K110520 · Tokyo Boeki Medisys, Inc. · CEM · Mar 23, 2012 · Clinical Chemistry
Device Facts
Record ID
K110520
Device Name
BIOLIS 12I
Applicant
Tokyo Boeki Medisys, Inc.
Product Code
CEM · Clinical Chemistry
Decision Date
Mar 23, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1600
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Biolis 12i is a discrete chemistry analyzer with ion-selective electrode (ISE), with direct quantitative measurement of sodium, potassium, chloride, and glucose in serum. It is a device intended for the in-vitro, spectrophotometric determination of general chemistry assays for clinical use. The Biolis 12i includes an optional lon Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. The Biolis 12i is not for Point-Of-Care testing. It is for vitro diagnostic use only. Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance. Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders. The Biolis 12i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic isle cell carcinoma.
Device Story
Biolis 12i is an automated, discrete, random-access clinical chemistry analyzer; processes serum samples for sodium, potassium, chloride, and glucose. System components include sampler, reagent/reaction trays, mixing units, cuvette washing, and spectrophotometer. ISE module uses potentiometry (ion-selective electrodes) for electrolytes; photometry (340 nm) for glucose hexokinase assay. Throughput: 90-100 tests/hour. Operated by laboratory personnel in clinical settings; not for point-of-care. Operator loads samples into cups on tray; system performs automatic aspiration and analysis. Results displayed to clinician for diagnosis/treatment of metabolic and electrolyte disorders. Benefits include automated, high-throughput quantitative analysis of patient serum.
Clinical Evidence
Bench testing only. Performance validated through method comparison studies against predicate devices (Sirrus and Prestige 24i). Correlation coefficients: Na (0.9872), K (0.9992), Cl (0.9922), Glucose (0.9975). Precision (CV%) for glucose within-run 1.04-1.31%, between-run 0.70-0.87%. ISE precision (CV%) within-run 0.61-0.94%, day-to-day 0.3-1.0%. Linearity ranges: Na 100-200 mmol/L, K 1-10 mmol/L, Cl 70-200 mmol/L, Glucose 25-540 mg/dL. No significant interference observed for bilirubin, hemoglobin, or lipemia at specified concentrations.
Technological Characteristics
Discrete, random-access analyzer. Photometric detection via photo-diode (340-800nm range) using tungsten halogen lamp. ISE module for potentiometry. Semi-disposable plastic cuvettes (8mm path length). Incubation at 37°C +/- 0.1°C. Throughput: 90-100 tests/hour. Connectivity: Analytical unit and control unit interface.
Indications for Use
Indicated for quantitative measurement of sodium, potassium, chloride, and glucose in human serum for diagnosis and treatment of electrolyte imbalances, metabolic disorders, diabetes mellitus, neonatal hypoglycemia, idiopathic glycemia, and pancreatic islet cell carcinoma. Not for point-of-care testing.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Predicate Devices
Tokyo Boeki Prestige 24i (ISE) (k040958)
Beckman CX-7 system (glucose) (k971467)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k110520
B. Purpose for Submission:
To obtain clearance for an automated clinical analyzer with ion selective electrode (ISE) and glucose reagent
C. Measurand:
Sodium, potassium, chloride and glucose
D. Type of Test:
Quantitative, photometry and potentiometry
E. Applicant:
Tokyo Boeki Medisys Inc.
F. Proprietary and Established Names:
Biolis 12i
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.2160, Discrete photometric chemistry analyzer for clinical use
21 CFR § 862.1665, Sodium test system
21 CFR § 862.1600, Potassium test system
21 CFR § 862.1170, Chloride test system
21 CFR § 862.1345, Glucose test system
2. Classification:
Class I for 21 CFR § 862.2160, Class II for all others
3. Product code:
JJE, JGS, CEM, CGZ, CFR
4. Panel:
75 (Clinical Chemistry)
{1}
H. Intended Use:
1. Intended use(s):
See indications for use statements below.
