K110754 · Cryopen, Inc. · GEH · Sep 9, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110754
Device Name
CRYOPEN CRYOSURGICAL SYSTEM, K102214
Applicant
Cryopen, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Sep 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cryosurgical unit used for ablative type surgical technique on multiple organ systems, wide range of disease, viral, premalignant and malignant tissue. Cryosurgical instrument used to necrose unwanted tissue, generally in the gynecological and dermatological practice
Device Story
CryoPen (CT-2000) is a cryosurgical system for tissue ablation; replaces traditional cryogenic gases (nitrous oxide/CO2) with a Stirling cooler-based system. System components: base unit, Stirling cooler, pen cores, and reusable aluminum tips (various sizes for dermatology/gynecology). Operation: Stirling cooler lowers pen core temperature to -105°C; heat rejected via cooler; temperature indicator signals readiness. Physician-operated in clinical settings. Device contacts tissue via chilled aluminum tip to induce necrosis. Benefits: avoids greenhouse gases/cryogenic liquids; provides controlled freezing for dermatologic/gynecologic procedures.
Clinical Evidence
Bench testing only. Comparison performed on ballistic gelatin to measure freezing volume and mass against the Wallach LL100 predicate. No clinical data provided.
Technological Characteristics
Stirling cooler system; 24V DC power supply (115/220V AC input). Aluminum cryotips. Operating temperature: -105°C. Sterilization: cold soak and autoclave. Connectivity: standalone. Form factor: pen-based system with interchangeable tips (3mm-25mm).
Indications for Use
Indicated for ablative surgical techniques on multiple organ systems to necrose unwanted tissue, including viral, premalignant, and malignant lesions in gynecological and dermatological practice.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
# <116756 510(k) Summary of Safety and Effectiveness
CryoPen, Inc. CryoPen Cryosurgical System Traditional 510(k)
A. General Information
| Applicant: | CryoPen, Inc. |
|----------------------|-------------------------------------------------------------|
| Address: | 2010 E. Continental Blvd.<br>Suite D<br>Southlake, TX 76092 |
| Contact Person: | Dr. Michael Haas |
| Telephone: | 888-246-3928 |
| Facsimile: | 817-549-0012 |
| Date Prepared: | April 25, 2011 |
| Device Trade Name: | CryoPen Cryosurgical System |
| Common Name: | Cryosurgical Unit and Accessories |
| Classification Name: | Unit, Cryosurgical, Accessories |
### B. Description of the Device
The CryoPen system provides a safe means of freezing tissue without the use of cryogenic liquids or greenhouse gases. The system consists of multiple pen cores, a Stirling cooler which rejects heat, and a variety of reusable tips for dermatologic and gynecology applications. When used properly, the system will deliver effective temperature for tissue ablation. The model number of this Cryosurgical system is CT-2000.
The Stirling cooler is used to lower the temperature of the CryoPen units to temperatures of -105 degrees Celsius. It operates on a 24 V DC external power supply powered by 115 (or 220) V AC obtained from conventional convenience outlets. The housing of the cooler provides necessary electrical safety features and electromagnetic compatibility. A temperature indicator indicates the readiness of the CryoPen units during cool down,
{1}------------------------------------------------
There are currently four different size tips for use in general surgery and dermatological applications as well as four different size tips for use in gynecology applications. They are:
| Application | Tip Sizes |
|---------------------|-----------------------|
| General/Dermatology | 3mm, 5mm, 7mm, 10mm |
| Gynecology | 3mm, 10mm, 19mm, 25mm |
### C. Intended Use Statement
Cryosurgical unit used for ablative type surgical technique on multiple organ systems, wide range of disease, viral, premalignant and malignant tissue.
