K110135 · Smith & Nephew, Inc. · LPH · Oct 14, 2011 · Orthopedic
Device Facts
Record ID
K110135
Device Name
POLARCUP DUAL MOBILITY SYSTEM
Applicant
Smith & Nephew, Inc.
Product Code
LPH · Orthopedic
Decision Date
Oct 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The POLARCUP® Dual Mobility System is indicated for: - Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or . rheumatoid arthritis. - Fracture or avascular necrosis of the femoral head . - Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement - . All forms of osteoarthritis - . Patients with hips at risk of dislocation - Femoral neck fracture or proximal fracture to hip joint . The POLARCUP® Dual Mobility System is intended for single use only and depending on its version is to be implanted either with or without bone cement.
Device Story
Acetabular component for total hip arthroplasty; dual mobility design features metal shell with polished inner surface and articulating plastic liner. Increases intra-prosthetic stability to reduce dislocation risk. Implanted by orthopedic surgeons in clinical/surgical settings. Subject device expands size range and adds highly crosslinked polyethylene liners to previously cleared system. Provides mechanical hip joint reconstruction; benefits patients requiring revision or at high risk of recurrent dislocation.
Clinical Evidence
Bench testing only. Evaluated via wear testing, coating characterization, range of motion testing, stress analysis, and fatigue properties to confirm ability to withstand in vivo loading.
Technological Characteristics
Metal shell with polished inner surface; highly crosslinked polyethylene liner. Dual mobility articulation. Uncemented or cemented versions. Single-use. Class II device.
Indications for Use
Indicated for patients with advanced hip degeneration (arthritis, post-traumatic), femoral head fracture/avascular necrosis, failed prior hip surgery, osteoarthritis, or high risk of dislocation (including elderly).
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
# K110135
OCT 1 4 2011
# > We are smith&nephew
### Smith & Nephew, Inc. Summary of Safety and Effectiveness
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Date of Summary: | October 10, 2011 |
| Contact Person<br>Manager | Theresa Leister, Regulatory Affairs Project |
| | T (901) 399-5899<br>F (901) 566-7816 |
| Name of Device: | POLARCUP® Dual Mobility System |
| Common Name: | Acetabular Component |
| Device Classification Name and Reference: | 21 CFR 888.3358 Hip joint<br>metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis |
| | 21 CFR 888.3390 Hip joint femoral (hemi-hip)<br>metal/polymer cemented or uncemented<br>prosthesis. |
| | 21 CFR 888.3350 Hip joint metal/polymer<br>semi-constrained cemented prosthesis. |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | LPH, KWY, JDI |
| Predicate Devices: | POLARCUP Dual Mobility System<br>510(k): K070278<br>Product Codes: LPH, KWY |
| | VERSAFIT Double Mobility System<br>510(k): K083116 |
#### Device Description
The POLARCUP Dual Mobility System consists of a metal shell and plastic liner (or insert). The inside of the metal shell is polished, and the outside of the plastic liner is able to articulate against this polished surface. This dual mobility design results in higher intra-prosthetic stability to address the treatment of patients with a high risk of dislocation (especially for elderly patients) or patients with
Product Code: MEH
{1}------------------------------------------------
K110135 (pg. 2 of 2)
recurrent dislocation. The subject device is identical to the previously cleared POLARCUP® Dual Mobility System with the exception of an increase in the size range and the addition of Highly Crosslinked Polyethylene Liners.
#### Technological Characteristics
A review of the mechanical data indicates that the POLARCUP® Dual Mobility System is capable of withstanding expected in vivo loading without failure.
#### Intended Use
The POLARCUP® Dual Mobility System is indicated for:
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or . rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head .
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, � hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
- . All forms of osteoarthritis
- . Patients with hips at risk of dislocation
- Femoral neck fracture or proximal fracture to hip joint .
The POLARCUP® Dual Mobility System is intended for single use only and depending on its version is to be implanted either with or without bone cement.
#### Substantial Equivalence Information
The overall design, materials, and indications for use for the POLARCUP® Dual Mobility System are substantially equivalent to the commercially available predicate devices identified.
Wear Testing, Coating Characterization, Range of Motion Testing, Stress Analysis and Faligue Properties were evaluated for the determination of substantial equivalence.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
OCT 1 4 2011
Smith & Nephew, Inc. Orthopaedic Division % Ms. Theresa Leister Regulatory Affairs Project Manager 1450 East Brooks Road Memphis. Tennessee 38116
Re: K110135 Trade/Device Name: POLARCUP Dual Mobility System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, KWY, JDI Dated: October 10, 2011 Received: October 13, 2011
Dear Ms. Leister:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Theresa Leister
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin D. Keith
Jor Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Premarket Notification Indications for Use Statement
510(k) Number (if known): _K110135
Device Name: POLARCUP® Dual Mobility System
Indications for Use:
The POLARCUP® Dual Mobility System is indicated for:
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or . rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head ●
- Failure of previous hip surgery: joint reconstruction, internal fixation, . arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
- All forms of osteoarthritis .
- Patients with hips at risk of dislocation .
- Femoral neck fracture or proximal fracture to hip joint .
The POLARCUP® Dual Mobility System is intended for single use only and depending on its version is to be implanted either with or without bone cement.
Prescription Use X AND/OR Over-the-Counter Use (Part 21 CFR 801.109) (Optional Format 1-2-96)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael Quay for MXM
of Surgical, Orthopedic,
K n 0135
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.