K110753 · Dentsply International, Inc. · EKX · Jun 6, 2011 · Dental
Device Facts
Record ID
K110753
Device Name
MIDWEST RDH FREEDOM CORDLESS PROPHY SYSTEM
Applicant
Dentsply International, Inc.
Product Code
EKX · Dental
Decision Date
Jun 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The MIDWEST® RDH Freedom™ Cordless Prophy System is a high-performance cordless prophylaxis handpiece with a wireless foot pedal for use with Freedom disposable prophylaxis angles in a hygiene operatory to perform cleaning and polishing procedures on teeth.
Device Story
Cordless, electric motor-driven handpiece; used with disposable prophylaxis angles (soft cup, hard cup, or brush). Operates via wireless foot pedal; vertical pedal displacement controls handpiece speed. Internal motion sensor activates device when picked up. No direct operator controls on handpiece. Used in dental hygiene operatory by dental professionals. Autoclavable aluminum outer sheath for infection control. Battery-powered; rechargeable via AC base. Benefits include cordless mobility, reduced noise compared to traditional air-powered handpieces, and improved ergonomics for cleaning/polishing procedures.
Clinical Evidence
No clinical data. Safety and efficacy verified through non-clinical bench testing, including design verification, speed/torque control, battery life, noise testing, software validation, sterilizability, biocompatibility, and electrical safety.
Technological Characteristics
Cordless electric motor handpiece; aluminum autoclavable outer sheath; plastic/rubber disposable prophy angles. Powered by rechargeable Li-ion battery in foot pedal. Wireless communication between pedal and handpiece. Internal motion sensor for activation. Speed/torque ranges comparable to traditional air-driven handpieces.
Indications for Use
Indicated for use by dental professionals on patients of all ages in a professional dental operatory to perform tooth cleaning and polishing procedures using prophy pastes or polishes.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
# K110753
JUN - 6 2011
### DENTSPLY International
World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872 18001 877-0020 Fax (717) 849-4343 www.dentsply.com
### SECTION 5
### 510(k) SUMMARY for MIDWEST® RDH Freedom™ Cordless Prophy System
#### 1.0 Submitter Information
DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number: | 717-849-4343 |
Date Prepared: 17 March 2011
#### 2.0 Device Name
Proprietary Name: Common Name: Classification Name: CFR Number: Device Class: Product Code:
MIDWEST® RDH Freedom™ Cordless Prophy System Prophy System Handpiece, Direct Drive, AC Powered 872.4200 I EKX
#### 3.0 Predicate Device
Zen Cordless Prophy System by Discus Dental, product code EKX, K101612. Indications for Use - The Zen Cordless Prophy System is a high-performance cordless prophylaxis handpiece with a wireless foot pedal for use with Zen's disposable prophylaxis angles to clean and polish teeth.
{1}------------------------------------------------
#### 4.0 Description of Device
The MIDWEST® RDH Freedom™ Cordless Prophy System is comprised of a cordless, electric motor-driven handpiece, an AC powered battery charging base and a wireless foot pedal. The handpiece utilizes Freedom disposable prophylaxis angles (single use) that are available with soft cup, hard cup or brush features. The handpiece has been designed for comfort and control Sont vap, had bap or oriece has no direct operator controls, and has an internal motion sensing by the ason." The mailpropare for operation when the handpiece is picked up. The handpiece can only be operated by the cordless foot pedal, and the amount of vertical displacement on the foot pedal corresponds to the speed of the handpiece. The handpiece is quieter than traditional hygiene and low speed handpieces and features an autoclavable outer sheath for infection hygiene and low speed inancipies and m Cordless Prophy System handpiece battery has control. The MID w BOY's worth of procedures on a single charge, and the system is sumerent power to completing one cleaning procedure after only fifteen minutes of charging. capable or completing one typlem sidning charger base and the foot pedal are all housed in various plastics. The autoclavable outer sheath is made from aluminum, and the single use disposable prophy angles are made from plastic (and rubber cups).
#### Indications for Use 5.0
The MIDWEST® RDH Freedom™ Cordless Prophy System is a high-performance cordless The With a wireless foot pedal for use with Freedom disposable prophylaxis angles in a hygiene operatory to perform cleaning and polishing procedures on teeth.
The device is a cordless version of a traditional corded, air powered, low speed handpiece that is r he device is a cordical velsion cleaning and polishing teeth. This system is used in conjunction with prophy pastes or polishes when the procedure is being performed. This is a general hygiene with propis painter of performed on people of all ages in a professional dental operatory.
The indications for use of the MIDWEST® RDH Freedom™ Cordless Prophy System and the The murchions Prophy System are essentially the same. They are both cordless, low speed handpieces that are used by a dental professional for cleaning and polishing teeth.
