ICEROD PLUS 17G CRYOABLATION NEEDLE, ICEROD PLUS 17G 90* CRYOABLATION NEEDLE
K110946 · Galil Medical , Ltd. · GEH · Jun 21, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110946
Device Name
ICEROD PLUS 17G CRYOABLATION NEEDLE, ICEROD PLUS 17G 90* CRYOABLATION NEEDLE
Applicant
Galil Medical , Ltd.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jun 21, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Galil Medical Cryoablation Systems are intended for cryoablative destruction of tissue during surgical procedures. The cryoablation systems is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
Device Story
IceRod Plus 17G Cryoablation Needles are sterile, single-use, disposable components for Galil Medical Cryoablation Systems; used during surgical procedures to destroy tissue via extreme cold or warm thawing. Needles feature sharp cutting tip, shaft, color-coded handle, gas tube, and connector; include 5mm markings for positioning. Key design feature: inner vacuum-insulated tube for enhanced shaft insulation. Operated by clinicians in surgical settings; device converts high-pressure gas into freezing/thawing applications at the needle tip. Output is localized tissue destruction; aids in treating tumors, lesions, and other specified pathologies.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via dimensional, functional, freezing, and stability testing per international standards and in-house protocols. Biocompatibility confirmed per ISO 10993.
Technological Characteristics
17-gauge disposable cryoablation needle; includes inner vacuum-insulated tube for thermal insulation. Features sharp cutting tip, color-coded handle, gas tube, and connector. 5mm depth markings. Biocompatible materials per ISO 10993. Operates with high-pressure gas via Galil Medical Cryoablation System.
Indications for Use
Indicated for cryoablative tissue destruction in general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. Specific applications include prostate cancer, BPH, malignant/benign tumors, skin lesions, warts, gynecological dysplasia, hemorrhoids, and cardiac arrhythmia ablation.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" with a circle at the end of one of its arms, followed by the text "GALIL MEDICAL" in bold, sans-serif font. Below the company name is the tagline "A New Vision of Precision" in a smaller, regular font.
#### Section 5. 510(k) Summary JUN 2 1 2011
K110946
### 510(k) SUMMARY TEMPLATE
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Galil Medical Ltd.<br>Tavor 1 Building<br>Shaar Yokneam<br>Yokneam Industrial Park 20692<br>Israel |
|------------|----------------------------------------------------------------------------------------------------|
|------------|----------------------------------------------------------------------------------------------------|
| Company Contact<br>Person: | Ms. Lynne Davies<br>Sr. Regulatory Affairs Advisor<br>Galil Medical Inc.<br>Phone: 651-287-5098 |
|--------------------------------|-------------------------------------------------------------------------------------------------|
| Fax:<br>Email: | 651-287-5099<br>lynne.davies@galilmedical.com |
| Device Name: | IceRod Plus 17G Cryoablation Needles |
| Device Classification<br>Name: | Cryosurgical unit and accessories (GEH)<br>21 CFR 878.4350 |
| Predicate Device: | IceRod 17G Cryoablation Needle (K051052)<br>IceRod 17G 90° Cryoablation Needle (K060144) |
#### Device Description:
The IceRod Plus 17G Cryoablation Needles are sterile, single use, disposable components used in conjunction with a Galil Medical Cryoablation System when performing cryoablative destruction of tissue. They are intended to convert high-pressure gas to either a very cold freezing application or to a warm thawing application. The IceRod Plus needles are available in two configurations, straight and angled 90°. Each 17-gauge (17G) disposable cryoablation needle has a sharp cutting tip, a shaft, a color-coded handle, a gas tube, and a connector. Additionally, each needle exhibits markings every 5 mm to aid in positioning the needle in tissue. The IceRod Plus needles differ from the predicate device in that they contain an inner vacuum insulated tube to provide additional shaft insulation during a cryoablation procedure.
#### Intended Use:
Galil Medical Cryoablation Systems are intended for cryoablative destruction of tissue during surgical procedures. The cryoablation systems is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is
Special 510(k) - IceRod™ Plus Cryoablation Needles
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Galil Medical. The logo features a stylized letter "G" with a small circle inside, followed by the text "GALILMEDICAL" in bold, sans-serif font. Below the company name is the tagline "A New Vision of Precision" in a smaller font size.
designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
I Cruschlation Systems have the following oppositions:
| • | Urology | Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) |
|---|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Oncology | Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention |
| • | Dermatology | Ablation or freezing of skin cancers and other cutaneous disorders<br>Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin |
| • | Gynecology | Ablation of malignant neoplasia or benign dysplasia of the female genitalia |
| • | General surgery | Palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas |
| • | ENT | Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth |
| • | Thoracic surgery | Ablation of arrhythmic cardiac tissue cancerous lesions |
| • | Proctology | Ablation of benign or malignant growths of the anus or rectum and hemorrhoids |
#### Summary of Performance Data and Substantial Equivalence:
The IceRod Plus Cryoablation Needles have the same intended use and method of operation as compared to the predicate devices. The IceRod Plus needles differ in design in that they contain an inner vacuum insulated tube to enhance shaft insulation during the cryoablation procedure. Both the new and predicate devices consist of a sharp cutting tip, a shaft, a color-coded handle, a gas tube, and a connector. Both the new and predicate devices are comprised of similar materials and serve as conduits for high pressure gas during a cryoablation procedure.
Performance testing was conducted on the IceRod Plus Cryoablation Needles to verify safety and performance characteristics and to establish substantial equivalence. Testing was conducted according to protocols based on international standards and in-house requirements and included dimensional testing, functional testing, freezing performance
Special 510(k) - IceRod™ Plus Cryoablation Needles
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Galil Medical. The logo features a stylized letter "G" with a small circle inside, followed by the text "GALILMEDICAL" in bold, uppercase letters. Below the company name is the tagline "A New Vision of Precision" in a smaller font size.
and stability. Additionally, the needles meet the biocompatibility requirements outlined in ISO 10993. Test results demonstrated that the IceRod Plus needles meet defined specifications and do not raise any new safety or effectiveness issues.
#### Conclusion:
The information and data provided in this 510(k) Notification establish that the IceRod Plus 17G Cryoablation Needles are substantially equivalent to the legally marketed predicate devices.
Special 510(k) - IceRod™ Plus Cryoablation Needles
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Galil Medical Ltd. % Ms. Lynne Davies Sr. Regulatory Affairs Advisor Tavor 1 Building, Shaar Yokneam Yokneam Industrial Park 20692 Israel
JUN 2 1 2011
Re: K110946
Trade/Device Name: IceRod™ Plus Cryoablation Needles Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: May 21, 2011 Received: May 25, 2011
Dear Ms. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric Keith
C / Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
# 510(k) Number (if known):
Device Name: IceRodTM Plus Cryoablation Needles
#### Indications For Use:
Galil Medical Cryoablation Systems are intended for cryoablative destruction of tissue during surgical procedures. The cryoablation systems is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
Galil Medical Cryoablation Systems have the following specific indications:
- Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
- . Oncology (ablation of cancerous or malignant tissue and benign turnors, and palliative intervention)
- . Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell turnors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
- Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia) .
- . General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma)
- . ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
- . Thoracic surgery (ablation of arrhythmic cardiac tissue cancerous lesions)
- . Proctology (ablation of benign or malignant growths of the anus or rectum, and hemorrhoids)
Prescription Use 8 (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nailre Ogle for nkn
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.