K110709 · Arkray Factory USA, Inc. · NBW · May 19, 2011 · Clinical Chemistry
Device Facts
Record ID
K110709
Device Name
ARK CARE DIABETES MANAGMENT SYSTEM
Applicant
Arkray Factory USA, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
May 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The ARKRAY Diabetes Management Software is an optional accessory for use with compatible blood glucose meters, such as ARKRAY Glucocard Vital Blood Glucose Meter with data management capabilities. The ARKRAY Diabetes Management Software transfers data from the meter's memory into a secured sever for enhanced data management. ARKRAY ARK Care™ Diabetes Management System is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support improved diabetes management.
Device Story
ARK Care is a web-based diabetes management system; acts as interface between personal blood glucose monitors and secure health management database. Input: blood glucose data uploaded from compatible meters (ARKRAY, Roche, Bayer, LifeScan, Abbott) via USB cable; user-entered health data (weight, blood pressure, lab values, exercise). Processing: data storage, organization, and analysis. Output: logbooks, charts, graphs, and reports accessible via secure web portal. Used in home and clinical settings; operated by patients, family members, and healthcare professionals. Function: facilitates remote monitoring, data sharing, and communication between patients and care teams via password-protected portal. Benefit: improved diabetes treatment regulation and health outcomes through centralized data tracking and trending.
Clinical Evidence
Usability study conducted with 38 diabetic adult consumers and 6 healthcare professionals. Participants performed standardized tasks (enrollment, data upload, logbook entry, chart viewing). Results showed >99% pass rate for ease-of-use and performance questionnaires. 68% of personal users and 100% of professional users indicated intent to use the system for diabetes management. Pass/fail criteria exceeded 70-80% thresholds. Bench testing confirmed accurate data transmission from compatible meters, including memory capacity and rollover synchronization.
Technological Characteristics
Web-based diabetes management software; utilizes USB data transfer cable for meter connectivity. Compliant with HIPAA and HITECH standards for data security. Operates via internet-connected browser. No specific materials or ASTM standards cited for the hardware interface.
Indications for Use
Indicated for people with diabetes and their healthcare professionals in home and clinical settings to manage blood glucose data and other biological statistics.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
MyCare Team, MCT Diabetes (k073699)
Submission Summary (Full Text)
{1}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k110709
B. Purpose for Submission:
New 510(k) for diabetes data management software accessory for use with compatible glucose meters such as ARKRAY Glucocard Vital Blood Glucose Meter (k091102).
C. Measurand:
Whole blood glucose
D. Type of Test:
Diabetes data management system
E. Applicant:
ARKRAY Factory USA, Inc.
F. Proprietary and Established Names:
ARK Care™ Diabetes Management System
G. Regulatory Information:
1. Regulation section:
21CFR Sec.-862.1345-Glucose test system.
21CFR Sec.-862.2100 - Calculator/data processing module for clinical use.
2. Classification:
Class II and I, respectively
3. Product code:
NBW - System, Test, Blood Glucose, Over the Counter
CGA - Glucose Oxidase, Glucose
JQP-Calculator/Data Processing Module, for Clinical Use
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below
2. Indication(s) for use:
The ARKRAY Diabetes Management Software is an optional accessory for use with compatible blood glucose meters, such as ARKRAY Glucocard Vital Blood Glucose Meter with data management capabilities. The ARKRAY Diabetes Management Software transfers data from the meter's memory into a secured sever for enhanced data management. ARKRAY ARK Care™ Diabetes
{2}
Management System is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support improved diabetes management.
3. Special conditions for use statement(s): Over the Counter (OTC) use
4. Special instrument requirements: ARKRAY Glucocard Vital Blood Glucose Meter
I. Device Description:
The ARK Care Diabetes Management system serves as an interface between the software in compatible glucose meters and a general purpose health management database to assist in the review, analysis and evaluation of blood glucose test results. ARK Care is a secure system designed for home use and professional healthcare settings.
