The Venner™ Tracheal Seal Monitor (TSM) is indicated for use to monitor, maintain and regulate the pressure within the PneuX P.Y.TM Endotracheal or Tracheostomy Tube Cuff in adult patients who have been confined to hospital ICU units where intubation is expected to be more than 24 hours, but less than or equal to 30 days.
Device Story
Venner™ Tracheal Seal Monitor (TSM) is an electronic, software-driven automatic inflation cuff controller. It connects to PneuX P.Y.™ endotracheal or tracheostomy tubes via a sterile, single-patient-use extension tube. The device uses internal pressure sensors and a pump to continuously measure and regulate cuff-to-tracheal seal pressure, maintaining a default of 30 cm H2O (range 10-50 mmHg). Used in ICU settings by healthcare professionals, the device provides a digital display of cuff pressure, user-adjustable settings, and alarms for pressure, air leaks, malposition, and blockages. It includes a mechanical safety relief valve that opens if pressure exceeds 57 mmHg. By maintaining stable cuff pressure, the device aims to prevent aspiration pneumonia and tracheal necrosis caused by inadequate or excessive cuff pressure.
Clinical Evidence
Bench testing only. Performance testing verified the device met its ± 5% accuracy specification for maintaining cuff pressure. Testing confirmed the device's ability to regulate pressure across various settings and its rapid response to environmental changes. The safety over-pressure valve was validated to open at >57 mmHg and reset at 27 mmHg. System-level software verification and validation were completed. Electrical and electromagnetic safety testing (EN 60601-1, 60601-2, 60601-4) and environmental stress testing (vibration/bump per IEC 60068 series) were successfully performed.
Technological Characteristics
Electro-mechanical control unit; software-driven; pressure sensors and pump; digital display with pushbuttons. Electrical power source. Connectivity: connects to airway tubes via sterile extension tube with non-luer proximal and luer-slip distal connectors. Safety features: electronic alarms and mechanical pressure relief valve (57 mmHg). Standards: EN 60601-1, 60601-2, 60601-4 (electrical/EMC); IEC 60068-2-6, 60068-2-27, 60068-2-64 (environmental).
Indications for Use
Indicated for adult patients requiring intubation in hospital ICU settings for durations between 24 hours and 30 days to monitor, maintain, and regulate endotracheal or tracheostomy tube cuff pressure.
Regulatory Classification
Identification
An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.
{0}------------------------------------------------
< 11063 / JUL 28 2011
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### l GENERAL INFORMATION
#### 1.1 Submitter and Owner of the 510(k)
Venner. Medical (Singapore) Pte Ltd 35 Joo Koon Circle Singapore 629110
#### 1.2 Official Correspondent
Christine L. Brauer, PhD Regulatory Affairs Consultant 7 Trail House Court Rockville, MD 20850
Telephone: (301) 545-1990 Fax: (301) 545-1992 E-mail: chrisbrauer@earthlink.net
#### 1.3 Date of Preparation
February 25, 2011
#### NAME OF THE DEVICE 2
#### 2.1 Trade/Proprietary Name
Venner™ Tracheal Seal Monitor
#### Common/Usual Name 2.2
Cuff pressure regulator and monitor
#### 2.3 Classification Information
Classification Name: Inflatable Tracheal Tube Cuff
Classification Regulation: 21 CFR § 868.5750
Class: ||
Product Code: BSK
Anesthesiology Panel:
000010
{1}------------------------------------------------
#### 3 PREDICATE DEVICES
The predicate devices are as follows:
- 1) the Lanz Pressure Valve cleared under premarket notification K885216; and,
- 2) the PYTON Cuff Pressure Regulator cleared under premarket notification K092733.
The predicate devices are cuff pressure regulators, have been cleared by FDA and classified under 21 CFR § 868.5750, as Class II medical devices.
