← Product Code [LLZ](/productcode/LLZ) · K110875

# CENTRICITY PACS (K110875)

_GE Healthcare · LLZ · May 5, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110875

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** May 5, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

Centricity PACS software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital images and data acquired from diagnostic imaging devices. The Centricity PACS Workstation software is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility.

## Device Story

Centricity PACS is an enterprise-grade software system for managing digital medical images and associated data. It ingests images from various diagnostic imaging modalities (CT, MR, CR, DX, MG, US, NM, PET, XA, IO, ES) via DICOM and HL7 standards. The system provides infrastructure for database management, image storage, retrieval, distribution, and printing. Clinicians and radiologists use the workstation interface to display, manipulate, annotate, and perform primary diagnostic interpretation of images. The system supports clinical workflows across healthcare facilities. By centralizing image access and providing tools for analysis and review, the device facilitates efficient diagnostic decision-making and patient care. It is a software-only product designed for integration into existing hospital IT environments.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including risk analysis, requirements reviews, design reviews, module verification, system integration testing, performance verification, safety testing, and simulated use validation.

## Technological Characteristics

Software-only PACS; supports DICOM and HL7 standards; integrates with IHE profiles. No specific materials, chemical composition, or energy sources as it is a software product. Operates on standard IT infrastructure. Connectivity via networked DICOM/HL7 interfaces. No specific algorithm class stated.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Centricity™ PACS ([K043415](/device/K043415.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

110875

## 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                        | February 11, 2011                                                                                         |
|------------------------------|-----------------------------------------------------------------------------------------------------------|
| Submitter:                   | GE Healthcare<br>540 W. Northwest Highway<br>Barrington, IL 60010                                         |
| Primary Contact Person:      | Cheryl Bork<br>Regulatory Affairs Leader<br>GE Healthcare<br>Phone: 847-277-6038<br>Fax: 847-277-5240     |
| Secondary Contact<br>Person: | Jeme Wallace<br>Director, Regulatory Affairs<br>GE Healthcare<br>Phone: 847-277-4468<br>Fax: 847-277-5240 |
| Device:<br>Trade<br>Name:    | Centricity PACS                                                                                           |
| Common/Usual<br>Name:        | PACS, Picture Archiving and Communication System                                                          |
| Classification Names:        | 21 CFR 892.2050 System, Image Processing, Radiological                                                    |
| Product Code:                | LLZ                                                                                                       |
| Secondary Product<br>Code:   | GCJ<br>876.1500 Endoscope and accessories                                                                 |
| Predicate Device(s):         | K043415 Centricity™ PACS                                                                                  |

{1}------------------------------------------------

| <b>Device Description:</b> | Centricity PACS is an enterprise grade Picture Archiving and<br>Communications System (PACS) for managing digital medical<br>images and associated data. Centricity PACS enables the<br>storage, retrieval, distribution, printing, and presentation of<br>images acquired from diagnostic imaging modalities.<br><br>Centricity PACS is a standards-based, customizable, and<br>scalable solution supporting several of the Integrating the<br>Healthcare Enterprise (IHE) profiles, Digital Imaging and<br>Communications in Medicine (DICOM), and the Health Level<br>Seven (HL7) protocol standards for managing digital medical<br>images and patient data. Centricity PACS supports<br>radiographic imaging-as in clinical radiography, cardiology,<br>dentistry, and mammography and non-radiologic imaging,<br>including video support. |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Intended Use:</b>       | Centricity PACS software product is intended for the storage,<br>reading, diagnostic review, analysis, annotation, distribution,<br>printing, editing, and processing of digital images and data<br>acquired from diagnostic imaging devices.<br><br>The Centricity PACS Workstation software is intended for use<br>as a primary diagnostic and analysis tool for diagnostic<br>images by trained healthcare professionals, including<br>radiologists, physicians, technologists, clinicians and nurses. It<br>is also intended for use as a clinical review workstation<br>throughout the healthcare facility.                                                                                                                                                                                                                                |

{2}------------------------------------------------

. . . . . . .

| Technology: | Centricity PACS employs the same fundamental scientific |
|-------------|---------------------------------------------------------|
|             | technology as its predicate devices.                    |

