REVEAL XT INSERTABLE CARDIAC MONITOR, MODEL 9529 AND REVEAL DX INSERTABLE CARDIAC MONITOR, MODEL 9528
K103764 · Medtronic, Inc. · DSI · May 4, 2011 · Cardiovascular
Device Facts
Record ID
K103764
Device Name
REVEAL XT INSERTABLE CARDIAC MONITOR, MODEL 9529 AND REVEAL DX INSERTABLE CARDIAC MONITOR, MODEL 9528
Applicant
Medtronic, Inc.
Product Code
DSI · Cardiovascular
Decision Date
May 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The Reveal XT/DX Insertable Cardiac Monitor is an implantable patient-activated and automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias. Patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain, that may suggest a cardiac arrhythmia. The Reveal XT/DX Patient Assistant is intended for unsupervised patient use away from a hospital or clinic. The Patient Assistant activates one or more of the data management features in the Reveal Insertable Cardiac Monitor: To verify whether the implanted device has detected a suspected arrhythmia or device related event. To initiate recording of cardiac event data in the implanted device memory.
Device Story
Leadless, subcutaneous implantable cardiac monitor (ICM) continuously monitors ECG via two integrated electrodes; automatically detects and records arrhythmias (AT/AF, bradyarrhythmia, asystole, ventricular tachyarrhythmia). Hand-held, battery-operated patient assistant enables patient-triggered event recording and status queries (XT model only) outside clinical settings. Device stores ECG and marker channel data for physician review. Output informs clinical decision-making regarding arrhythmia diagnosis and management. Benefits include long-term monitoring for transient symptoms and automated detection for asymptomatic patients.
Clinical Evidence
No clinical data provided. Evidence consists of bench and animal performance validation testing, system verification, and human factors testing to demonstrate safety and effectiveness.
Technological Characteristics
Leadless, subcutaneous implantable monitor; two-electrode sensing; battery-operated; telemetry-enabled via hand-held patient assistant. No changes to materials, hardware, or sterilization from previous cleared versions.
Indications for Use
Indicated for patients with clinical syndromes/situations at increased risk of cardiac arrhythmias or those experiencing transient symptoms (dizziness, palpitation, syncope, chest pain) suggesting cardiac arrhythmia.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Instromedix King of Hearts Express AF Recorder (K020825)
Submission Summary (Full Text)
{1}------------------------------------------------
K103764
p. 1/3
MAY - 4 2011
# 510(k) Summary
| Date Prepared: | 22 December 2010 |
|-----------------------|-----------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Cardiac Rhythm Disease Management<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112 |
| Contact: | Yukiko Corson<br>Regulatory Affairs Specialist |
| Telephone: | (763)526-3409 |
| Fax: | (651)367-0603 |
| E-mail: | yukiko.m.corson@medtronic.com |
| Proprietary Name: | Reveal XT Insertable Cardiac Monitor,<br>Model 9529<br>Reveal XT Patient Assistant 9539 |
| | Reveal DX Insertable Cardiac Monitor,<br>Model 9528<br>Reveal DX Patient Assistant 9538 |
| Common Name: | Insertable Cardiac Monitor |
| Device Classification | Class II, 21 CFR 870.1025, Arrhythmia<br>detector and alarm |
| Product Code: | DSI |
#### Summary of Substantial Equivalence
The intended use, design, materials and performance of the Reveal XT Model 9529 and Reveal DX Model 9528 Insertable Cardiac Monitors, Reveal XT Patient Assistant 9539and Reveal DX Patient Assistant 9538 are substantially equivalent to the following predicate devices:
- Medtronic Reveal XT Model 9529 Insertable Cardiac Monitor (ICM) originally cleared via . 510(k) reference K.071641 and the Reveal DX Model 9528 Insertable Cardiac Monitor originally cleared via 510(k) reference K071655 on November 21, 2007 (including patient assistants). The Reveal XT Model 9529 and Reveal DX Model 9528 enhancements previously cleared via 510(k) reference K082475 on November 7, 2008 (including patient assistants).
- . Instromedix King of Hearts Express AF Recorder - K020825, Cleared April 5,2002
#### Device Description
The Reveal XT Model 9529 and Reveal DX Model 9528 Insertable Cardiac Monitors (ICM) are designed to automatically record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial tachyarrhythmia/atrial fibrillation (AT/AF), bradyarrhythmia, asystole, or
{2}------------------------------------------------
K103764
p 2/3
(fast) ventricular tachyarrhythmia. The Reveal XT ICM provides storage of ECG and Marker Channel during patient-activated and automatically-detected (auto-activated) events. Auto activation may help to detect abnormal heart rhythms in patients who may not activate/trigger the ICM.