#### DESCRIPTION OF THE DEVICE 4
The Venner™ Tracheal Seal Monitor consists of an automatic inflation cuff controller (control unit) that connects to the Venner™ PneuX P.Y.TM Endotracheal or Tracheostomy Tube using a single patient use, sterile extension tube. The Venner™ Tracheal Seal Monitor regulates and maintains cuff pressure in endotracheal and tracheostomy tubes in adult patients in the intensive care unit (ICU) setting. The device is intended to measure, monitor and maintain a stable cuff pressure in endotracheal and tracheostomy tubes.
During intubation, variations in cuff pressure may occur for multiple reasons, including the transfusion of gases through the cuff, changes in tracheal compliance, modifications in the location of the cuff within the airway, and other reasons. Maintaining a consistent and stable cuff pressure is intended to help minimize complications. such as excessive tracheal wall pressure that may lead to necrosis, and problems associated with inadequate tracheal wall pressure, such as aspiration pneumonia.
The Venner™ Tracheal Seal Monitor consists of an electronic automatic pressure controller with a pressure sensors and a pump, and provides a user interface with adjustable settings, indicators and alarms. The Venner Tracheal Seal Monitor is intended to be used with the following commercially available devices:
- 1. The Venner™ Pneux P.Y.™ Endotracheal Tube: This endotracheal tube was cleared by FDA under premarket notification K093135. It is a straight, cuffed, wire reinforced tracheal tube. It is packaged separately and provided as a sterile, single use device.
- 2. The Venner™ PneuX P.Y.TM Tracheostomy Tube: The tracheostomy tube was cleared by FDA under premarket notification K100950. It is a straight, flexible, cuffed tracheostomy tube with an adjustable neck plate. It is packaged separately and provided as a sterile, single use device.
- 3. Extension Tube: The extension tube is a class I. medical device exempt from premarket notification under 21 CFR § 868.5860. The extension tube connects the Venner Tracheal Seal Monitor to the cuff of an endotracheal or tracheostomy tube. It uses a non-luer connector at the proximal end to connect to the control unit and a patented luer slip connector at the distal end with a unique design feature to prevent
{2}------------------------------------------------
accidental misconnection of the air line to a standard luer lock connector (i.e., standard infusion sets and catheters with female lugged luer lock connectors). It is packaged separately and provided as a sterile, single patient use device.
#### 5 INDICATIONS FOR USE AND INTENDED USE
The Venner™ Tracheal Seal Monitor has the following indications for use:
"The Venner™ Tracheal Seal Monitor (TSM) is indicated for use to monitor, maintain and regulate the pressure within the PneuX P.Y.TM Endotracheal or Tracheostomy Tube Cuff in adult patients who have been confined to hospital ICU units where intubation is expected to be more than 24 hours, but less than or equal to 30 days."
This is the same intended use as the predicate devices.
The Venner™ Tracheal Seal Monitor and the predicate devices share the same intended use namely, to maintain and regulate pressure in the cuffs of endotracheal and tracheostomy tubes in patients who require intubation. Two of the devices (the Venner™ Tracheal Seal Monitor and the PYTON Cuff Pressure Regulator) also measure and monitor cuff pressure.
Each device is intended to ensure proper inflation of the cuff to provide a seal between the airway tube and the patient's trachea by maintaining and regulating cuff pressure. The Venner™ Tracheal Seal Monitor and the predicate devices all share the same primary function; that is, to obtain a cuff-to-tracheal seal pressure of approximately 25-30 cm H2O to ensure that the cuff provides a proper seal between the tube and the patient's trachea to prevent the leakage of liquids and risk of aspiration, and to minimize the risk of overpressure and damage to the trachea. The devices are intended for use for the same patient population: patients who require intubation. The table on the following page establishes the Venner™ Tracheal Seal Monitor and the predicate devices have the same intended use.
