VISUCON-F FROZEN COAG SCREEN N, VISUCON-F FROZEN COAG SCREEN ABN

K110432 · Affinity Biologicals, Inc. · GGN · Oct 20, 2011 · Hematology

Device Facts

Record IDK110432
Device NameVISUCON-F FROZEN COAG SCREEN N, VISUCON-F FROZEN COAG SCREEN ABN
ApplicantAffinity Biologicals, Inc.
Product CodeGGN · Hematology
Decision DateOct 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Indications for Use

VisuCon-F Frozen Coag Screen N: The VisuCon-F Frozen Coag Screen N plasma is an assayed normal control plasma intended for use in the quality control of quantitative coagulation assays, including Prothrombin Time (PT), Activated partial thromboplastin time (APTT) and fibrinogen, in the normal range. The VisuCon-F Frozen Coag Screen N plasma may be used with mechanical and photo-optical coagulation instruments in conjunction with appropriate commercial reagents. VisuCon-F Frozen Coag Screen ABN: The VisuCon-F Frozen Coag Screen ABN plasma is an assayed abnormal control plasma intended for use in the quality control of quantitative coagulation assays, including Prothrombin Time (PT) and Activated partial thromboplastin time (APTT), in the mid-level abnormal range. The VisuCon-F Frozen Coag Screen ABN plasma may be used with mechanical and photo-optical coagulation instruments in conjunction with appropriate commercial reagents.

Device Story

VisuCon-F Frozen Coag Screen N and ABN are assayed human plasma controls used to monitor performance of coagulation analyzers (mechanical and photo-optical). N is pooled normal citrated human plasma; ABN is pooled normal citrated human plasma diluted to a defined concentration. Both are buffered with 0.02 M HEPES and rapidly frozen. Used in clinical laboratories to verify accuracy/precision of PT, APTT, and fibrinogen assays. Healthcare providers run these controls alongside patient samples; results falling within assigned ranges confirm system performance. If results fall outside ranges, clinical decision-making regarding patient coagulation status is suspended until system calibration or reagent issues are resolved.

Clinical Evidence

No clinical data. Bench testing only. Precision studies performed on STA Compact and BCS systems using three lots of control plasma over 20 days (n=80 determinations per lot). Acceptance criteria (CV ≤ 5% for PT/APTT; CV ≤ 15% for fibrinogen) were met. Stability studies confirmed 8-hour open-vial stability and 30-36 month closed-vial shelf life at ≤ -60°C.

Technological Characteristics

Pooled citrated human plasma (min 20 donors), 0.02 M HEPES buffer. Form factor: 1 mL or 4 mL frozen vials. Storage: ≤ -60°C. Compatibility: Mechanical and photo-optical coagulation analyzers. No software or electronic components.

