HL568 Self-Monitoring Blood Glucose (SMBG) System is self-test medical device and intended for single patient home-use to monitor the blood glucose (β-D-glucose) levels in quantitative measurement from fresh capillary whole blood obtained from the finger tip. It is intended for use outside the body (in vitro diagnostic use) by diabetics at home to measure the glucose concentration for aiding diabetes management. It is not intended for the diagnosis of or screening for diabetes mellitus. HL568 Self-Monitoring Blood Glucose (SMBG) System is intended to be used by a single person and should not be shared and it is also not intended for use on neonates and should be used with HL568 Blood glucose test strip and HL568 Control Solution. HL568 Blood glucose test strips are for use with the HL568 Blood glucose meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip. HL568 Control Solutions are for use with the HL568 Blood glucose meter and HL568 Blood glucose test strips to check that the meter and test strips are working together properly and that the test is performing correctly. The device is capable of transferring the storage data to the connected personal computer (PC) via USB cable. In addition, HL568 Self-Monitoring Blood Glucose System is featured with audio function; it could help the user to know the measured result by hearing but is not intended for visually impaired users.
Device Story
System includes portable, battery-operated glucose meter, test strips, control solutions, lancing device, and lancets. User applies capillary whole blood from fingertip to test strip; electrochemical biosensor (glucose oxidase) generates current proportional to glucose concentration. Meter measures current, displays result on LCD, and provides audio output of results. Data can be transferred to PC via USB. Intended for single-patient home use to aid diabetes management. Device requires manual coding via test strip label. Benefits include convenient, quantitative glucose monitoring for diabetes management.
Clinical Evidence
Bench testing only. Device demonstrated compliance with ISO 15197:2003 requirements for blood-glucose monitoring systems. Additional testing included precision (CLSI EP05-A2), linearity (CLSI EP06-A), interference (CLSI EP07-A2), and method comparison/bias estimation (CLSI EP09-A2).
Technological Characteristics
Electrochemical biosensor; glucose oxidase method; ferricyanide mediator. Test range 20-600 mg/dL; test time 5 seconds; sample volume >1µL. Battery-operated; LCD display; USB connectivity. Compliant with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for use by diabetics at home for quantitative measurement of blood glucose levels from fresh capillary whole blood obtained from the fingertip to aid in diabetes management. Not for use on neonates, not for diagnosis or screening of diabetes, and not intended for visually impaired users.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch Ultra Blood Glucose Monitoring System (k062195)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k110074
B. Purpose for Submission:
New device
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative, amperometric, electrochemical biosensor, Glucose Oxidase
E. Applicant:
Health and Life Co., Ltd.
F. Proprietary and Established Names:
HL568 Self-Monitoring Blood Glucose System
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGA
glucose oxidase, glucose | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| NBW
system, test, blood glucose, over the counter | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| JJX
single (specified) analyte controls
(assayed and unassayed) | Class I, reserved | 21 CFR § 862.1660 | Clinical Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
HL568 Self-Monitoring Blood Glucose (SMBG) System is self-test medical device and intended for single patient home-use to monitor the blood glucose (β-D-glucose) levels in quantitative measurement from fresh capillary whole blood obtained from the finger tip. It is intended for use outside the body (in vitro diagnostic use) by diabetics at home to measure the glucose concentration for
{1}
aiding diabetes management. It is not intended for the diagnosis of or screening for diabetes mellitus.
HL568 Self-Monitoring Blood Glucose (SMBG) System is intended to be used by a single person and should not be shared and it is also not intended for use on neonates and should be used with HL568 Blood glucose test strip and HL568 Control Solution.
HL568 Blood glucose test strips are for use with the HL568 Blood glucose meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertip.
HL568 Control Solutions are for use with the HL568 Blood glucose meter and HL568 Blood glucose test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
The device is capable of transferring the storage data to the connected personal computer (PC) via USB cable. In addition, HL568 Self-Monitoring Blood Glucose System is featured with audio function; it could help the user to know the measured result by hearing but is not intended for visually impaired users.
