PRESSURE RIGHT, SINGLE-USE, DISPOSABLE, PRESSURE-SENSITIVE EMETIC-MANAGEMENT WRIST STRIP

K110563 · Therapeutics 101, Inc (Dba Pressure Point, Inc.) · MVV · Apr 8, 2011 · NE

Device Facts

Record IDK110563
Device NamePRESSURE RIGHT, SINGLE-USE, DISPOSABLE, PRESSURE-SENSITIVE EMETIC-MANAGEMENT WRIST STRIP
ApplicantTherapeutics 101, Inc (Dba Pressure Point, Inc.)
Product CodeMVV · NE
Decision DateApr 8, 2011
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is indicated for the relief of emetic (nausea and vomiting) symptoms associated with Post-Operative Anesthesia.

Device Story

Single-use, disposable acupressure wrist strip; features 5.50" x 1.0" band with affixed hard plastic button. Operates by applying mechanical pressure to Nei-Kuan (P6) acupressure point. Includes P6 locator template for accurate placement. Worn on each wrist by adult surgical patients. Provides non-pharmacological nausea and vomiting relief as part of multi-modal antiemetic strategy. Clinical benefit: reduction of postoperative emetic symptoms.

Clinical Evidence

Clinical study evaluated use of disposable acupressure device as part of multi-modal antiemetic strategy for reducing postoperative nausea and vomiting, including assessment of device adhesive performance.

Technological Characteristics

Mechanical acupressure device; consists of wrist band with affixed hard plastic button. Dimensions: 5.50" long x 1.0" wide. Single-use, disposable. Biocompatible materials for prolonged skin contact.

Indications for Use

Indicated for relief of nausea and vomiting associated with post-operative anesthesia in adults 18 years and older.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K110563 Therapeutics 101, Inc. (dba Pressure Point, Inc.) Traditional 510(k) Premarket Submission Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip ## · SECTION 5 - 510(k) SUMMARY APR - 8 2011 ## Submission Correspondent Emergo Group, Inc. www.emergogroup.com/ ## Address י ל 611 West Fifth Street Third Floor Austin, TX 78701 ## Phone (512) 327-9997 ## Fax (512) 327-9998 ## Contact Julie Powell, Vice President, Quality Assurance jpowell@emergogroup.com ## Submission Sponsor Therapeutics 101, Inc. (dba Pressure Point, Inc.) 36 Green Meadow Drive Tinton Falls, NJ 07724 Phone: 732-747-6049 FDA Establishment Registration #: none, they will obtain number prior to marketing device. ## Date Prepared February 21, 2011 ## Trade Name Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip ## Classification Name Device, Acupressure ## Regulation Number NA Product Code MVV ## Classification Panel Neurology {1}------------------------------------------------ K110563 Therapeutics 101, Inc. (dba Pressure Point, Inc.) Traditional 510(k) Premarket Submission Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip Device Class Unclassified ## Predicate Device Sea-Band, by Sea-Band UK LTD 510(k) # K033268 ## Indications for Use The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is indicated for the relief of emetic (nausea and vomiting) symptoms associated with Post-Operative Anesthesia. ## Device Description The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is a single-use, disposable acupressure strip that operates by exerting pressure on the Nei-Kuan (acupuncture) or (P6) pressure point. The Pressure Right® device has a three-quarter wrist size design (5.50″ long X 1.0″ with an affixed hard plastic button. Its pressure stimulation effect to the P6 acupressure point provides nausea and vomiting relief for surgical patients. Pressure Right® device is intended for use by adults, 18 years and older and intended to be worn on each wrist. The Pressure Right® device also comes with a P6 locator template to accurately determine an individual's P6 pressure point. ## Summary of Non-Clinical Data Submitted The following testing has been performed to support substantial equivalence: - Bench testing to evaluate the pressure equivalence of the Pressure Right® device compared . to the marketed predicate (Sea-Band) device. - . Materials with patient contact were evaluated for biocompatibility requirements with intact skin for prolonged use. ## Summary of Clinical Data Submitted The following clinical study has been performed on the Pressure Right® device to support substantial equivalence: - Use of disposable acupressure device as part of a multi-model antiemetic strategy for . reducing postoperative nausea and vomiting. Study includes evaluation of adhesive abilities of the Pressure Right® device. ## Safety and Effectiveness By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device. {2}------------------------------------------------ K110563 Therapeutics 101, Inc. (dba Pressure Point, Inc.) Traditional 510(k) Premarket Submission Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip . It has been shown in this 510(k) submission that the differences between the Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip and the predicate device do not raise any questions regarding its safety and effectiveness. The Pressure Right® device, as designed and manufactured, is therefore determined to be substantially equivalent to the referenced predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Therapeutics 101, Inc. (dba Pressure Point, Inc.) c/o Julie Powell Vice President, Quality Assurance Emergo Group. Inc 611 West Fifth Street Third floor Austin, TX 78701 APR = 8 2511 Re: K110563 Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip Regulatory Class: unclassified Product Code: MVV Dated: February 24, 2011 Received: February 28, 2011 Trade/Device Name: Dear Ms. Powell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Kesia Alexander Malviná B. Eydelman. M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Therapeutics 101, Inc. (dba Pressure Point, Inc.) Traditional 510(k) Premarket Submission Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip ## SECTION 4 - INDICATIONS FOR USE 510(k) Number (if known): K110563 Device Name Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip ## Indications for Use The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is indicated for the relief of emetic (nausea and vomiting) symptoms associated with Post-Operative Anesthesia. Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Seon. (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and This. . Section 4: Page 1 of 1 510(k) Number K 110 563
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...