ROG KNOTLESS SUTURE ANCHOR
K110230 · Rog Sports Medicine, Inc. · MBI · May 19, 2011 · Orthopedic
Device Facts
| Record ID | K110230 |
| Device Name | ROG KNOTLESS SUTURE ANCHOR |
| Applicant | Rog Sports Medicine, Inc. |
| Product Code | MBI · Orthopedic |
| Decision Date | May 19, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The RoG 5.5 mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows: Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, bicens tenodesis. Wrist/Hand indications:- Ulnar/Radial collateral ligament reconstruction, scapholunate ligament reconstruction. Foot/Ankle indications: - Achilles tendon repair/reconstruction, hallax valgus reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions. Elbow indications:- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair. Knee indications :- Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure.
Device Story
Screw-like knotless suture anchor; composed of PEEK plastic; 5.5mm diameter, 17mm length. Used by surgeons in clinical settings for soft tissue reattachment to bone. Device secures user-selected suture to anchor top. Supplied non-sterile; requires user-facility autoclave sterilization. Reusable taps and guides provided. Benefits patient by facilitating tendon/ligament repair and stabilization.
Clinical Evidence
Bench testing only. Pull and fatigue testing performed in accordance with FDA Guidance Document for Testing Bone Anchor Devices (4/20/96).
Technological Characteristics
Material: PEEK (ISO 10993 compliant). Form factor: Screw-like, 5.5mm diameter, 17mm length. Sterilization: Autoclave (user-facility). No software or electronics.
Indications for Use
Indicated for soft tissue reattachment in shoulder, elbow, wrist/hand, foot/ankle, and knee. Procedures include Bankart, rotator cuff, SLAP, capsulolabral, acromio-clavicular, deltoid, biceps, ligament, and tendon repairs/reconstructions, hallux valgus, and joint capsule closure.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- ALLthread™ PEEK Suture Anchor (K070389)
Related Devices
- K110229 — ROG SUTURE ANCHOR 5.5MM, 2.9MM · Rog Sports Medicine, Inc. · May 18, 2011
- K113299 — ROG SUTURE ANCHOR · Rog Sports Medicine, Inc. · Dec 15, 2011
- K111590 — ROG SUTURE ANCHOR, ROG KNOTLESS SUTURE ANCHOR · Rog Sports Medicine, Inc. · Aug 8, 2011
- K230517 — Legends Orthopedics Suture Anchors (Legends Orthopedics 5.5 mm Suture Anchors, Legends Orthopedics 2.9 mm Suture Anchor, and Legends Orthopedics 5.5 mm Knotless Suture Anchor) · Jarmon Medical, LLC (Dba Legends Orthopedics) · Sep 11, 2023
- K112991 — MODIFIED ROG SUTURE ANCHOR · Rog Sports Medicine, Inc. · Dec 8, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
Premarket Notification Number: K110230
DEVICE: RoGTM Sports Medicine Knotless Suture Anchor
SPONSOR/MANUFACTURER: RoG Sports Medicine, Inc. 16450 S. 104th Ave. Orland Park, IL 60467
### SUMBITTER/REGULATORY CONTACT:
Curtis Raymond Orchid Design 80 Shelton Technology Ctr. Shelton, CT 06484
### FDA ESTABLISHMENT REGISTRATION NUMBER: (pending)
### TRADE NAME, COMMON NAME, CLASSIFICATION:
TRADE NAME: RoG Sports Medicine Knotless Suture Anchor
COMMON NAME: Suture Anchor
CLASSIFICATION: Class II (ref .: 21 CFR 888.3040); Product Code MBI
## PREDICATE DEVICE(S):
K070389 - ALLthread™ PEEK Suture Anchor (Biomet Sports Medicine)
### DESCRIPTION OF SUBJECT DEVICE:
The subject device is screw-like in shape and composed exclusively of PEEK plastic. It is available in diameter of 5.5mm lengths of 17mm. It is supplied non-sterile and is intended for sterilization by the user facility. It allows the user to secure a suture of his/her selection to the top of the anchor. The anchor is supplied with reusable taps and guides of corresponding size.
#### INTENDED USE:
The RoG 5.5 mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:
- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, capsule ● repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, bicens tenodesis.
- . Wrist/Hand indications:- Ulnar/Radial collateral ligament reconstruction, scapholunate ligament reconstruction.
{1}------------------------------------------------
K110230*#2/2
- . Foot/Ankle indications: - Achilles tendon repair/reconstruction, hallax valgus reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- . Elbow indications:- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair.
- . Knee indications :- Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure.
### PERFORMANCE CHARACTERISTICS:
The subject was evaluated in accordance with FDA's Guidance Document for Testing Bone Anchor Devices (4/20/96). Testing consisted of pull testing and fatigue testing with comparison to the predicate device. The subject device was found to be equivalent to the predicate device in such performance.
### SAFETY CHARACTERISTICS:
The device is composed exclusively of polyetheretherketone (PEEK). A Master File demonstrating safety of the material has been supplied by the PEEK supplier and shows compliance to the requirements of ISO 10993. Additionally, validation testing was conducted to show the sterility of the device following autoclave sterilization.
## CONCLUSION(S):
The subject device has the same design considerations, assembly configurations, performance characteristics and indications for use as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 9 2011
RoG Sports Medicine, Inc. % Mr. Curtis Raymond Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484
Re: K110230
Trade/Device Name: RoG Knotless Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 12, 2011 Received: May 13, 2011
Dear Mr. Raymond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
### Page 2 - Mr. Curtis Raymond
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
A. Ky B. n h
fen
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
(Page 1 of 1)
# Indications for Use
510(k) Number (if known): K110230
Device Name: RoG Knotless Suture Anchor
The RoG 5.5mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:
- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/Hand indications:- Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.
- Foot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- Elbow indications : Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction, lateral epicondylitis repair.
- Knee indications:- Lateral collateral ligament repair, medial collateral ligament ● repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off) for M. Mulkerson
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110230