← Product Code [MBI](/productcode/MBI) · K110229

# ROG SUTURE ANCHOR 5.5MM, 2.9MM (K110229)

_Rog Sports Medicine, Inc. · MBI · May 18, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K110229

## Device Facts

- **Applicant:** Rog Sports Medicine, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** May 18, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The RoG 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows: Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis. Wrist/hand indications:- Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction. Foot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions. . Elbow indications .- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair. Knee indications:- Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure. The RoG 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows: Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis. Wrist/hand indications:- Scapholunate ligament reconstruction. ● Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction. . Knee indications:- Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure. extracapsular repair, vastus medialis obliquus (VMO) muscle advancement.

## Device Story

Screw-like suture anchor composed of PEEK plastic; available in 5.5mm and 2.9mm diameters. Used by surgeons in orthopedic procedures to secure sutures to bone for soft tissue reattachment. Supplied non-sterile; requires user-facility autoclave sterilization. Includes reusable taps and guides. Device functions as mechanical fixation fastener. Benefits patient by providing secure anchor point for tendon/ligament repair, facilitating healing.

## Clinical Evidence

Bench testing only. Testing conducted in accordance with FDA's Guidance Document for Testing Bone Anchor Devices (4/20/96), including tapping and pull testing compared to predicate devices. No clinical data presented.

## Technological Characteristics

Material: PEEK (polyetheretherketone) compliant with ISO 10993. Form factor: Screw-like anchor, 5.5mm/2.9mm diameters, 10mm/17mm lengths. Energy source: None (mechanical). Sterilization: Autoclave (user-performed).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Helix™ PEEK Anchor ([K071481](/device/K071481.md))
- Hitch™ PEEK Suture Anchor ([K060693](/device/K060693.md))
- CorkScrew® PEEK [5.5mm] Suture Anchor ([K061863](/device/K061863.md))
- SutureTak® PEEK [3.0mm] Suture Anchor ([K061863](/device/K061863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 1/2

## 510(k) SUMMARY K110229

MAY 18 2011

May 16, 2011

DEVICE: RoGTM Sports Medicine Suture Anchor

SPONSOR/MANUFACTURER: RoG Sports Medicine, Inc. 16450 S. 104th Ave. Orland Park, IL 60467

### SUMBITTER/REGULATORY CONTACT:

Curtis Raymond Orchid Design 80 Shelton Technology Ctr. Shelton, CT 06484

### FDA ESTABLISHMENT REGISTRATION NUMBER: (pending)

### TRADE NAME, COMMON NAME, CLASSIFICATION:

TRADE NAME: RoGTM Sports Medicine Suture Anchor

COMMON NAME: Suture Anchor

CLASSIFICATION: Class II (ref .: 21 CFR 888.3040); Product Code MBI

### PREDICATE DEVICE(S):

K071481 - Helix™ PEEK Anchor (Depuy Mitek) K060693 - Hitch™ PEEK Suture Anchor (Biomet Sports medicine) K061863 - CorkScrew® PEEK [5.5mm] Suture Anchor (Arthrex, Inc.) K061863 -- SutureTak® PEEK [3.0mm] Suture Anchor (Arthrex, Inc.)

### DESCRIPTION OF SUBJECT DEVICE:

The subject device is screw-like in shape and composed exclusively of PEEK plastic. It is available in diameters of 5.5mm and 2.9mm and lengths of 10mm and 17mm. It is supplied non-sterile and is intended for sterilization by the user facility. It allows the user to secure a suture of his/her selection to the top of the anchor. The anchor is supplied with reusable taps and guides of corresponding sizes.

### INTENDED USE:

The RoG 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- . Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair.

{1}------------------------------------------------

# KII0229 ** 12

capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.

- Wrist/hand indications:- Ulnar/Radial collateral ligament reconstruction. . scapholunate ligament reconstruction.
- Foot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- t Elbow indications:- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair.
- Knee indications:- Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

The RoG 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows:

- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications:- Scapholunate ligament reconstruction. .
- . Elbow indications:- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction.
- Knee indications:- Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure. extracapsular repair, vastus medialis obliquus (VMO) muscle advancement.

### PERFORMANCE CHARACTERISTICS:

Both sizes of subject were evaluated in accordance with FDA's Guidance Document for Testing Bone Anchor Devices (4/20/96). Testing consisted of tapping and pull testing with comparison to the corresponding size of predicate devices. The subject device was found to be equivalent to the corresponding sizes of the predicate devices in such testing.

### SAFETY CHARACTERISTICS:

The device is composed exclusively of polyetheretherketone (PEEK). A Master File demonstrating safety of the material has been supplied by the PEEK supplier and shows compliance to the requirements of ISO 10993. Additionally, validation testing was conducted to show the sterility of the device following autoclave sterilization.

### CONCLUSION(S):

The subject device has the same design considerations, assembly configurations, performance characteristics and indications for use as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

RoG Sports_Medicine,_Inc.__ % Mr. Curtis Raymond Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484

MAY 1 8 2011

Re: K110229

Trade/Device Name: RoG Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 12, 2011 Received: May 13, 2011

Dear Mr. Raymond:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

### Page 2 - Mr. Curtis Raymond

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

A.J. B. R.ta

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

510(k) Number (if known): K110229

Device Name: RoG Suture Anchor

The RoG 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications:- Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.
- Foot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- . Elbow indications .- Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair.
- Knee indications:- Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure.

The RoG 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows:

- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications:- Scapholunate ligament reconstruction. ●
- Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction.
- . Knee indications:- Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair. iliotibial band tenodesis, joint capsule closure. extracapsular repair, vastus medialis obliquus (VMO) muscle advancement.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |
| AND/OR Over-The-Counter Use |   |
| (21 CFR 801 Subpart C)      |   |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

D.M.Melkman

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K110229

Page 112 of 112

---

**Source:** [https://fda.innolitics.com/device/K110229](https://fda.innolitics.com/device/K110229)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