{3}------------------------------------------------
| Device Characteristic | A Comparison of the Venner™ Tracheal Seal Monitor to the Predicate Devices | | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Venner™ Tracheal Seal Monitor and<br>Extension Tube | Lanz Pressure Valve<br>Mallinckrodt, Inc.<br>K8852161 | PYTON Cuff Pressure Regulator<br>ARM Medical Devices, Inc.<br>K0927332 |
| Indications for Use | ... is indicated for use to monitor, maintain<br>and regulate the pressure within the PneuX<br>P.Y.T™ Endotracheal or Tracheostomy Tube<br>Cuff in adult patients who have been<br>confined to hospital ICU units where<br>intubation is expected to be more than 24<br>hours, but less than or equal to 30 days | ... is indicated for use with cuffed<br>tracheal and tracheostomy tubes when<br>intracuff pressures of approximately 30<br>cm H₂O are judged by the clinician to be<br>appropriate | ... to measure and regulate intra-cuff<br>pressures of endotracheal supraglottic<br>airways or tracheostomy tubes...for<br>patients who are intubated... in hospitals,<br>pre-hospital (EMS), extended care<br>facilities and outpatient clinics... |
| Intended Use | To regulate and maintain cuff pressure<br>To measure and monitor cuff pressure | To regulate and maintain cuff pressure | To regulate and maintain cuff pressure<br>To measure and monitor cuff pressure |
| Intended Patient Population | Adult patients requiring intubation | Patients requiring intubation | Patients requiring intubation |
| Intended Users | Prescription device - healthcare<br>professionals | Prescription device - healthcare<br>professionals | Prescription device - healthcare<br>professionals |
| Operational Environment | Intensive care units (ICU) | Unspecified | Multiple settings where a patient may be<br>intubated, including the ICU |
| Primary Device Functions | To continuously measure cuff pressure<br>To regulate and maintain cuff pressure | To regulate and maintain cuff pressure | To continuously measure cuff pressure<br>To regulate and maintain cuff pressure |
| Cuff-Tracheal Seal Pressure<br>Range | Default: 30 cm H₂O (22 mmHg)<br>Range: 10 - 50 mmHg in 10 mmHg<br>intervals | Recommended setting: 30 cm H₂O | Default: 25 cm H₂O<br>Range: 0 - 85 cm H₂O |
| Pressure Monitor and Control<br>Mechanisms | Control unit consists of an electronic,<br>software-driven device with pressure<br>sensors and stepping motor to adjust<br>pressure | Elastic balloon aids in regulating pressure<br>by expanding and contracting | Control unit consists of an electronic,<br>software-driven device with pressure<br>sensors and pump to adjust pressure |
| Pressure Accuracy | $\pm$ 5% of set value | Not available | $\pm$ 2.0 cm H₂O |
| System Components | Control unit (capital equipment)<br>Extension tube (single patient use) to<br>connect control unit to endotracheal or<br>tracheostomy tube | Single patient use device attached to the<br>endotracheal or tracheostomy tube | Control unit (capital equipment)<br>Plastic tubing set (single patient use) to<br>connect control unit to endotracheal or<br>tracheostomy tube |
| User Interface | Digital display with cuff pressure and<br>pushbuttons | Visual inspection of balloon | LED back-lit LCD digital display with cuff<br>pressure and pushbuttons |
| Device Characteristic | Venner™ Tracheal Seal Monitor and Extension Tube | Lanz Pressure Valve<br>Mallinckrodt, Inc.<br>K885216¹ | PYTON Cuff Pressure Regulator<br>ARM Medical Devices, Inc.<br>K092733² |
| Alarms | Multiple alarms, including alarms for pressure, air leak, malposition, blockage | One alarm when pressure exceeds 25 cm H₂O | Yes although details not provided |
| Safety Valve | Incorporates a pressure relief valve if cuff-tracheal seal pressure were to exceed 57 mmHg | No | Incorporates a pressure relief valve if cuff-tracheal seal pressure were to exceed 100 cm H₂O |
| Power source | Electrical | Manual | Electrical with battery backup |
<sup>1</sup> Information obtained from the purged 510(k) application K885216.