3. Special conditions for use statement(s):
- For Over-the-Counter Use only
- Not for neonatal use
- Not for screening or diagnosis of diabetes mellitus
- Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar patients
- Not intended for alternative site testing
- For single-patient use only
4. Special instrument requirements: HL568 Blood Glucose Meter
I. Device Description:
The HL568 Self-Monitoring Blood Glucose System is comprised of the HL568 Blood Glucose Meter, HL568 Blood Glucose Test Strip, HL568 Control Solution (2 levels), a lancing device and lancets. All the measured values can be read in one LCD panel. The blood glucose meter is portable and battery operated. The device is a self-monitoring IVD medical device which is intended for quantitative measurement of blood glucose (β-D-glucose) levels from fresh capillary whole blood obtained from the fingertip and used outside the body only (in vitro diagnostic use). It is intended for over-the-counter, home use by a single patient to measure the glucose concentration for aiding diabetes management.
2
{2}
J. Substantial Equivalence Information:
| Predicate device name | Predicate 510(k) number |
| --- | --- |
| OneTouch Ultra Blood Glucose Monitoring System | k062195 |
Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Proposed Device HL568 SMBG System | Predicate Device (k062195) |
| Indications for Use | Same | For the quantitative measurement of glucose in fresh capillary whole blood. |
| Intended User | For single patient, home use. | Healthcare professional and home use. |
| Enzyme | Same | glucose oxidase |
| Detection Method | Same | Amperometric method |
| Measurement Range | Same | 20-600 mg/dL |
| Hematocrit Range | Same | 30-55% |
| Sample Type | Same | Capillary whole blood |
| Sample Sites | Fingertip. | Fingertip, forearm or palm. |
| Sample Volume | Same | >1 μl |
| Test Time | Same | 5 seconds |
| Altitude Limit | Same | 10,000 feet |
| Operating Temperature Range | 50-104°F | 43-111°F |
| Operating Humidity | 10-80% | 10-90% |
| Memory | 500 measurements | 150 measurements |
| Audio Function | Voice Adding | No |
| Data Transmission | USB Interface | No |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods.
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement.
CLSI EP07-A2: Interference Testing in Clinical Chemistry.
CLSI EP09-A2: Method Comparison and Bias Estimation Using Patient Samples.
FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005.
FDA Guidance: Total Product Cycle Life for Portable Invasive Blood Glucose Monitoring System, October 24, 2006.
IEC 60601-1:1988 + A1:1991 + A2:1995 Medical electrical equipment - Part 1: General requirements for safety.
IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
{3}
ISO 15197:2003 In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
L. Test Principle:
The test principle is based on electrochemical biosensor technology using glucose oxidase methodology. Each test strip reacts with glucose in the blood sample to produce a current proportional to the blood glucose level. This reaction is measured by the meter and displayed as the blood glucose result.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
To evaluate the precision of the HL568 Self-Monitoring Blood Glucose System, within-run and total-run studies were performed. For the within-run precision study, ten measurements were obtained with ten glucose meters using two test strip lots on six levels of venous blood samples (n = 100 measurements on each test strip at each level).
Within-Run Precision Results
| Concentration (mg/dL) | Venous Whole Blood Samples | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Strip Lot 1 | | Test Strip Lot 2 | | Overall | |
| | SD | % CV | SD | % CV | SD | % CV |
| 40 | 4.6 | 9.7 | 4.6 | 10.7 | 4.6 | 10.3 |
| 80 | 3.2 | 3.9 | 3.3 | 4.0 | 3.3 | 4.0 |
| 130 | 4.8 | 3.8 | 5.1 | 4.1 | 4.9 | 3.8 |
| 200 | 7.6 | 3.8 | 7.5 | 3.9 | 7.6 | 3.8 |
| 325 | 12.8 | 3.8 | 13.0 | 3.8 | 13.1 | 3.9 |
| 500 | 18.0 | 3.4 | 19.6 | 3.8 | 18.9 | 3.6 |
For the total-run precision study, ten measurements were obtained with ten glucose meters using two test strip lots on four levels of glucose control solutions during the ten day study (n = 100 measurements on each test strip at each level).
