The Wireless Monitoring System is intended for use by health care professionals for routine surveillance of patient physiological parameters to include, pulse rate, respiratory rate and axillary temperature, in a hospital setting. Data is transmitted wirelessly to a central location. Notifications can be prospectively set to notify healthcare professionals to excursions outside of selected parameters. The device is not intended to be used on critical care patients and is intended to supplement vital signs monitoring by healthcare professionals, not to replace current standards of care. The device is intended for use on general care patients and on patients who are 18 years of age or older.
Device Story
Wireless Monitoring System consists of a disposable patient patch with integrated temperature sensor, electronic bridge, and server software. Patch attaches to patient, connects to existing leads to capture heart rate, respiration, and axillary temperature. Data transmits wirelessly to bridge; multiple bridges installed in hospital environment to support patient mobility. Bridge transmits data via hard-wire to nurses' station for surveillance. Healthcare practitioners set parameter limits; system triggers alerts for excursions. Used in hospital settings by healthcare professionals to supplement standard vital signs monitoring. Benefits include continuous surveillance of general care patients, enabling timely response to physiological changes.
Clinical Evidence
No clinical trials performed. Evidence consists of usability testing at a user site (University of Texas at Arlington) and wireless performance testing at a hospital (Palomar Pomerado Hospital). Usability study confirmed proof of concept, operating principles, and positive staff feedback. Wireless testing confirmed system operated as designed without interference from or to hospital equipment. Bench testing included biocompatibility (ISO 10993), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2).
Technological Characteristics
Disposable patient patch with integrated temperature sensor; electronic bridge; server software. Sensing: heart rate, respiration, axillary temperature. Connectivity: wireless transmission (ISM bands, 911.38-918.59 MHz) to bridge, hard-wired to central server. Power: 3.0V battery (patch), AC main (bridge). Standards: ASTM 2475-05 (packaging), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60529 IP 64 (protection), ISO 10993 series (biocompatibility).
Indications for Use
Indicated for routine surveillance of pulse rate, respiratory rate, and axillary temperature in non-critical, general care hospital patients aged 18 years or older. Not for use on critical care patients.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
K110809
# 510(k) SUMMARY
A Summary of 510(k) Safety and Effectiveness Information in Accordance with the
Requirements of 21 CFR 807.92.
| Submitter Information | |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Frank Pokrop |
| Address | CareFusion Inc., 3750 Torrey View Court. San Diego, Ca 92130 |
| Phone number | Work - (858) 617-4364.<br>Cell: (858) 245-8197 |
| Fax number | (858) 617-6735 |
| Establishment<br>Registration<br>Number | 2016493 |
| Name of contact<br>person | Frank Pokrop |
| Date prepared | March 15, 2011 |
| Name of device | |
| Trade or proprietary<br>name | Trade name is still being determined |
| Common or usual<br>name | Wireless Monitoring System |
| Classification name | Cardiovascular Monitoring Devices |
| Classification panel | Cardiovascular |
| Regulation | 21 CFR Part 870.2910 |
| Product Code(s) | DRG |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | (1) Philips Intellivue Patient Monitor. 510(k) #: K052961<br>(2) Nihon Koden Multiple Patient Receiver. 510(k) #: K071058, |
| Reason for 510(k)<br>submission | New device. |
| Device description | The proposed device consists of a patient patch with integrated<br>temperature sensor, an electronic bridge and server software. The<br>patch is attached to the patient and connected to existing monitoring<br>leads to capture heart rate, respiration and body temperature. This<br>data is wirelessly transmitted to a bridge. Multiple bridges can be<br>installed in a hospital setting to capture signals in case the patient is<br>moved or becomes ambulatory.<br>Data from the patch is transmitted through a hard-wire connection to<br>the nurses' station for surveillance. The healthcare practitioner can<br>set limits on the patient data which in turn may trigger an alert. |
| Intended use of the device | General surveillance of non-critical acre patients who are at least 18 years of age. |
| Indications for use | The Wireless Monitoring System is intended for use by health care professionals for routine surveillance of patient physiological parameters to include, pulse rate, respiratory rate and axillary temperature, in a hospital setting. Data is transmitted wirelessly to a central location. Notifications can be prospectively set to notify healthcare professionals to excursions outside of selected parameters.<br>The device is not intended to be used on critical care patients and is intended to supplement vital signs monitoring by healthcare professionals, not to replace current standards of care. The device is intended for use on general care patients and on patients who are 18 years of age or older. |
.
