K110829 · Barosense, Inc. · OCW · Jun 3, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K110829
Device Name
ACE STAPLER AND CARTRIDGE
Applicant
Barosense, Inc.
Product Code
OCW · Gastroenterology, Urology
Decision Date
Jun 3, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The BaroSense ACE Stapler and Cartridge are indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.
Device Story
The ACE Stapler is a surgical device used in hospitals or surgery centers for tissue closure in the stomach or GI tract. It consists of a reusable, flexible handle and a single-patient-use, disposable stapler head fitted with a sterile, single-use staple cartridge. Operated by a physician, the device is introduced trans-orally; a flexible endoscope is passed through the stapler for visualization. The device uses a vacuum pump to create a tissue plication, which is then compressed and secured with a double, circular row of titanium staples and a non-absorbable reinforcement ring. Tissue compression and stapling are controlled via external inflation syringes. The device may be used with a guidewire or an endogastric overtube for protection during repeated insertions. By enabling endoluminal tissue approximation, the device provides a minimally invasive alternative to traditional surgical approaches for GI tract procedures.
Clinical Evidence
No clinical data provided. Safety and effectiveness were established through bench testing and animal studies. All patient-contacting components were tested for biocompatibility per ISO 10993 requirements. All testing yielded acceptable results.
Technological Characteristics
Materials: Biocompatible per ISO 10993. Sensing/Actuation: Vacuum-assisted tissue plication; mechanical compression and stapling. Energy: Manual (inflation syringes). Form factor: Flexible, endoluminal, trans-oral. Connectivity: None. Sterilization: Radiation (ISO 11137-1, ISO 11137-2). Components: Reusable handle, disposable head, single-use cartridge, titanium staples, non-absorbable reinforcement ring.
Indications for Use
Indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract for patients requiring such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K082044 — ACE STAPLER, MODEL F0031 · Barosense, Inc. · May 11, 2010
K120147 — ACE STAPLER AND CARTRIDGE · Barosense, Inc. · Feb 17, 2012
K091832 — ENDOGASTRIC SOLUTIONS STOMAPHYX SYSTEM WITH SEROSAFUSE FASTENER · Endogastric Solutions, Inc. · Jul 22, 2009
K024275 — AUTO SUTURE PREMIUM PLUS CEEA DISPOSABLE STAPLER · United States Surgical, A Division of Tyco Healthc · Mar 14, 2003
K192566 — Reach and CQ'ENCE Circular Staplers · Reach Surgical, Inc. · Oct 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
#### III. 510(k) Summary
JUN - 3 2011
# BaroSense ACE™ Stapler and Cartridge
Image /page/0/Picture/4 description: The image shows the logo for Barosense Inc. The logo features a stylized figure with arms raised inside of a circle. The text "BAROSENSE" is to the right of the figure, with "INC." below it.
# General Information
| Criteria | Information |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | ACE™ Stapler<br>(Note: the trademark name is still being finalized<br>and may change from that listed above) |
| Product Name | ACE Stapler and Cartridge |
| Catalog/Model Number | F0084 ACE Stapler Reusable Handle<br>F0085 ACE Stapler Head<br>F0087 ACE Stapler Cartridge |
| Common Name | surgical stapler and cartridge |
| Classification | 21 CFR 876.1500- Endoscope and Accessories;<br>Class II; Product code: OCW |
| 510(k) Owner | BaroSense, Inc.<br>250 Chesapeake Drive<br>Redwood City CA 94063 |
| Contact Person | Sheila Stevens, PhD<br>Director Clinical and Regulatory Affairs BaroSense, Inc.<br>sstevens@barosense.com<br>650-362-6016 (phone) 650-362-0070 (fax) |
## Summary of Substantial Equivalence
The BaroSense, Inc., ACE Stapler and Cartridge (component models F0084, F0085 and F0086) are substantially equivalent to the BaroSense ACE Stapler and Cartridge (component models F0031 and F0007).
Date: March 22, 2011
{1}------------------------------------------------
# Predicate Devices
| Manufacturer | Predicate device | 510(k) |
|-------------------------------------|----------------------------------------------------------------|---------|
| BaroSense, Inc.<br>Redwood City, CA | ACE Stapler, model F0031<br>ACE Stapler Cartridge, model F0007 | K082044 |
## Device Description
The ACE™ Stapler is a surgical stapler used in hospitals or surgery centers for staple closure on the wall of the stomach or gastrointestinal tract.