Total-Run Precision Results
| Concentration (mg/dL) | Control Solutions | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Strip Lot 1 | | Test Strip Lot 2 | | Overall | |
| | SD | % CV | SD | % CV | SD | % CV |
| 32 | 3.3 | 10.3 | 3.0 | 9.8 | 3.3 | 10.4 |
| 105 | 3.9 | 3.7 | 4.3 | 4.1 | 4.3 | 4.1 |
| 313 | 11.9 | 3.8 | 12.6 | 4.0 | 12.3 | 3.9 |
| 436 | 13.7 | 3.1 | 14.0 | 3.2 | 14.3 | 3.3 |
{4}
b. Linearity/assay reportable range:
The linearity of the device was evaluated using a protocol based on CLSI EP6-A. Testing was performed using venous blood samples at ten different blood glucose levels, ranging from 0 to 600 mg/dL, on two lots of test strips, by five meters. The YSI 2300 glucose analyzer was used as the reference method. The sponsor claimed a measuring range of 20 to 600 mg/dL for their proposed device. The results were reviewed and found to be acceptable.
Linear Regression Analysis:
Lot 1: y = 1.0728x - 2.4977, r² = 0.9936.
Lot 2: y = 1.0231x + 2.4535, r² = 0.9964.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Two levels of Control Solution (Level 1 and 2) are included with each test kit. For each glucose control solution, a reference measurement was taken using an YSI 2300 analyzer. Forty measurements were performed (4 bottles * 10 measurements/per bottle) for each glucose control solution and the mean value, SD and CV were calculated. Level 1 and 2 glucose solutions were found to have ranges of 89-111 mg/dL, mean of 100 mg/dL (YSI Value: 76 mg/dL) and 247-305 mg/dL, mean of 280 mg/dL (YSI Value: 248 mg/dL), respectively.
Stability characteristics of both levels of Glucose Control Solutions were determined using real time studies at 45°C and 90% humidity. The sponsor claimed open and unopened vial stabilities are 3 and 20 months, respectively.
For the test strips, open vial (stored at 40°C and 90% humidity) and unopened shelf-life (stored at 40°C and 90% humidity) stability were determined using real-time studies. The sponsor claimed an open vial stability of 90 days and an unopened shelf-life stability of 9 months for the test strips.
d. Detection limit:
Please see the linearity study data in Section M1b above.
e. Analytical specificity:
Interference testing was assessed by spiking various endogenous and exogenous substances at therapeutic and toxic levels into whole blood samples at three glucose concentrations (60, 250 and 500 mg/dL) that were prepared and divided into a test (dosed) pool and a control pool. Studies were conducted using five HL568 Blood Glucose Meters and two lots of test strips. Significant interference is defined as a bias ≥ 10% from the control group as measured on YSI 2300 analyzer. Interference was observed for the substances and concentrations shown in the table below:
{5}
| Compound | Concentration with > 10% interference |
| --- | --- |
| Acetaminophen | > 5 mg/dL |
| Ascorbic Acid | > 3 mg/dL |
| Dopamine | Not observed up to 0.1 mg/dL |
| Gentisic Acid | Not observed up to 2 mg/dL |
| Ibuprofen | Not observed up to 40 mg/dL |
| L-Dopa | > 2 mg/dL |
| Methyl-Dopa | > 2.5 mg/dL |
| Sodium Salicylate | Not observed up to 50 mg/dL |
| Tetracycline | Not observed up to 1.5 mg/dL |
| Tolbutamide | Not observed up to 100 mg/dL |
| Galactose | Not observed up to 20 mg/dL |
| Maltose | Not observed up to 10 mg/dL |
| Xylose | Not observed up to 20 mg/dL |
| Hydroxyurea | > 3.8 mg/dL |
| Bilirubin (unconjugated) | > 20 mg/dL |
| Cholesterol | > 500 mg/dL |
| Creatinine | Not observed up to 30 mg/dL |
| Triglycerides | Not observed up to 3,000 mg/dL |
| Uric acid | > 14 mg/dL |
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor conducted an accuracy study, HL568 Self-Monitoring Blood Glucose System versus YSI-2300 Glucose Analyzer, at two sites with a 127 specimens over 10 days. A healthcare professional collected fingertip capillary blood and tested the sample with HL568 Blood Glucose Meter. A venous blood sample was collected by the trained operator for hematocrit determination and analysis on the YSI 2300 analyzer. Samples spanned 23.7 to 571 mg/dL including 13 altered samples (five samples below 50 mg/dL and eight samples above 400 mg/dL).
