The PAINfusor Catheter, along with related accessory devices, is intended for use to provide continuous or intermittent delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for preoperative, perioperative and postoperative pain management. Routes of administration may be intraoperative or percutaneous.
Device Story
PAINfusor Catheter is a closed-tip, multi-holed, radiopaque catheter with a fenestrated length featuring a helical micro-hole pattern; designed for homogenous delivery of local anesthetics or narcotics to surgical wound sites. Used in clinical settings (intraoperative or percutaneous) for pain management; operated by healthcare professionals. Kit includes peel-away introducer needle, occlusive wound dressing, and catheter securement adhesive. Device facilitates targeted medication delivery to surgical sites, potentially improving postoperative pain control. Clinicians use graduated markings for positional reference during placement.
Clinical Evidence
Bench testing only. Validation included biocompatibility, sterilization, packaging, and shelf-life testing per ISO 10555-1:1995. Flow performance testing demonstrated <1% flow reduction across various elastomeric pumps. Kink testing confirmed functional diameter ranges. Peel testing verified sterilization did not impact adhesive strength of wound dressings.
Technological Characteristics
Closed-tip, multi-holed, radiopaque catheter; fenestrated length with helical micro-hole pattern. Constructed of Polyamide Pebax. Includes graduated markings. Supplied as kit with introducer needle, dressing, and adhesive. Sterilization and biocompatibility validated per ISO 10555-1 and ISO 10993.
Indications for Use
Indicated for patients requiring continuous or intermittent delivery of local anesthetics or narcotics to or around surgical wound sites for preoperative, perioperative, and postoperative pain management.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
{0}------------------------------------------------
# SECTION 2.
# 510(K) SUMMARY OF SAFETY AND EFFECT
JUL - 8 2011
#### K) SUMMARY OF SAFETY AND EFFECTIVENESS 2.510(
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Plan 1 Health, S.r.l.<br>Via Solari 5, 33020 Amaro (UD) Italy<br>info@p1h.it, www.plh.eu<br>Tel. +39 0433.468376<br>Fax +39 0433.468383 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| OFFICIAL<br>CORRESPONDENT | Sigi Caron<br>MedTech Consultants, Inc.<br>2400 Via Carrillo<br>Palos Verdes Estates, CA 90274<br>sigi@medtechconsultants.com<br>Tel: (310) 377-3069<br>Fax: (310) 265-7618 |
| TRADE NAME | PAINfusor Catheter |
| COMMON NAME | Anesthesia Conduction Catheter |
| CLASSIFICATION NAME | Catheter, conduction, anesthetic |
| DEVICE CLASSIFICATION | 21 CFR §868.5120 |
| PRODUCT CODES | 73 BSO |
| PREDICATE DEVICE | I-Flow Corporation's On-Q SliverSoaker Catheter,<br>(K051401), and Pajunk's Wound Infiltration Catheter<br>Kit (K080675). |
#### SUBSTANTIALLY EQUIVALENT TO:
The PAINfusor Catheter is substantially equivalent to the I-Flow Corporation's On-Q SliverSoaker Catheter, (K051401), and Pajunk's Wound Infiltration Catheter Kit (K080675). Both predicate devices are Class II devices, and are substantially equivalent to the PAINfusor Catheter in intended use and technological features.
## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The PAINfusor Catheter is a closed tip multi-holed catheter with a fenestrated length that incorporates micro-holes in a helical pattern that allow for homogenous flow of analgesia into a surgical wound. The PAINfusor Catheter is radiopaque and has graduated markings along the length of the catheter for positional reference.
{1}------------------------------------------------
It is supplied and packaged as a kit with a peel-away introducer needle, occlusive wound dressing and catheter securement adhesive.
#### INDICATIONS FOR USE:
The PAINfusor Catheter, along with related accessory devices, is intended for use to provide continuous or intermittent delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for preoperative, perioperative and postoperative pain management. Routes of administration may be intraoperative or percutaneous.
#### TECHNICAL CHARACTERISTICS:
The PAINFusor Catheter delivers local anesthetics or narcotics through a fenestrated section at the distal end of a closed catheter. The materials incorporated in the PAINfusor Catheter are substantially equivalent to those contained in the predicate devices.
#### PERFORMANCE DATA:
Safety and performance of the PAINfusor Catheter have been validated through bench, biocompatibility, sterilization, packaging, and shelf-life testing. Bench testing confirms that the PAINfusor Catheter can be used according to its intended use and in an equivalent manner to the predicate devices. Testing was conducted in accordance with ISO 10555-1:1995 "Sterile, Single Use Intravascular Catheters- Part 1: General Requirements." Flow performance testing confirmed a low rate of flow reduction (<1%) when the PAINfusor catheter is used with different size elastomeric infusion pumps. Kink testing verified that the kinking diameter for the catheter is within the specified acceptable range for functional use of the catheter. A comparison between peel testing results for the occlusive wound dressing verified the sterilization process did not affect the adhesive strength. Sterilization and biocompatibility testing requirements listed in ISO 10555-1 were met.
#### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The PAINFUSOR Catheter is an anesthesia conducting catheter substantially equivalent in intended use and mechanism of action to I-Flow Corporation's On-Q SliverSoaker Catheter, (K051401) and Pajunk's Wound Infiltration Catheter Kit (K080675). All three devices are intended to provide delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for pain management. All three devices deliver medication through multiple holes in a fenestrated length of a closed tip catheter. The differences in fenestration pattern did not affect flow characteristics, as demonstrated by flow performance testing. Biocompatibility testing per ISO10993 ensures that, although the PAINfusor is constructed of different materials (Polyamide Pebax® vs. Polyamide (Nylon) for the On-Q SliverSoaker and Polyamide with an Internal stainless steel coil for the Pajunk's Wound Infiltration Catheter Kit), the PAINfusor Catheter is as safe as the predicate devices for its intended use. All tests indicate that the PAINfusor Catheter functions in an equivalent manner to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight or flowing ribbons.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Plan 1 Health, S.R.L. C/O Ms. Sigi Caron Regulatory and Clinical Consultant Medtech Consultants, Incorporated 2400 Via Carillo Palos Verdes Estates, California 90274
JUL - 8 2011
Re: K111031 Trade/Device Name: PAINfusor Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: April 12, 2011 Received: April 14, 2011
Dear Ms. Caron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Caron
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices /CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH1's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. Newton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### 1. INDICATIONS FOR USE STATEMENT
| 510(k) Number (if known): | K111031 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: PAINfusor Catheter
Indications for Use:
The PAINfusor Catheter, along with related accessory devices, is intended for use to provide continuous or intermittent delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for preoperative, perioperative and postoperative pain management. Routes of administration may be intraoperative or percutaneous.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ohm Cilla 2/1/14
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page of of
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.