Indications for Use

Indicated for quality control of quantitative coagulation assays (PT, APTT, fibrinogen) in normal and mid-level abnormal ranges. For prescription use only.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k110432 B. Purpose for Submission: Clearance of a new device C. Measurand: PT, APTT, and Fibrinogen D. Type of Test: Quality Control Material, Assayed E. Applicant: Affinity Biologicals Inc. F. Proprietary and Established Names: VisuCon-F Frozen Coag Screen N VisuCon-F Frozen Coag Screen ABN G. Regulatory Information: 1. Regulation section: 21 CFR §864.5425, Multipurpose system for in vitro coagulation studies 2. Classification: Class II 3. Product code: GGN, Plasma, Coagulation Control 4. Panel: 81 (Hematology) H. Intended Use: 1. Intended use(s): VisuCon-F Frozen Coag Screen N: The VisuCon-F Frozen Coag Screen N plasma is an assayed normal control plasma intended for use in the quality control of quantitative coagulation assays, including Prothrombin Time (PT), Activated partial thromboplastin time (APTT) and fibrinogen, in the normal range. The VisuCon-F Frozen Coag Screen N plasma may be used with mechanical and photo-optical coagulation instruments in conjunction with appropriate commercial reagents. VisuCon-F Frozen Coag Screen ABN: The VisuCon-F Frozen Coag Screen ABN plasma is an assayed abnormal control plasma intended for use in the quality control of quantitative coagulation assays, including Prothrombin Time (PT) and Activated partial thromboplastin time (APTT), in the mid-level abnormal range. The VisuCon-F Frozen Coag Screen ABN plasma may be used with mechanical and photo-optical coagulation instruments in conjunction with appropriate commercial reagents. 2. Indication(s) for use: Same as Intended use 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: {2} Diagnostica Stago Inc. STA Compact Hemostasis System and Dade Behring Inc. BCS™ System. I. Device Description: VisuCon-F Frozen Coag Screen N is a pool of normal citrated human plasma collected from a minimum of 20 donors, buffered with 0.02 M HEPES buffer, dispensed and rapidly frozen. The product is packaged in box containing 25 x 1 mL vials or 81 x 1 mL vials or 81 x 4 mL vials. VisuCon-F Frozen Coag Screen ABN is a pool of normal citrated human plasma collected from a minimum of 20 donors, diluted to a defined concentration, buffered with 0.02 M HEPES buffer, dispensed and rapidly frozen. The product is packaged in box containing 25 x 1 mL vials or 81 x 1 mL vials or 81 x 4 mL vials. J. Substantial Equivalence Information: 1. Predicate device name(s) and 510(k) number(s): HemosIL™ Routine Control Level 1, 2, and 3, k082859 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device VisuCon-F Frozen Coag Screen N and ABN | Predicate HemosIL™ Routine Control Level 1, 2, and 3 | | Intended Use | Quality controls of Prothrombin Time (PT), Activated partial thromboplastin time (APTT), and fibrinogen in the normal and abnormal ranges. | Same | | Matrix | Citrated normal human plasma | Same | | Differences | | | | --- | --- | --- | | Item | Device VisuCon-F Frozen Coag Screen N and ABN | Predicate HemosIL™ Routine Control Level 1, 2, and 3 | | Manufacturing process | Frozen pooled normal citrated human plasma buffered with 0.02 M HEPES buffer. The plasma is diluted to defined concentration in the creation of the abnormal control. | Lyophilized pooled human citrated plasma from healthy donors containing buffers and stabilizers. Normal plasma is modified to simulate an abnormal sample. | | Storage | ≤ -60°C | 2-8°C | | Open vial stability | 8 hours at 2-8°C or on-board the coagulation instrument (15 - 22°C) | 8 hours at 2-8°C in the original vial | | Analyzers | STA Compact Hemostasis System; BCS™ System | ACL Classic System (100 – 7000) | | Final Product Form (per control level) | 25 x 1 mL or 81 x 1 mL or 81 x 4 mL vials Two levels: normal and low abnormal range | 10 x 1 mL vials Three levels: normal, low abnormal, and high abnormal range | K. Standard/Guidance Document Referenced (if applicable): Not applicable {3} # L. Test Principle: Not applicable # M. Performance Characteristics: # 1. Analytical performance: # a. Precision/Reproducibility: System #1: Precision studies on two STA Compact analyzers were performed using three lots for both the VisuCon-F Frozen