← Product Code [KPS](/productcode/KPS) · K110507

# CARDIAC SCANNER AND PROCESSING SYSTEM (K110507)

_Spectrum Dynamics (Israel) , Ltd. · KPS · Jul 13, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110507

## Device Facts

- **Applicant:** Spectrum Dynamics (Israel) , Ltd.
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Jul 13, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The D-SPECT™ Cardiac Scanner System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. The D-SPECT™ Cardiac Scanner System is primarily intended for cardiac applications. The D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40 -170 keV. The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and is featuring low count rate (low dose) acquisition, with no lose in image quality. The D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion of CCT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).

## Device Story

D-SPECT™ Cardiac Scanner System performs myocardial perfusion imaging; utilizes detector head, gantry, and patient chair; detector modules rotate within closed head; no external moving parts surrounding patient. System uses RFID tags on patient wrist for positive identification of patient and radiopharmaceutical. Acquisition station console controls operation. Processing workstation performs data reconstruction, archiving, display, communication (DICOM), and analysis. Includes automated review and quantification software (QPS, QGS, QBS). Used in clinical settings by healthcare professionals. Benefits include shorter scan times and low-dose acquisition without loss of image quality; automated tools assist clinicians in cardiac SPECT data interpretation and decision-making.

## Clinical Evidence

No clinical data provided; substantial equivalence supported by technological comparison and bench testing of system modifications.

## Technological Characteristics

Emission computed tomography system; detector head with rotating internal modules; supports 40-170 keV radionuclides. Connectivity via NEMA DICOM protocol for PACS integration. Includes RFID-based patient/agent identification. Software includes automated quantification algorithms (QPS, QGS, QBS).

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- D-SPECT™ Cardiac Scanner System ([K072468](/device/K072468.md))
- eNTEGRA workstation ([K003264](/device/K003264.md))
- AutoQUANT ([K980715](/device/K980715.md))
- WBR Half Dose ([K080784](/device/K080784.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for Spectrum Dynamics. The logo consists of a triangular shape on the left and the words "SPECTRUM DYNAMICS" on the right. Below the words "SPECTRUM DYNAMICS" is the phrase "HIGH SPEED FUNCTIONAL IMAGING".

JUL 1 3 2011

110567

## 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92

## Submitter Information

Spectrum Dynamics (Israel) Ltd. 22 Bareket Street North Industrial Park POB 3033 Caesarea 30889, Israel Tel: (+972) 73 7374500, Fax: (+972) 73 737450 Submission contact person: Ilan Sharon P.O.Box 4262, Zichron Yaacov 30900, Israel TEL: (+972) 52 8704904, Fax: (+972) 73 737450

#### Device Classification

|  | Proprietary Device Name:   | D-SPECT™ Cardiac Scanner System                                                               |
|--|----------------------------|-----------------------------------------------------------------------------------------------|
|  | Common name:               | Emission Computed Tomography System                                                           |
|  | Product Code:              | KPS (subsequent code LLZ)                                                                     |
|  | Classification Name:       | Emission computed tomography system (Subsequent – Picture archiving and communications system |
|  | Classification Regulation: | 21 CFR §892.1200 (subsequent 21 CFR §892.2050)                                                |
|  | Regulatory Class:          | II                                                                                            |

## Identification of Legally Marketed Predicate Devices

| Spectrum Dynamics: D-SPECT™ Cardiac Scanner System - | K072468 |
|------------------------------------------------------|---------|
| ELGEMS: eNTEGRA workstation -                        | K003264 |
| ADAC Laboratories: AutoQUANT -                       | K980715 |
| UltraSPECT: WBR Half Dose -                          | K080784 |