The purpose of the electronic diabetes management system is to help users and healthcare teams manage blood glucose information to help with diabetes management. After the user transmits and stores blood glucose data to the secure database, the ARK Care system allows family members and/or healthcare professionals to view and monitor the user's data and reports. Family and healthcare team members must receive permission from the primary user and create a password protected login before viewing data and allows for secure communications. The user and approved healthcare team members can view the blood glucose data in different formats such as logbooks, charts, and graphs. The data can be viewed through selected time intervals and these intervals can be compared over time to track disease management. The subject can also enter and track other health-related information such as body weight, blood pressure, lab values, and exercise activities. Groups of profiles can be queried for bulk reporting on data related to tracking and trending of outcomes, supporting diabetes disease management in individuals and in managed care organizations.
J. Substantial Equivalence Information:
1. Predicate device name(s): MyCare Team, MCT Diabetes
2. Predicate 510(k) number(s): k073699
3. Comparison with predicate:
| Device Characteristic | New Device: ARK Care Diabetes Management | Predicate: MCT Diabetes |
| --- | --- | --- |
{3}
| Device Characteristic | New Device: ARK Care Diabetes Management | Predicate: MCT Diabetes |
| --- | --- | --- |
| Indications for Use Statement | ARKRAY ARK Care Diabetes Management System™ is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support diabetes management. | Same |
| Intended Use | Home use or Clinical Assist diabetics, families, and professionals in management of blood glucose Support diabetes management Single or multi-patient use Same | Same |
| Data Source | Compile data from multiple different brands of glucose meters for display in one place on user's personal computer | Same |
| Software Use | Single (individual) or multiple user (clinical) settings | Same |
| Use in Clinic | Patient list available | Same |
| Report types | Logbook Readings Line Chart Average Bar Chart Percentage Pie Chart Percentage Reading by Time of Day | Same |
| Special Glucose Monitor Instrument Requirements | Wide Range of Supported meters are available; As additional software protocols are obtained and devices are validated, capabilities will be broadened to include other blood glucose meters. | Same |
| User's Personal Computer (PC) | Windows Vista (32 and 64 bit) Windows XP-XP Home Edition & XP Professional (32 bit) Windows 7 (32 and 64 bit) | Windows 98 Second Edition (SE), Windows 2000 and Windows XP Home and Professional |
| Requirements | • 600 MHz Intel Pentium II or equivalent • Minimum 128MB RAM 100-200MB RAM used during installations and 100MB used after installation. | Same Same |
{4}
| Device Characteristic | New Device: ARK Care Diabetes Management | Predicate: MCT Diabetes |
| --- | --- | --- |
| | • CD ROM Drive
• BGM Manufacturer’s required cable.
• Installed from website | Same
Same
Same |
| Technical Support | Yes | Same |
| User’s Manual | Available on Internet while using program (“Help System”) Hard copy user manual available | Same |
| Auto-detect COM port | Yes | Same |
| Installation of program | Installed from internet link | Same |
| Required data on patient entry includes patient ID, name, address, ect. | Yes | Same |
| Physician Information | Multiple individuals | Same |
| Diabetes educator information | Multiple individuals | Same |
| Types of information that can be manually entered | Insulin list, medication list, exercise, blood pressure data, laboratory results, meals | Same |
K. Standard/Guidance Document Referenced (if applicable):
IS014971 - Medical Devices-Application of Risk Management to Medical Devices
L. Test Principle:
ARK Care Diabetes Management software is an accessory to compatible meters, which use specific test principles such as in the ARKRAY Glucocard Vital Blood Glucose Meter (k091102)
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Presented in k091102
b. Linearity/assay reportable range:
Presented in k091102
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Presented in k091102
d. Detection limit:
Presented in k091102
e. Analytical specificity:
4 of 8
{5}
Presented in k091102
f. Assay cut-off:
Presented in k091102
2. Comparison studies:
a. Method comparison with predicate device:
Presented in k091102
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Thirty eight (38) diabetic adult consumers and six (6) healthcare professionals were recruited and screened for study inclusion criteria. Inclusion criteria for participants included that the user was capable of using a computer independently and had at minimum basic computer skills. Personal Users must be a Type 1 or Type 2 diabetic who is actively testing their blood glucose. Professional Users needed to have a degree in a healthcare field that may require interaction with diabetic persons.
