← Product Code [JJY](/productcode/JJY) · K110904

# RANDOX LIQUID URINE CONTROL LEVELS 2 AND 3 (K110904)

_Randox Laboratories, Ltd. · JJY · Sep 30, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110904

## Device Facts

- **Applicant:** Randox Laboratories, Ltd.
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Sep 30, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Randox Liquid Urine Controls Level 2 and Level 3 are liquid controls containing amylase, calcium, chloride, cortisol, creatinine, glucose, hCG pregnancy, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, and uric acid. They have been developed for in vitro diagnostic use in the quality control of amylase, calcium, chloride, cortisol, creatinine, glucose, hCG pregnancy, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, and uric acid assays on various clinical chemistry systems. This in vitro diagnostic device is intended for prescription use only.

## Device Story

Randox Liquid Urine Controls (Level 2 and Level 3) are liquid, ready-to-use quality control materials. They contain specific analytes (Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG, Magnesium, Microalbumin, Osmolality, pH, Inorganic Phosphate, Potassium, Total Protein, Sodium, Specific Gravity, Urea, Uric Acid). Used in clinical laboratories to monitor the performance and precision of clinical chemistry analyzers. Laboratory personnel process these controls alongside patient samples to verify assay accuracy and identify potential systematic errors. The output is a measured value compared against established target ranges; if results fall outside expected limits, the clinician or technician identifies the need for system recalibration or troubleshooting before reporting patient results. Benefits include ensuring reliable diagnostic testing and patient safety.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies confirmed 2-year shelf life and 30-day open-vial stability at 2-8°C. Value assignment validated through in-house analysis and external studies across 9 instrument families in 203 laboratories.

## Technological Characteristics

Ready-to-use liquid human urine matrix. Analytes: amylase, calcium, chloride, cortisol, creatinine, glucose, hCG, magnesium, microalbumin, osmolality, pH, inorganic phosphorus, potassium, total protein, sodium, specific gravity, urea, uric acid. Storage: 2-8°C. Compatible with various clinical chemistry analyzers (Abbott, Beckman, Roche, Siemens, etc.).

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k110904

B. Purpose for Submission:
New device

C. Measurand:
Urine quality control materials containing the following: amylase, calcium, chloride, cortisol, creatinine, glucose, hCG, magnesium, microalbumin, osmolality, pH, inorganic phosphorus, potassium, total protein, sodium, specific gravity, urea, and uric acid.

D. Type of Test:
Not applicable

E. Applicant:
Randox Laboratories, Ltd.

F. Proprietary and Established Names:
Randox Liquid Urine Controls, Level 2 and Level 3

G. Regulatory Information:
1. Regulation section:
21 § 862.1660
2. Classification:
Class I, reserved
3. Product code:
JJY

{1}

4. Panel:
75, Chemistry

H. Intended Use:

1. Intended use(s):
See indications for use below.

2. Indication(s) for use:
The Randox Liquid Urine Controls Level 2 and Level 3 are liquid controls containing amylase, calcium, chloride, cortisol, creatinine, glucose, hCG pregnancy, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, and uric acid. They have been developed for in vitro diagnostic use in the quality control of amylase, calcium, chloride, cortisol, creatinine, glucose, hCG pregnancy, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, and uric acid assays on various clinical chemistry systems. This in vitro diagnostic device is intended for prescription use only.

3. Special conditions for use statement(s):
The labeling contains a statement that all human source material was tested by FDA-approved methods and found to be non-reactive for the presence of HBsAg, HCV and antibody to HIV 1/2.

In-vitro diagnostic use. Prescription use.

4. Special instrument requirements:
Performance was evaluated on the following instrument systems: Abbott Architect c/ci systems, Beckman Coulter AU400/500/600/800, Beckman CX4/5/7/9/LX20/DxC600/800, Cobas Integra 400/800, Hitachi Series 717/7150/902/904/911/912/917, Johnson and Johnson Ortho Vitros 250/350/500/700/950/5.1FS, Roche Cobas 6000/c501/e601, Siemens Advia 1200/1650/2400, Siemens/Dade Dimension RxL/Max/Xpand, pH meter, freezing point depression osmometers.

I. Device Description:
The Randox Liquid Urine Controls Level 2 and Level 3 are ready-to-use, liquid controls derived from human urine. The source material has been tested by FDA approved methods and found non-reactive for Hepatitis B surface antigen, Hepatitis C and the antibody to HIV-1 and HIV-2. Each control level is purchased separately and

2

{2}

is packaged in 10x10 mL bottles intended to be stored at 2-8°C.

## J. Substantial Equivalence Information:

1. Predicate device name(s):
BioRad Liquichek Urine Chemistry Control, Levels 1 and Level 2

2. Predicate K number(s):
k020817

3. Comparison with predicate:
|  Comparison Table  |   |   |
| --- | --- | --- |
|  Characteristics | Device | Predicate k020817  |
|  Intended use | Intended for in vitro diagnostic use in the quality control of diagnostic assays | Same  |
|  Matrix | Human urine | Same  |
|  Format | Liquid | Same  |
|  Vial size | 10 mL | Same  |
|  Storage (unopened) | 2-8°C | Same  |
|  Open vial stability | 30 days at 2-8°C | Same  |
|  Shipping temperatures | 2-8°C | Same  |
|  Analytes | amylase, calcium, chloride, cortisol, creatinine, glucose, hCG pregnancy, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, and uric acid. | amylase, calcium, chloride, cortisol, creatinine, glucose, magnesium, microalbumin, osmolality, pH, inorganic phosphate, potassium, total protein, sodium, specific gravity, urea, urea nitrogen, and uric acid.  |

## K. Standard/Guidance Document Referenced (if applicable):

None were referenced

## L. Test Principle:

Not applicable.

{3}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability and Value Assignment: Control solutions are derived from gravimetrically prepared stock solutions and analyzed in-house. Value assignment was determined through multiple analyses in-house on instrumentation calibrated with a master lot of calibrator. Control values were also confirmed in external studies with 9 different instrument families in 203 laboratories. Assigned control ranges are listed in the labeling by instrument family and analyte methods.
Stability: Real time stability studies were conducted for closed and open vials. Shelf-Life and Open Vial Stability testing protocols and acceptance criteria were described and found to be adequate. Shelf-Life – The sponsor provided real time stability data demonstrating that the Randox Liquid Urine Controls, Levels 2 and 3, can be stored for 2 years at 2-8 °C. Open-vial – The sponsor provided data demonstrating that upon opening a vial, the material is stable for 30 days at 2-8 °C.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:

{4}

Not applicable.

b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Expected values are provided in the value assignment sheets provided with the package insert.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K110904](https://fda.innolitics.com/device/K110904)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