The single-patient-use, disposable stapler head is supplied non-sterile and is fitted with a sterile, single-use staple cartridge. The stapler head is attached to a reusable, flexible stapler handle that controls the position and articulation of the stapler head. In use, the stapler is introduced into the patient through the mouth. A flexible endoscope passes through the stapler for gastric tissue visualization. The stapler works in conjunction with a vacuum pump to create a plication [tissue fold] in the GI tract, which is then compressed. The stapler then places a double, circular row of titanium staples. A non-absorbable ring helps reinforce the staple placement in the tissue. The tissue compression and stapling functions are controlled by commercially available inflation syringes.
The stapler may be introduced over a guidewire. If multiple plications are required, an endogastric overtube may be used to protect the esophageal tissues during repeated insertions of the device. The guidewire, overtube, flexible endoscope, vacuum pump and inflation syringes used with the stapler are all commercially available medical devices, not the subject of this 510(k), and are not supplied with the stapler.
The predicate device has identical technological characteristics. However, the predicate stapler device is provided as a single-patient-use, non-sterile, disposable unit. The predicate device is introduced through an overtube, as it does not include a membrane designed for guidewire introduction.
### Indications for Use
The BaroSense ACE Stapler is indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.
### Bench/Animal Testing
All patient contacting components of the ACE Stapler are composed of materials of known biocompatibility tested to the requirements of ISO 10993. The safety and effectiveness of the device was further established through a series of bench and animal tests. All testing yielded acceptable results.
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120; Expiration Date: 8/31/10 | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|-------------------------------------------------------|
| Department of Health and Human Services<br>Food and Drug Administration | | |
| STANDARDS DATA REPORT FOR 510(k)s<br>( <i>To be filled in by applicant</i> ) | | |
| This report and the Summary Report Table are to be completed by the applicant when submitting a 510(k) that refer-<br>ences a national or international standard. A separate report is required for each standard referenced in the 510(k). | | |
| <b>TYPE OF 510(K) SUBMISSION</b> | | |
| <label><input type="checkbox"/> Traditional</label> | <label><input checked="" type="checkbox"/> Special</label> | <label><input type="checkbox"/> Abbreviated</label> |
| <b>STANDARD TITLE</b> <sup>1</sup> | | |
| ISO 11137-1:2006 Sterilization of Health Care Products - Radiation- Part 1: | | |
| <b>Please answer the following questions</b> | | |
| Is this standard recognized by FDA <sup>2</sup> ? | <label><input checked="" type="checkbox"/> Yes</label> | <label><input type="checkbox"/> No</label> |
| FDA Recognition number <sup>3</sup> | # 14-297 | |
| Was a third party laboratory responsible for testing conformity of the device to this standard identified in the 510(k)? | <label><input checked="" type="checkbox"/> Yes</label> | <label><input type="checkbox"/> No</label> |
| Is a summary report <sup>4</sup> describing the extent of conformance of the standard used included in the 510(k)?<br>If no, complete a summary report table. | <label><input checked="" type="checkbox"/> Yes</label> | <label><input type="checkbox"/> No</label> |
| Does the test data for this device demonstrate conformity to the requirements of this standard as it pertains to this device? | <label><input checked="" type="checkbox"/> Yes</label> | <label><input type="checkbox"/> No</label> |
| Does this standard include acceptance criteria?<br>If no, include the results of testing in the 510(k). | <label><input checked="" type="checkbox"/> Yes</label> | <label><input type="checkbox"/> No</label> |
| Does this standard include more than one option or selection of tests?<br>If yes, report options selected in the summary report table. | <label><input type="checkbox"/> Yes</label> | <label><input checked="" type="checkbox"/> No</label> |
| Were there any deviations or adaptations made in the use of the standard?<br>If yes, were deviations in accordance with the FDA supplemental information sheet (SIS) <sup>5</sup> ? | <label><input type="checkbox"/> Yes</label> | <label><input checked="" type="checkbox"/> No</label> |
| Were deviations or adaptations made beyond what is specified in the FDA SIS?<br>If yes, report these deviations or adaptations in the summary report table. | <label><input type="checkbox"/> Yes</label> | <label><input checked="" type="checkbox"/> No</label> |
| Were there any exclusions from the standard?<br>If yes, report these exclusions in the summary report table. | <label><input type="checkbox"/> Yes</label> | <label><input checked="" type="checkbox"/> No</label> |
| Is there an FDA guidance <sup>6</sup> that is associated with this standard?<br>If yes, was the guidance document followed in preparation of this 510k? | <label><input type="checkbox"/> Yes</label> | <label><input checked="" type="checkbox"/> No</label> |
| Title of guidance: | | |
| <sup>1</sup> The formatting convention for the title is: [SDO] [numeric identifier]<br>[title of standard] [date of publication]<br><sup>2</sup> Authority [21 U.S.C.,360d], www.fda.gov/cdrh/stdsprog.html<br><sup>3</sup> http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm<br><sup>4</sup> The summary report should include: any adaptations used to adapt to the device under review (for example, alternative test methods); choices made when options or a selection of methods are described; deviations from the standard; requirements not applicable to the device; and the name and address of the test laboratory or certification body involved in conformance assessment to this standard. The summary report Includes information on all standards utilized during the development of the device.<br><sup>5</sup> The supplemental Information sheet (SIS) is additional information which is necessary before FDA recognizes the standard. Found at http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm<br><sup>6</sup> The online search for CDRH Guidance Documents can be found at www.fda.gov/cdrh/guidance.html | | |
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120; Expiration Date: 8/31/10
Department of Health and Human Services
Food and Drug Administration
**STANDARDS DATA REPORT FOR 510(k)s**
(*To be filled in by applicant*)
This report and the Summary Report Table are to be completed by the applicant when submitting a 510(k) that refer- ences a national or international standard. A separate report is required for each standard referenced in the 510(k).
