POWDERED VINYL PATIENT EXAMINATION GLOVES, LIGHT YELLOW COLOR (NON-COLORED)
K110966 · Ppp Medical and Safety Products Company Limited · LYZ · Jul 21, 2011 · General Hospital
Device Facts
Record ID
K110966
Device Name
POWDERED VINYL PATIENT EXAMINATION GLOVES, LIGHT YELLOW COLOR (NON-COLORED)
Applicant
Ppp Medical and Safety Products Company Limited
Product Code
LYZ · General Hospital
Decision Date
Jul 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
Powdered Vinyl Patient Examination Gloves, Light Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable powdered vinyl patient examination gloves; light yellow color. Worn on examiner's hand or finger; provides barrier protection to prevent contamination between patient and examiner. Used in clinical or medical settings. Device meets ASTM D5250-06e1 standards for dimensions and physical properties. Non-sterile.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including physical property testing (ASTM D5250-06e1), powder residue testing (ASTM D6124-06), freedom from pinholes (21 CFR 800.20), and biocompatibility testing (primary skin irritation and dermal sensitization per ISO 10993-10).
Indicated for use as a non-sterile, disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powdered Vinyl Patient Examination Gloves, Clear(Non-colored), ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. (K091662)
{0}------------------------------------------------
# Section C 510(k) Summary (21 CFR 807.92)
# 510(k) Summary
JUL 2 1 2011
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
KIIO96 (applicant leave blank) "The assigned 510(k) number is: _
Premarket Notification [510(k)] Summary
## [{a)(1)]. The summary contains on the first page, preferably on your letterhead paper. the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | PPP Medical and Safety Products Company Limited |
|--------------------------------|-------------------------------------------------|
| Submitter's address : | Xinlian,Houqiao Town, Wuxi,Jiangsu,214106,China |
| Phone number : | 0086-510-88725378 |
| Fax number : | 0086-510-88725378 |
| Name of contact person: | Mr. Zou Xiaojun |
| Date the summary was prepared: | Feb.9, 2011 |
## [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
Device Name:
Powdered Vinyl Patient Examination Gloves, Light Yellow Color
Proprietary/Trade name:
Powdered Vinyl Patient Examination Gloves, Light Yellow Color Other clients private labeling
Common Name: Classification Name: Device Classification: Regulation Number: Panel: Product Code:
Exam gloves Patient examination glove I 21 CFR 880.6250 General Hospital (80) LYZ
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* Powdered Vinyl Patient Examination Gloves, Light Yellow Color that meets all of the requirements of ASTM standard D 5250-06e1.
{1}------------------------------------------------
Powdered Vinyl Patient Examination Gloves. Predicate device: Clear(Non-colored) ,ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. K091662.
# [(a)(4)] A description of the device
Device Description: Powdered Vinyl Patient Examination Gloves, Light Yellow Color that meets all of the requirements of ASTM standard D 5250-06e1.
#### [(a)(5)] The summary describes the intended use of the device
Device Intended Use: Powdered Vinyl Patient Examination Gloves, Light Yellow Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
## [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The Powdered Vinyl Patient Examination Gloves, Light Yellow Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|-----------------------|----------------------------------------------------------------------|-----------------------------------------|
| Dimension | ASTM standard D 5250-06 el. | Meets |
| Physical Properties | ASTM standard D 5250-06 el. | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-06 el<br>and D6124-06 | Meets<br><10mg/dm² |
| Biocompatability | Primary Skin Irritation in rabbits<br>AAMI / ANSI / ISO 10993-10 | Passes<br>Not a Primary Skin Irritation |
| | Dermal sensitization in the guinea pig<br>AAMI / ANSI / ISO 10993-10 | Passes<br>Not a Dermal sensitization |
## [(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Powdered Vinyl Patient Examination Gloves, Light Yellow Color, meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-152: AAMI / ANSI / ISO 10993-10:2002/Amd. 1:2006.
#### [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
## [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe.as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powdered Vinyl Patient Examination Gloves, Light Yellow Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AOL., meet labeling claims .
It can be concluded that the Powdered Vinyl Patient Examination Gloves, Light Yellow Color is as safe, as effective, and performs as well as the predicate device, Powdered Vinyl Patient Examination Gloves, Clear(Non-colored) , ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. K091662.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with two distinct parts. On the left, there is a circular seal with text arranged around the perimeter, reading "DEPARTMENT OF HEALTH & HUMAN". To the right of the seal, there is a stylized graphic consisting of three curved, parallel lines that resemble a symbol or abstract representation of a flag.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
PPP Medical and Safety Products Company Limited C/O Mr. Chu Xiaoan Official Correspondent Beijing Easy-Link Company, Limited Room 1606 Building 1 Jian Xian Yuan NO.209 Bei Si Huan Zhong RD. Hai Di. Beijing China 100083
JUL 2 1 2011
Re: K110966
Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Light Yellow Color
Regulation Number: 21 CFR 880.6250
Regulation Name: Patient Examination Glove Regulatory Class: I
Product Code: LYZ
Dated: July 1, 2011 Received: July 12, 2011
Dear Mr. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Winter
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section B Indications for Use
#### INDICATIONS FOR USE
PPP Medical and Safety Products Company Limited Applicant:
110966 510(k) Number (if known): *
Powdered Vinyl Patient Examination Gloves, Light Yellow Color Device Name:
#### Indications For Use:
Powdered Vinyl Patient Examination Gloves, Light Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Runningham (for ECW)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110966
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.