{1}------------------------------------------------
#### Summary of the technological characteristics of the device compared to the predicate device
| Characteristic | New Device | Predicate [Device Name]<br>[510(k) number] |
|------------------------------------------------------------|------------|-----------------------------------------------|
| Central monitoring of remotely<br>transmitted patient data | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Uses existing sources of<br>physiological data | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Programmable alert settings<br>and limits | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Data sent to central server | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Monitors heart rate | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Monitors respiration | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Monitors body or axillary<br>temperature | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Can be used with a general<br>hospital patient population | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
| Monitors multiple patients | Yes | Philips Intellivue Patient Monitor. K052961 |
| | | Nihon Koden Multiple Patient Monitor. K071058 |
'
{2}------------------------------------------------
| Uses both AC Main and battery power | Yes | Philips Intellivue Patient Monitor. K052961 Nihon Koden Multiple Patient Monitor. K071058 | |
|-------------------------------------|-----|-------------------------------------------------------------------------------------------|--|
|-------------------------------------|-----|-------------------------------------------------------------------------------------------|--|
PERFORMANCE DATA
SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF
SUBSTANTIAL EQUIVALENCE*
Performance Test Summary-New Device
| Characteristic | Standard/Test/FDA Guidance | Results Summary |
|---------------------------------------------|----------------------------|----------------------------------------------------------------|
| 1. Packaging - biocompatibility | ASTM 2475-05 | The proposed device passes the applicable tests and standards. |
| 2. Basic safety | IEC 60601-1 | |
| 3. EMC Compatiblity | IEC60601-1-2 | |
| 4. Collateral Safety | IEC 60601-1-1 | |
| 5. Electrocardiographic equipment | IEC 60601-2-27 | |
| 6. Patient Monitoring | IEC 60601-2-49 | |
| 7. Degrees of Protection | IEC 60529 IP 64 | |
| 8. Human exposure to electromagnetic fields | IEC 62311 | |
| 9. IT Equipment | IEC 60950-1 | |
| 10. Animal welfare | ISO 10993-2 | |
| 11. Interactions with Blood | ISO 10993-4 | |
| 12. In Vitro Cytotoxicity | ISO 10993-5 | |
| 13. Irritation and Skin Sensitization | ISO 10993-10 | |
| 14. Systemic Toxicity | ISO 10993-11 | |
| 15. Sample Preparation | ISO 10993-12 | |
| 16. Leachable Substances | ISO 10993-17 | |
| 17. Chemical Characterization of Materials | ISO 10993-18 | |
| 18. Risk Management | ISO 14971 | |
## Comparative Performance Information Summary
| Characteristic | Requirement | New Device | Predicate Device |
|------------------|----------------------------------|-----------------|------------------|
| FCC Approval | Letter of approval | Yes | Yes |
| Frequency Range | 911.38-918.59 (Mhz). FCC Part 15 | Same or similar | Same or similar |
| Type of Device | Near-patient data collection | Disposable | Non-disposable |
| Heart Rate | 30 to 200 bpm | Same or similar | Same or similar |
| Axillary temp | 89.6° F – 111.2°F | Same or similar | Same or similar |
| Respiratory rate | 5 - 60 resp/min | Same or similar | Same or similar |
| HL7 Protocol | Communicate with HIS using HL7 | Yes | Yes |
{3}------------------------------------------------
| Interval reporting | Data reporting can be programmed | Range: 2-30 minute<br>increments in 2 minute<br>increments | Same or similar |
|--------------------|--------------------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Battery power | Low power battery for near patient device | 3.0 volts | Same or similar |
| Power for bridges | Bridges use AC Main for power | yes | Same or similar |
| Notifications | System has programmable notifications | Yes | Yes |
| Telemetry | Transmissions between bridge and patch shall<br>use ISM bands | Yes | Same or similar |
| Vital signs data | Provides: (1) heart rate, (2) Respiration rate,<br>(3) Axillary temperature | Yes | - Same or similar<br><br>Predicate devices<br>have these<br>capabilities and<br>more |
| Operating Paradigm | Uses: (1) Patient data collection device, (2)<br>bridges, (3) Hospital servers | Yes | Same or similar |
#### SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information
- (1) Clinical testing was not performed with this device.
- (2) Usability testing at a user site.
The University of Texas at Arlington showed that the proof of concept and the proof of operating principles were both found to be valid, the system operated and functioned as planned, staff opinions were positive with participants showing an interest in using the proposed device.
- a. "The study execution went smoothly and it afforded the opportunity to conduct a realistic and rather rigorous test of system usability.
- b. Based on the collective data analysis the system is considered to have good usability."
- (3) Performance testing a user site.
Wireless testing at Palomar Pomerado Hospital showed that the device's wireless features operated as designed and as intended. Testing for EMI and other parameters were monitored and recorded. Attached reports show safe and effective levels of operation without harming the patient along with acceptable levels of performance. From the report:
{4}------------------------------------------------
- a. "The System Patch and Bridge did not interfere with any wireless equipment provided by the hospital for the testing.
- b. Further no in hospital equipment noise or interference had any impact on the operation of the System." .
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The device was found to be safe and effective and it operated as intended in user environments.
*Although a table format is shown and recommended, use whatever format is appropriate for your medical device or in vitro diagnostic product.
** Insert as many lines as necessary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three curved lines that resemble an abstract representation of a human figure or a symbol related to health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL - 5 2011
CareFusion, Inc. c/o Mr. Frank Pokrop Director, Regulatory Affairs 3750 Torrey View Court San Diego, CA 92130
Re: K110809
> Trade/Device Name: Wireless Monitoring System Regulatory Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: II (two) Product Code: DRG Dated: June 3, 2011 Received: June 6, 2011
Dear Mr. Pokrop:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
Page 2 - Mr. Frank Pokrop
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MédicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Wireless Monitoring System
The Wireless Monitoring System is intended for use by health care professionals for routine surveillance of patient physiological parameters to include, pulse rate, respiratory rate and axillary temperature, in a hospital setting. Data is transmitted wirelessly to a central location. Notifications can be prospectively set to notify healthcare professionals to excursions outside of selected parameters.
The device is not intended to be used on critical care patients and is intended to supplement vital signs monitoring by healthcare professionals, not to replace current standards of care. The device is intended for use on general care patients and on patients who are 18 years of age or older.
Prescription Use _ XXX_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter use (21 CFR Part 801 Subpart C)
(PLEASE DO NO WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEFDED)
| Division Sig |
|--------------------------------------------------------|
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
| Division of Cardiovascular Devices |
| K110809 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.