TYPE OF 510(K) SUBMISSION
| | Traditional | <input type="checkbox"/> | Special | <input checked="" type="checkbox"/> | Abbreviated | <input type="checkbox"/> |
|--|-------------|--------------------------|---------|-------------------------------------|-------------|--------------------------|
|--|-------------|--------------------------|---------|-------------------------------------|-------------|--------------------------|
| STANDARD TITLE¹ | ISO 11137-2:2006 Sterilization of Health Care Products - Radiation- Part 2: |
|-----------------|-----------------------------------------------------------------------------|
|-----------------|-----------------------------------------------------------------------------|
Please answer the following questions
| | Yes | No |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------------------|
| Is this standard recognized by FDA²? | <input checked="" type="checkbox"/> | <input type="checkbox"/> |
| FDA Recognition number³ | # 14-225 | |
| Was a third party laboratory responsible for testing conformity of the device to this standard identified in the 510(k)? | <input checked="" type="checkbox"/> | <input type="checkbox"/> |
| Is a summary report⁴ describing the extent of conformance of the standard used included in the 510(k)? If no, complete a summary report table. | <input checked="" type="checkbox"/> | <input type="checkbox"/> |
| Does the test data for this device demonstrate conformity to the requirements of this standard as it pertains to this device? | <input checked="" type="checkbox"/> | <input type="checkbox"/> |
| Does this standard include acceptance criteria? If no, include the results of testing in the 510(k). | <input checked="" type="checkbox"/> | <input type="checkbox"/> |
| Does this standard include more than one option or selection of tests? If yes, report options selected in the summary report table. | <input type="checkbox"/> | <input checked="" type="checkbox"/> |
| Were there any deviations or adaptations made In the use of the standard? If yes, were deviations in accordance with the FDA supplemental information sheet (SIS)⁵? | <input type="checkbox"/> | <input checked="" type="checkbox"/> |
| Were deviations or adaptations made beyond what is specified in the FDA SIS? If yes, report these deviations or adaptations in the summary report table. | <input type="checkbox"/> | <input checked="" type="checkbox"/> |
| Were there any exclusions from the standard? If yes, report these exclusions in the summary report table. | <input type="checkbox"/> | <input checked="" type="checkbox"/> |
| Is there an FDA guidance⁶ that is associated with this standard? If yes, was the guidance document followed in preparation of this 510k? Title of guidance: | <input type="checkbox"/> | <input checked="" type="checkbox"/> |
¹ The formatting convention for the title is: [SDO] [numeric Identifier] [title of standard] [date of publication]
² Authority [21 U.S.C. 360d], www.fda.gov/cdrh/stdsprog.html
³ http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/ search.cfm
⁴ The summary report should include: any adaptations used to adapt to the device under review (for example, alternative test methods); choices made when options or a selection of methods are described; deviations from the standard; requirements not applicable to the device; and the name and address of the test laboratory or certification body involved in conformance assessment to this standard. The summary report includes information on all standards utilized during the development of the device.
⁵ The supplemental information sheet (SIS) Is additional information which la necessary before FDA recognizes the standard. Found at http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/ search cfm
⁶ The online search for CDRH Guidance Documents can be found at www.fda.gov/cdrh/guidance.html
FORM FDA 3654 (9/07)
Page 1
PSC Graphics (301) 443-1020
:
・
{4}------------------------------------------------
| Form Approved: OMB No. 0910-0120; Expiration Date: 8/31/10 | |
|------------------------------------------------------------------------------|--|
| Department of Health and Human Services<br>Food and Drug Administration | |
| STANDARDS DATA REPORT FOR 510(k)s<br>( <i>To be filled in by applicant</i> ) | |
This report and the Summary Report Table are to be completed by the applicant when submitting a 510(k) that references a national or international standard. A separate report is required for each standard referenced in the 510(k).
