← Product Code [HGX](/productcode/HGX) · K110343

# TOMMEE TIPPEE CLOSER TO NATURE ELECTRIC BREAST PUMP (K110343)

_Mayborn Baby and Child / Jackel International, Ltd. · HGX · Mar 11, 2011 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110343

## Device Facts

- **Applicant:** Mayborn Baby and Child / Jackel International, Ltd.
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Mar 11, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Tommee Tippee® Closer to Nature® Electric Breast Pump is used to express and collect milk from the breast of a lactating woman.

## Device Story

Electric breast pump; powered by mains or 4 AA batteries; utilizes reciprocating diaphragm pump mechanism to create vacuum suction. User selects from 3 suction levels (light, medium, high). Device includes breast cup, milk storage pots, and feeding components. Used by lactating women for expression, storage, and feeding. Vacuum levels measured and compared to predicate; device performance verified through fluid displacement testing. Materials meet FDA food contact and ISO 10993 biocompatibility standards.

## Clinical Evidence

Bench testing only. Comparative performance testing against Medela Swing included vacuum measurement (bare and installed), suction rate (fluid displacement), and material biocompatibility (ISO 10993). Results showed no significant difference in performance or safety between subject and predicate devices.

## Technological Characteristics

Reciprocating diaphragm pump; 6V DC power (mains or 4 AA batteries). Materials: Soft silicone (Wacker R401) for breast interface; all milk-contact materials compliant with 21 CFR 176, 177, 178. Biocompatibility per ISO 10993. No software/algorithm.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Medela Swing ([K053052](/device/K053052.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kilo343 page 1 of 5

MAR 1 1 2011

# SECTION 5: 510(k) Summary

#### Submitter

Mayborn Group Limited Northumberland Business Park (West) Dudley Lane Cramlington Northumberland United Kingdom NE23 7RH

Tel: +44 191 250 1864 Fax: +44 191 250 1727 Owner/Operator Number: Establishment Registration No.: 10034113

#### Contact Person

Neil Hewitt Senior Quality Manager, Mayborn Baby & Child Division Northumberland Business Park (West) Dudley Lane Cramlington Northumberland United Kingdom NE23 7RH

Phone Number + 44 1912501864 Fax Number: +44 1912501727 E-mail Address nhewitt@jackel.co.uk

#### Date Prepared

December 2010

#### Device Information

Trade name: closer to nature® Common name: electric breast pump Classification Name: pump, breast, powered Review Panel: Obstetrics/Gynecology Product Code: HGX Device Class: 2

### Indication for Use

Tommee Tippee Closer to Nature® Electric Breast Pump is used to express and collect milk from the breast of a lactating woman.

Image /page/0/Picture/15 description: The image shows the logo for Tommee Tippee. The logo consists of a five-pointed star on the left and the words "tommee tippee" on the right. The word "tommee" is on the top line and "tippee" is on the bottom line. There is a registered trademark symbol next to the word "tippee".

{1}------------------------------------------------

#### Device Description

This electric breast pump product is designed to provide everything a mother will need to start sterilizing, expressing, storing and feeding. The pump unit has 3 alternate settings: light, medium and high suction levels during expression. The collect and protect breast milk storage system enables the user to express directly into the small graduated 2 oz pot, which fits into "Closer to Nature" bottles for feeding. The product is designed to use, one pot size for expressing, storing and feeding. This device is Mains or Battery powered.

Included with the pump and vacuum system is:

#### Accessories

Sterilizer box - This doubles up as a storage case

Feeding Components

Feeding bottle with slow flow teat 2 Milk storage pots Milk storage lid 6 Breast pads

Devices to which substantial equivalence is claimed:

| <br>LACK PS PROMOULL LE LE LE LEASE LEA<br>------------------------------------------------------------------------------------------------ | 24 1 1 1 7 6 - 11<br><br>PIS MALL M<br>.<br>FARSE . FRAN<br><br>.<br><br><br><br>1140.<br>.<br>A<br>******************************************************************************************************************************************************************************<br>.<br> | 4 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>4 2 11 81 5 5 111 5 10 4 10 11<br><br><br><br><br><br>AN HONGA OF A<br>152525 255 945 121 1<br>Car Can Isa.<br> |
|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| WAN                                                                                                                                         | A BOOK con Company of                                                                                                                                                                                                                                                                       | SWITZ<br>----------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1                                             |

#### Device Comparison to Legally Marketed Device

Below is a comparison chart outlining differences and similarities between the Tommee Tippee® Closer to Nature® Electric Breast Pump, and the Medela AG, SwingTM electric breast pump.

|                | New Product                         | The Predicate |
|----------------|-------------------------------------|---------------|
| Device Name    | Tommee Tippee® Closer to<br>Nature® | SwingTM       |
| Manufacturer   | Mayborn Group Limited               | Medela AG     |
| 510(k) number  | To be assigned                      | K053052       |
| Product Code   | HGX                                 | HGX           |
| Classification | 2                                   | 2             |