2. Indication(s) for use:
The Biolis 12i Clinical Chemistry analyzer is a discrete photometric chemistry analyzer with ion-selective electrode (ISE), with direct quantitative measurement of sodium, potassium, chloride, and glucose in serum. It is a device intended for the in-vitro, spectrophotometric determination of general chemistry assays for clinical use. The Biolis 12i includes an optional Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. The Biolis 12i is not for Point-Of-Care testing. It is for vitro diagnostic use only.
Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.
Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.
The Biolis 12i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Biolis 12i Clinical Chemistry Analyzer
I. Device Description:
The Biolis 12i is an automated clinical analyzer which uses a discrete, single line random access, multiple analysis method. The throughput is 90 tests/hour or 100
{2}
tests/hour with ISE synchronization. The system is composed of the sampler, sample delivery, reagent tray, reagent delivery, reaction tray, mixing units, cuvette washing unit, spectrophotometer and other sections.
The Ion-Selective Electrode (ISE) module includes a sodium electrode, a potassium electrode, a chloride electrode, a reference electrode, a cleaning solution and two calibrators (level 1 and 2). The sodium electrode membrane is a crown ether liquid-membrane. The potassium electrode membrane is a valinomycin liquid-membrane. The chloride electrode membrane is a quaternary ammonium salts polymer membrane. The ISE calibrator 1 and 2 contains the following chemicals: sodium, potassium, chloride, sodium formate, triethanolamine,
Glucose reagent kit contains 3x8 mL bottles of R1 reagent. The R1 reagent contains the following: ATP (1.1 mmol/L), NAD (2.7 mmol/L), magnesium (2 mmol/L), hexokinase (yeast) (>2,000 IU/L), G-6-PD (>4,000 IU/L), preservatives and stabilizers.
ISE calibrators have been previously cleared in k050958. Glucose calibrator has been previously cleared in k070202.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Tokyo Boeki Prestige 24i (ISE)
Beckman CX-7 system (glucose)
2. Predicate K number(s):
k040958, k971467
3. Comparison with predicate:
For ISE:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device (Prestige 24i, k040958) | Candidate Device |
| Intended Use/Indications for use | Intended to measure the concentration of the electrolytes, sodium, potassium and chloride in serum, using indirect potentiometry. Measurement of sodium, | Same |
{3}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device (Prestige 24i, k040958) | Candidate Device |
| | potassium and chloride in serum to is monitor electrolyte balance. | |
| Instrument | Prestige 24i | Biolis 12i |
| Instrument throughput and principle | 400 tests including ISE Discrete, random access, multi-test analysis | 100 tests including ISE Discrete, single-line random access, multi-test analysis |
| Test Principle | Ion Selective Electrode | Same |
| Reaction time | Maximum 10 minutes | Same |
| Analysis time | 100 seconds | Same |
| Measurands | Sodium, Potassium, Chloride | Same |
| Sample Type | Serum | Same |
For glucose:
| Item | Predicate device (Beckman CX-7, k971467) | Candidate device |
| --- | --- | --- |
| Similarities and Difference | | |
| Intended Use/Indications for use | For quantitative measurements of glucose. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma. | Same |
| Measured parameters | glucose | Same |
| Sample type | Serum | Same |
| Test principle | hexokinase | Same |
| Calibrations materials | Aqueous, 2 levels | Aqueous, 1 level |
{4}
5
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
There are two different measuring principles employed, potentiometry and photometry/colorimetry.
Potentiometry: A potential is recorded using a voltmeter, which relates to the concentration of the sample. A reference electrode is used to provide a stable, fixed potential against which other potential differences can be measured. This measurement technique is used for sodium, potassium, and chloride.