The Reveal XT ICM Model 9529 and Reveal DX Model 9528 are small, leadless devices that are typically implanted under the skin, in the chest. Two electrodes on the body of the device continuously monitor the patient's subcutaneous ECG.
The Reveal XT Patient Assistant Model 9539 and the Reveal DX Patient Model 9538 are hand-held, battery-operated telemetry devices that enable the patient to start recording cardiac information in the Reveal XT and DX ICMs, respectively after experiencing symptoms of a possible cardiac event. The Reveal XT Patient Assistant query function enables the patient to check the status of physician-programmed parameters. A query function enables the patient to check his or her device and receive notification when an arrhythmia has occurred or when the device status has changed.
#### Indications for Use
The Reveal XT and DX Insertable Cardiac Monitors are implantable patient-activated and automatically-activated monitoring systems that record subcutaneous ECG and are indicated in the following cases:
- � Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms such as dizziness, palpitation, syncope and chest . pain, that may suggest a cardiac arrhythmia.
The Patient Assistant is intended for unsupervised patient use away from a hospital or clinic. The Patient Assistant activates one or more of the data management features in the Reveal Insertable Cardiac Monitor:
- . To verify whether the implanted device has detected a suspected arrhythmia or device related event. *
- � To initiate recording of cardiac event data in the implanted device memory.
*only applicable to the Reveal XT Patient Assistant Model 9539
### Technological Characteristics
Intended use, design, materials, performance and technological characteristics are substantially equivalent to the predicate devices referenced.
## Summary of Testing
There are no changes to the hardware, materials, processes, packaging, shelf life and sterilization of the Reveal XT and Reveal DX Systems included in this submission. Therefore, electromagnetic compatibility (EMC), electrical safety, mechanical performance, component, biocompatibility, sterilization and packaging testing were not repeated since the 510(k) clearance via K071641, K071655.
{3}------------------------------------------------
K102764
p 3/3
Software and system verification testing, system validation testing, electrical design verification testing, human factors testing, and performance validation testing (bench and animal) were performed to demonstrate that the Reveal XT Model 9529 Insertable Cardiac Monitor met established performance criteria and to support equivalency with the reference predicate devices.
The results of the testing indicate that the Reveal XT/DX Insertable Cardiac Monitors, Model 9529 and 9528, Reveal XT/DX Patient Assistants Model 9539 and 9538 perform as intended and are safe for their intended use.
#### Conclusion
Medtronic considers the Reveal XT and Reveal DX Insertable Cardiac Monitor systems to be substantially equivalent to legally marketed predicate devices through the data and information presented. No safety or effectiveness issues were identified.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic, Inc. Cardiac Rhythm Disease Management (CRDM) c/o Ms. Yukiko Corson Regulatory Affairs Specialist Mounds View MN 55112
- 4 2011 MAY
Re: K103764
Reveal XT Insertable Cardiac Monitor, Model 9529; Trade/Device Name: Reveal XT Patient Assistane, Model 9539 Reveal DX Insertable Cardiac Monitor, Model 9528; Reveal DX Patient Assistane, Model 9538
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: Class II (two) Product Code: DSI Dated: March 31, 2011 Received: April 1, 2011
Dear Ms. Corson
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{5}------------------------------------------------
#### Page 2 - Ms. Yukiko Corson
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indication for Use
510(k) Number (if known): K10376 4 Device Name: Reveal XT Insertable Cardiac Monitor (9529) Reveal XT Patient Assistant (9539) Reveal DX Insertable Cardiac Monitor (9528) Reveal DX Patient Assistant (9538) Indications for Use: The Reveal XT/DX Insertable Cardiac Monitor is an implantable patientactivated and automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: . Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias. Patients who experience transient symptoms such as dizziness, . palpitation, syncope, and chest pain, that may suggest a cardiac arrhythmia. The Reveal XT/DX Patient Assistant is intended for unsupervised patient use away from a hospital or clinic. The Patient Assistant activates one or more of the data management features in the Reveal Insertable Cardiac Monitor: To verify whether the implanted device has detected a suspected . arrhythmia or device related event. * To initiate recording of cardiac event data in the implanted device . memory. *only applicable to the Reveal XT Patient Assistant Model 9539 Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(Division Sign-Off) Division of Cardiovascular Devices
K103764 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.