CONNEX VITAL SIGNS MONITOR 6000 SERIES

K110516 · Welch Allyn, Inc. · MWI · Jun 1, 2011 · Cardiovascular

Device Facts

Record IDK110516
Device NameCONNEX VITAL SIGNS MONITOR 6000 SERIES
ApplicantWelch Allyn, Inc.
Product CodeMWI · Cardiovascular
Decision DateJun 1, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Intended Use

The VSM 6000 Series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of neonatal, pediatric, and adult patients for: - noninvasive blood pressure. - pulse rate, - noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and - body temperature in normal and axillary modes. The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. The optional Masimo Rainbow SET® and accessories are indicated for the continuous noninvasive monitoring of total hemoglobin concentration of adult, pediatric, and neonatal patients during both motion and no motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities. Optional compatible weight scales (e.g., Health o meter®) can be used for height, weight, and BMI input. This product is available for sale only upon the order of a physician or licensed health care professional.

Device Story

Connex® Vital Signs Monitor 6000 Series is a modular, scalable system for spot-check and continuous patient monitoring. Inputs include noninvasive blood pressure, pulse rate, SpO2, body temperature, and total hemoglobin (via integrated Masimo Rainbow SET® module); device also accepts manual or electronic data entry (weight, height, BMI) from external accessories (e.g., Health o meter scales). Device features touch screen interface, silicone light bar for alarms, and internal battery. Used in general medical/surgical floors and hospital environments by clinicians. Data transferred via USB, Ethernet, or wireless. Output displayed on screen to assist clinicians in patient assessment and decision-making. Benefits include consolidated vital signs and patient data management.

Clinical Evidence

No clinical studies were utilized. Safety and effectiveness were established through bench testing, including verification and validation of the Masimo Rainbow SET® integration and external scale data transfer. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, IEC 60601-2-30, ISO 9919, and ISO 14971 standards was demonstrated.

Technological Characteristics

Enclosure: engineering plastics with internal steel members. Interface: touch screen, silicone alarm light bar. Connectivity: USB, wired Ethernet, wireless. Power: internal AC supply, Lithium Ion battery. Integration: Masimo Rainbow SET® module for SpHb; external accessory interface for weight/height/BMI. Standards: IEC 60601-1, 60601-1-2, 60601-1-4, 60601-1-8, 60601-2-30, ISO 9919, ISO 14971.

