K110919 · Carestream Health, Inc. · LLZ · Oct 20, 2011 · Radiology
Device Facts
Record ID
K110919
Device Name
CARESTREAM PACS
Applicant
Carestream Health, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems. The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates the review, dictation and reporting tools that creates a productive work environment for the radiologists and physicians. The CARESTREAM PACS Lightweight Viewer software program is used for patient management by the referral community in order to access and display patient data, medical reports, and medical images from different modalities including CR, DR, CT, MR, NM and US after the primary reading has been completed on dedicated diagnostic workstations. The CARESTREAM PACS Lightweight Viewer provides wireless and portable access to medical images for referral purposes. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography.
Device Story
CARESTREAM PACS is a scalable image management system for hospitals/clinics; archives, distributes, retrieves, and displays medical images/data from various modalities. System integrates review, dictation, and reporting tools for radiologists/physicians. Includes 'LightWeight Viewer' for web-based, portable access via HTML5-enabled devices (e.g., iPad) for referral community. Inputs include multi-modality image data; system performs 3D registration (rigid space transformation) and automatic vessel segmentation (user-adjustable). Outputs include reformatted, side-by-side, or superimposed images. Used in clinical environments; Lightweight Viewer used for remote/portable referral access when primary workstations are unavailable. Healthcare providers use output for patient management and referral purposes; does not drive therapy decisions directly. Benefits include improved workflow, rapid image distribution, and remote access to patient records.
Clinical Evidence
Bench testing only. Display performance testing conducted on iPad per IEC 62563-1 and AAPM TG18 guidelines (luminosity, resolution, noise). Simulated use studies performed by qualified radiologists across multiple modalities (CR, DR, CT, MR, NM, US) under varying environmental conditions to validate diagnostic viewing capabilities.
Technological Characteristics
Software-only PACS solution; open architecture. Supports 3D registration (rigid transformation) and vessel segmentation. Lightweight Viewer uses HTML5 for browser-based access on portable hardware. Connectivity via industry-standard interfaces. No hardware components manufactured by Carestream; relies on off-the-shelf hardware.
Indications for Use
Indicated for hospital and clinical institutions to archive, distribute, retrieve, and display medical images/data from all modalities (CR, DR, CT, MR, NM, US). Lightweight Viewer indicated for referral community access to patient data/images after primary diagnostic reading. Contraindicated for mammography in the US and not intended to replace full diagnostic workstations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K110919
Image /page/0/Picture/1 description: The image contains the word "Carestream" in a stylized font. The letters are outlined with a slightly rough or textured edge, giving them a distinct appearance. The word is presented in a single line and appears to be the primary focus of the image.
OCT 2 0 2011
Carestream Health Inc.
150 Verona Street Rochester, NY 14608
# "510(k) Summary"
.
| 510(k) Owner Name: | Carestream Health, Inc. |
|------------------------|-------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner Address: | 150 Verona Street<br>Rochester, New York 14608 |
| 510(k) Owner Phone: | 585 627-6543 |
| 510(k) Owner Fax: | 585 454-1894 |
| Contact Name & Info: | John Pardo<br>Director, Regulatory Affairs and Quality Systems<br>John.pardo@carestreamhealth.com<br>585-627-6543 |
| Date Summary Prepared: | 7/15/2011 |
| Device Trade Name: | Carestream PACS |
| Device Common Name: | PACS |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Name: | Picture Archiving and Communication System |
| Device Class: | Class II |
| Device Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Predicate Device: | Carestream PACS<br>Manufactured by Carestream Health, Inc.<br>510(k) No. - K083673 (December 30, 2008) |
.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "Carestream" in a stylized font. The letters are outlined with a double border, giving them a distinct appearance. The word is presented in a horizontal orientation, filling the majority of the frame. The font choice and outlining contribute to the visual prominence of the text.
Carestream Health Inc. 150 Verona Street Rochester, NY 14608
#### Device Description:
CARESTREAM PACS is an image management system whose intended use is to provide completely scalable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute. retrieve and display images and data from all hospital modalities and information systems.
It is a software only solution that contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates the review, dictation and reporting tools that creates a productive work environment for the radiologists and physicians. It offers an intuitive user interface which includes a tab concept, grouping features, and functions with the same context together such as Image display, Monitor layout, Comparison, Wide area user tab, etc.
The device provides functionality to allow remote site access to image and patient data enabling diagnostic reading through industry standard interfaces. It is designed using an open architecture that allows for various proprietary and off the shelf software components to be integrated with off the shelf hardware components and configured meeting the user's specific needs in a single-site or multi-site environment.
