← Product Code [LYR](/productcode/LYR) · K110899

# GOLD STANDARD DIAGNOSTICS H. PYLORI ELISA IGA TEST KIT (K110899)

_Gold Standard Diagnostics · LYR · Feb 1, 2012 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110899

## Device Facts

- **Applicant:** Gold Standard Diagnostics
- **Product Code:** [LYR](/productcode/LYR.md)
- **Decision Date:** Feb 1, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3110
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K110899 · Feb 1, 2012 | GOLD STANDARD DIAGNOSTICS H. PYLORI ELISA IGA TEST KIT | Gold Standard Diagnostics | Routine clinical serum samples | A clinical correlation study was conducted using 625 patient samples routinely tested for H. pylori to compare the performance of the Gold Standard Diagnostics H. pylori ELISA IgA assay against a predicate device. | Clinical correlation study; Routine clinical samples; Comparative performance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical correlation study; Retrospective clinical correlation study | Patients with gastrointestinal symptoms undergoing routine H. pylori testing; Sample Size: 625 | Micro Detect Inc. Pylori Detect IgA (K003794) | Percent positive agreement, percent negative agreement, and overall agreement |

## Indications for Use

The Helicobacter pylori (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H. pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

## Device Story

The device is an ELISA (Enzyme-Linked Immunosorbent Assay) kit for detecting IgA antibodies against H. pylori in human serum. It utilizes microtiter wells coated with purified H. pylori antigen. Patient serum is added; if IgA antibodies are present, they bind to the antigen. After washing, an HRP-conjugated anti-human IgA is added, followed by a TMB substrate. The resulting color intensity is measured spectrophotometrically at 450nm. The assay is performed in a clinical laboratory setting by trained technicians. The output is a quantitative optical density (OD) value, which is interpreted to provide a qualitative result. This result assists clinicians in diagnosing H. pylori infection in symptomatic patients when used alongside clinical findings and an IgG antibody test.

## Clinical Evidence

Method comparison study conducted with 628 clinical samples across three sites. Compared subject device to predicate. Results: 94.5% positive agreement, 93.8% negative agreement, 94.0% overall agreement. Discrepant analysis performed using a second commercial assay. No clinical sensitivity/specificity studies against gold standard (biopsy/culture/urea breath test) were performed.

## Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA). Components: 96-well polystyrene microtiter plate coated with purified H. pylori antigen, wash buffer, diluent, controls, IgA conjugate, TMB substrate, and citrate stopping solution. Detection via spectrophotometry at 450 nm. Standardized via CLSI guidelines (EP17-A, EP7-A2).

## Regulatory Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

## Predicate Devices

- Micro Detect, Inc. Pylori Detect IgA ELISA ([K003794](/device/K003794.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K110899

B. Purpose for Submission:
To obtain a substantial equivalence determination for *Helicobacter pylori* (H. pylori) ELISA IgA test Kit

C. Measurand:
IgA antibodies to *H. pylori*

D. Type of Test:
Enzyme linked immunosorbent assay

E. Applicant
Gold Standard Diagnostics

F. Proprietary and Established Names:
*Helicobacter pylori* IgA test kit

G. Regulatory Information:
1. Regulation section:
26 CFR 866.3110 – Campylobacter fetus serological reagents
2. Classification:
Class I
3. Product code:
LYR – Campylobacter pylori

{1}

4. Panel:
83 - Microbiology

H. Intended Use:

1. Intended use:
The *Helicobacter pylori* (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H. pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

2. Indications for use:
The *Helicobacter pylori* (H. pylori) ELISA IgA test kit is intended for the qualitative detection of IgA antibodies to H. pylori in human serum in the adult population. This test is intended as a second test to aid in the diagnosis of H. pylori in patients with gastrointestinal symptoms, in conjunction with clinical findings. It should be performed and interpreted with another assay for detection of IgG antibodies to H. pylori.

