K110006 · Ophthalmic Imaging Systems · NFJ · May 18, 2011 · Radiology
Device Facts
Record ID
K110006
Device Name
OIS SYMPHONY IMAGE MANAGMENT SYSTEM
Applicant
Ophthalmic Imaging Systems
Product Code
NFJ · Radiology
Decision Date
May 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
OIS Symphony is a comprehensive software platform for the import, integration and review of patient data and clinical information in an eye care environment. OIS Symphony allows for the collection, management, enhancement and review of the patient demographics, image data, diagnostic data and clinical reports from a variety of medical devices through either a direct connection with the instruments or through computerized networks.
Device Story
OIS Symphony is a software platform for managing ophthalmic clinical data. It imports images, diagnostic data, and reports from various medical devices via direct connection or network. The system integrates this data into a centralized database, allowing clinicians to view, manage, and enhance patient information. Used in hospitals, ophthalmology clinics, and medical facilities, it is operated by physicians, optometrists, and technicians. The system provides a real-time viewing environment across network-connected workstations. By consolidating data from disparate instruments into a single interface, it facilitates clinical decision-making and patient management. Security is maintained via authorized user login and SSL encryption.
Clinical Evidence
No clinical data. The device underwent extensive bench testing and software validation to ensure functional specifications were met.
Technological Characteristics
Software-based image management system. Connectivity via network or direct instrument interface. Supports HL7 and DICOM standards for intercommunication with hospital systems. Security features include authorized user login and 256-bit or greater SSL encryption. Operates as a centralized platform for data display, management, and archiving.
Indications for Use
Indicated for use in eye care environments to collect, manage, enhance, and review patient demographics, diagnostic data, clinical reports, and images from various medical devices via direct connection or network. Intended for use by ophthalmologists, optometrists, and ophthalmology technicians/photographers.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
.
# 510(K) SUMMARY
# OIS EYESCAN PORTABLE MODULAR IMAGING SYSTEM
# 510(k) Number K_ / / / 00 / __________________________________________________________________________________________________________________________________________________
| Applicant's Name: | Ophthalmic Imaging Systems<br>221 Lathrop Way<br>Suite I<br>Sacramento, CA 95815<br>USA<br>Tel: (916) 646-2020<br>Fax: (916) 646-0207<br>e-mail:aambrose@oisi.com |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Andrea Ambrose / Gil Allon<br>221 Lathrop Way<br>Suite I<br>Sacramento, CA 95815<br>USA<br>Tel: (916) 646-2020<br>Fax: (916) 646-0207<br>e-mail:aambrose@oisi.com or gallon@oisi.com |
| Manufacturing Site: | 221 Lathrop Way<br>Suite I<br>Sacramento, CA 95815<br>USA |
| Date Prepared: | December 31, 2010 |
| Trade Name: | OIS Symphony |
| Classification Name: | CFR classification section 892.2050 (product Code NFJ). |
| Classification: | Class II Medical Device |
i
:
{1}------------------------------------------------
## Predicate Device:
OIS Symphony Image Management System is substantially equivalent to the following predicate devices:
(1) Zeiss Forum (K090439) Manufactured by Carl Zeiss Meditec Inc. is intended as a software system in storing, managing, and displaying patient data, diagnostic data, videos, and images from computerized diagnostic instruments through networks.
(2) Nidek Advanced Vision Information System (NAVIS) (K013694) manufactured by Nidek is a computer technology software that collects, stores, and maintains patient data information; providing real-time diagnostic patient information from a number of diagnostic ophthalmic instruments at any PC workstation. Navis contains core software that can be augmented by additional software modules, supporting the various applications used clinically.
Synergy (K093313) Manufactured by Topcon Medical Systems Inc. is a (3) comprehensive software platform intended for use in acquisition or importing. processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
## Comparison of Technology with Predicates:
The OIS Symphony employs the same, or similar, technological characteristics as the predicate devices.
#### Device Description:
The OIS Symphony System is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. OIS Symphony consists of image display via Symphony Web, Image Import via Symphony Import and integration of imaging performed in OIS WinStation.
{2}------------------------------------------------
OIS Symphony is a software solution consistent with the predicate devices listed herein, using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment.
Symphony provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.
## Intended Use / Indication for Use: -
OIS Symphony is a comprehensive software platform for the import, integration and review of patient data and clinical information in an eye care environment. OIS Symphony allows for the collection, management, enhancement and review of the patient demographics, image data, diagnostic data and clinical reports from a variety of medical devices through either a direct connection with the instruments or through computerized networks.
## Performance Standards:
None. There are no mandatory performance standards for this type of device.
## Test Data:
The OIS Symphony System has been subjected to extensive performance testing and validation before release. Final testing of OIS Symphony included various performance tests and software validation tests, designed to ensure that the device met all its functional specifications. See Section 9 for statement and declaration of conformity.
{3}------------------------------------------------
#### Substantial Equivalence:
OIS Symphony is a comprehensive platform that collects, processes, measures, analyzes, and manages patient data and clinical information in an eye care environment. OIS Symphony provides a real time viewing environment for the collection and management of the patient demographics, image data, and clinical reports from a variety of approved medical devices, either directly or via a computer network. Similarly the predicate devices, including Zeiss Forum, NAVIS, and Topcon Synergy, are platforms for the import, acquisition, storage, and review of clinical images and tests of the eye, produced via a network or direct connection to the clinical instrument. The intended operators of OIS Symphony, similar to those of the predicate devices, are physicians (ophthalmologists), optometrists, ophthalmology technicians/ photographers. OIS Symphony and the predicate devices are used in environments that include hospital departments, ophthalmology clinics, and related medical facilities.
## Conclusions:
The conclusions drawn from the above comparison to predicate devices is that OIS Symphony is substantially equivalent in safety and effectiveness to the predicate devices listed above.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is written in bold, black letters. A line is drawn underneath the text. To the left of the text is a graphic of a symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ophthalmic Imaging Systems % Ms. Andrea Ambrose Quality Manager 221 Lanthrop Way, Suite I Sacramento, CA 95864
MAY 18 2011
Re: K110006
Trade/Device Name: OIS Symphony Image Management System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: NFJ Dated: May 11, 2011 Received: May 16, 2011
Dear Ms. Ambrose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Debra Falls
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): ★ // 0 0 0 k
Device Name: OIS Symphony
Indications for Use:
OIS Symphony is a comprehensive software platform for the import, integration and review of patient data and clinical information in an eye care environment. OIS Symphony allows for the collection, management, enhancement and review of the patient demographics, image data, diagnostic data and clinical reports from a variety of medical devices through either a direct connection with the instruments or through computerized networks.
Prescription Use V (Per 21 C.F.R. 801 Subpart D) OR
Over-The-Counter Use (Optional Format Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number k110006
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.