ENDOBON XENOGRAFT GRANULES
K110449 · Biomet 3i · LYC · May 13, 2011 · Dental
Device Facts
| Record ID | K110449 |
| Device Name | ENDOBON XENOGRAFT GRANULES |
| Applicant | Biomet 3i |
| Product Code | LYC · Dental |
| Decision Date | May 13, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3930 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Endobon® Xenograft Granules are used in the following dental and/or surgical procedures: . Alveolar ridge augmentation/reconstruction. . . Filling of defects after root resection, apicoectomy, and cystectomy. . . Filling alveoli after tooth extraction. . . Sinus elevation. This product should not be used in non-peridontal mandibular applications.
Device Story
Endobon Xenograft Granules are bovine-derived hydroxyapatite ceramic granules used as an implantable, non-resorbable osteoconductive scaffold in dental and surgical procedures. The material is processed via high-temperature pyrolysis (>900°C) and sintering (>1200°C) to ensure complete deproteinization and elimination of biological contaminants (bacteria, viruses, prions). The device is intended for use by dental professionals to fill bone defects, augment alveolar ridges, or perform sinus elevations. It provides a structural scaffold to support bone healing. The product is not for use in non-periodontal mandibular applications.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on chemical composition analysis (EDS) comparing Endobon to the predicate device and literature review.
Technological Characteristics
Material: Bovine-derived hydroxyapatite ceramic [Ca5(PO4)3OH]. Manufacturing: Two-step high-temperature process (pyrolysis >900°C, sintering >1200°C). Function: Non-resorbable osteoconductive scaffold. Form: Granules.
Indications for Use
Indicated for dental and surgical procedures including alveolar ridge augmentation/reconstruction, sinus elevation, filling of bone defects (root resection, apicoectomy, cystectomy), and filling alveoli post-tooth extraction. Contraindicated for non-periodontal mandibular applications.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- BioOss (Xenograft Bone Granules (Bovine)) (K980679)
Reference Devices
- Jensen et al. 1996 (Literature study)
Related Devices
- K980679 — ENDOBON · Biomet, Inc. · Apr 3, 1998
- K211551 — Wishbone HA · Wishbone SA · Aug 20, 2021
- K191737 — The Graft Bone Substitute · Purgo Biologics, Inc. · Feb 14, 2020
- K981182 — OSTEOGRAF/N-300 · Ceramed Dental, LLC · Jul 2, 1998
- K240133 — Xenograft Bovine Bone Particulate · Collagen Solutions, LLC · Aug 16, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 1 3 2011
K110449
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92
| Submitter: | BIOMET 3i |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 4555 Riverside Drive Palm Beach Gardens, FL 33410 |
| Establishment Registration Number: | 1038806 |
| Contact Person: | Martha I. Garay Sr. Regulatory Affairs Specialist BIOMET 3i |
| Telephone/Fax/Email: | Tel. 561-776-6923 Fax. 561-514 6316 Email martha.garay@biomet.com |
| Date Prepared: | 02/14/2011 |
| Trade/Proprietary Name: | Endobon Xenograft Granules |
| Address of Sponsor: | BIOMET 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410 |
| Common/Usual Name: | Xenograft Hydroxyapatite Bovine Derived Granules |
| Classification Name: | Class II 21 CFR §872.3930/LYC |
| Device Classification: | The Office of Device Evaluation Center for Devices and Radiological Health - Dental Devices Branch Food and Drug Administration, has classified Bone Grafting Material as Class II |
{1}------------------------------------------------
K110449
| Biomet 3i |
|------------------------------------------------------------------------|
| Traditional 510(k) Premarket Notification - Endobon Xenograft Granules |
device pursuant to 21 C.F.R. § 872.3930 regulations. Legally Marketed Name of Manufacturer: GEISTLICH PHARMA AG CH -Predicate Devices: 6110 Wolhusen Switzerland Washington, DC 20036 Device Name: Xenograft Bone Granules (Bovine) Device Trade Name: BioOss Endobon Xenograft Granules is a modification to Endobon Purpose of the Traditional 510(k) Xenograft Granules, K980679. The purpose of this 510K is to notice: add Sinus Elevation to the current Indication for Use. Device Description: Endobon Xenograft Granules consist of bovine derived hydroxyapatite ceramic granules. Endobon Xenograft Granules are used in the following dental Intended Use: and/or surgical procedures: . Alveolar ridge augmentation/reconstruction, . Sinus elevation, . Filling of resection defects in benign bone tumor, bone cysts, or other defects in the alveolar ridge or wall, . Filling of bone defects after apicectomy, ● Filling alveoli after tooth extraction. This product should not be used in non-peridontal mandibular applications. Endobon - hydroxyapatite xenograft granules are derived from Technological Characteristics: cancellous bovine bone that are used as an implantable material to function as a non-resorbable osteoconductive scaffold for dental applications. A two-step, high-temperature manufacturing process (pyrolysis at a temperature above 900 ℃ and sintering at a temperature above 1200ºC) allows complete deproteinization as well as destruction of potential residual bacteria, virus and prions. The slow resorption rate attributed to Endobon and other similar materials processed in the same manner relates to the crystalline-like structure of 95% of the hydroxyapatite (HA) in Endobon. The resorption rate of hydroxyapatite (HA) increases as crystallinity decreases, and the crystallinity is highly dependent on the sintering temperature. High sintering
{2}------------------------------------------------
K110449
Biomet 3i Traditional 510(k) Premarket Notification - Endobon Xenograft Granules
increase in crystallinity.