3. Special conditions for use statement:
For Prescription Use

4. Special instrument requirements:
N/A

I. Device Description:
The *Helicobacter pylori* (H. pylori) ELISA IgA test is an enzyme linked immunosorbent assay. The device consists of a kit containing wash buffer, diluent, a negative control, positive control, and a cut-off control, IgA conjugate, TMB solution and Citrate - Stopping solution. The kit contains a microtiter plate consisting of 96 antigen coated, breakable single wells. The reagents are sufficient for 96 determinations.

J. Substantial Equivalence Information:

1. Predicate device name:
Micro Detect Inc. Pylori Detect IgA

{2}

2. Predicate K number(s):

K003794

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | For the detection of IgA antibodies to H. pylori in human serum | Same  |
|  Methodology | Enzyme linked immunosorbent assay | Same  |
|  Matrix | Serum | Same  |
|  Reader/Wavelength | Spectrophotometer /450 nm | Same  |
|  Reagents | Substrate (TMB), positive and negative controls | Substrate (TMB), positive and negative controls  |
|  Indications for Use | Use as a second test. To be performed in conjunction with IgG | Use as a second test. to be performed in conjunction with IgG  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Incubation time | 90 minutes | 55 minutes  |
|  Reagents | Wash solution-20x Diluent – ready to use | Diluent wash concentrate – 25x  |
|   |  Kit contains H. pylori cut off control | Kit contains H. pylori calibrator  |

# K. Standard/Guidance Documents Referenced :

EP 17-A Protocols for determination of Limits of Detection and Limits of Quantitation; Approved Guideline, CLSI Vol 24, No.34, 2004

EP 7-A2 Interference Testing in Clinical Chemistry, Approved Guideline,  $2^{\mathrm{nd}}$  ed., CLSI Vol 25, No 27, 2005

# L. Test Principle:

The Helicobacter pylori (H. pylori) ELISA IgA test is an enzyme-linked immunosorbent assay. Purified antigen is bound to microwells in a polystyrene

{3}

microtiter plate. Serum is added to each well and incubated for 30 minutes at 37°C. If H. pylori IgA antibodies are present they will bind to the antigen in the well. Unbound serum is removed by washing the wells three times. An HRP-conjugated anti-human IgA is then added to each well and incubated for 30 minutes at 37°C. The wells are washed three times to remove any unbound conjugate. A TMB substrate is added to each well and the plate is incubated for 30 minutes at 37°C. This reaction generates a blue color produced by the bound enzyme (peroxidase). The color changes to yellow when the stopping solution is added and the color intensity is measured spectrophotometrically

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

**Precision**

Intra and inter assay precision were calculated by running six patient sera (four positives and two negatives at three different sites. Results are summarized in the table below:

|   |   | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Site 1 | Ave: | 0.945 | 0.803 | 0.860 | 0.498 | 0.091 | 0.115  |
|   |  SD: | 0.024 | 0.034 | 0.039 | 0.025 | 0.006 | 0.008  |
|   |  CV: | 2.5% | 4.2% | 4.5% | 5.0% | 6.3% | 7.1%  |
|  Site 2 | Ave: | 1.030 | 0.737 | 0.716 | 0.579 | 0.091 | 0.137  |
|   |  SD: | 0.060 | 0.057 | 0.041 | 0.046 | 0.004 | 0.009  |
|   |  CV: | 5.9% | 7.7% | 5.7% | 7.9% | 4.8% | 6.9%  |
|  Site 3 | Ave: | 0.996 | 0.667 | 0.768 | 0.579 | 0.091 | 0.151  |
|   |  SD: | 0.108 | 0.035 | 0.057 | 0.042 | 0.010 | 0.009  |
|   |  CV: | 10.8% | 5.3% | 7.4% | 7.3% | 12.2% | 6.1%  |
|  Inter-Assay | Ave: | 0.964 | 0.770 | 0.812 | 0.527 | 0.090 | 0.126  |
|   |  SD: | 0.062 | 0.063 | 0.074 | 0.049 | 0.007 | 0.017  |
|   |  CV: | 6.4% | 8.1% | 9.1% | 9.4% | 7.4% | 13.7%  |