Cryosurgical instrument used to necrose unwanted tissue, generally in the gynecological and dermatological practice
## D. Components
| CONTENTS: | ITEM NUMBER: |
|------------------------------------------------|--------------|
| Base Unit | |
| CryoPen® Cooling System Assembly (1ea.) | CT-2000 |
| Reservoir Solution 500ml (1ea.) | CT2-RS-1001 |
| Reservoir Tube- Plastic (23ml) | CT2-RS-1002 |
| Reservoir Tube Cap (1ea.) | CT2-RS-1003 |
| Cleaning Swabs -Single end 14"(6) | CT2-SW-1000 |
| Transfer Pipette- ½ dozen | CT2-PP-1000 |
| CryoPen® Operators Packet | CT2-OP1000 |
| | |
| General/Dermatological Configuration | |
| Non-Sterile re-usable CryoPen® tip (1ea.) 3mm | CT2-T-5003 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 5mm | CT2-T-5005 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 7mm | CT2-T-5007 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 10mm | CT2-T-5010 |
| Pen Core - Blue (4) | CT2-C-1000 |
| Temperature Indicator (Grey) (1ea.) | CT2-T1-1000 |
| | |
| Gynecological Configuration | |
| Non-Sterile GYN Handle (1 ea.) | GY2-1100 |
| GYN Pen Cores - Green (2 ea.) | GY2-2201 |
| Non-Sterile Re-usable CryoPen® tip (1ea.) 3mm | GY2-2303 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 10mm | GY2-2310 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 19mm | GY2-2319 |
| Non-Sterile re-usable CryoPen® tip (1ea.) 25mm | GY2-2324 |
| Blank Core (1ea.) | GY2-3001 |
| Temperature Indicator (Green) (1ea.) | GY2-4100 |
{2}------------------------------------------------
### E. Substantial Equivalence
The CryoPen cryosurgical system is substantially equivalent to the Wallach LL100 Predicate Device, approved on February 4, 1981, (reference K803311) based on the nonclinical bench testing performed on ballistic gelatin. Data generated from bench testing shows the CryoPen is able to freeze the same volume and mass as the Wallach LL100.
The CryoPen and the Wallach LL100 destroy tissue using cryogenic temperatures. However, the CryoPen does not rely on a cryogenic gas such as nitrous oxide or carbon dioxide for cryoablation. Instead, the CryoPen uses a chilled copper mass fashioned into a treatment medium. The contacting metal incorporated in the cryotips used with the CryoPen are made of aluminum, which differs from the predicate contact material of stainless steel. Appropriate biocompatibility and sterility testing was performed on the aluminum cryotips used in the CryoPen and accessories.
The modified CryoPen and accessories have the following similarities to those which previously received 510(k) concurrence:
- · have the same indicated use,
- · destroy unwanted tissue by cryoablation, and
- · method of sterilization
The Wallach LL100 510(k) K803311 has indications of "a cryosurgical instrument used to necrose unwanted tissue, generally in the gynecological and dermatological practice". The CryoPen and Accessories has clearance as a Cryosurgical unit used "for ablative type surgical technique" with use on "multiple organ systems, wide range of disease, viral, premalignant and malignant tissue." Additionally, the CryoPen system is adding the same intended use as the LL100 for use in the gynecological and dermatological practices.
Additionally, the CryoPen and Accessories uses the same sterilization methods as the Wallach LL100, cold soak and autoclave.
In summary, the CryoPen and Accessories described in this submission are substantially equivalent to the predicate device as the safety, efficacy, and intended use are equivalent,
73
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The seal is in black and white and appears to be a scanned image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cyrol'en, Inc. % Michael Haas, M.D. 2010 East Continental Boulevard Suite D Southlake, Texas 76092
- 9 2011
Re: K110754
Trade/Device Name: CyroPen and Accessories (Model: CT-2000) Regulation Number: 21 CFR 878.4350 Regulation Name: Cyrosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: August 31. 2011 Received: September 01, 2011
Dear Dr. Haas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Michael Haas, M.D.
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practive requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For Roblino mpit
DEP CLint Dir
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number
K110754
Device Name
CryoPen and Accessories (Model: CT-2000)
Indications for Use
Cryosurgical unit used for ablative type surgical technique on multiple organ systems, wide range of disease, viral, premalignant and malignant tissue.
Cryosurgical instrument used to necrose unwanted tissue in the gynecological and dermatological practice
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Kiri RP Duke Sherman
(Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110754
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.