Linear regression results were as follows: $y = 0.9869x + 3.1630$ , $r^2 = 0.9882$ , $n = 127$ .
{6}
| For glucose concentrations <75 mg/dL | | | |
| --- | --- | --- | --- |
| User | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| Trained Operator | 10/19 (53%) | 16/19 (84%) | 19/19 (100%) |
| For glucose concentrations >75 mg/dL | | | | |
| --- | --- | --- | --- | --- |
| User | Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| Trained Operator | 64/108 (59%) | 96/108 (89%) | 104/108 (96%) | 108/108 (100%) |
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): A user performance study was performed to compare the lay user self-test results (with fingertip) and the YSI method. The study was conducted at three sites with 205 subjects. Each lay user participant performed their own finger stick and tested their blood on the HL568 Blood Glucose meter using only the instructions in the Owner's Booklet and test strip insert. A trained operator then performed a second finger stick and tested the blood on the same meter. A venous blood sample was collected by the trained operator for hematocrit determination and analysis on the YSI 2300 analyzer. The total range of samples tested was 50 to 415 mg/dL. Linear regression results are presented below:
Linear Regression Analysis:
Lay user vs. YSI $y = 1.01x + 1.03, r = 0.982, n = 205$
Professional vs. YSI $y = 0.99x + 3.89, r = 0.988, n = 205$
Lay user vs. Professional $y = 0.96x + 4.64, r = 0.993, n = 205$
The study results met the ISO 15197 accuracy criteria where ninety-five percent (95%) of the individual glucose results fell within ±15mg/dL of the YSI results at glucose concentrations <75mg/dL and within ±20% at glucose concentrations ≥75mg/dL.
{7}
| For glucose concentrations <75 mg/dL | | | |
| --- | --- | --- | --- |
| User | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| Lay User | 9/15 (60%) | 15/15 (100%) | 15/15 (100%) |
| Professional | 9/15 (60%) | 13/15 (87%) | 15/15 (100%) |
| For glucose concentrations >75 mg/dL | | | | |
| --- | --- | --- | --- | --- |
| User | Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| Lay User | 77/190 (41%) | 135/190 (71%) | 164/190 (86%) | 185/190 (97%) |
| Professional | 84/190 (44%) | 146/190 (77%) | 168/190 (88%) | 187/190 (98%) |
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected glucose results for non-diabetic, non-pregnant adults are as follows:
| Time | Range (mg/dL) |
| --- | --- |
| Normal Fasting Range | 70 to 100 |
| Two hours after meals | < 140 |
Reference:
American Diabetes Associations: Standards of Medical Care in Diabetes-2010, Diabetes Care, Vol. 33, Suppl. 1, 2010, p.S13.
N. Instrument Names:
HL568 Blood Glucose Meter
O. Systems Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes ☐ or No ☐ X
{8}
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the fingertip, which is directly applied to the test strip.
5. Calibration:
The device must be coded with the code found on the current test strip label. No further calibration is required.