TYPE OF 510(K) SUBMISSION
| <span style="display:inline-block; width: 100px;"></span> Traditional | <span style="display:inline-block; width: 100px;"></span> Special | <span style="display:inline-block; width: 100px;"></span> Abbreviated |
|-----------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------|
|-----------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------|
| STANDARD TITLE <sup>1</sup> |
|-----------------------------------------------------------------------------------------------------------|
| 10993-1:2009 Biological evaluation of medical devices -- Part 1:Evaluation and testing (Biocompatibility) |
**Please answer the following questions**
| Is this standard recognized by FDA <sup>2</sup> ? | Yes <img alt="Checked" src="checkbox_checked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|--------------------------------------------------------|
| FDA Recognition number <sup>3</sup> | #2-156 | |
| Was a third party laboratory responsible for testing conformity of the device to this standard identified in the 510(k)? | Yes <img alt="Unchecked" src="checkbox_unchecked.png"/> | No <img alt="Checked" src="checkbox_checked.png"/> |
| Is a summary report <sup>4</sup> describing the extent of conformance of the standard used included in the 510(k)?<br>If no, complete a summary report table. | Yes <img alt="Checked" src="checkbox_checked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Does the test data for this device demonstrate conformity to the requirements of this standard as it pertains to this device? | Yes <img alt="Checked" src="checkbox_checked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Does this standard include acceptance criteria?<br>If no, include the results of testing in the 510(k). | Yes <img alt="Checked" src="checkbox_checked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Does this standard include more than one option or selection of tests?<br>If yes, report options selected in the summary report table. | Yes <img alt="Checked" src="checkbox_checked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Were there any deviations or adaptations made in the use of the standard?<br>If yes, were deviations in accordance with the FDA supplemental information sheet (SIS) <sup>5</sup> ? | Yes <img alt="Unchecked" src="checkbox_unchecked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Were deviations or adaptations made beyond what is specified in the FDA SIS?<br>If yes, report these deviations or adaptations in the summary report table. | Yes <img alt="Unchecked" src="checkbox_unchecked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Were there any exclusions from the standard?<br>If yes, report these exclusions in the summary report table. | Yes <img alt="Unchecked" src="checkbox_unchecked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
| Is there an FDA guidance <sup>6</sup> that is associated with this standard?<br>If yes, was the guidance document followed in preparation of this 510k?<br>Title of guidance: | Yes <img alt="Unchecked" src="checkbox_unchecked.png"/> | No <img alt="Unchecked" src="checkbox_unchecked.png"/> |
<sup>1</sup> The formatting convention for the title is: [SDO] [numeric Identifier] certification body involved in conformance assessment to this
(title of standard) [date of publication] standard. The summary report includes information on all standards
<sup>2</sup> Authority (21 U.S.C. 360d), www.fda.gov/cdrh/stdsprog.html utilized during the development of the device.
<sup>3</sup> http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/ <sup>5</sup> The supplemental information sheet (SIS) is additional information
search.cfm which is necessary before FDA recognizes the standard. Found at
<sup>4</sup> The summary report should Include: any adaptations used to adapt http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/
to the device under review (for example, alternative test methods); search.cfm
choices made when options or a selection of methods are described; <sup>6</sup> The online search for CDRH Guidance Documents can be found at
deviations from the standard: requirements not applicable to the www.fda.gov/cdrh/guidance.html
device; and the name and address of the test laboratory or
| FORM FDA 3654 (9/07)<br>Page 1 | PSC Graphics (301) 443-6500 |
|--------------------------------|-----------------------------|
|--------------------------------|-----------------------------|
·
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring. MD 20993-0002
Sheila S. Stevens, Ph.D. Director, Clinical and Regulatory Affairs BaroSense, Inc. 250 Chesapeake Drive REDWOOD CITY CA 94063
JUN - 3 2011
Re: K110829
Trade/Device Name: ACE Stapler and Cartridge, Models F0084, -85 and -86 Regulation Number: 21 CFR 8876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCW Dated: April 28, 2011 Received: May 6, 2011
Dear Dr. Stevens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{6}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K110829 page loft
#### Statement of Indications for Use II.
510(k) Number (if known): K // 0829
Device Name: ACE™ Stapler and Cartridge
Indications for Use: The BaroSense ACE Stapler and Cartridge are indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helid Reum
Page 1 of 1
uctive, Gastro-Renal, and
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.