Coag Screen N plasma and the VisuCon-F Frozen Coag Screen ABN plasma. The control plasmas were assayed in duplicate, with 2 runs per day for 20 days, giving a total of 80 determinations per lot of control plasma. HemosIL™ PT-Fib, HemosIL™ APTT-SP and Stago Fib-5 reagents were used for the precision studies. The acceptance criteria, set by the sponsor, are as follows: $\% \mathrm{CV}\leq 5$ for PT and APTT $\% \mathrm{CV}\leq 15$ for fibrinogen (Coag Screen N plasma only). The following table summarizes precision results for the VisuCon-F Frozen Coag Screen N plasma: | Coag Screen N Control | | PT | APTT | Fibrinogen | | --- | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 2.3% | 2.4% | 3.2% | | Intra-Assay Precision (Repeatability) | | 1.7% | 1.2% | 2.2% | | Within-device Precision (Reproducibility) | Lot #2 | 2.4% | 3.7% | 3.9% | | Intra-Assay Precision (Repeatability) | | 1.2% | 2.0% | 2.9% | | Within-device Precision (Reproducibility) | Lot #3 | 1.9% | 1.9% | 4.2% | | Intra-Assay Precision (Repeatability) | | 1.1% | 1.2% | 3.4% | The following table summarizes all precision results for the VisuCon-F Frozen Coag Screen ABN: | Coag Screen ABN Control | | PT | APTT | | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 2.2% | 4.3% | | Intra-Assay Precision (Repeatability) | | 1.4% | 1.9% | | Within-device Precision (Reproducibility) | Lot #2 | 1.6% | 4.5% | | Intra-Assay Precision (Repeatability) | | 0.8% | 1.8% | | Within-device Precision (Reproducibility) | Lot #3 | 1.9% | 4.5% | | Intra-Assay Precision (Repeatability) | | 0.9% | 2.1% | {4} System #2: Precision studies on two STA Compact analyzers were performed using three lots for both the VisuCon-F Frozen Coag Screen N plasma and the VisuCon-F Frozen Coag Screen ABN plasma. The control plasmas were assayed in duplicate, with 2 runs per day for 20 days, giving a total of 80 determinations per lot of control plasma. Stago Neoplastine $\mathrm{Cl}^+$ and APP-Automate reagents were used in the precision studies. The acceptance criteria, set by the sponsor, are as follows: $\% \mathrm{CV}\leq 5$ for PT and APTT The following table summarizes precision results for the VisuCon-F Frozen Coag Screen N: | Coag Screen N Control | | PT | APTT | | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 1.6% | 2.3% | | Intra-Assay Precision (Repeatability) | | 1.0% | 1.5% | | Within-device Precision (Reproducibility) | Lot #2 | 1.8% | 2.7% | | Intra-Assay Precision (Repeatability) | | 1.2% | 1.4% | | Within-device Precision (Reproducibility) | Lot #3 | 1.5% | 2.2% | | Intra-Assay Precision (Repeatability) | | 1.0% | 1.1% | The following table summarizes all precision results for the VisuCon-F Frozen Coag Screen ABN: | Coag Screen ABN Control | | PT | APTT | | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 2.4% | 4.6% | | Intra-Assay Precision (Repeatability) | | 1.2% | 1.8% | | Within-device Precision (Reproducibility) | Lot #2 | 2.5% | 4.1% | | Intra-Assay Precision (Repeatability) | | 1.7% | 1.3% | | Within-device Precision (Reproducibility) | Lot #3 | 3.0% | 3.8% | | Intra-Assay Precision (Repeatability) | | 1.9% | 1.3% | System #3: Precision studies on two BCS™ System analyzers were performed using three lots for both the VisuCon-F Frozen Coag Screen N plasma and the VisuCon-F Frozen Coag Screen ABN plasma. The control plasmas were assayed in duplicate, with 2 runs per day for 20 days, giving a total of 80 determinations per lot of control plasma. Siemens Innovin and Actin FSL reagents were used in the precision studies. The acceptance criteria, set by the sponsor, are as follows: {5} - \% \mathrm{CV} \leq 5 for PT and APTT The following table summarizes precision results for the VisuCon-F Frozen Coag Screen N: | Coag Screen N Control | | PT | APTT | | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 1.3% | 1.5% | | Intra-Assay Precision (Repeatability) | | 0.4% | 0.8% | | Within-device Precision (Reproducibility) | Lot #2 | 2.2% | 2.1% | | Intra-Assay Precision (Repeatability) | | 0.7% | 1.3% | | Within-device Precision (Reproducibility) | Lot #3 | 1.5% | 1.4% | | Intra-Assay Precision (Repeatability) | | 0.7% | 0.9% | The following table summarizes all precision results for the VisuCon-F Frozen Coag Screen ABN: | Coag Screen ABN Control | | PT | APTT | | --- | --- | --- | --- | | Within-device