ERBEFLO CLEVERCAP HYBRID TUBING/CAP SETS FOR OLYMPUS MODELS 160 AND 180 SERIES SCOPES

K103696 · Erbe USA, Inc. · FEQ · Mar 22, 2011 · Gastroenterology, Urology

Device Facts

Record IDK103696
Device NameERBEFLO CLEVERCAP HYBRID TUBING/CAP SETS FOR OLYMPUS MODELS 160 AND 180 SERIES SCOPES
ApplicantErbe USA, Inc.
Product CodeFEQ · Gastroenterology, Urology
Decision DateMar 22, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ERBEFLO CleverCap Hybrid Tubing/Cap Sets provide sterile water and air from a single source to an endoscope for endoscopic procedures.

Device Story

The ERBEFLO CleverCap Hybrid Tubing/Cap Set is a sterile, disposable accessory for Olympus 160 and 180 series endoscopes. It consists of a cap that seals to a sterile water bottle and two tubing segments: an irrigation line and an air/water line. The cap connects to a water source; the irrigation tubing interfaces with an irrigation pump to deliver lavage fluid to the scope; the air/water tubing connects to the scope's air/water port to facilitate lens cleaning and air insufflation. The air/water tubing utilizes a tube-within-a-tube design where endoscope processor pressure drives water flow. The system includes backflow valves and clamps to manage flow. Used in clinical settings by medical professionals during endoscopic procedures, the device simplifies setup by consolidating water and air delivery from a single bottle. It provides improved flow rates and equivalent pressure performance compared to predicate systems, supporting effective visualization and irrigation during endoscopy.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility studies, flow rate measurements, backflow pressure testing, internal pressure testing, and durability assessments.

Technological Characteristics

Disposable, sterile tubing/cap set. Materials include PVC and silicone. Features include backflow valves, locking clamps, and tubing weight inserts. Designed for 24-hour use. Sterilized via Ethylene Oxide. Connects to designated irrigation pumps and Olympus 160/180 series endoscopes. Mechanical operation; no software or electronic components.

