This device is intended to measure/ display collected urine volume. It detects fill level and temperature.
Device Story
iBag™ is a urine management system for catheterized patients; provides real-time urine volume and bladder temperature data. System includes a calibrated drainage bag integrated with a wireless sensor (Wi-Mitter™). Wi-Mitter™ transmits data via 418 MHz wireless signal to a base unit (Netbook PC) via USB. Data is time/date stamped; system calculates fill rates and triggers threshold-based alarms. Alerts (email/text) notify patients, caregivers, or family members. Used in clinical or home settings; operated by patients or caregivers. Output displayed on laptop/monitors to assist in clinical decision-making regarding patient hydration and bladder health.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), FCC compliance, RF dosimetric assessment, transmitter range, fill level accuracy (±5% for 70-100% fill), temperature measurement accuracy (±0.2 °F), battery life, leak testing, and sterility testing. A user feedback study and simulated use testing were also conducted.
Technological Characteristics
System consists of a calibrated drainage bag, wireless Wi-Mitter™ transmitter, and PC-based display. Power: 3V lithium coin-cell (CR2032) for transmitter; 120 VAC for display unit. Connectivity: 418 MHz wireless link to USB-connected PC. Temperature sensing: 50-110 °F range, 0.1 °F increments, ±0.2 °F accuracy. Volume sensing: 1000/2000 ml capacity, ±5% accuracy at 70-100% fill. Software: Threshold-based alarming and data logging.
Indications for Use
Indicated for catheterized patients requiring urine volume and bladder temperature monitoring.
Regulatory Classification
Identification
A urine collector and accessories is a device intended to collect urine. The device and accessories consist of tubing, a suitable receptacle, connectors, mechanical supports, and may include a means to prevent the backflow of urine or ascent of infection. The two kinds of urine collectors are:(1) A urine collector and accessories intended to be connected to an indwelling catheter, which includes the urinary drainage collection kit and the closed urine drainage system and drainage bag; and (2) A urine collector and accessories not intended to be connected to an indwelling catheter, which includes the corrugated rubber sheath, pediatric urine collector, leg bag for external use, urosheath type incontinence device, and the paste-on device for incontinence.
Special Controls
*Classification* —(1)*Class II (special controls) for a urine collector and accessories intended to be connected to an indwelling catheter.* The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.(2)
*Class I (general controls).* For a urine collector and accessories not intended to be connected to an indwelling catheter, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
{0}------------------------------------------------
510(k) Summary 510(k) Number K110518 Future Path Medical, LLC 3020 Hinsel Rd Columbus, Ohio 43232 Office: 614-863-5651 Fax: 614-573-6636 Date Prepared: April 18. 2011 Contact: Ty Bryant, Chairman
- l . Identification of the Device: Proprietary-Trade Name: iBag™ Urine Management System Device: Urinometer, electrical Regulation Description: Urine flow or volume measuring system. Regulation Medical Specialty: Gastroenterology/Urology Review Panel: Gastroenterology/Urology Product Code: EXS Submission Type: 510(K) Exempt subject to 21CFR876.9 Regulation Number: 876.1800 Device Class: 2
- 2. Equivalent legally marketed devices: BARD CritiCore (K924436) Bard Urological 510(k); UROTRACK II Monitoring System
- 3. Indications for Use (intended use). This device is intended to measure/ display collected urine volume. It detects fill level and temperature.
- 4. Description of the Device: iBag™ is a urine management system that includes bladder and temperature measurement for catheterized patients who desire to be fully mobile. · iBag™ provides accurate, real-time assistive data. It does this through use of a calibrated drainage bag integrated with a Future Path Medical sensor, a base transmitter unit called Wi-Mitter™ for wireless transmission of urine bag fill and bladder temperature information. Once the Wi-Mitter™ is snapped in place, the container is polled and the measurement of fluid level begins and this data is then transmitted to the patient, the care provider (an aid) a family member, or even to the hospital's base station. Data is transmitted via traditional data networking and may be interpreted and displayed on a user's laptop computer, and other types monitors. Since the data from the Wi-Mitter™ is time and date stamped, time of fill level may be automatically calculated, noted and threshold values used for fill level alarming. Features:
- Automatically measures and displays urine fill level and bladder temperature.
- · Re-usable wireless transmitter is inexpensive and has a one-year battery life
- · Performs functions such as email or text alerts on fill levels.
- · Software is designed to be used with little or no training.
- · Transmits data in real time.
- 5. Safety and Effectiveness, Summary of Testing and Comparison to Predicate Device. Bench and test laboratory results indicate that the new device is as safe and effective as the predicate device. The following testing has been successfully conducted:
JUN 2 4 2011
{1}------------------------------------------------
- Electrical Safety to IEC 60601-1 ●
- Electromagnetic Compatibility to IEC 60601-1-2 .
- FCC Compliance Testing .
- . RF Dosimetric Assessment Testing
- . Transmitter effective range.
- User Feedback Study .
- Fill level accuracy .
- Temperature measurement accuracy .
- . Battery life computations
- Simulated Use Testing .
- Leak Testing .
- Sterility Testing .
## 6. Substantial Equivalence Chart
| Characteristic | BARD CritiCore (K924436) Bard<br>Urological 510(k): UROTRACK II<br>Monitoring System | iBag™ Urine Management<br>System |
|----------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | This device is intended to measure/<br>display collected urine volume. It<br>detects fill level and temperature. | SAME |
| Configuration | Portable unit with integrated display | Wireless communication (418<br>mHz) to Netbook type PC via<br>USB |
| Power Source | 6 "D" cell alkaline batteries | 3 volt coin-cell size, lithium<br>battery. (e.g., CR2032); 120<br>VAC for the Netbook display<br>unit. |
| Urine Bag Capacity | 0-2100 ml | 1000 and 2000 ml |
| Volume Accuracy | Not stated | ±5% of the actual fill for 70 –<br>100% fill level readings. |
| Temperature | 50°F to 110°F (10°C to 43.3°C) | Bladder core temperature is<br>measured in degrees Fahrenheit,<br>°F, units from 50 to 110 °F in<br>increments of 0.1 °F. |
| Temperature accuracy | ± 0.2 °F. | Temperature measurement<br>accuracy is ± 0.2 °F. |
| Electrical safety | Electrical Safety per IEC-60601. UL<br>listed | SAME |
## 7. Conclusion
After analyzing bench, simulated use, and compliance to standards testing using external laboratory testing to applicable standards, it is the conclusion of Future Path LLC that the iBag™ Urine Management System is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60" Silver Spring, MD 20993-0002
Future Path Medical. LLC c/o Daniel Kamm, P.E. Principal Consultant .Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114
JUN 2 4 2011
· Re: K110518
Trade/Device Name: iBag™ Urine Management System Regulation Number: 21 CFR \$876.5250 Regulation Name: Urine Collector and Accessories Regulatory Class: Class II Product Code: KNX and EXS Dated: June 12, 2011 Received: June 16, 2011
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability i warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Par1 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
CAN
Sincerely yours,
Huchert Hemer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K105 G
Device Name: iBag™ Urine Management System
Indications For Use: This device is intended to measure/ display collected urine volume. It detects fill level and temperature.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hogm 722
(Division Sign-Oft)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K110518
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.