## Reproducibility

The reproducibility of the assay was done by testing three samples in triplicate (a high negative, low positive and a moderate positive) for five days, twice a day, at three sites with two technicians per site. The results are summarized in the following table:

{4}

|   |   | 5 day average: | Sample 1 | Sample 2 | Sample 3  |
| --- | --- | --- | --- | --- | --- |
|  Site 1 | Tech 1 | OD: | 0.614 | 0.724 | 1.545  |
|   |   |  SD: | 0.045 | 0.066 | 0.091  |
|   |   |  CV: | 7.3% | 9.1% | 5.9%  |
|   |  Tech 2 | OD: | 0.588 | 0.717 | 1.494  |
|   |   |  SD: | 0.059 | 0.073 | 0.146  |
|   |   |  CV: | 10.1% | 10.2% | 9.8%  |
|  Site 2 | Tech 1 | OD: | 0.594 | 0.704 | 1.546  |
|   |   |  SD: | 0.034 | 0.045 | 0.082  |
|   |   |  CV: | 5.7% | 6.4% | 5.3%  |
|   |  Tech 2 | OD: | 0.618 | 0.845 | 1.741  |
|   |   |  SD: | 0.036 | 0.044 | 0.071  |
|   |   |  CV: | 5.8% | 5.2% | 4.1%  |
|  Site 3 | Tech 1 | OD: | 0.356 | 0.480 | 1.046  |
|   |   |  SD: | 0.048 | 0.087 | 0.128  |
|   |   |  CV: | 13.4% | 18.0% | 12.2%  |
|   |  Tech 2 | OD: | 0.478 | 0.636 | 1.323  |
|   |   |  SD: | 0.086 | 0.110 | 0.148  |
|   |   |  CV: | 18.0% | 17.3% | 11.2%  |

b. Linearity/assay reportable range:

N/A

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Controls provided are ready to use. Units of the cut-off controls are defined as 10 units. The calculated units of the positive controls should be within the ranges mentioned in the quality control certificate. If those requirements (OD values, units) are not fulfilled, the test has to be repeated.

d. Detection limit:

The limit of detection was determined by testing the kit controls along with a pooled positive sample and several dilutions of the pooled sample. The limit of detection was determined at a 3.3 fold dilution of the pooled sample giving an OD value of 0.558 with a corresponding unit value of 11.3

e. Analytical specificity:

Cross reactivity

An adsorption study was performed to evaluate cross reactivity. Briefly, sera with different levels of antibodies to $H.$ pylori were adsorbed with either of the following organisms; $H.$ pylori, Candida albicans, E. coli, Borrelia burgdorferi, Clostridium spp., Campylobacter, Bacillus, Enterobacter, Pseudomonas, Proteus, or two different strains of $H.$ influenza. The identity of the bacteria used were identified by the ATCC and confirmed by mass

{5}

spectrometry. The bacteria were evaluated at a concentration of  $10^{7}$  CFU/ml or higher. The adsorbed samples were compared to the untreated samples and the mean percent inhibition was calculated. The results are summarized in the following table:

|  Organism | Concentration (CFU/ml) | Mean percent inhibition  |
| --- | --- | --- |
|  Helicobacter pylori |  | 96%  |
|  Candida albicans | 2.40x107 | 0.8%  |
|  Escherichia coli | 6.90x107 | 2.4%  |
|  Borrelia burgdorferi | 1.00x108 | 4.3%  |
|  Clostridium spp. | 1.20x107 | 1.7%  |
|  Campylobacter | 1.50x109 | 3.3%  |
|  Bacillus | 4.40x107 | 10.9%  |
|  Enterobacter | 1.80x108 | 4.1%  |
|  Pseudomonas | 1.45x108 | 3.8%  |
|  Haemophilus Influenza | 7.90x107 | 5.8%  |
|  Proteus | 1.40x108 | 3.1%  |

The mean percent inhibition for  $H.$  pylori was  $96\%$ , and  $0.8\%$  to  $10.9\%$  with the other organisms. Overall no effects on the analytical specificity were seen with the Gold Standard Diagnostics  $H.$  pylori ELISA IgA assay.

