K110243 · Medart A/S · GEX · Jun 28, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110243
Device Name
MEDART DIODE LASER SYSTEM & ACCESSORIES
Applicant
Medart A/S
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedArt 720 Diode Laser System & Accessories is a prescription only device intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The MedArt 720 Diode Laser System & Accessories is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, and cardiothoracic surgery and ophthalmology. The MedArt 720 Diode Laser System & Accessories is specifically indicated for laser assisted lipolysis
Device Story
MedArt 720 is a Class 4 diode laser system emitting 980 nm invisible radiation. Device delivers continuous or pulsed laser energy to soft tissue via contact or non-contact modes, including endoscopes, introducers, or catheters. Energy is primarily absorbed by melanin, hemoglobin, and dark tissue; limited water absorption enables rapid heating at low power. Used by physicians in clinical/surgical settings for soft tissue management (incision, excision, ablation, hemostasis, coagulation). Output allows precise tissue interaction, facilitating surgical procedures across multiple specialties. Benefits include controlled tissue destruction and hemostasis, reducing bleeding and improving surgical outcomes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to the predicate device.
Technological Characteristics
Class 4 diode laser; 980 nm +/- 10 nm wavelength; 15W output power; continuous or pulsed wave operation. Dimensions: 10.9" x 13.0" x 6.0". Delivery via contact/non-contact modes, endoscopes, introducers, or catheters. System is standalone.
Indications for Use
Indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue across various surgical specialties including ENT, dental, arthroscopy, GI, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery, pulmonary, cardiothoracic, ophthalmology, and laser-assisted lipolysis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
K115243 Page U + U
Image /page/0/Picture/1 description: The image shows the logo for "MedArt", with the tagline "an ENERGIST - Group Company" underneath. The word "MedArt" is written in a bold, stylized font, with the letters "M", "e", and "d" appearing to have a textured or outlined effect, while the letters "A", "r", and "t" are solid black. The tagline is in a smaller, simpler font, centered below the main logo.
# 5. 510 (k) Summary (21 CFR 807.92)
| 510 (k) Submitter's Name: | MedArt A/S |
|------------------------------------|------------------------------------------------------------------------------------------------------|
| Address: | MEDART A/S Industriholmen 15a Hvidovre, Denmark DK-2650 |
| Contact Person: | Joanne Davies, Regulatory Affairs Manager, Energist Limited joanne.davies@energist-international.com |
| Telephone: | +44(0)1792 798768 |
| Fax: | +44(0)1792 762099 |
| Date: | 20 <sup>th</sup> January 2010 |
| Device Name: | MedArt® 720 |
| Trade/Proprietary Name: | MedArt® Diode Laser System & Accessories |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Confidentiality Preference: | No preference |
| Establishment Registration Number: | 8021543 |
| 510 (K) Type: | Traditional |
| Class: | 2 |
| Panel: | General & Plastic Surgery |
| Product Code: | GEX |
| Submission Basis: | New Device |
| Description of Device: | |
{2}------------------------------------------------
11043 page (2) of (4)
Image /page/1/Picture/1 description: The image shows the logo for MedArt, with the words "MedArt" in a stylized font. The first three letters, "Med", are in a lighter color and have a textured, almost sketched appearance, while the last three letters, "Art", are in a solid, darker color. Below the logo, in smaller font, it says "an ENERGIST * Group Company".
The MedArt® 720 laser unit is a Class 4 laser that emits invisible laser radiation. The unit is fed by a diode laser module providing continuous or pulsed wave laser beam operating at a wavelength of 980 nm that is primarily absorbed in melanin, haemoglobin and dark tissue. A limited amount of light is absorbed by water enabling fast increase in irradiated areas leading to effective heating of tissue at low power outputs.
### Intended Use:
The MedArt 720 Diode Laser System & Accessories is a prescription only device intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters.
The MedArt 720 Diode Laser System & Accessories is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, and cardiothoracic surgery and ophthalmology.
The MedArt 720 Diode Laser System & Accessories is specifically indicated for laser assisted lipolysis
| Ear Nose, Throat &<br>Oral Surgery | Hemostasis, incision, excision, ablation, coagulation &<br>vaporization of tissue from the ear, nose, throat & adjacent<br>areas including soft tissue in the oral cavity. |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dental Applications | Intraoral and extra-oral soft tissue (including marginal and<br>interdental gingiva and epithelial lining of free gingiva |
| Arthroscopy | Hemostasis, incision, excision, coagulation, vaporization and<br>ablation of joint tissues during arthroscopic surgery |
| Gastroenterology | Hemostasis, incision, excision, ablation, coagulation &<br>vaporization of tissue in the upper and lower<br>gastrointestinal tracts and also with endoscopic procedures |
| General surgery,<br>Dermatology,<br>Plastic Surgery &<br>Podiatry | Excision, ablation, vaporization and photocoagulation of<br>skin lesions, hemostasis, incision, excision, vaporization,<br>ablation and debulking of soft tissue, abdominal, rectal,<br>skin, fat or muscle tissue and dermabrasion |
| Urology | Excision, vaporization, incision, coagulation, ablation and<br>hemostasis of urological tissues |
| Gynecology | Ablation, excision, incision, coagulation, hemostasis and<br>vaporization of gynecological tissue |
#### Indications For Use:
{3}------------------------------------------------
K110243
:
" "
Page 3 of 4
Image /page/2/Picture/2 description: The image shows the logo for MedArt, which is written in a stylized font. Below the company name, it says "an ENERCIST Group Company". The text is black and white.