Precision (Reproducibility) | Lot #1 | 2.6% | 2.2% | | Intra-Assay Precision (Repeatability) | | 1.1% | 2.0% | | Within-device Precision (Reproducibility) | Lot #2 | 2.1% | 2.5% | | Intra-Assay Precision (Repeatability) | | 1.0% | 2.4% | | Within-device Precision (Reproducibility) | Lot #3 | 2.0% | 2.5% | | Intra-Assay Precision (Repeatability) | | 0.9% | 2.0% | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Value assignment: For each of three lots of the VisuCon-F Frozen Coag Screen N and VisuCon-F Frozen Coag Screen ABN controls the mean values were assigned to fibrinogen, PT, and APTT by testing 20 vials over 4 days with 5 vials tested per day $(n = 20)$ . Fibrinogen values were generated for the VisuCon-F Frozen Coag Screen N control only. Ranges were then assigned by calculating the mean $\pm 2$ standard deviations (SD). System #1 (STA Compact, HemosIL™ PT-Fib, HemosIL™ APTT-SP, Stago Fib-5), System #2 (STA Compact, Stago Neoplastine Cl+ and APP-Automate) and System #3 (BCS™ System, Siemens Innovin, Actin FSL) were used and all testing was performed internally. Acceptance criteria set by the sponsor, for the mean value for each assay within each platform/reagent system are as follows: {6} 6 | System #1 | | | | --- | --- | --- | | VisuCon-F Frozen Coag Screen N | | | | PT (sec) | APTT (sec) | Fibrinogen (g/L) | | 8 -16 | 20 - 39 | 1.5 – 4.5 | | VisuCon-F Frozen Coag Screen ABN | | | | PT (sec) | APTT (sec) | Fibrinogen (g/L) | | ≥ 19 | ≥ 60 | not applicable | | System #2 | | | | VisuCon-F Frozen Coag Screen N | | | | PT (sec) | APTT (sec) | | | 8 -16 | 20 - 39 | | | VisuCon-F Frozen Coag Screen ABN | | | | PT (sec) | APTT (sec) | | | ≥ 19 | ≥ 60 | | | System #3 | | | | VisuCon-F Frozen Coag Screen N | | | | PT (sec) | APTT (sec) | | | 8 -11 | 20 - 39 | | | VisuCon-F Frozen Coag Screen ABN | | | | PT (sec) | APTT (sec) | | | ≥ 12 | ≥ 60 | | All mean values for the three lots of VisuCon-F Frozen Coag Screen N and Frozen Coag Screen ABN fell within the acceptance criteria listed for each platform/reagent system. **Open Vial Stability**: Open-vial stability of the VisuCon-F Coag Screen N and VisuCon-F Coag Screen ABN control plasmas was performed using System #1 (STA Compact, HemosIL™ PT-Fib, HemosIL™ APTT-SP, Stago Fib-5), System #2 (STA Compact, Stago Neoplastine Cl+ and APP-Automate) and System #3 (BCS™ System, Siemens Innovin, Actin FSL). After initial thawing for the recommended time at 37°C the control plasmas were tested for PT, APTT and fibrinogen then stored at 2-8°C and on-board the instrument (15-22°C) in the original vial and re-tested at 4 hours and 8 hours. Acceptance criteria are as follows: - Results at 4 and 8 hour time points: - Within ±10% of the time 0 result for PT and APTT - Within ±15% of the time 0 result for Fibrinogen A minimum of two lots of control plasma were included in the study. All results fell within the above stated acceptance criteria. The data supports an open-vial 8 hour stability at 2-8°C or on-board the instrument (15-22°C). **Closed Vial Stability**: The closed vial shelf-life stabilities of the VisuCon-F Frozen Coag Screen N and VisuCon-F Frozen Coag Screen ABN controls were assessed through real-time stability studies. The instrument and reagents used for the closed vial stability testing were as follows: System #1 (STA Compact with reagents HemosIL™ PT-Fibrinogen, HemosIL™ APTT-SP, and Stago Fib-5). The VisuCon-F Frozen Coag Screen N study included {7} five lots of control and the VisuCon-F Frozen Coag Screen ABN tested three lots. All controls were stored at their recommended storage temperature of ≤ -60°C and tested internally at the indicated time points for PT and APTT. Fibrinogen was assessed on the normal control only. Acceptance criteria set by the sponsor, are as follows: - ±10% of the original assigned value for PT and APTT - ±15% of the original assigned value for Fibrinogen The study data for the VisuCon-F Frozen Coag Screen N and VisuCon-F Frozen Coag Screen ABN supports a closed vial stability of 30 and 36 months respectively when stored at ≤ -60°C. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data: Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected values are provided in the Package Insert accompanying the product. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...