### Identification of Risk Analysis Method 6.0
Risk analysis was performed on the MIDWEST® RDH Freedom™ Cordless Prophy System Risk analysis was performed on the MD . Le 207. The results of the risk analysis performed unitizang an FMEA process based on ISO 11971 12001 System concluded that all device design on the MIDWEST® RDTI Recom - Ocreases Proper potential failures and effects. In controls and piocess confors will be able to mangers and electrical safety and electromagnetic addition; performain were performed to mitigate other potential risks.
PERSONAL
{2}------------------------------------------------
#### Description of Safety and Substantial Equivalence 7.0
#### Technological Characteristics . 7.1
The technological characteristics of the MIDWEST® RDH Freedom™ Cordless Prophy System are very similar to the Zen Cordless Prophy System in that they are both cordless devices used for prophylaxis cleaning and powered by a cordless foot pedal. The foot pedal in the MIDWEST® RDH Freedom™ Cordless Prophy System has a Lithium-Ion battery and is able to be recharged multiple times by inclusion of an AC/DC power supply. The foot pedal in the Zen Cordless Prophy System does not have an power supply and takes 1.5V AAA batteries which does not allow for the foot pedal battery to be recharged internally. The MIDWEST® RDH Freedom™ Cordless Prophy System and the Zen Cordless Prophy System both have proprietary Disposable Prophy Angles that work exclusively with their own systems. Both systems include removable, autoclavable outer sheaths for infection control. The Zen Prophy System has a handpiece with two options for operation; one is using it with the wireless foot pedal and the other uses a power button located on the handpiece for a constant speed mode. The MIDWEST® RDH Freedom™ Cordless Prophy System has a handpiece that does not have any user interface, and the handpiece is controlled exclusively by the cordless foot pedal. The MIDWEST® RDH Freedom™ Cordless Prophy System and the Zen Prophy System have similar speeds and similar torques, which are well within the range of traditional air-driven hygiene handpieces.
#### Non-Clinical Performance Data 7.2
Performance testing focused on verification of design, speed, torque, function and safety of the MIDWEST® RDH Freedom™ Cordless Prophy System. The testing included verification of component specifications, speed and torque control, battery life, noise testing, software validation, sterilizability, biocompatibility and electrical equipment safety. Tests were also performed to compare the speed and torque ranges of the MIDWEST® RDH Freedom™ Cordless Prophy System to the Zen Cordless Prophy System to verify substantial equivalence. The results of these performance tests Bystem to that the MIDWEST® RDH Freedom™ Cordless Prophy System is safe and effective for its intended use as a high-performance cordless prophylaxis handpiece with a wireless foot pedal for use with Freedom disposable prophylaxis angles in a hygiene operatory to perform cleaning and polishing procedures on teeth. This testing also operatory to perioding and Freedom™ Cordless Prophy System is substantially verifications the MD MD Cordless Prophy System in design, component specifications, equivalent to the 201 Control, battery life, sterilizability, biocompatibility and electrical equipment safety.
DENTSPLY International
{3}------------------------------------------------
#### Clinical Performance Data 7.3
K110753
Due to the low risk nature of the MIDWEST® RDH Freedom™ Cordless Prophy System, clinical performance data was not applicable for verification of safety and efficacy.
#### Conclusion as to Substantial Equivalence 7.4
The similarities in design, function, safety and intended use of the MIDWEST® RDH Freedom™ Cordless Prophy System with the legally marketed device, the Zen Cordless Prophy System, provide evidence that these devices are substantially equivalent. The I ropiny System, provide our support the safety and effectiveness of the MIDWEST® RDH Freedom™ Cordless Prophy System for its indicated use.
MIDWEST® RDH Freedom™ Cordless Prophy System
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Helen Lewis Director Corporate Compliance and Regulatory Affairs Dentsply International, Incorporated 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17404
JUN - 6 2011
Re: K110753
Trade/Device Name: MIDWEST RDH Freedom Cordless System Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: May 25, 2011 Received: May 26, 2011
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{6}------------------------------------------------
### Section 4. INDICATIONS FOR USE STATEMENT
KU0753 510(k) Number (if known):
Device Name: MIDWEST® RDH Freedom™ Cordless Prophy System
Indications for Use:
1.
The MIDWEST® RDH Freedom™ Cordless Prophy System is a high-performance cordless prophylaxis handpiece with a wireless foot pedal for use with Freedom disposable prophylaxis angles in a hygiene operatory to perform cleaning and polishing procedures on teeth.
Prescription Use ___X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suor Ruann
(División Sign-Off) (Division Sign-Chi)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110753
## 0000003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.