| Neurosurgery | Vaporization, coagulation, excision, incision ablation and hemostasis of soft tissue |
|-----------------------------|------------------------------------------------------------------------------------------------------------------|
| Pulmonary Surgery | Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system |
| Cardiothoraic<br>Surgery | Incision, excision, vaporization, ablation, hemostasis, and coagulation of soft tissue including cardiac tissues |
| Laser Assisted<br>Lipolysis | Laser Assisted Lipolysis |
| Opthalmology | The ablation, vaporization, excision, incision, and coagulation of soft tissue in opthalmology |
{4}------------------------------------------------
11/0243
. :
Page (4) of (4)
Image /page/3/Picture/2 description: The image shows the logo for MedArt, which is written in a stylized font. Below the logo, it says "an ENERGIST • Group Company". The logo is black and white and appears to be slightly pixelated.
# Predicate Device: K081015
# Technological Characteristic Comparison
| Characteristic | MedArt® 720 | Cerelas D980 | Cerelas D980 |
|------------------|----------------------|---------------|----------------|
| 510 (k) | K110243 | K081015 | K081015 |
| CFR | 878.4810 | 878.4810 | 878.4810 |
| Laser Class | 4 | 4 | 4 |
| Laser Wavelength | 980 nm +/- 10 nm | 980 nm | 980 nm |
| Output Power | 15W | 15W | 25W |
| Size | 10.9" x 13.0" x 6.0" | 7" x 9" x 14" | 7" x 15" x 16" |
| | II | II | II |
| Product Code | GEX | GEX | GEX |
- This summary only includes information that is covered in the body of the 510 (k) .
- Ths summary does not contain any unsubstantiated labeling claims o
- This summary includes summary data only ●
- This summary does not contain any trade secrets or confidential commercial o information
- The summary does not contain any patient identifiction information �
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a representation of a human figure embracing or protecting another, conveying a sense of care and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MedArt A/S % Energist, Ltd. Ms. Joanne Davies Clos Llyn Cwm Swansea, Wales United Kingdom SA6 8QY
JUN 2 8 2011
Re: K110243
Trade/Device Name: MedArt® 720 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 16, 2011 Received: June 21, 2011
Dear Ms. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Joanne Davies
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office ' of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ernst Keith
- Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K1||0243 Page① of ②
Image /page/6/Picture/1 description: The image shows the logo for MedArt, with the words "MedArt" in a stylized font. Below the company name, in a smaller font, it says "an ENERGIST • Group Company". The logo is simple and professional, and the text is easy to read.
4. Indications for Use Statement
510{k} Number (if known):
Device Name: MedArt® 720 Intended Use:
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110243
The MedArt 720 Diode Laser System & Accessories is a prescription only device intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters.
The MedArt 720 Diode Laser System & Accessories is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, and cardiothoracic surgery and ophthalmology.
The MedArt 720 Diode Laser System & Accessories is specifically indicated for laser assisted lipolysis
| Ear Nose, Throat &<br>Oral Surgery | Hemostasis, incision, excision, ablation, coagulation &<br>vaporization of tissue from the ear, nose, throat & adjacent<br>areas including soft tissue in the oral cavity. |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dental Applications | Intraoral and extra-oral soft tissue (including marginal and<br>interdental gingiva and epithelial lining of free gingiva |
| Arthroscopy | Hemostasis, incision, excision, coagulation, vaporization and<br>ablation of joint tissues during arthroscopic surgery |
| Gastroenterology | Hemostasis, incision, excision, ablation, coagulation &<br>vaporization of tissue in the upper and lower gastrointestinal<br>tracts and also with endoscopic procedures |
| General surgery,<br>Dermatology,<br>Plastic Surgery &<br>Podiatry | Excision, ablation, vaporization and photocoagulation of skin<br>lesions, hemostasis, incision, excision, vaporization, ablation<br>and debulking of soft tissue, abdominal, rectal, skin, fat or<br>muscle tissue and dermabrasion |
| Urology | Excision, vaporization, incision, coagulation, ablation and<br>hemostasis of urological tissues |
| Gynecology | Ablation, excision, incision, coagulation, hemostasis and<br>vaporization of gynecological tissue |
| Neurosurgery | Vaporization, coagulation, excision, incision ablation and |
### Indications for Use:
Industriholmen 15A DK-2650 Hvidovre Denmark
{8}------------------------------------------------
K110243
Page (2) of (2)
Image /page/7/Picture/2 description: The image shows the logo for MedArt, with the words "MedArt" in large, bold letters. Below the logo, in smaller letters, it says "an ENERGIST Group Company". The logo is simple and professional, and the text is clear and easy to read.
| | hemostasis of soft tissue |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------|
| Pulmonary Surgery | Hemostasis, vaporization, coagulation, incision, excision and<br>ablation of soft tissue in the pulmonary system |
| Cardiothoracic<br>Surgery | Incision, excision, vaporization, ablation, hemostasis, and<br>coagulation of soft tissue including cardiac tissues |
| Laser Assisted<br>Lipolysis | Laser Assisted Lipolysis |
| Opthalmology | The ablation, vaporization, excision, incision, and coagulation<br>of soft tissue in opthalmology |
Prescription Use _____X__ AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1/1
Neil RP Oshn for mkm
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110243
Industriholmen 15A DK-2650 Hvidovre Denmark
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.