Carestream PACS provides support for 3D registration of studies taken at different times or by different modalities for reading of CT/MRI or PET-CT images. The volumetric data sets are synchronized allowing the user to view reformatted series side by side and superimposed images. In all methods the algorithm is only using a rigid space transformation. Automatic vessel segmentation suggests a segmentation that that can either be accepted, ignored or fine tuned by the user.
The CARESTREAM PACS LightWeight Viewer feature addresses the need for a fast and simple web based tool to access patient records and images. It allows for high speed distribution of image data to users in a wide area setup. The software technology uses HTML5 which allows browser enabled devices to run the application without local software installation. The LightViewer has a simple GUI for viewing including zoom, pan, windowing, basic measurements, cine, etc.
The CARESTREAM PACS Lightweight Viewer provides wireless and portable access to medical images for referral purposes. It provides a diagnostic viewer of medical images substantially equivalent to the CARESTREAM PACS software, with portable device characteristics and functionality. This device is not intended to replace full workstations and should be used only when there is no access to a workstation. The CARESTREAM PACS LightWeight Viewer software is not to be used for mammography in the US.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Carestream" in a stylized font. The letters are outlined with a double border, giving them a distinct appearance. The text is horizontally oriented and appears to be a logo or brand name.
Carestream Health Inc. 150 Verona Street Rochester, NY 14608
The CARESTREAM PACS LightWeight Viewer operates on "off-the-shelf 'portable hardware devices and is therefore subject to factors not typical for reading room workstations (e.g. screen size, environmental variability, network dependencies, etc.). It is therefore required that the user follows the operating instructions and adhere to the risk mitigation guidelines.
CARESTREAM PACS does not drive or influence the use of the source device. It does not directly drive a decision regarding treatment or therapy without the intervention of subsidiary means. Images must be evaluated and the decision regarding treatment or therapy is determined by the user based on their standard procedures. CARESTREAM PACS is a complement to these standard procedures.
## Intended Use:
The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.
The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates the review, dictation and reporting tools that creates a productive work environment for the radiologists and physicians.
The CARESTREAM PACS Lightweight Viewer software program is used for patient management by the referral community in order to access and display patient data. medical reports, and medical images from different modalities including CR, DR, CT, MR, NM and US after the primary reading has been completed on dedicated diagnostic workstations.
The CARESTREAM PACS Lightweight Viewer provides wireless and portable access to medical images for referral purposes. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography,
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Carestream" in a stylized font. The letters are outlined in black, giving them a distinct appearance. The word is presented in a clear and legible manner, making it easily readable.
Carestream Health Inc. 150 Verona Street Rochester. NY 14608
## Comparison of Technological Characteristics:
The modifications to the CARESTREAM PACS do not alter the fundamental scientific technology of the device. The only device modification was to the software. No new image manipulation tools are implemented that do not currently exist in the Carestream PACS device.
The differences in the Indication Statements between Carestream PACS and CARESTREAM PACS LightWeight Viewer describe specific restrictions on how CARESTREAM PACS LightWeight Viewer is to be used, given the hardware and portability differences between these two devices. CARESTREAM PACS LightWeight Viewer adds the explicate requirement that it should only be used when there is no access to a workstation, and that it is not to be used for mammography in the US.
#### Discussion of Testing
Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Testing was conducted under simulated use conditions. Predefined acceptance criteria was met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
Carestream Health conducted display performance testing using the CARESTREAM PACS Lightweight Viewer software on the iPad device. Testing measured contrast response and evaluated test patterns for luminosity, resolution, and noise according to IEC 62563-1 and AA PM TG18 guidelines. Carestream Health also performed multiple studies with qualified radiologists using a variety of modalities, specifically CR, DR, CT, MR, NM and US under different environmental conditions. Results of these studies affirm the diagnostic image viewing capabilities of CARESTREAM PACS LightWeight Viewer when used as indicated.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
OCT 2 0 2011
Mr. John Pardo Senior Director, Regulatory Affairs and Quality. Systems Carestream Health, Inc. 150 Vernon Street ROCHESTER NY 14608
Re: K110919
Trade/Device Name: CARESTREAM PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 29, 2011 Received: September 30, 2011
Dear Mr. Pardo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Statement of Intended Use
510(k) Number (if known): K110919
Device Name:
CARESTREAM PACS
#### Indications for Use:
The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.
The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates the review, dictation and reporting tools that creates a productive work environment for the radiologists and physicians.
The CARESTREAM PACS Lightweight Viewer software program is used for patient management by the referral community in order to access and display patient data, medical reports, and medical images from different modalities including CR, DR, CT, MR, NM and US after the primary reading has been completed on dedicated diagnostic workstations.
The CARESTREAM PACS Lightweight Viewer provides wireless and portable access to medical images for referral purposes. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography.
Prescription Use
AND/OR
Over-The-Counter Use_
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Marr Stadel
Office of In Vitro
510K K110919
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.