| Summary of<br>Technological<br>Characteristics: |                         | Predicate Device<br>Centricity PACS<br>K043415 (v2.1)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Proposed Device<br>Centricity PACS 3.2.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-------------------------------------------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                 | DESIGN                  | Centricity PACS is a software<br>product with optional turnkey<br>solution for hardware<br>components.<br><br>Centricity PACS is a standards-<br>based, customizable, and<br>scalable solution supporting<br>several of the Integrating the<br>Healthcare Enterprise (IHE)<br>profiles, Digital Imaging and<br>Communications in Medicine<br>(DICOM), and the Health Level<br>Seven (HL7) protocol standards<br>for managing digital medical<br>images and patient data.<br>Centricity PACS supports<br>radiographic imaging-as in<br>clinical radiography,<br>cardiology, dentistry, and<br>mammography and non-<br>radiologic imaging, including<br>video support. | Centricity PACS v3.2.1 employs<br>the same fundamental<br>scientific technology as its<br>predicate devices, however<br>v3.2.1 is a software only<br>product.<br><br>Centricity PACS is a standards-<br>based, customizable, and<br>scalable solution supporting<br>several of the Integrating the<br>Healthcare Enterprise (IHE)<br>profiles, Digital Imaging and<br>Communications in Medicine<br>(DICOM), and the Health Level<br>Seven (HL7) protocol standards<br>for managing digital medical<br>images and patient data.<br>Centricity PACS supports<br>radiographic imaging-as in<br>clinical radiography,<br>cardiology, dentistry, and<br>mammography and non-<br>radiologic imaging, including<br>video support. |
|                                                 | Material                | Not Applicable - Centricity<br>PACS is a software product,<br>and therefore, has no specific<br>materials of manufacture.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Not Applicable - Like the<br>predicate product, the<br>Centricity PACS v3.2.1, is a<br>software only product, and<br>therefore, has no specific<br>materials of manufacture.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                 | Chemical<br>Composition | Not Applicable - Centricity<br>PACS is a software product,<br>and therefore, has no specific<br>chemical composition.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not Applicable - Like the<br>predicate product, the<br>Centricity PACS v3.2.1, is a<br>software only product, and<br>therefore, has no specific<br>chemical composition.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                 | Energy<br>Source        | Not Applicable - Centricity<br>PACS is a software product,<br>and therefore, has no Energy<br>Source.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not Applicable - Like the<br>predicate product, the<br>Centricity PACS v3.2.1, is a<br>software only product, and<br>therefore, has no Energy<br>Source.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{3}------------------------------------------------

## Determination of Substantial Equivalence:

Summary of Non-Clinical Tests:

Centricity PACS comply with voluntary standards as detailed in Section 9 of this premarket submission. The following quality assurance measures were applied to the development of the system:

- Risk Analysis
- Requirements Reviews
- · Design Reviews
- · Testing on unit level (Module verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- · Safety testing (Verification)
- · Simulated use testing (Validation)

# Summary of Clinical Tests:

The subject of this premarket submission, Centricity PACS, did not require clinical studies to support substantial equivalence.

GE Healthcare considers Centricity PACS to be as safe, as effective, Conclusion: and performance is substantially equivalent to the predicate device(s).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

GE Healthcare % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062

MAY - 5 2011

Re: K110875

Trade/Device Name: Centricity PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ and GCJ Dated: March 29, 2011 Received: March 30, 2011

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of

{5}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 5.10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Form

FDA CDRH DMC K56

| 510(k) Number (if known): |                 |
|---------------------------|-----------------|
| Device Name:              | Centricity PACS |

|  | MAR 3 0 2011 |
|--|--------------|
|  | Received     |

### Indications for Use:

The Centricity PACS provides scalable image and data management solutions for medical imaging modalities, such as Computed Tomography (CT). Magnetic Resonance (MR), Computed Radiography (CR), Digital X-Ray (DX), Digital Mammography (MG), Ultrasound (US), Nuclear Medicine (NM), Positron Emission Tomography (PET), X-Ray Angiography (XA), Oral X-Ray (IO), Endoscopic Video (ES), and any other DICOM devices.

The workstation interface software provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure.

To be viewed for primary interpretation, the digital mammography images must be acquired from an FDA approved Full Field Digital Mammography (FFDM) device for primary interpretation. Furthermore, the FFDM must be able to provide, to the Centricity PACS, a viewable DICOM 'for presentation' mammography image as approved by the FDA for primary interpretation. Images that are printed to film must be printed using a FDA approved printer for the diagnosis of digital mammography images.

To be viewed for primary diagnosis, digital mammography images must be viewed on a display system that has been cleared by the FDA for the diagnosis of digital mammography images.

The Centricity PACS allows integration with other open interfaces, such as DICOM. to web client products and archive devices.

The Centricity Infrastructure software provides for the system's database and image management, printing, HL-7 interfacing, and all DICOM services including but not limited to, Store, Print, Query/Retrieve, and Send.

It is the user's responsibility to ensure quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 801 Subpart C)

Page 1 of 2

નું

ﻟﻤﻌﺔ

{7}------------------------------------------------

# Indications for Use Form

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary S Patel

Division Sign Off Office of In Vitro Diagnostic Device Evaluation and Safety

**510(k)** K110875

---

**Source:** [https://fda.innolitics.com/device/K110875](https://fda.innolitics.com/device/K110875)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