Indications for Use

Indicated for use in endoscopic procedures requiring the delivery of sterile water and air to an endoscope from a single source. Intended for use with designated Olympus Model 160 and 180 series scopes.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### ERBE USA Incorporated Traditional 510(k); ERBEFLO CleverCap™ Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes | | 510(k) SUMMARY | |------------------------------------|----------------------------------------------------------------------------------------------| | | MAR 2 2 2011 | | Submitted By: | ERBE USA, Inc. 2225 Northwest Parkway Marietta, GA 30067 Tel: 770-955-4400 Fax: 770-955-2577 | | Contact Person: | John Tartal QA/RA Manager | | Date Prepared: | December 16, 2010 | | Common Name: | Endoscopic Irrigation Tubing Set and Water Bottle Adaptor | | Trade/Proprietary Name: | ERBEFLO CleverCap™ Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes | | Classification Name: | Endoscopes And/or Accessories (21 CFR Part 876.1500) | | Product Code: | KOG | | Legally Marketed Predicate Device: | EndoGator® System, 510(k) Number K092429 and Endo SmartCap®, K093665 | EndoGator and Endo SmartCap are registered trademarks of Byrne Medical. Inc. ### Device Description: The ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes consist of two (2) tubing segments and a cap. The cap of a Set attaches with an air-tight seal to a water source (i.e., a sterile water bottle). Then from the water bottle, irrigation tubing of a Set interfaces with a designated pump and via ERBEFLO 2 connector accessories to a specified Olympus Scope for endoscopic lavage. The other segment, the air/water tubing, also coming from the same water bottle; connects to an air/water port of a specified Olympus® Scope for air intubation as well as lens cleaning (Note: The air/water tubing is a tube within a tube in which the endoscope's air/water processor pressures the bottle for water flow.). Both the irrigation as well as the air/water tubing segments have a backflow valve and a clamp to pinch off the tubing while not in use. Each Set is designed for use with designated irrigation pumps and have a air/water connector for specified Olympus® Scopes. The Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes are provided sterile and are disposable. {1}------------------------------------------------ # ERBE USA Incorporated Traditional 510(k): ERBEFLO CleverCap™ Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes Olympus is a registered trademark of the Olympus Corporation ### Intended Use: The ERBEFLO CleverCap Hybrid Tubing/Cap Sets provide sterile water and air from a single source to an endoscope for endoscopic procedures. ## Similarities and Differences of the Proposed Device to the Current Devices (Predicate Comparison/Substantial Equivalence): Similarities The ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes have the same basic intended use of both of the predicates combined. The proposed Sets have the same thread connections, cap, tubing segments, and back flow valve placement. The tubing of the Hybrid Sets also has comparable Durometer, Inner Diameters (I.D.s), and Outer Diameters (O.D.s) as the predicate devices. The proposed and predicate devices also both have the same locking clamp and tubing weight insert. The length of the air/water segment is the same with the Hybrid Sets as compared to the Endo SmartCap. The Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes have the same duration of use as the EndoGator System as well as Endo SmartCap (24 hour use). The proposed Sets and predicate devices use the same types of water bottles. pumps, endoscope connection accessories, and endoscope. Finally, the proposed and predicate devices are sterilized via Ethylene Oxide and disposable. Differences The predicate devices (i.e., the EndoGator System and Endo SmartCap) can also be used with additional pumps and endoscopes as well as have capability of using CO2 instead of air for intubation. To address these differences, the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes Notes On Use specifies designated pumps and Olympus® Scopes for use as well as only has instructions for use with the air/water processor of the Scopes. The types of materials used for the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes are similar to the predicates (EndoGator System and Endo SmartCap) but specific materials are slightly different. Therefore. biocompatibility of the specific materials for the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes was evaluated. See Section III, Product Data - Biocompatibility Study. The predicates are two separate products used to do the same function of the proposed device. The Hybrid Sets also use only one water source as compared to the predicates each having a respective water source. Additionally, the irrigation tubing of the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes is a longer single PVC piece as compared to EndoGator Tubing being a shorter 3segment piece with a silicone portion interfacing with a pump head (Note: The segment of the proposed device also has a clamp and the predicate Tubing does not.). Also, the air/water tubing of the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes has a back flow valve as compared {2}------------------------------------------------ ### ERBE USA Incorporated Traditional 510(k): ERBEFLO CleverCap™ Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes to the predicate Endo SmartCap that does not have a valve. Nonetheless with the differences in the physical and dimensional aspects of the proposed Sets as compared to the combined predicate devices; performance testing demonstrated improved flow rates as well as equivalent back flow pressure, internal pressure, and durability. See Section III, Product Data - Performance Testing. #### Conclusion: The ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes intended use is a part of both of the predicate's indications in the previously cleared 510(k)s. The proposed Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes have the same principles of operation and technological characteristics as the predicate devices combined. The duration of use for the proposed and predicates is the same. As compared to the predicates, the proposed Sets are constructed with the same type of materials as well as have enhanced flow rates as well as comparable back flow pressure, internal pressure, and durability characteristics. In conclusion, all the differences were verified or validated. As a result, the ERBEFLO CleverCap Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes did not adversely affect safety or effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center is an abstract symbol that resembles an eagle or a stylized human figure, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Mr. John Tartal QA/RA Manager ERBE USA, Inc. 2225 Northwest Parkway MARIETTA GA 30067. MAR 2 2 2"11 Re: K103696 Trade/Device Name: ERBEFLOW CleverCap™ Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FAJ Dated: March 3, 2011 Received: March 4, 2011 Dear Mr. Tartal: : : We have reviewed your Section 510(k) premarket notification of intent to market the device. referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898: In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {4}------------------------------------------------ ## Page 2 adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Hubert Lemmer, MD Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Ko35686 Device Name: ERBE USA, Inc.'s ERBEFLO CleverCap™ 2 Hybrid Tubing/Cap Sets for Olympus® Model 160 and 180 Series Scopes Indications For Use: The ERBEFLO CleverCap™ Hybrid Tubing/Cap Sets provide sterile water and air from a single source to an endoscope for endoscopic procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Helentenn Division Sign-Off Division Sign-Off) Di sion of Reproductive, Gastro-R l, and Urological Devices 510(k) Number Page 1 of 1 ,
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...