ATLANTIS ABUTMENT FOR CAMLOG IMPLANT, DUPLICATE ABUTMENT, LATER ABUTMENT
K110640 · Astra Tech, Inc. · NHA · Jun 10, 2011 · Dental
Device Facts
Record ID
K110640
Device Name
ATLANTIS ABUTMENT FOR CAMLOG IMPLANT, DUPLICATE ABUTMENT, LATER ABUTMENT
Applicant
Astra Tech, Inc.
Product Code
NHA · Dental
Decision Date
Jun 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. This device is compatible with the following manufacturers' implant systems: The titanium abutments are compatible with the Camlog 3.3mm, 4.3mm, 5.0mm, 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants. The zirconia abutments are compatible with the Camlog 3.8mm, 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants. Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments on small diameter implants are intended for the anterior region only.
Device Story
Atlantis Abutment provides interface between endosseous dental implant and prosthetic restoration; supports single/multiple tooth prostheses via cement or screw retention. Device consists of abutment and titanium screw; fabricated from Titanium alloy (Ti-6Al-4V ELI) or yttria-stabilized tetragonal zirconia polycrystals (Y-TZP). Placed by dental professionals in clinical settings; secures to implant shoulder. Provides structural support for dental prosthetics; enables functional restoration of edentulous sites. Compatibility varies by material and specific Camlog implant system dimensions.
Clinical Evidence
Bench testing only. Static and fatigue compression testing performed on worst-case implant assemblies using angled titanium and zirconia abutments with Camlog implants. Results demonstrated compatibility and that the system supports appropriate loads.
Technological Characteristics
Materials: Titanium alloy (ASTM F-136) and yttria-stabilized tetragonal zirconia polycrystals (ISO 6972 & 13356). Form factor: Custom abutment for endosseous dental implants. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for partially or completely edentulous patients requiring single or multiple tooth prosthetic support in the mandible or maxilla via endosseous dental implants. Highly angled abutments on small diameter implants restricted to anterior region.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
K110640
JUN 1 0 2011
Premarket Notification Section 5: Page - 4
#### 510(k) Summary
### Astra Tech Inc. Atlantis™ Abutment for Camlog Implant
# ADMINISTRATIVE INFORMATION
| 510K Summary preparation date: | Februaury 1, 2011 |
|--------------------------------|-----------------------------------------------------------------------|
| Manufacturer Name: | Astra Tech Inc.<br>590 Lincoln Street<br>Waltham, Massachusetts 02541 |
| Telephone: | 781-810-6462 |
| Fax: | 781-810-6719 |
| Official Contact: | Franklin Uyleman |
| Representative/Consultant: | Betsy A. Brown |
Betsy A. Brown B.A. Brown and Associates Inc. Telephone: 847-560-4406 847-677-0177 Fax:
#### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: Common Name:
Product Code: Classification Panel: Reviewing Branch:
Atlantis™ Abutment for Camlog Implant Endosseous dental implant abutment 21 CFR 872.3630 NHA Dental Products Panel Dental Devices Branch
#### INTENDED USE
The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
{1}------------------------------------------------
Premarket Notification Section 5: Page - 5
## INTENDED USE (continued)
This device is compatible with the following manufacturers' implant systems:
The titanium abutments are compatible with the Camlog 3.3mm, 4.3mm, 5.0mm, 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants.
The zirconia abutments are compatible with the Camlog 3.8mm, 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants.
Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.
Highly angled abutments on small diameter implants are intended for the anterior region only.
#### DEVICE DESCRIPTION
The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented or screw retained restorations. The Atlantis™ Abutment for Camlog Implant and abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136) for the 3.3mm. 3.8mm. 4.3mm. 5.0mm and 6.0mm sizes. In addition. the Atlantis™ Abutment for Camlog Implant for the 3.8mm. 4.3mm. 5.0mm and 6.0mm sizes are also made of biocompatible material, yttria-stabilized tetragonal zirconia polycrystals (Y -TZP) and meets ISO Standards 6972 & 13356). Zirconia may have variation in shade. The titanium and zirconia abutments are placed over the implant shoulder and are mounted into the implant with a titanium screw.
### EQUIVALENCE TO MARKETED DEVICE
Astra Tech Inc. demonstrated that, for purposes of the FDA's regulations of medical devices, the Atlantis™ Abutment for Camlog Implant is substantially equivalent in indications and design principles to Altatec GmbH predicate devices: Screwline Implant cleared under K022425. Cylinder Implant cleared under K000065 and Rootform Implant cleared under K000100 each of which has been determined by FDA to be substantially equivalent to preamendment devices.
{2}------------------------------------------------
Premarket Notification Section 5: Page - 6
| Technological<br>Characteristics | Atlantis™ Abutment for<br>Camlog Implant | Altatec Screwline,Cylinder<br>and Rootform Implants |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | -Titanium Alloy<br>-Biocompatible ceramic<br>material | -Titanium Alloy |
| Performance characteristics | Allows the prosthesis to be<br>cemented or screw retained to<br>abutment. While the abutment<br>screw is intended to secure the<br>abutment to the endosseous<br>implant. | Allows the prosthesis to be<br>cemented or screw retained to<br>abutment. While the abutment<br>screw is intended to secure the<br>abutment to the endosseous<br>implant. |
| Intended Use | Intended for use with an<br>endosseous implant to support<br>a prosthetic device in a<br>partially or completely<br>endentulous patient. Intended<br>for use to support single or<br>multiple tooth prosthesis, in<br>mandible or maxilla. | Intended for use with an<br>endosseous implant to support<br>a prosthetic device in a<br>partially or completely<br>endentulous patient. Intended<br>for use to support single or<br>multiple tooth prosthesis, in<br>mandible or maxilla. |
### Table 1: Substantial Equivalence Summary
# Summary of Non-clinical Testing
Static and fatigue compression testing was conducted on "worst case scenario" implant assemblies using Atlantis angled titanium and zirconia abutments with the Camlog implants. Test results demonstrated that the Atlantis Abutments are compatible with the Camlog implants and the implant system supported appropriate static and fatigue test loads demonstrating that the implant system performs as intended.
# Conclusion for Substantial Equivalence:
The Atlantis™ Abutment for Camlog Implant is substantially equivalent to Altatec's predicate devices: : Screwline Implant cleared under K022425, Cylinder Implant cleared under K000065 and Rootform Implant cleared under K000100, based on noted similarities in indication, manufacturing material, generated design principle and performance characteristics data.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected, flowing lines, possibly representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 0 2011
Astra Tech. Incorporated C/ O Ms. Betsy A. Brown Regulatory Consultant B.A. Brown & Associates 8944 Tamaroa Terrace Skokie, Illinois 60076
Re: K110640
Trade/Device Name: AtlantisTM Abutment for Camlog Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 3, 2011 Received: June 8, 2011
Dear Ms. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification Section 4: Page - 3
#### Indications for Use
510(k) Number (if known)_
Device Name: Atlantis™ Abutment for Camlog Implant
Indication for Use:
The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
This device is compatible with the following manufacturers' implant systems:
The titanium abutments are compatible with the Camlog 3.3mm, 4.3mm, 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants.
The zirconia abutments are compatible with the Camlog 3.8mm. 5.0mm and 6.0mm Root-Line, Cylinder-Line, Cylinder Line TPS, Screw-Line, Promote and Promote Plus K-Series Implants.
Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.
Highly angled abutments on small diameter implants are intended for the anterior region only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
510(k) Number:
(Division Sigm-Uff) Division of Anesthesiology, General Hospital infection Control, Dental Devices
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.