← Product Code [EMA](/productcode/EMA) · K110508

# SUGLUE-10 (K110508)

_3M Espe AG · EMA · Jun 3, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110508

## Device Facts

- **Applicant:** 3M Espe AG
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Jun 3, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis) Final cementing of posts and screws . Final cementation of all-ceramic, or composite veneers ● Final cementation of all-ceramic, composite, or metal restorations on implant . abutments

## Device Story

Suglue-10 is a dual-curing, resin-based dental cement supplied in an automix syringe. Used by dental professionals for adhesive cementation of indirect restorations (inlays, onlays, crowns, bridges, veneers, posts, screws, implant abutments). Operates via chemical and light-curing mechanisms; contains bi-functional (meth)acrylate and inorganic fillers (43% volume). Applied in combination with an adhesive (e.g., Adhesive EXL 759) using either 'Total Etch' or 'Self-Etch' procedures. Provides mechanical fixation and adhesive bonding to tooth structure or implant abutments, improving restoration retention and clinical outcomes.

## Clinical Evidence

Bench testing only. Performance evaluated against ISO 4049, ISO 9917, and ISO 2039-1 standards. Metrics included film thickness, working/setting time, radiopacity, flexural strength, compressive strength, and surface hardness. Results for Suglue-10 were comparable to the predicate device Malta (K100756). Biocompatibility testing was performed.

## Technological Characteristics

Dual-curing, resin-based dental cement. Composition: bi-functional (meth)acrylate, ~43% inorganic fillers by volume (D90% ~13 µm). Mixing ratio: 1:1 base to catalyst. Form factor: automix syringe. Standards: ISO 4049 (film thickness, working/setting time, radiopacity, flexural strength), ISO 9917 (compressive strength), ISO 2039-1 (surface hardness).

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- Malta by 3M ESPE, Germany ([K100756](/device/K100756.md))
- Panavia F 2.0 by Kuraray Medical Inc., Japan ([K032455](/device/K032455.md))
- Nexus 3 by Kerr Corporation, U.S.A. ([K062519](/device/K062519.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS: K110508

## Submitter

JUN - 3 2011

| Company:                          | 3M ESPE AG                                              |
|-----------------------------------|---------------------------------------------------------|
| Street:                           | ESPE Platz                                              |
| ZIP-Code, City:                   | D-82229 Seefeld                                         |
| Federal State:                    | Bavaria                                                 |
| Country:                          | Germany                                                 |
| Establishment Registration Number | 9611385                                                 |
| Official Correspondent:           | Dr. Desi W. Soegiarto,<br>Regulatory Affairs Specialist |
| Phone:                            | +49-8152-700 1169                                       |
| Fax:                              | +49-8152-700 1869                                       |
| E-mail:                           | desi.soegiarto@mmm.com                                  |
| Date:                             | May 04, 2011                                            |

### Name of Device

………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

| Proprietary Name:    | Suglue-10                            |
|----------------------|--------------------------------------|
| Classification Name: | Dental cement other than zinc oxide- |
|                      | eugenol                              |
| Common Name:         | Adhesive resin cement                |

## Predicate Devices

| Malta by 3M ESPE, Germany                    | K100756 |
|----------------------------------------------|---------|
| Panavia F 2.0 by Kuraray Medical Inc., Japan | K032455 |
| Nexus 3 by Kerr Corporation, U.S.A.,         | K062519 |

{1}------------------------------------------------

K110508

### Description for the Premarket Notification

Suglue-10 is classified as a Dental cement other than zinc oxide-eugenol (21 C.F.R. §872.3275 [b]) because it is a device composed of various materials other than zinc oxide-eugenol.

Suglue-10, manufactured by 3M ESPE, is a dual-curing resin cement supplied in an automix syringe. It is intended to be used for the adhesive cementation of indirect restorations and will be available in various shades.