# Interfering Substances

# Interference

The effect of potential interfering substances on samples using the Gold Standard Diagnostics H. pylori ELISA IgA assay was evaluated. High levels of hemoglobin, bilirubin, cholesterol and triglycerides in serum samples were tested. The concentrations, recommended in the guideline "Interference Testing in Clinical Chemistry" from the Clinical and Laboratory Standards Institute, were used. The tested substances did not affect the performance of the Gold Standard Diagnostics H. pylori ELISA IgA assay.

{6}

7

|  Substance | Concentration | H. pylori concentration | Mean Percent Inhibition  |
| --- | --- | --- | --- |
|  Hemoglobin | 2 g/L | 9-11 units | 9%  |
|  Bilirubin | 342 μmol/L | 9-11 units | 7%  |
|  Cholesterol | 13 mmol/L | 9-11 units | 0%  |
|  Triglyceride | 37 mmol/L | 9-11 units | -32%  |

f. Assay cut-off:

The cut-off was determined by testing normal sera, clinical defined samples, proficiency samples, along with sera positive for H. pylori, borderline positive for H. pylori, and high negative for H. pylori for a total of 254 samples. The cutoff was determined by taking the negative average plus three standard deviations from normal sera samples. The cutoff was further adjusted so that the clinically defined samples were positive, the proficiency samples met their criteria, the positive H. pylori sera were positive, and the H. pylori negative sera were negative.

2. Comparison studies:

a. Method comparison with predicate device:

The performance of the Gold Standard Diagnostics H. pylori ELISA IgA assay was determined by conducting a correlation study using 628 samples being routinely tested for H. pylori. Testing was conducted at three geographically diverse clinical sites. The samples were tested with both the Gold Standard Diagnostics H. pylori ELISA IgA assay and the predicate device. The results are summarized in the following table:

|  Gold Standard Diagnostics IgA ELISA | Predicate Device IgA ELISA  |   |   |
| --- | --- | --- | --- |
|   |  Positive | Borderline | Negative  |
|  Positive | 121 | 31 | 26  |
|  Borderline | 13 | 8 | 23  |
|  Negative | 7 | 5 | 394  |

%Positive Agreement = 94.5% (C.I. 76.0 - 100%)
%Negative Agreement = 93.8% (C.I. 88.8% - 99.5%)
Overall Agreement = 94.0% (C.I. 85.9% - 100%)

The discrepant samples were further tested with a second commercially available assay. Of the seven samples, that were negative by the Gold Standard assay and positive by the predicate device, five samples were positive by the second assay. Of the 26 Gold Standard Diagnostics positive samples, which were negative by

{7}

the predicate device, two samples were borderline, one sample was negative and 23 samples were positive by the second assay.

b. Matrix comparison:

N/A

3. Clinical studies:

a. Clinical Sensitivity:

N/A (comparison was not done to the gold standard e.g. biopsy, culture and/or histological examination, or the urea breath test.

b. Clinical specificity:

N/A (comparison was not done to the gold standard.)

c. Other clinical supportive data (when a. and b. are not applicable):

N/A

4. Clinical cut-off:

N/A

5. Expected values/Reference range:

H. pylori is universally distributed and affects all genders, ages and races. The prevalence of infection with H. pylori is higher in underdeveloped countries and in the communities with a low standard of living and poor hygiene. Studies in asymptomatic Caucasians in the United States show that there is an increase in the prevalence of H. pylori infection with age. An expected values study was performed with the H. pylori ELISA Ig A assay testing 628 patients from various populations. Results showed that a total of 121 patients were positive with the assay. The expected values result seen in this study for the H. pylori IgA assay is 19%.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K110899](https://fda.innolitics.com/device/K110899)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