{2}------------------------------------------------

|                                    | New Product                                           | The Predicate                                      |
|------------------------------------|-------------------------------------------------------|----------------------------------------------------|
| Intended use                       | As described in 21 CFR 884.5160                       | As described in 21 CFR 884.5160                    |
| Power Source                       | 6V DC<br>four (4) AA batteries or electric<br>adaptor | 6V DC<br>four (4) AA batteries or electric adaptor |
| Pump type                          | Reciprocating Diaphragm                               | Reciprocating Diaphragm                            |
| Single or Double<br>Pumping        | Single                                                | Single                                             |
| Adjustable Suction<br>Levels       | Yes - 4 Separate Speed Rates                          | Yes                                                |
| Adjustable Cycle<br>Speed          | No                                                    | Yes                                                |
| Overflow<br>Protection             | No                                                    | No                                                 |
| Highest Vacuum<br>Setting [in Hg]  | 10.6                                                  | 11.8                                               |
| Lowest Vacuum<br>Setting [in Hg.]  | 5.3                                                   | 1.7                                                |
| Range of Cycle<br>Speeds           | 41 Fixed                                              | 45-72                                              |
| Breast Cup-to-<br>Breast Interface | Soft Silicone Wacker R401                             | Rigid Plastic with Soft TPE section                |

Both The Medala Swing (K053052) and the Closer to Nature Electric Breast Pump are identified by the FDA as Product code HGX under CFR section 884.5160.

Both Units use a standard 6v battery supply or mains adaptor.

Both units use a single diaphragm pump.

Both units generate a vacuum equivalent to approx 11 inches of Mercury on the maximum setting.

#### Key Areas

Mayborn Baby and Child during the development of the product identified five key comparative areas which would ensure the safety and integrity of the Closure to Nature Electric Breast Pump would match or exceed the ability of the Medela Swing (K053052)

These cover the following:

- 1. Pump vacuum direct measurement of the pumps used
- 2. Pump vacuum when installed in the housing
- 3. Lowest suction rate
- 4. Suction rate
- 5. Materials

The first key comparative area identified was the bare vacuum capability of the pumps when removed from the housing. A direct comparison of the pumps was carried out and the results

:

:

{3}------------------------------------------------

K110343 page 4 of 5

showed the Closer to Nature pump drawing a vacuum of -40kpa and the Medala Swing pump drawing a vacuum of -38kpa.

#### This shows there is no significant difference in the devices.

The second key comparative area identified was the vacuum capability of the pumps when installed in the housing. A direct comparison of the pumps was carried out and the results showed the Closer to Nature Electric Breast Pump was capable of drawing a vacuum up to -38kpa and the Medala Swing pump was capable of drawing a vacuum up to -36kpa.

#### This shows there is no significant difference in the devices.

The third key comparative area identified was the lowest vacuum capability of the pumps when installed in the housing and the pump is set up for simulated use. A direct comparison of the pumps was carried out and the results showed the Closer to Nature pump drawing a vacuum of -20kpa and the Medala Swing pump drawing a vacuum of -29kpa.

### This shows the Closer to Nature Electric Breast Pump can draw a slightly lower vacuum during use by the mother.

The forth key comparative area identified was the suction rate of the pump in terms of its ability to flow liquid. The comparison of the pumps identified the following:

Closer to Nature Electric Breast Pump: Time period = 2 minutes (120seconds) Fluid Displaced = 180millilitres

Medala Swing Electric Breast Pump: Time period = 2 minutes (120seconds) Fluid Displaced = 150millilitres

#### This shows there is no significant difference in the devices.

The fifth key area identified was material selection; all the materials specified and tested for use in the Closer to Nature Electric Breast Pump have been selected to ensure they meet the appropriate FDA regulations concerning food contact and or bio compatibility to ensure maximum protection of the infant from contaminated food and maximum comfort for the mother.

All milk and human contact components are manufactured from materials that meet FDA food additive criteria as set forth in Part 21 Code of Federal Regulations Parts 176, 177 and 178.

The skin contacting materials associated with the Electric Breast Pump will satisfy all required biocompatibility testing conducted in accordance with ISO 10993 for skin contact.

This shows there is no significant difference in the devices.

{4}------------------------------------------------

# K110343 page 5 of 5

#### Conclusion

Tommee Tippee® Closer to Nature® Electric Breast Pump, subject of this submission, constitutes a safe, reliable and effective medical device, meeting all the declared requirements of its intended use. Device presents no adverse health effects or safety risks when used as intended.

Tommee Tippee® Closer to Nature® Electric Breast Pump has the same intended use and fundamental scientific technology as its predicate device - the Medela® SwingTM Breast Pump (K053052).

Tommee Tippee® Closer to Nature® Electric Breast Pump was tested against its predicate, and was found to be substantially equivalent based on the summary information above and the detailed evaluation within this submission.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mayborn Group Limited c/o Ms. Paula Wilkerson Senior Staff Engineer - Medical Devices Intertek 2307 E Aurora Road, Unit B7 TWINSBURG OH 44087

MAR 1 1 20il

Re: K110343

Trade Name: Closer to Nature® Electric Breast Pump Regulation Number: 21 CFR \$884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product code: HGX Dated: March 2, 2011 Received: March 4, 2011

Dear Ms. Wilkerson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898: In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{6}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

i

Sincerely yours,

Hubert Lenoir MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{7}------------------------------------------------

## SECTION 4: Indications for Use

510(k) Number: FBA KII0343

Device Name: Closer to Nature® Electric Breast Pump

Indications for Use:

Tommee Tippee® Closer to Nature® Electric Breast Pump is used to express and collect milk from the breast of a lactating woman.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| Page 1 of 1                                            |

  

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices

| 510(k) Number | K110343 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K110343](https://fda.innolitics.com/device/K110343)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