2 Information obtained from the 510(k) Summary for K092733 dated February 26, 2010.
{4}------------------------------------------------
{5}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE 6 DEVICES
The Venner™ Tracheal Seal Monitor and the predicate devices share many of the same technological characteristics as summarized in the previous table. All of the devices share the same primary function: that is, to obtain a cuff-to-tracheal seal pressure of approximately 25-30 cm H2O to ensure that the cuff provides a proper seal between the tube and the patient's trachea to prevent the leakage of liquids and risk of aspiration, and to minimize the risk of over-pressure and damage to the trachea.
Two of the devices (the Venner™ Tracheal Seal Monitor and the PYTON Cuff Pressure Regulator) accomplish this goal by using a software-based, electro-mechanical control unit with similar accuracies in their measurements and other features, including alarms, indicators and safety features. Please refer to the table for additional details.
# 7 PERFORMANCE TESTING
The 510(k) submission provided performance data to establish the substantial equivalence of the Venner™ Tracheal Seal Monitor to the predicate devices. Performance testing was performed to demonstrate that the Venner™ Tracheal Seal Monitor met its specifications, and is suitable and ready for commercial distribution.
These tests evaluated the ability of the Venner™ Tracheal Seal Monitor to maintain and regulate cuff pressure in a variety of settings, and often compared its performance to other devices. These studies established the accuracy the Venner™ Tracheal Seal Monitor to maintain the selected pressure; the device met its specified accuracy of ± 5%. Other studies evaluated the ability of the Venner™ Tracheal Seal Monitor to regulate and maintain cuff pressure in a variety of physical settings, and the reaction time of the device to change the pressure in the cuff. In each study, the Venner™ Tracheal Seal Monitor was able to maintain and regulate the pressure of a cuff of an airway tube, and to quickly adapt to changes in the environment or changes programmed by the user. These studies also evaluated the performance of the safety over-pressure valve to demonstrate that the valve automatically opened with an audible "click" when the calculated cuff-tracheal seal wall pressure exceeded 57 mm Hg, and subsequently closed (or reset) when the calculated cuff-tracheal seal wall pressure returned to 27 mm Hg for safe operation. The safety valve performed within specifications and met the acceptance criteria.
The Venner™ Tracheal Seal Monitor underwent system level software verification and validation testing to demonstrate the device performs as intended, and meets its requirements. The device passed all testing and met its specifications.
The Venner™ Tracheal Seal Monitor met international safety standards, including electrical and electromagnetic safety testing (EN 60601-1, 60601-2 and 60601-4).
The sponsor also performed a series of tests to demonstrate that its device can withstand exposure to environmental stresses (i.e., random vibration, sinusoidal vibration and bump) and continue to perform as intended. In these studies, the Venner Tracheal Seal Monitor was
{6}------------------------------------------------
tested according to the following standards: IEC 60068-2-64: Random Vibration, IEC 60068-2-6: Sinusoidal Vibration; IEC 60068-2-64: Random Vibration; and IEC 60068-2-27: Bump. The device successfully completed the testing.
#### CONCLUSIONS 8
This 510(k) submission demonstrates that the Venner™ Tracheal Seal Monitor is substantially equivalent to the predicate devices.
{7}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Venner Medical (Singapore) PTE Limited C/O Christine Brauer, Ph.D. Regulatory Affairs Consultant Brauer Device Consultants 7 Trail House Court Rockville, Maryland 20850
JUL 28 2011
Re: K110631
Trade/Device Name: Venner™ Tracheal Seal Monitor Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: II Product Code: BSK Dated: June 22, 2011 Received: June 24, 2011
Dear Dr. Brauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
## Page 2 - Dr. Brauer
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act`s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDR /CDRHOffices/ucm1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Ams
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{9}------------------------------------------------
510(k) Number (if known):
Venner™ Tracheal Seal Monitor Device Name:
Indications for Use:
The Venner™ Tracheal Seal Monitor (TSM) is indicated for use to monitor, maintain and requlate the pressure within the Pneux P. Y. ™ Endotracheal or Tracheostomy Tube Cuff in adult patients who have been confined to hospital ICU units where intubation is expected to be more than 24 hours, but less than or equal to 30 days.
Prescription Use __ x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Schutthelm
(Division Sign-Ott) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110631
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.