← Product Code [KCT](/productcode/KCT) · K110898

# S.C.O.R.E.S. UNIT (K110898)

_Ammed Surgical Equipment, LLC · KCT · Jul 26, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110898

## Device Facts

- **Applicant:** Ammed Surgical Equipment, LLC
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Jul 26, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 27.0° and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray. The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm. Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.

## Device Story

S.C.O.R.E.S. Unit is a stainless steel sterilization container and delivery system; holds multiple instrument trays for single surgical cases. System includes cabinet with internal shelves, filtration system for steam penetration, and adjustable transport cart. Used in clinical settings; operated by healthcare staff. Loaded unit transported via cart into prevacuum steam autoclave; processed in standard cycle; dried in chamber. Maintains sterility of contents for up to 30 days post-sterilization. Eliminates need for individual tray wrapping. Benefits include streamlined instrument delivery to surgical suite and maintained sterile integrity.

## Clinical Evidence

Bench testing only. Validation performed using a hospital steam sterilizer (Amsco Model 3043) with prevacuum steam half-cycles. Testing confirmed successful sterilization and drying of contents. Sterile maintenance validated for up to 30 days. Validated for specific lumen dimensions (≥3.8mm ID, ≤370mm length) and weight limits (max 25 lbs/tray).

## Technological Characteristics

Stainless steel sterilization cabinet with filtration system. Wheeled form factor for transport via adjustable cart. Operates in prevacuum steam sterilizers. Mechanical design; no electronic or software components.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Allegiance Genesis Container System

## Reference Devices

- Amsco (Steris) Model 3043

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUL 26 2012

# 510(k) Summary Pursuant to 21 CFR 807.92

## K110898

## Date of preparation: 5 July 2012

- Submitted By: 1.
AmMed Surgical Equipment, LLC 3801 Corporex Park Drive, Suite 120 Tampa, Florida 33619

2. Contact: David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733 512-906-9654

3. Product: S.C.O.R.E.S Units. Self Contained Operating Room Equipment Sterilization Containers Regulation: 880.6850. Sterilization Wrap Class II Product Code: KCT

#### 4. Common/Trade Name:

S.C.O.R.E.S Units

Sterilization Wrap/Container

#### Description:

The S.C.O.R.E.S. Unit (self contained operating room equipment sterilization unit) is a sterilization container and delivery system capable of holding multiple instrument sterilization trays. The S.C.O.R.E.S. container can be loaded with multiple surgical instrument trays needed for a single surgery, steam sterilized, dried in the chamber drying cycle, removed from the autoclave on a transport cart, stored for up to 30 days, and delivered to the surgical suite for opening and use of the contents. The system includes the S.C.O.R.E.S. cabinet with wheels for loading into sterilizers, internal shelves for placement of sterilization trays, a filter system to allow steam penetration and maintain sterility, and an adjustable transport cart to allow for movement of the loaded system.

{1}------------------------------------------------

Instrument trays within the S.C.O.R.E.S Unit do not need to be placed in sterilization containers or wrapped to maintain sterile integrity as the filtered Unit does so. Once opened in the surgical suite and the contents are to be used immediately. The subject device has been extensively tested, and verified to perform in a safe and effective manner.

## Intended Use:

The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks.

The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 27.0° and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.

The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.

Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.

# Technological Characteristics:

The S.C.O.R.E.S. Unit is a sterilization cabinet constructed from stainless steel and provided with a filtration system.

The Unit can be loaded with all of the instrumentation needed for a surgical case. Instruments are placed into trays, which need not be wrapped. Individual trays of instruments should weigh no more than 251bs. and up to 12 trays can be arranged on the shelves of the S.C.O.R.E.S. Unit. The loaded unit has a filtration system to allow for steam penetration.

{2}------------------------------------------------

The wheeled Unit is transported on an adjustable cart, which can be adjusted to the height of a prevacuum autoclave chamber. The sealed, loaded unit can be rolled into the chamber and processed in a normal cycle.

The S.C.O.R.E.S. Unit has been validated in accordance with relevant industry standards using a hospital steam sterilizer. Validation was done with prevacuum steam half cycles and drying was also accomplished. Validation was done with an Amsco (Steris) Model 3043.

The product has also been shown to be capable of maintaining sterility of the contents for up to 30 days.

## Substantial Equivalence:

The S.C.O.R.E.S. Unit has been determined to be substantially equivalent to the Allegiance Genesis Container System. For prevacuum sterilization applications, the predicate device has similar technical features, indications for use, and the safety and effectiveness of the devices is equivalent.

# Conclusions:

The predicate devices and the AmMed Surgical S.C.O.R.E.S Unit container system share similar indications, technology, and application. Although the S.C.O.R.E.S. unit holds more equipment, the device is essentially equivalent to the predicate device products in key areas of performance that affect safety and effectiveness.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake entwined around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL 2 6 2012

Ammed Surgical Equipment, LLC C/O Mr. David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733

Re: K110898

Trade/Device Name: Scores Unit Sterilization Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: July 5, 2012 Received: July 13, 2012

Dear Mr. Furr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Mr. Furr

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{5}------------------------------------------------

# Indications for Use Statement

K110898 . 510(k) Number:

Device Name: SCORES® Unit Sterilization Container

Indications for Use:

The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks.

The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270° and exposure time of 4 minutes. Use no more than 3 travs per shelf or 25 lbs. per tray.

The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.

Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.

| Prescription Use  |  |
|-------------------|--|
| (per CFR 801.109) |  |

or

| Over-the-counter use | X |
|----------------------|---|
|----------------------|---|

Concurrence of CDRH

E. Mitchell F. O'Quinn - Williams

Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, Deneral Devices Division of Arcel. Dental Devices

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K110898](https://fda.innolitics.com/device/K110898)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