#### Device Description

Spectrum Dynamics' D-SPECT™ Cardiac Scanner System is a SPECT device, which is designed to perform myocardial perfusion imaging. The device is comprised of a detector head, gantry, and patient chair. Device operation is controlled from an acquisition station console. The system is supported by use of data-transfer accessories (RFID tags), which are attached to the patient's wrist, for patient and radiopharmaceutical agent positive identification. The cardiac gamma camera is designed such that there are no external moving parts that surround the patient. Detector modules rotate within the closed detector head. The special scanning geometry and detector technology enable shorter scan times. Additionally, a new revision of software was developed, adding Archiving, Display and Communication capabilities (NEMA standard DICOM protocol) with external processing stations PACS systems. The new software package also features correction of insignificant bugs discovered since last release. The modified D-SPECT™ Cardiac Scanner System include a Processing workstation. The Processing Workstation, which is part of the imaging system, is equipped with software package for Processing, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including the Cedars Sinai Quantitative Perfusion SPECT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS) software.

{1}------------------------------------------------

### Intended Use of Device

The D-SPECT™ Cardiac Scanner System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data.

The D-SPECT™ Cardiac Scanner System is primarily intended for cardiac applications.

The D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40 -170 keV. The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and is featuring low count rate (low dose) acquisition, with no lose in image quality.

The D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion of CCT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).

## Safety & Effectiveness

The intended use and indications of the submitted modified D-SPECT™ Cardiac Scanner is identical to the legally marked devices: Spectrum Dynamics: D-SPECT™ Cardiac Scanner System (K072468), ELGEMS: eNTEGRA workstation (K003264), ADAC Laboratories: AutoQUANT (K980715) and UltraSPECT: WBR Half Dose (K080784).

Similarity of intended use, indications and technological features demonstrated, therefore, the risks and benefits are comparable to the predicate devices.

We therefore believe that there are no new questions of safety or effectiveness raised by the introduction of modified D-SPECTTM Cardiac Scanner System.

#### Rational for Substantial Equivalency

Similarities:

The intended use and indications of the proposed modified D-SPECT™ Cardiac Scanner System are similar to the legally marketed predicate devices.

The technology, performance and most of the specifications of the proposed modified D-SPECT™ Cardiac Scanner System are similar to the legally marketed predicate devices Differences:

The proposed modified system includes wider indications in reference to legally marked D-SPECT™ Cardiac Scanner System (K072468) indication for use that was limited to Cardina SPECT aquicition and reconstruction.

The proposed System intended use and indications for use are similar to the above mentioned predicate devices (Adding PACS and fully automated review and quantification of Cardiac SPECT data).

## Substantial Equivalence Statement

Based on the above, it is Spectrum Dynamics's opinion that the proposed modified D-SPECT™ Cardiac Scanner System is substantially equivalent in terms of design, functional features and safety & effectiveness to the unmodified D-SPECT™ Cardiac Scanner System (K072468) legally marketed device and to the legally marked predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Spectrum Dynamics (Israel) Ltd. % Mr. Ilan Sharon QA & RA Consultant Ilan Sharon - Regulatory Affairs Consultant 1 Ha'rakefet St. Zichron Yaacov, 30900 ISRAEL

JUL 1 3 2011

Re: K110507

Trade/Device Name: D-SPECT™ Cardiac Scanner System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: February 20, 2011 Received: February 22, 2011

## Dear Mr. Sharon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

#### Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: D-SPECTTM Cardiac Scanner System

Indications for Use:

The D-SPECT™ Cardiac Scanner and System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. The D-SPECTTM Cardiac Scanner System includes display equipment, patient and equipment supports, component parts, and accessories.

D-SPECT™ Cardiac Scanner System is primarily intended, but not limited to, cardiac applications.

D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40-170 keV.

The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and features low count rate (low dose) acquisition, with no lose in image quality.

D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of SPECT data, using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion SPECT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S. Patel

Division of Radiologica Vitro Diagnostic D

510K K110507

Page 1 of 1

: " | |

时 ﮯ

---

**Source:** [https://fda.innolitics.com/device/K110507](https://fda.innolitics.com/device/K110507)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