Photometry/colorimetry: The change of absorbance measured at 340 nm is monitored for the glucose reaction. The amount of absorbance change is proportional to the concentration of the glucose being present in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-run precision for ISE and glucose was determined on a single Biolis 12i using three serum based control materials. 20 replicates of each of the control level were evaluated with the following results:
| Analyte | Parameter | Control 1 | Control 2 | Control 3 |
| --- | --- | --- | --- | --- |
| Sodium | Mean (mmol/L) | 146.7 | 125.5 | 138.9 |
| | SD | 1.254 | 0.761 | 1.298 |
| | %CV | 0.85 | 0.61 | 0.93 |
| Potassium | Mean (mmol/L) | 2.81 | 5.79 | 4.20 |
| | SD | 0.025 | 0.053 | 0.039 |
| | %CV | 0.90 | 0.91 | 0.94 |
| Chloride | Mean (mmol/L) | 105.9 | 87.6 | 97.1 |
| | SD | 0.841 | 0.729 | 0.673 |
| | %CV | 0.79 | 0.83 | 0.69 |
| Glucose | Mean (mg/dL) | 61.9 | 207.9 | 364.0 |
| | SD | 0.81 | 2.48 | 3.80 |
| | %CV | 1.31 | 1.19 | 1.04 |
Day-to-day precision for ISE was determined from one run per day for 15 days on a single Biolis 12i. Day to day precision for glucose was determined from 5 different days with runs of five replicates on a single Biolis 12i. Both precision studies used three serum based control materials and results were
{5}
summarized below:
| Analyte | Parameter | Control 1 | Control 2 | Control 3 |
| --- | --- | --- | --- | --- |
| Sodium | Mean (mmol/L) | 147.1 | 125.4 | 139.5 |
| | SD | 0.534 | 0.518 | 0.442 |
| | %CV | 0.4 | 0.4 | 0.3 |
| Potassium | Mean (mmol/L) | 2.81 | 5.79 | 4.23 |
| | SD | 0.023 | 0.049 | 0.016 |
| | %CV | 0.8 | 0.9 | 0.4 |
| Chloride | Mean (mmol/L) | 105.9 | 87.6 | 97.5 |
| | SD | 1.036 | 0.587 | 0.385 |
| | %CV | 1.0 | 0.7 | 0.4 |
| Glucose | Mean (mg/dL) | 65.7 | 221.1 | 382.2 |
| | SD | 0.57 | 1.73 | 2.67 |
| | %CV | 0.87 | 0.87 | 0.70 |
# b. Linearity/assay reportable range:
A linearity study for ISE was performed using a high serum sample mixed with a reference solution (with known conc. of NaCl or KCl) to create various target concentrations of analytes. 6 levels of each of the analytes measured were prepared and tested in triplicate on one Biolis 12i analyzer. A linearity study for glucose was performed using commercially available serum linearity standards with target range from 25 to $725\mathrm{mg / dL}$ . 6 levels of different concentrations of glucose were measured in triplicate on one Biolis 12i analyzer. The observed values were plotted against the expected values and an appropriate line fitted by standard linear regression. Results were summarized below:
| | Sodium | Potassium | Chloride | Glucose |
| --- | --- | --- | --- | --- |
| Correlation (r) | 0.9996 | 0.9995 | 0.9984 | 0.9997 |
| Slope | 1.0913 | 0.9710 | 0.9667 | 0.9633 |
| Intercept | -9.1421 | 0.1184 | 6.9739 | -2.6641 |
| Range | 100-200 (mmol/L) | 1-10 (mmol/L) | 70-200 (mmol/L) | 25-725 (mg/dL) |
| Levels tested | 100, 120, 140, 160, 180, 200 (mmol/L) | 1.0, 2.3, 5.0, 7.0, 8.5, 10.0 (mmol/L) | 70, 100, 125, 150, 175, 200 (mmol/L) | 25, 200, 375, 550, 725 (mg/dL) |
| % recovery range | 96% to 101% | 97 to 103% | 96 to 101% | 100 to 109.3% |
{6}
Results of the study support the sponsor claims for the following measuring/linearity ranges:
| Sodium | Potassium | Chloride | Glucose |
| --- | --- | --- | --- |
| 100 – 200 mmol/L | 1 – 10 mmol/L | 70 – 200 mmol/L | 25 – 540 mg/dL |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
All the analytes in the calibration solutions are traceable to a reference method and the analyte targets are assigned based on the reference methods or other commercially available methods. See table below for traceability.