Indications for Use

Indicated for neonatal, pediatric, and adult patients in hospital/alternate care settings for monitoring noninvasive blood pressure, pulse rate, SpO2, body temperature, and (via optional module) total hemoglobin concentration. Compatible with external weight/height/BMI scales. Prescription use only.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K11051.6 JUN - 1 2011 510(k) Summary [As described in 21 CFR 807.92] Submitted by: Contact Person: Welch Allyn Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 Fred Schweitzer Vice President, QA/RA Phone: (315) 554-4001 Fax: (315) 685-2532 E-mail: schweitzerf@welchallyn.com Welch Allyn Connex® Vital Signs Monitor 6000 Series Monitor, Physiological, Patient (without Arrhythmia Class II, 870.2300, Cardiovascular Monitoring Devices Welch Allyn Vital Signs Monitor -- VSM 6000 Series Masimo Rainbow SET® Radical 7 Pulse CO-Oximeter Patient scales marketed pursuant to 21 CFR 880.2700 and 880.2720, Class I, 510(k) Exempt (e.g. Health o meter) Vital Signs Monitor, CVSM 6000 Series, CVSM Masimo Rainbow SET® Radical 7R Pulse CO- Feb 18, 2011 Detection or Alarms) Product Code - MWI Welch Allyn, Inc. and Accessories Masimo Corporation 510(k) Number K080238 Masimo Corporation 510(k) Number K100428 510(k) Number K101445 Pulse Oximeter and Sensor Oximeter and Accessories Pulse Oximeter and Sensor Date Prepared: Trade Name: Common Name: Classification Reference: Predicate Devices: Special 510(k) – Connex® Vital Signs Monitor 6000 Series Page 1 of 4 {1}------------------------------------------------ ## Welch Allyn: ### Description of the Device: The Welch Allyn Connex® Vital Signs Monitor 6000 Series is designed to provide a scalable, modular system of components that can be configured to address the needs for vital signs spot check and continuous monitoring. The CVSM 6000 Series is intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), Total Hemoglobin (SpHb and SpHbv), and body temperature in normal and axillary modes of neonatal, pediatric, and adult patients. The CVSM can also display and transfer patient data that is electronically or manually entered from external and accessory devices, e.g., weight and height data, barcode, IR temperature, and other patient or facility information. Data can be transferred electronically via USB, wired Ethernet, or wireless communications. The monitor has an enclosure constructed of engineering plastics with internal steel members for strengthening. A silicone light bar is prominent in a carry handle on the top of the device and illuminates for different alarm conditions. A power button is located on the side of the device. A touch screen display is prominent on the front of the device and provides the primary interface for the user to interact with the device. Internal and external communications are primarily by USB. External host USB connections for accessories are tool accessible. A USB connection for data transfer is on the side of the device, as is a connection to an internal relay for use with nurse call systems. The device contains an internal AC power supply for operating the device and charging the internal Lithium Ion battery. The monitor can be configured for use in different workflows including desktop, affixed to a mobile stand, or on a wall mount. #### Indications for Use: Indications for Use: The VSM 6000 Series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of neonatal, pediatric, and adult patients for: - noninvasive blood pressure. - pulse rate, - noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and - l body temperature in normal and axillary modes. The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. {2}------------------------------------------------ # Welch Allyn The optional Masimo Rainbow SET® and accessories are indicated for the continuous noninvasive monitoring of total hemoglobin concentration of adult, pediatric, and neonatal patients during both motion and no motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities. Optional compatible weight scales (e.g.,Health o meter®) can be used for height, weight, and BMI input. This product is available for sale only upon the order of a physician or licensed health care professional. #### Technological Characteristics: The subject device has the same technological characteristics and indications for use as the predicate devices; minor modifications and additions to software and hardware were made to the CVSM to enable acquisition and/or display of total hemoglobin utilizing the existing Masimo Rainbow SET® module and the transfer and display of weight, height, and BMI from OEM devices (e.g., Health o meter®). #### Non-Clinical Tests: Verification and validation were conducted to ensure expected performance of the CVSM 6000 Series with enabled Masimo Rainbow SET® and data transfer from accessory patient scales. The Connex® Vital Signs Monitor 6000 Series was tested to evaluate its safety and effectiveness based on the following standards: - . IEC 60601-1:Ed. 2: 1988 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (with A1: 1991+A2:1995) - . IEC 60601-1-2: Ed. 3: 2007 - Medical Electrical Equipment - Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility -Requirements and Test - IEC 60601-1-4: Consolidated Ed. 1.1: 2000 General Requirement for Safety: . Collateral Standard: Programmable Electrical Medical Systems - . IEC 60601-1-8: Ed. 1: 2003 - General requirements for safety - Collateral Standard: Alarm Systems - Requirements, tests and guidances - General requirements and guidelines for alarm systems in medical equipment (with A1:2006) - IEC 60601-2-30: Ed. 2: 1999 Manual, electronic or automated sphygmomanometers . (with A1:2003, A2:2006) - ISO 9919: Ed. 2: 2005 - Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use - . ISO 14971: Ed. 2: 2007 - Medical devices - Application of risk management to medical devices {3}------------------------------------------------ # Welch Allyn #### Clinical Performance Data: No clinical studies were utilized for the purpose of obtaining safety or effectiveness data. ### Conclusion: Based on the information presented in this 510(k) premarket notification, Welch Allyn's Connex® Vital Signs Monitor 6000 Series is considered substantially equivalent (as safe, as effective and performs as well as) the currently marketed devices cited in this submission. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN - 1 2011 Welch Allyn, Inc. c/o Mr. Fred Schweitzer Vice President, Regulatory Affairs 4341 State Street Road Skaneateles Falls, NY 13153-0220 Re: K110516 Trade/Device Name: Connex® Vital Signs Monitor 6000 Series, with models 6300, 6400 and 6500 Regulatory Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (Two) Product Code: MWI Dated: May 5, 2011 Received: May 6, 2011 Dear Mr. Schweitzer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {5}------------------------------------------------ Page 2 - Mr. Fred Schweitzer found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fdai.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K (( 0 د ک Device Name: Welch Allyn Connex® Vital Signs Monitor 6000 Series Indications for Use: The VSM 6000 Series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of neonatal, pediatric, and adult patients for: - noninvasive blood pressure. ー - pulse rate, l - noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and - - body temperature in normal and axillary modes. The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. The optional Masimo Rainbow SET® and accessories are indicated for the continuous noninvasive monitoring of total hemoglobin concentration of adult, pediatric, and neonatal patients during bath motion and no motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities. Optional compatible weight scales (e.g., Health o meter®) can be used for height, and BMI input. This product is available for sale only upon the order of a physician or licensed health care professional. | Prescription Use<br>(Part 21 CFR 801 Subpart D) | <b>X</b> AND/OR | |-------------------------------------------------|-----------------| |-------------------------------------------------|-----------------| Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) *(Division Sign-Off)* (Division Sign-Off) Division of Cardiovascular Devices K110516 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...