The precise chemical name of the hydroxyapatite is pentacalcium hydroxide (tris) phosphate [Ca5 (PO4)3 OH]. Semiquantitative Energy-Dispersive Spectrometry (EDS) analysis of the chemical composition of both Endobon and Bio-Oss grafting materials from Jensen et al. 1996 are presented in Table 1. The data demonstrates similarities in both weight percentage and atomic percentage of the elements.
| | Ca | | P | | Na | | Si | | Mg | | Cl | |
|---------|------|------|------|------|------|------|------|------|------|------|------|------|
| | Wt % | At % | Wt % | At % | Wt % | At % | Wt % | At % | Wt % | At % | Wt % | At % |
| Endobon | 68.2 | 61.3 | 26.4 | 30.6 | 1.6 | 2.5 | 0.6 | 0.7 | 3.3 | 4.9 | 0 | 0 |
| Bio-Oss | 64.4 | 59.3 | 27.2 | 31.4 | 2.9 | 4.5 | 1.6 | 2.0 | 1.6 | 2.4 | 0.4 | 0.4 |
Table 1
| Nonclinical<br>Performance Data: | N/A |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Data: | N/A |
| Performance<br>Standards: | Though there is a Guidance Document for this type of product:<br>"Guidance for Industry and FDA Staff - Class II Special Controls<br>Guidance Document: Dental Bone Grafting Material Devices" it<br>was not utilized as part of this submission. |
| Substantial<br>Equivalence | Endobon Xenograft Granules has the same intended use and,<br>physiochemical properties as BioOss. Any differences in<br>material, chemical composition, or physical structure, do not raise<br>any new questions of safety or effectiveness. Extensive<br>equivalency assessments demonstrates that Endobon Xenograft<br>Granules is as safe and effective as BioOss. Thus; the Endobon<br>Xenograft Granules is substantially equivalent to its predicate<br>device. |
| Conclusion: | Base on the predicate comparison, literature review and<br>equivalency assessment, Endobon Xenograft Granules is<br>substantially equivalent to the predicate device. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Martha I. Garay Senior Regulatory Affairs Specialist Biomet 31 4555 Riverside Drive Palm Beach Gardens, Florida 33410
MAY 1 3 2011
Re: K110449
Trade/Device Name: Endobon Xenograft Granules Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: May 6, 2011 Received: May 9, 2011
Dear Ms. Garay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2- Ms. Garay
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{5}------------------------------------------------
Biomet 3i Traditional 510(k) Premarket Notification - Endobon Xenograft Granules
K110449
Indications for Use
510(k) Number (if known): K110449
Device Name: Endobon Xenograft Granules
Indications for Use:
Endobon® Xenograft Granules are used in the following dental and/or surgical procedures:
- Alveolar ridge augmentation/reconstruction. .
- . Filling of defects after root resection, apicoectomy, and cystectomy.
- . Filling alveoli after tooth extraction.
- . Sinus elevation.
This product should not be used in non-peridontal mandibular applications.
Prescription Use X
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of De ee Evaluation (ODE)
> (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110640