| Analyte | Traceability |
| --- | --- |
| Sodium | Flame photometry |
| Potassium | Flame photometry |
| Chloride | Coulometric titration |
| Glucose | NIST SRM 965b |
ISE calibrators- Calibration solutions 1 and 2 had been previously cleared in k040958 (Toyko Boeki Prestige 24i)
Glucose reagents- previously cleared in k971467 (Carolina Liquid Chemistries)
Stability of the ISE calibration solutions: A real-time stability study was performed to demonstrate a maximum shelf life of 18 months at storage temperatures between 15 and 30 degrees C and an open vial stability of 31 days after first opened. Protocols and acceptance criteria has been provided and found to be adequate.
ISE calibrators have been previously cleared in k050958. Glucose calibrator has been previously cleared in k070202.
d. Detection limit:
The sponsor determined that the detection limit was defined by the linear range study. The minimum detection limit is defined as the lowest analyte concentrations over the range that the assay has been shown to be acceptably linear by using serum control samples at 6 different concentrations and analyzed by comparing results with the theoretical values. The stated measuring ranges of the candidate device are as follows. These ranges were verified by the linearity study (see section M.1.b. above).
| Sodium | Potassium | Chloride |
| --- | --- | --- |
| 100 – 200 mmol/L | 1 – 10 mmol/L | 70 – 200 mmol/L |
In addition, the sponsor performed a limit of detection study for the glucose
{7}
assay according to a modified protocol based on the CLSI EP17-A guideline. LoB was determined by running 5 low samples multiple times on 2 different lots of reagents with one instrument to generate 80 measurements. LoD and LoQ were determined by running 4 low samples four times in two runs with two different lots of reagents on one instrument generating a total of 64 measurements. LoB was calculated to be 2.48 mg/dL and LoD was calculated to be 3.81 mg/dL. LoQ was determined to be 7.83 mg/dL based on inter-assay precision of 9.3%.
The sponsor claimed that the glucose assay has a measuring range of 25 to 540 mg/dL.
e. Analytical specificity:
An interference study was conducted to evaluate the potential endogenous and exogenous interfering substances. For endogenous interfering substances, the sponsor tested intralipid, bilirubin, and hemoglobin. For exogenous interferencing substances, the sponsor tested lithium, bromide, salicylate, and thiocyanate. Interference studies were performed by spiking serum pool sample at clinically significant medical decision points for each measurand at various concentrations of interfering compounds normally found in serum. Measurements (N=3) obtained from serum containing each potential interfering substance was evaluated and compared against measurements with the unspiked samples. The study was conducted for the analytes Na, K, and CL, and glucose at two analyte concentrations (normal and abnormal). The sponsor states that interferences are considered to be non-significant if the bias between the tested and control samples are within ±10% for all the analytes.
For endogenous interfering substances, the results demonstrate no significant interference with the sodium, potassium, and chloride when serum contains up to 19.7 mg/dL of unconjugated bilirubin, 21 mg/dL of conjugated bilirubin, 500 mg/dL of intralipid, or 488 mg/dL hemoglobin, however potassium was affected by hemolysis. For glucose, the results demonstrate no significant interference when serum contains up to 20 mg/dL bilirubin, 500 mg/dL hemoglobin, and 1000 mg/dL of Intralipid.
For exogenous interfering substances, the results demonstrate no significant interference with the Na, K, and Cl tests when serum contains up to 3.20 mmol/dL lithium, 37.60 mmol/dL bromide, 4.34 mmol/dL salicylate, and 6.88 mmol/L thiocyanate.
Based on the hemolysis interference, the sponsor has the following limitations in their labeling:
{8}
"Do not use hemolyzed samples for potassium since significant hemolysis may increase K concentration because of high levels of K in erythrocytes".
