ZOLL AED PLUS WITH 2010 AHA GUIDLINES SOFTWARE UPDATE
K110154 · Zoll Medical Corporation, World Wide Headquarters · MKJ · Feb 17, 2011 · Cardiovascular
Device Facts
Record ID
K110154
Device Name
ZOLL AED PLUS WITH 2010 AHA GUIDLINES SOFTWARE UPDATE
Applicant
Zoll Medical Corporation, World Wide Headquarters
Product Code
MKJ · Cardiovascular
Decision Date
Feb 17, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Intended Use
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by: - Unconsciousness and - Absence of normal breathing and - Absence of a pulse or signs of circulation. When a victim is less than 8 years of age, or weighs less thank 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
Device Story
Portable, battery-powered semi-automatic external defibrillator; uses voice prompts/visual icons to guide cardiac arrest rescue. Inputs: ECG rhythm via electrodes; CPR compression depth via accelerometer in electrodes. Operation: Analyzes ECG for shockable/non-shockable rhythms; prompts user to deliver energy if shockable; prompts CPR if non-shockable. Provides real-time CPR feedback: metronome for 100 compressions/min frequency; visual depth indicator for compression depth. Updated software aligns with 2010 AHA Guidelines: increases minimum compression depth threshold to 2.0 inches; adds 'Fully Release' chest prompt. Used by trained responders in clinical/field settings. Output: Defibrillation energy, audio/visual guidance. Benefits: Standardized, guideline-compliant CPR and defibrillation therapy.
Clinical Evidence
Bench testing only. Performance and safety testing verified that the software update meets functional requirements and industry safety standards, demonstrating substantial equivalence to the predicate device.
Technological Characteristics
Portable, battery-powered semi-automatic external defibrillator. Uses Rectilinear Bi-Phasic defibrillation waveform. Includes accelerometer-integrated electrodes for CPR depth monitoring. Software updated to reflect 2010 AHA Guidelines (2.0-inch depth threshold, chest recoil prompt).
Indications for Use
Indicated for suspected cardiac arrest victims (unconscious, no breathing, no pulse/signs of circulation). Pediatric use (under 8 years or <55 lbs) requires pediatric electrodes.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
K120406 — ZOLL FULLY AUTOMATIC AED PLUS · Zoll Medical Corporation, World Wide Headquarters · Oct 26, 2012
K110526 — ZOLL AEDPRO WITH 2010 AHA GUIDELINES SOFTWARE UPDATE · ZOLL Medical Corporation · Mar 25, 2011
K110168 — ZOLL E SERIES WITH 2010 AHA GUIDLINES SOFTWARE UPDATE · Zoll Medical Corporation, World Wide Headquarters · Feb 17, 2011
K142709 — samaritan PAD 450P · Heartsine Technologies, Inc. · Mar 27, 2015
K041892 — ZOLL AED PRO EXTERNAL DEFIBRILLATOR · ZOLL Medical Corporation · Feb 4, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for ZOLL Medical Corporation. The logo features the word "ZOLL" in large, bold, black letters. Below the word "ZOLL" is the tagline "Advancing Resuscitation. Today." in a smaller font. There is a curved line below the tagline.
K110154
ZOLL Medical Corporation Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824 U.S.A
FFB 17 201
## 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Eileen M. Boyle (978) 421-9171
Date Summary Prepared:
January 14, 2011
Device:
ZOLL AED Plus with 2010 AHA Guidelines Software Update
## Classification:
Automatic External Defibrillators: Class III (21 CFR 870,5310) Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200)
## Description:
The ZOLL AEDPlus was cleared by the agency under 510(k) application K033474. The device is a lightweight, portable, battery-powered semi-automatic external defibrillator that uses voice prompts and visual icons to guide a user through a cardiac arrest rescue. The device utilized the ZOLL Rectilinear Bi-Phasic defibrillation waveform. The device is designed to be used by trained responders for the treatment of cardiac arrest.
When connected with ZOLL AED Plus defibrillation electrodes to a patient, the device will analyze the electrocardiographic (ECG) rhythm of the patient and detect whether the rhythm is shockable or non-shockable. If the device detects a shockable rhythm, the device charges the capacitor, enables the treatment button and prompt the user to deliver the defibrillation energy to the patient. If the device detects a non-shockable rhythm, the device will prompt the user to begin CPR. The electrodes used with the device incorporates an accelerometer that mesures the depth of CPR compressions. This information is used by the device to provide feedback to the user and encourage
{1}------------------------------------------------
the user to administer CPR in compliance with the American Heart Association (AHA) Guidelines. The device provides feedback in the form of a metronome (to encourage the proper CPR frequency of 100 compressions per minute) and a visual depth indicator on the display (to encourage the recommended compression depth).
The recent American Heart Association (AHA) 2010 Guidelines (see attached) changed it's previous recommendation for a minimum depth of CPR compressions from 1.5 inches to a depth of at least 2.0 inches. The Guidelines also advise caregivers to allow full recoil of the patient's chest during CPR. As a result we are modifying the device's software from it's current CPR depth monitoring of 1.5" to at least 2.0" and provide an additional text user prompt to remind users to fully release the patient's chest during CPR.
The specific changes include:
- changing the depth indicator on the device screen from 1.5 inches to 2.0 inches -.
- prompting the user to "push harder" when compressions are less than 2.0 inches . instead of less than 1.5 inches in the previous release
- adding a text prompt to remind the user to "Fully Release" the patient's chest . during CPR
#### Intended Use:
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:
- Unconsciousness and
- Absence of normal breathing and .
- Absence of a pulse or signs of circulation ●
When a victim is less than 8 years of age, or weighs less thank 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
#### Substantial Equivalence:
The features and functions of the proposed ZOLL AED Plus (with 2010 AHA Guidelines for CPR software update) are substantially equivalent to the currently marketed ZOLL AED Plus (K033474, cleared for use on 5/21/2004).
#### Comparison of Technological Characteristics
The technological characteristics of the proposed ZOLL AED Plus (with 2010 AHA Guidelines software update) are substantially equivalent to the currently marketed ZOLL AED Plus (K033474, cleared for use on 5/21/2004).
## Performance Testing:
Extensive performance testing ensures that the ZOLL AED Plus (with 2010 AHA Guidelines software update) performs as well as the indicated predicate device and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards:
{2}------------------------------------------------
# Conclusion
Performance and safety testing of the ZOLL AED Plus (with 2010 AHA · Guidelines software update) demonstrates that its features and functions are substantially equivalent to those of the indicated commercially distributed predicate device with regard to performance, safety and effectiveness.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Eileen M. Boyle Regulatory Affairs Specialist Zoll Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105
FEB 17 20:1
Re: K110154
Trade/Device Name: Zoll AED Plus with 2010 AHA Guidelines Software Update Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: January 14, 2011 Received: January 19, 2011
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be omrois:
{4}------------------------------------------------
Page 2 - Ms. Eileen M. Boyle
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W.M.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Section 4 - Indications for Use
KITOS
510(k) Number (if known): Device Name: ZOLL AED Plus
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:
- · Unconsciousness and
- · Absence of normal breathing and
- · Absence of a pulse or signs of circulation
When a victim is less than 8 years of age, or weighs less thank 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
W.Moe
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K110154
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.