Users were given the use of a laptop with the ARK Care website open, a preloaded meter with glucose and control solution values, the User Guide and a set of user (Personal and Professional) instructions to complete study tasks. Program evaluation was measured by completing ease of use and performance questionnaires
The Personal User participants were asked to:
1. Enroll as a patient
2. Navigation Overview
3. Create a Care Team
4. Glucose Meter Upload
5. Manually add Medication, Lab results, Meals and Insulin using My Data Logbooks
6. View Detail Readings chart in My Charts
7. View and Print Average Readings charts in My Charts
8. View Total View chart in My Charts
9. Sign Out of ARK Care site
10. Rate the program by completing questionnaires
The Professional participants were asked to:
1. Enroll from an email invitation
2. Register as a Care Team Member
5 of 8
{6}
3. From My Data -> Alerts view your team member's information
4. Navigation overview
5. Display Blood Glucose Logbook
6. Display Medication Log
7. Display and Print Lab Test Results
8. Display Meal Log
9. Display Insulin Log
10. Display Total View interactive chart
11. Under My Information -> Notifications set to Daily updates and save
12. Sign Out of ARK Care site
13. Rate the program by completing questionnaires
The scores of "Ease of Use" and "Performance" questionnaires were both over 99% passing. Sixty-eight percent (68%) of Personal Users would use this software program to help manage their diabetes and one-hundred percent (100%) of Professional Users would use this program to assist in managing their patient's diabetes.
The scores for pass/fail criteria of both groups exceeded 80% and 70% required for the acceptance criteria. The clinical data demonstrated that the ARK Care Internet-Based Diabetes Management System for diabetics is easy to use for both personal and professional users. The software performed as expected during the course of the study.
4. Clinical cut-off: Not Applicable
5. Expected values/Reference range: Presented in k091102
N. Instrument Name:
ARKRAY Diabetes Management Software
O. System Descriptions:
1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ☐ X or No ☐
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes ☐ or No ☐ X
6 of 8
{7}
Windows Vista (32 & 64 bit) Windows XP-XP Home Edition & XP Professional (32 bit) Windows 7 (32 & 64 bit)
600 MHz Intel Pentium II or equivalent
Minimum 128MB RAM 100-200MB RAM used during installations and 100MB used after installation.
CD ROM Drive
Internet web application which provides the specified features in the familiar form of a web site accessible via the Internet Explorer web browser or Firefox web browser using the IE Tabs plug-in
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Meter controlled, by time and date stamp
4. Specimen Sampling and Handling:
Not Applicable
5. Calibration:
Not Applicable
6. Quality Control:
Not Applicable
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
For each meter that will be claimed as compatible with the ARK Care system a demonstration of accurate data transmission must be performed. This testing includes meter load up and meter memory rollover testing to confirm data synchronization upon after rollover.
Meters are loaded to memory capacity using the low and high controls, in addition a
7 of 8
{8}
mid level control solution was created by a technician using equal parts of the low and high control solutions. The assigned meter came with a data log sheet. Each result and its details were logged after each individual test. This methodology also prevented duplicate test results in the event that the same dose level was used successively and potentially skewing the data set. In addition the technicians varied the testing of control solutions, low, middle, and high to reduce potential misanalysis of duplicate results on the meter system.
After the meters were loaded to capacity they were passed off to the software validation group. Each meter system was then plugged into the ARK Care system, correct meter platform selected in ARK Care and then meter memory was uploaded and capacity verified against claim.
After initial memory capacity was confirmed ten more results were added to the meters. Results of the data additions were compared against the original load up data.
The above testing confirms the ARK Care system can accurately acquire blood glucose data from the tested glucose meter system.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.