"Drugs and other substances may affect sodium, potassium, chloride, and glucose determinations. See Young, D.S.¹ for a compilation of reported interferences"
1. Young, D.S., Effects of Drugs on Clinical Laboratory Tests, third edition, AACC Press, Washington D.C., 1990.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
i.) For ISE method comparison study:
A total of 140 patient serum samples were assayed on one Biolis 12i (candidate device) and one Prestige 24i (predicate device) analyzer at Tokyo Boeki Medical System's laboratory. Some samples were diluted and spiked to cover the hard-to-find sample range (no more than 20%). The linear regression results are as follows:
Sodium:
| Correlation Coefficient | 0.9872 |
| --- | --- |
| Slope | 1.0204 |
| Intercept | -3.1903 |
| Sample range tested | 100 – 195.9 mmol/L |
Potassium:
| Correlation Coefficient | 0.9992 |
| --- | --- |
| Slope | 1.0255 |
| Intercept | -0.1457 |
| Sample range tested | 1.0 – 10.0 mmol/L |
Chloride:
| Correlation Coefficient | 0.9922 |
| --- | --- |
| Slope | 0.9736 |
| Intercept | 1.8580 |
| Sample range tested | 70 - 200 mmol/L |
{9}
ii.) For glucose method comparison study:
A total of 124 patient serum samples were evaluated using the candidate device (Biolis 12i) and the predicate device (Beckman CX7). Some samples were spiked and diluted to obtain the hard-to-find sample range $(13\%)$ . The linear regression results were summarized below:
Glucose:
| Correlation Coefficient | 0.997 |
| --- | --- |
| Slope | 1.045 |
| Intercept | -4.858 |
| Sample range tested | 29-540 mg/dL |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Reference values may vary with age, sex, diet, geographic location and instrumentation. Good laboratory practice dictates that each laboratory establishes its own expected values to reflect its patient population. Suggested guidelines are as follows:
{10}
Sodium:
Adult: 136 – 145 mmol/L
Kathleen Deska Pagana and Timothy James Pagana, Mosby’s Diagnostic and Laboratory Test Reference, 2nd ed. (St. Louis: Mosby, 1995), p 744.
Potassium:
Adult: 3.5 – 5.0 mmol/L
Kathleen Deska Pagana and Timothy James Pagana, Mosby’s Diagnostic and Laboratory Test Reference, 2nd ed. (St. Louis: Mosby, 1995), p 638.
Chloride:
Adult/elderly: 90 – 110 mmol/L
Kathleen Deska Pagana and Timothy James Pagana, Mosby’s Diagnostic and Laboratory Test Reference, 2nd ed. (St. Louis: Mosby, 1995), p 197.
Glucose: 70-105 mg/dL
Tietz, N. W., Clinical Guide to Laboratory Tests, W.B. Saunders Co., Philadelphia, PA, 1983.
N. Instrument Name:
Biolis 12i
O. System Descriptions:
1. Modes of Operation:
Single sample
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Manual sample identification or Barcode reader
4. Specimen Sampling and Handling:
Automatic sample aspiration: samples need to be pour over into a sample cup and put onto the sample tray.
11
{11}
5. Calibration:
Operator must initiate a calibration sequence of the calibration solutions (2 levels) at the start of the morning and every 24 hours subsequently.
6. Quality Control:
Control materials are supplied by Microgenics, Chem Trak H, 3 levels of serum based liquid controls (k030942).
Operator must initiate an external quality control procedure the quality control materials. Recommendations for frequency of testing are including in the labeling. As a minimum, users are to follow regulatory guidelines for testing quality control materials.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
To assess the effect of temperature on test results across the recommended range of ambient temperature (15 – 30°C), the sponsor conducted an ambient temperature study using 30 native patient samples. Calibrations were performed at each of these temperatures (15°C, 23°C, and 30°C) and testing was performed at all the three temperatures for each calibration temperature. Protocols and acceptance criteria provided are found to be adequate. Based on the ambient temperature study, the sponsor claimed that ambient temperature between 15 - 30°C will not affect the performance of the device.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.