← Product Code [LYZ](/productcode/LYZ) · K110338

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES (K110338)

_Zibo Yuechuan Plastic Products Co., Ltd. · LYZ · Apr 19, 2011 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110338

## Device Facts

- **Applicant:** Zibo Yuechuan Plastic Products Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Apr 19, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

## Device Story

Disposable synthetic vinyl patient examination gloves; worn by healthcare personnel to prevent cross-contamination between examiner and patient; non-sterile, non-colored, powder-free; used in clinical or medical settings; provides a physical barrier against fluids, waste, and environmental contaminants; device is donned manually by the user; benefits include protection of both patient and healthcare provider during examinations.

## Clinical Evidence

No clinical data. Evidence consists of bench testing including physical dimensions, water-tightness (FDA 1000 ml water fill test), and biocompatibility (primary skin irritation and skin sensitization testing) in accordance with ASTM D5250-06.

## Technological Characteristics

Synthetic vinyl material; powder-free (contains no more than 2 mg powder per glove); non-sterile; non-colored; conforms to ASTM D5250-06; physical testing conducted at Inspection Level S-2, AQL 2.5; water-tightness testing conducted at Inspection Level I, AQL 2.5.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free ([K992821](/device/K992821.md))

## Reference Devices

- Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K110338

### APR 1 9 2011

## 510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

#### 1, Submitter's Identification:

Zibo Yuechuan Plastic Products Co., Ltd. 32 Yuming Road, Zhangdian District Zibo, Shandong, China

Contact Person: Cao, Yanan, General Manager Phone Number: 011-86-5333819566

Date summary prepared: February 1, 2011

#### 2. Name of the Device:

Zibo Yuechuan Plastic Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free

#### 3. Predicate Device Informaton:

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)

#### 4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I. 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.

#### ડ. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

{1}------------------------------------------------

#### 6. Comparison to Predicate Devices:

Zibo Yuechuan Plastic Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. and Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves. Powder free.

### 7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:

The standards used for Zibo Yuechuan Plastic Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "bowder-free" claims (contain no more than 2 mg powder per glove).

#### 8. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic claim.

#### 9. Conclusions:

Zibo Yuechuan Plastic Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR 1-9 2011

Zibo Yuechuan Plastic Products, Company, Limited C/O Mr. Frank Liu President Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710

Re: K110338

Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile, Non Colored Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 2, 2011 Received: February 4, 2011

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo of Zibo Yuechuan Plastic Products Co., Ltd. The logo includes a symbol that looks like a caduceus, which is a symbol of medicine. The text "Zibo Yuechuan Plastic Products Co., Ltd." is written in a bold, sans-serif font and is underlined. The logo is simple and professional, and it clearly identifies the company.

32 Yuming Road, Zhangdian District Zibo, Shandong, China

# INDICATIONS FOR USE

Zibo Yuechuan Plastic Products Co., Ltd. Applicant:

510(k) Number:

K110338

Device Name:

Patient Vinyl Examination Gloves, Powder free, Non-Sterile, Non Colored

Indications of Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Prescription Use _____________________________________________________________________________________________________________________________________________________________

Over the Counter Use _____X

Factory Initials

Elizabeth F. Clawie-William

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K110338

---

**Source:** [https://fda.innolitics.com/device/K110338](https://fda.innolitics.com/device/K110338)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com