ENSITE VELOCITY SYSTEM

K110549 · St Jude Medical · DQK · May 5, 2011 · Cardiovascular

Device Facts

Record IDK110549
Device NameENSITE VELOCITY SYSTEM
ApplicantSt Jude Medical
Product CodeDQK · Cardiovascular
Decision DateMay 5, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2

Indications for Use

The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. When used with the EnSite Array™ Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR When used with the EnSite NavX Surface Electrode Kit, the EnSite System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. When used with the v.1.5.1 MediGuide gMPS System Interface Module, the EnSite Velocity System interfaces to the MediGuide Guided Medical Positioning System II to combine and display gMPS II processed patient positioning and orientation Mapping information.

Device Story

EnSite Velocity System is an electrophysiology mapping system used in clinical settings by physicians. It processes inputs from diagnostic catheters and surface electrode kits to map cardiac electrical activity and display catheter positioning. The v.1.5.1 update adds an interface module to integrate with the MediGuide gMPS II system, allowing the display of processed patient positioning and orientation data alongside cardiac maps. The system consists of a display workstation and amplifier. Physicians use the visual output to identify complex arrhythmias and guide catheter placement during procedures. The integration improves clinical workflow by combining positioning data from the MediGuide system with the EnSite mapping environment.

Clinical Evidence

Bench testing only. Verification and integrated system testing were performed using a gMPS II emulator to confirm data packet transmission, error condition handling, and system performance (map/waveform display refresh rates and data rates). No clinical data was required.

Technological Characteristics

Electrophysiology mapping system comprising a display workstation and amplifier. Integrates with MediGuide gMPS II via a software interface module. Connectivity includes data exchange with external positioning systems. Software-based processing for map and waveform display.

Indications for Use

Indicated for patients requiring electrophysiology studies. Used in the right atrium for complex arrhythmias (with EnSite Array Catheter) or to display position of conventional EP catheters (with NavX Surface Electrode Kit).

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ MAY - 5 2011 # 12 510(k) Summary for Public Disclosure ## 12.1 Submitter's Name/Contact Person Donna R. Lunak St. Jude Medical One St. Jude Medical Drive St. Paul. MN 55117 USA The Establishment Registration Number is 2184149. ### 12.2 Predicate Device EnSite Velocity System with PRT (K101667). ### 12.3 Common or Usual Name Electrophysiology Mapping System with console and catheter ### 12.4 Proprietary Name EnSite Velocity System Consisting of: EnSite Multi-electrode Diagnostic Catheter (EnSite Array™ - Model EC1000) EnSite Electrophysiology Workstation – Model EE3000 EnSite NavX Surface Electrode Kit - Model EN0010 ## 12.5 Classification Name DQK, Programmable diagnostic computer (21 CFR 870.1425), Class II, Circulatory Systems Device Panel ### 12.6 Hardware Description The EnSite Velocity System consists of the following: - Display Workstation . - Amplifier . {2}------------------------------------------------ ## 12.7 Indications for Use The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. - t When used with the EnSite Array™ Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR - When used with the EnSite NavX Surface Electrode Kit, the EnSite . System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. ### 12.8 Intended Use When used with the EnSite Velocity System v.1.5.1 MediGuide aMPS System Interface Module, the EnSite Velocity System interfaces to the MediGuide gMPS System to combine and display SJM MediGuide gMPS System processed patient positioning and orientation information. ## 12.9 Device Comparison to the Predicate Device The EnSite Velocity System v.1.5.1 has the same intended use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity System v.1.5.1 are substantially equivalent to the predicate device. ## 12.10 Summary of Non-Clinical Testing Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product. As part of the testing of EnSite Velocity System v.1.5.1, a gMPS II MediGuide emulator was used to simulate the gMPS II (K102905). The aMPS II MediGuide emulator was used to verify packets sent by the EnSite Velocity System. In addition, the emulator allowed the testing of error conditions by simulating error conditions that would typically be sent by the gMPS II (K102905). The testing included testing of the EnSite Velocity System to ensure that is handles the error conditions as expected. The full software performance benchmarks which measure map and waveform display refresh rates, data rates, and time to display a frame of map and waveforms were conducted. The performance of the EnSite Velocity System was {3}------------------------------------------------ observed during verification and integrated system testing to ensure performance was not severely impacted by the integration with the gMPS II (K102905). ### 12.11 Summary of Design Control Activities The development of the EnSite Velocity System v.1.5.1 was performed in accordance with SJM's Quality System requirements, and in compliance with Quality System Requlation design controls requirements documented in 21 CFR 820.30. A Declaration of Conformity with Design Controls follows in section 13.4. ### 12.12 Conclusion The EnSite Velocity v.1.5.1 has the same indications for use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity System v.1.5.1 are substantially equivalent to the predicate device. Where operational and performance differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness. Therefore, St. Jude Medical considers the EnSite Velocity System v.1.5.1 to be substantially equivalent to the predicate device. {4}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY - 5 2011 St. Jude Medical c/o Donna R. Lunak Regulatory Specialist One St. Jude Medical Drive Saint Paul MN 55117 Re: K110549 Trade/Device Name: EnSite Velocity System V.1.5.1 with MediGuide gMPS System Interface Module Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: March 29, 2011 Received: March 30, 2011 Dear Ms. Lunak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ Page 2 – Ms. Donna Lunak comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # 7 Indications for Use Device Name: EnSite Velocity System Indications for Use: The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. - When used with the EnSite Array™ Catheter, the EnSite System is intended to . be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR - When used with the EnSite NavX Surface Electrode Kit, the EnSite System is . intended to display the position of conventional electrophysiology (EP) catheters in the heart. Intended Use: When used with the v.1.5.1 MediGuide gMPS System Interface Module, the EnSite Velocity System interfaces to the MediGuide Guided Medical Positioning System II to combine and display gMPS II processed patient positioning and orientation Mapping information. | Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | |-------------------------------------------------------|---| | AND/OR Over-The-Counter Use<br>(21 CFR 801 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices | 510(k) Number | K110549 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...