6. Quality Control:
There are two levels of glucose control solution (Level 1 and 2), which are included with the test kit, which can be run with this device. Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above:
- A usability study was performed to assess the readability of the labeling by recruiting 100 lay users (aged 18-70 yrs old) who were provided with the test kit containing labeling for the US market. Participants varied in age, education, country of origin, and were about evenly divided between men and women. These lay users also completed a questionnaire in response to whether the device is easy to use and whether the Instructions for use were written in a way that makes it easy to use. The majority of the users responded that the device is very easy to use.
- Flesch-Kincaid readability assessment was conducted and the results showed that the labeling (User Manual, test strip package insert and control solution package insert) were written at less than an 8th grade level.
- A sample volume study was performed to verify the test strip sample volume requirement and the test strip fill error requirement established for the HL568 BGMS. Whole blood adjusted to four glucose concentrations (50, 120, 350 and 550 mg/dL) were measured with five meters and two test strip lots. Blood
9
{9}
at each concentration was applied to strips at sample volumes of 0.6 μL, 0.8 μL and 1.0 μL. Protocols and acceptance criteria were provided and found to be acceptable. Insufficient volume of less than 1.0 μL will produce an error message which alerts the user to an insufficient sample volume.
- The device system is intended for single-patient use only. Disinfection efficacy studies using Hepatitis B surface antigen were performed on the materials comprising the meter and lancing device demonstrating complete inactivation of hepatitis B virus (HBV) using Clorox Germicidal Wipes (EPA registration # 67619-12). The sponsor demonstrated that there was no change in performance or in the external materials of the meter or the lancing device after 8,000 cleaning and 8,000 disinfection cycles designed to simulate 5 years of device use. Labeling has been reviewed for adequate instructions in validated cleaning and disinfection procedures.
- The data transmission capability, data transmission port, H &L Glucose Data Transmission Software and USB connection cable was evaluated in usability study and found to be acceptable. Additionally verification and validation of these functions were conducted and found to be acceptable.
- Electromagnetic Compatibility (EMC) testing was performed, all requirements were met and a certificate was issued to Health and Life Co., Ltd.
- **Hematocrit Study**: Venous blood samples with varying hematocrit levels (29%, 42% and 56%) were tested at four glucose concentrations (50, 120, 350 and 550 mg/dL) against the YSI-2300 Glucose Analyzer. Each glucose level/hematocrit combination was tested on five meters and two test strip lots. The differences of the glucose meter results at each hematocrit/glucose combination were calculated against YSI-2300 Glucose Analyzer results at 42% hematocrit. The bias relative to YSI was that all results fell within ±15 %. The results were reviewed and found to be acceptable. The sponsor claimed a hematocrit range of 30% to 55% for their proposed device.
- **Altitude Study**: A study was conducted to evaluate the effect of altitude on the HL568 BGMS. Using ten glucose meters and one test strip lot, venous blood samples at five glucose concentrations (50, 165, 270, 380 and 500 mg/dL) were measured at five elevations (167; 6,450; 8,088; 9,082; 10,073 and 10,309 feet above sea level). Each venous blood sample was also tested by the YSI 2300 analyzer. The meter readings obtained were compared to the YSI measurement at the same altitude and the percent bias was determined at each level against the YSI results. Bias was within ± 10% for up to 10,073 feet. Based on the data, the sponsor claims that the HL568 BMGS can be used at altitude up to 10,000 feet.
10
{10}
- Temperature and humidity studies: Studies were performed to demonstrate that the HL568 meter can be used at combinations of extreme temperatures from 10 to 40 °C and humidity conditions ranging from 38 to 82 %. The performance of the proposed device was evaluated on three venous blood samples (120, 250 and 450 mg/dL) using 5 meters and 2 test strip lots. Duplicate measurements were taken for each condition. The sponsor claimed a temperature range of 10 to 40°C and humidity range of 10-80% for their proposed device. The YSI 2300 glucose analyzer was used to check the glucose concentrations of the samples. The bias relative to YSI was that all results fell within ±10%. The results were reviewed and found to be acceptable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.