← Product Code [FFZ](/productcode/FFZ) · K111013

# ACEM (K111013)

_Sandhill Scientific, Inc. · FFZ · Dec 19, 2011 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111013

## Device Facts

- **Applicant:** Sandhill Scientific, Inc.
- **Product Code:** [FFZ](/productcode/FFZ.md)
- **Decision Date:** Dec 19, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Plethysmography Module: The device provides non-invasive measurement of pulse waveform and heart rate by photoelectric plethysmography. Electrogastrography (EGG) Module: A device that receives, records, and produces a visual display of the myoelectrical signal produced by the stomach as an aid to diagnosis of various gastric disorders.

## Device Story

ACEM is an electronic accessory signal acquisition device; functions as a module for the Sandhill InSight System. Inputs: photoelectric plethysmography (red LED and sensor) for pulse/heart rate; myoelectrical signals from stomach via standard pad electrodes. Operation: signal conditioning circuits filter and digitize inputs; signals forwarded to InSight System for display. Used in clinical settings; operated by healthcare professionals. Output: visual display of pulse waveforms and gastric myoelectrical signals. Benefit: aids diagnosis of gastric disorders and provides non-invasive cardiovascular monitoring.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing confirmed that plethysmograph and EGG channels meet product specifications and that the device is electrically safe per IEC 60601-1 and IEC 60601-1-2 standards.

## Technological Characteristics

Electronic accessory signal acquisition device. Includes five signal conditioning circuits: respiration, plethysmograph, ECG, and two EGG channels. Plethysmograph uses red LED and photoelectric sensor. EGG uses standard pad electrodes. Signals are filtered, amplified, and digitized. Connectivity: interfaces with Sandhill InSight System for display. Safety standards: IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Meridian McPulse ([K023238](/device/K023238.md))
- 3CPM EGG machine ([K984637](/device/K984637.md))

## Reference Devices

- Sandhill InSight System ([K012232](/device/K012232.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KIII0 IS PIL

DEC 1 9 2011

・

## 510(k) Summary

| Submitted by                     | Sandhill Scientific, Inc.<br>9150 Commerce Center Circle, #500<br>Highlands Ranch, CO 80129                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                   | Lewis Ward<br>L.W. Ward and Associates, Inc.<br>4655 Kirkwood Court<br>Boulder, CO 80301<br>303-530-3279<br>lwward@qwest.net                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date Prepared                    | 3-25-11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Trade Name                | ACEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common Name                      | Sensor Pac, signal acquisition                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation Number                | 870.2780 Cardiovascular (plethysmograph)<br>876.1735 Gastroenterology (electrogastrography)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Name              | Plethysmograph, Product Code JOM<br>System Electrogastrography (EGG), Product Code MYE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended Use                     | Plethysmography Module:<br>The device provides non-invasive measurement of pulse<br>waveform and heart rate by photoelectric<br>plethysmography.<br><br>Electrogastrography (EGG) Module:<br>A device that receives, records, and produces a visual<br>display of the myoelectrical signal produced by the<br>stomach as an aid to diagnosis of various gastric<br>disorders.                                                                                                                                                                                                                                                                                                                                                                                                             |
| Technological<br>Characteristics | The ACEM is an electronic accessory signal acquisition<br>device. The ACEM includes five (5) signal conditioning<br>circuits: respiration, plethysmograph, ECG and two<br>channels of the EGG. Respiration and ECG have<br>previous clearance under the InSight 510(k) K012232.<br>The new plethysmograph accessory consists of a small,<br>red LED placed against the finger and a photoelectric<br>sensor to measure changes in the light intensity. The filter<br>signal is amplified and digitized.<br><br>The added EGG signal conditioning unit receives signals<br>from the body through standard pad electrodes. The<br>signals are filtered and digitized.<br><br>Both the plethysmograph and EGG signals are forwarded<br>to the Sandhill InSight System (K012232) for display. |

{1}------------------------------------------------

| Substantial Equivalence | The plethysmograph module is substantially equivalent to<br>the Meridian McPulse device K023238.  The EGG<br>module is substantially equivalent to the 3CPM EGG<br>machine K984637. Function, technology, and signals are<br>equivalent. The ACEM is as safe and effective as the<br>legally marketed devices. |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test Data               | The ACEM has been demonstrated electrically safe<br>through testing and meeting IEC 60601-1 and IEC 60601-<br>1-2 safety standards. The ACEM does not raise new<br>issues of safety or effectiveness. Bench data confirms the<br>plethysmograph and EGG channels meet the product<br>specifications.           |

.

.

:

・・

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Flampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 1 9 2011

Sandhill Scientific, Inc. c/o Mr. Lewis Ward L.W. Ward and Associates, Inc. 4655 Kirkwood Court Boulder, CO 80301

K111013 Re:

Trade/Device Name: ACEM Sensor Pac Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: Class II (two) Product Codes: FFZ, JOM, MYE Dated: November 30, 2011 Received: December 7, 2011

Dear Mr. Ward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Lewis Ward

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/Aboutf-DA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number (if known): KIII013

Device Name: ACEM

Indications for Use:

Plethysmography Module:

The device provides non-invasive measurement of pulse waveform and heart rate by photoclectric plethysmography.

Electrogastrography (EGG) Module:

A device that receives, records, and produces a visual display of the myoelectrical signal produced by the stomach as an aid to diagnosis of various gastric disorders.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

F.C. for bdz

pvascular Devices

510(k) Number K111013

・・・

---

**Source:** [https://fda.innolitics.com/device/K111013](https://fda.innolitics.com/device/K111013)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