Suglue-10 is intended to be used in combination with an adhesive, e.g. Adhesive EXL 759 (by 3M ESPE). The Adhesive EXL 759 can be used either as a "Total Etch" or "Self-Etch" procedure. The additional etching of the tooth structure increases the adhesive strength of the adhesive even further.

Suglue-10 contains bi-functional (meth)acrylate. The proportion of inorganic fillers is about 43% by volume; the grain size (D 90%) is about 13 µm. The mixing ratio, based on volume, is 1 part base paste : 1 part catalyst.

Predicate devices to which Suglue-10 has been compared are Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

#### Technological Characteristics

As its predicate devices, Suglue-10 is a dual-curing, resin-based cement. Suglue-10 cement is based on the Malta material (K100756, by 3M ESPE, Germany). As Suglue-10 is designed to compete with the latest state of the art adhesive resin cements, its in vitro performance was compared to the newer system Nexus 3 and the still widely used Panavia F2.0.

The intended use of Suglue-10 is comparable to the area of the intended use of the predicate devices of Suglue-10.

In this 510(k) premarket notification Suglue-10 has been compared to its predicate devices with regard to chemical composition, performance data and indications for use.

Biocompatibility testing was carried out.

{2}------------------------------------------------

K110508

## Performance Data

The following table shows the performance data of Suglue-10 and its predicate device Malta (K100756):

|                         |             |            |          | Suglue-10    | Malta        |
|-------------------------|-------------|------------|----------|--------------|--------------|
|                         |             | Method     | Limit    | Results      |              |
| Film thickness          |             | ISO 4049   | < 50 μm  | $10 \pm 2$   | $13 \pm 1$   |
| Working time            |             | ISO 4049   | > 60 sec | pass         | pass         |
| Setting time            |             | ISO 4049   | < 10 min | 03:30        | 03:30        |
| Radiopacity             |             | ISO 4049   | > 1.0 mm | 2.3          | 1.8          |
| Flexural<br>strength    | dark cured  | ISO 4049   | > 50 MPa | $77 \pm 5$   | $84 \pm 12$  |
|                         | light cured |            | > 50 MPa | $111 \pm 7$  | $111 \pm 16$ |
| Compressive<br>strength | dark cured  | ISO 9917   | na [MPa] | $278 \pm 9$  | $263 \pm 9$  |
|                         | light cured |            | na [MPa] | $286 \pm 36$ | $256 \pm 36$ |
| Surface<br>hardness     | dark cured  | ISO 2039-1 | na [MPa] | $187 \pm 12$ | $190 \pm 20$ |
|                         | light cured |            | na [MPa] | $217 \pm 83$ | $212 \pm 30$ |

In summary, it can be concluded that Suglue-10 is as safe and effective as the predicate devices: Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455), and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

#### Indications for Use:

- · Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis)
- Final cementing of posts and screws .
- Final cementation of all-ceramic, or composite veneers ●
- Final cementation of all-ceramic, composite, or metal restorations on implant . abutments

{3}------------------------------------------------

K110508

# Conclusion of Substantial Equivalence

The comparison for chemistry, performance data and indications for use shows that Suglue-10 is substantially equivalent to the predicate devices: Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Desi W. Soegiarto 3M ESPE AG ESPE Platz Seefeld, Bavaria Germany D-82229

Re: K110508

Trade/Device Name: Suglue-10 Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: May 16, 2011 Received: May 23, 2011

JUN - 3 2011

Dear Ms. Soegiarto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Ms. Soegiarto

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{6}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known):

Suglue-10 Device Name:

Indications For Use: • • Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis)

- Final cementing of posts and screws .
- Final cementation of all-ceramic, or composite veneers .
- Final cementation of all-ceramic, composite, or metal restorations on . implant abutments

× Prescription Use . {Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Dunne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of __1

510(k) Number: K110508

---

**Source:** [https://fda.innolitics.com/device/K110508](https://fda.innolitics.com/device/K110508)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
