, to provide illumination from the Light Emitting Diode (LED) of the AE-250L motor. This technology is substantially equivalent to the Aseptic AEU-5000 cleared under 510(k) number K103399.
#### **Reciprocating drive
Reciprocating drive is a forward and reverse cycle that can be selected as an alternate motion to simply driving forward or reverse. Bench testing of the AEU-26L establishes that the AEU-26L provides the same type of reciprocation as the ATR TECNIKA device, cleared under K000547.
** Technological characteristics from predicate devices other than Aseptico's Model AEU-925
#### Performance data
The AEU-26L was assessed to IEC 60601-1:1988-Medical electrical equipment-Part 1: General requirements for safety. This included thermal and electrical safety of the AE-250L-30 motor containing the Light Emitting Diode (LED).
The AEU-26L met electromagnetic compatibility requirements of IEC 60601-1-2—Medical electrical equipment-Part 1: General requirements for safety-2. Collateral standard: Electromannetic compatibility.
Specific performance testing relevant to the AE-250L-30 motor is met by application of ISO 11498 Dental handpieces -- Dental low-voltage electrical motors
All of the functional characteristics for the AEU-26L (e.g. directional and reciprocating drive, speed, torque control, ability to attached selected geared handpieces, etc) were established using comparative design verification methods between the predicate devices and the AEU-26L. Records of these design verification activities are maintained in the Design History File.
#### Conclusions
The AEU-26L bears all the features necessary to perform endodontic surgery of its predecessor AEU-925. using all of the same technological developments and feature that are in place for performing endodontic dental surgery. In addition, the AEU-26L includes the ability to utilize legally marketed geared handpieces that have optics to carry the light for illuminating the working area of the tooth. It also includes a reciprocating drive feature that is substantially equivalent to the ATR TECNIKA device, cleared under K000547.
{2}------------------------------------------------
### TABLE 1
### SUMMARY ONLY -- See Substantial Equivalence Discussion for more details
## Comparison chart of AEU-26L Essential Performance and Features
| | AEU-26L | AEU-925 | ATR TECNIKA | AEU-5000 |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------|--------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| a) Allows<br>adjustment of the<br>motor speed from<br>300 to 30,000 RPM | 300 to 30000 RPM<br>Configurable | 300 to 30000 RPM<br>Configurable | 1600 to 12800 RPM<br>Configurable | 2000 to 40000 RPM<br>Configurable |
| b) Allows setting<br>the torque applied<br>to the motor in<br>gram/cm | *Yes<br>Better than predicate<br>device AE- 925<br>Torque accuracy | *Yes<br>Torque accuracy | Units not in gram/cm.<br>Torque is configurable from<br>0 to 100<br>(no units specified) | Units not in gram/cm.<br>Torque is configurable from<br>5% to 100% |
| c) Allows<br>selection of gear<br>ratios for different<br>geared E-Type<br>handpieces | 1:5, 1:1<br>8:1, 16:1 Reduction | 1:5, 1:4, 1:3, 1:2, 1:1<br>35:1 to 3:1 Reduction | 1:1<br>8:1, 16:1, 18:1, 20:1<br>Reduction | 1:5, 1:2, 1:1<br>5:1 and 8:1 Reduction |
| d) Allows<br>selection of forward<br>or reverse drive<br>rotation | Yes | Yes | Yes | Yes |
| e) Allows<br>reciprocating drive<br>(forward/reverse<br>cycling) | Yes | No | Yes | No |
| f) Allows use of a<br>foot pedal control to<br>operate the<br>attached handpiece<br>motor | Yes, electronic foot control | Yes, electronic foot control | Yes, electronic foot control | Yes, pneumatic sensor |
| g) Allows the user<br>to define their own<br>presets for speed<br>and torque | Yes | Yes | Yes | Yes |
| h) Allows for<br>attaching geared<br>handpieces that<br>include optics for a<br>light. | Yes | No | No | Yes |
Page 18
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Aseptico, Incorporated C/O Mr. Jeff D. Rongero Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle, North Carolina 27709
JUN 1 4 2011
Re: K111078
Trade/Device Name: AEU-26L Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: II Product Code: EKX Dated: June 7, 2011 Received: June 8, 2011
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.tda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Anthony Di Roston
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: AEU-26L
Indications For Use:
For use in a wide range of dental procedures including; endodontic surgeries, such as drilling in to the tooth canal, and general dentistry, such as removing carious material from the dentin.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Ruane
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K11078
Page 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.