K103654 · Starmedtec GmbH · GEX · Apr 6, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103654
Device Name
VELA XL
Applicant
Starmedtec GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The vela® XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft and hard tissue in medical specialities including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Device Story
The vela® XL is a thulium laser system emitting radiation at approximately 1.9 µm. It consists of a laser unit with a control panel, a foot switch, and an optical application fiber. The device is operated by clinicians in surgical settings (open, laparoscopic, or endoscopic) to perform tissue management (incision, excision, resection, ablation, vaporization, coagulation, and hemostasis). Laser energy is delivered through the fiber to the target tissue. The system provides a surgical tool for various medical specialties, enabling precise tissue interaction to achieve clinical outcomes like tumor removal, lithotripsy, or hemostasis. It is a prescription-only device.
Clinical Evidence
Bench testing only. No clinical data required.
Technological Characteristics
Thulium laser system; wavelength ~1.9 µm; fiber optic delivery. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1, and IEC 62366. Software complies with IEC 62304. Regulatory class II.
Indications for Use
Indicated for surgical incision, excision, resection, ablation, vaporization, coagulation, and hemostasis of soft and hard tissue across multiple specialties including urology (e.g., BPH, tumors, lithotripsy), gastroenterology, thoracic/pulmonary, gynecology, ENT, dermatology, plastic surgery, general surgery, and arthroscopy/orthopedic surgery (including spinal).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lisa Laser Products – Revolix 120 Laser System (K070476)
Submission Summary (Full Text)
{0}------------------------------------------------
APR - 6 2011
Image /page/0/Picture/2 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol inside, followed by the word "starmedtec" in black text. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font size.
103654
510(k) summary of safety and effectiveness information according 21 CFR Part 807.87(h)
- 1. General Information:
- a. Applicant:
StarMedTec GmbH Kreuzstrasse 22 82319 Starnberg Germany +49815126861-0 (phone) +49815126861-35 (fax)
b. Contact: Gregor Weidemann November 16, 2010 Date Prepared: C.
- 2. Names:
- a. Device Name:
- vela® XL
- b. Common Name: Classification Name: C.
- d. Product code:
vela® XL Laser instrument, surgical, powered GEX
- 3. Predicate Device:
Lisa Laser Products – Revolix 120 Laser System (510(k) Number: K070476)
- 4. Product Description:
The vela® XL is thulium laser system, which emits laser radiation with a wavelength of approximately 1.9 µm. The laser power transfers an optical application fiber. The indications are dissection, ablation, resection and coagulation of tissue. The laser system consists of:
- laser system including control panel (user interface) -
- foot switch -
- application fiber
- 5. Indications for use:
The vela® XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft and hard tissue in medical specialities including:
Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery
### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coaqulation and hemostasis) including:
- Strictures of urethra and ureter -
- Bladder Neck Incisions (BNI) -
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for "starmedtec" with the tagline "innovative solutions." To the left of the company name is a black square containing a white target symbol. A horizontal line separates the company name from the tagline. The text is in a simple, sans-serif font.
1103654
- Ablation and resecetion of Bladder Tumors, Uretheral Tumors and Ureteral -Tumors
- Ablation of Benian Prostatic Hyperplasia (BPH) -
- Transurethral incision of prostate (TUIP) -
- Holmium Laser Resection of the prostate (HoLRP) -
- -Holmium Laser Enucleation of the prostate (HoLEP)
- Holmium Laser Ablation of the prostate (HoLap) -
- -Condvlomas
- ﮮ Lesions or external genitalia
#### Gastroenterology
Open and endoscopic gastroenterological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy -
- -Polyps
- Biopsy -
- Gall Bladder Calculi (Lithotripsy) -
- Biliary/Bile duct calculi (Lithotripsy) -
- -Ulcers
- Gastric ulcers -
- Duodenal ulcers -
- ー Non bleeding ulcers
- Pancreatitis -
- Hemorrhoids -
- Cholecystectomy -
- Benign and malignant neoplasm -
- -Angiodysplasia
- Colorectal cancer -
- -Telangiectasias
- Telangiectasias of Osler-Weber-Renu disease -
- Vascular malformation -
- Gastritis ー
- Esophagitis -
- -Esophageal ulcers
- Varices -
- Colitis -
- Mallory-Weiss tear -
- Gastric Erosions …
#### Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
- Laryngeal lesions ー
- Airway obstructions including carcinoma l
- Polyps and granuloma -
- Palliation of obstructing carcinoma of the tracheobroncial tree -
{2}------------------------------------------------
Image /page/2/Picture/38 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol inside, followed by the word "starmedtec" in a bold, sans-serif font. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller, lighter font. The phrase is underlined by a thin black line.
K103654
#### Gynaecology
Open and endoscopic gynaecological surgery (incision, excision, resection, ablation, vaporization, coaqulation and hemostasis):
- Intra-uterine treatment of sub mucous fibroids -
- Benign endometrial polyps and uterine septum by incision, excision and or ー vessel coagulation
- Soft tissue excision procedures such as excisional conisation of the cervix -
#### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coaqulation and hemostasis) including:
- Endonasal / sinus surgery —
- -Partial turbinectomy
- -Polypectomy
- Dacryocystorhinostomy —
- -Frontal sinustomy
- -Ethmoidectomy
- Maxillary antrostomy -
- Functional endoscopic sinus surgery -
- Lesions or tumours (oral, nasal, glossal, pharyngeal or laryngeal) -
- Tonsillectomy -
- -Adenoidectomy
#### Dermatology and Plastic Surgery
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue, muscosal tissue, fatty tissue and cartilaginous tissue (hard tissue) including:
- Basal Cell Carcinomas -
- Lesion of skin and subcutaneous tissue -
- -Skin tags
- -Plantar warts
#### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectomy 1
- Lysis of adhesion -
- Appendectomy ー
- Biopsv -
- -Skin incision
- Tissue dissection -
- Excision of external tumours and lesions -
- Complete or partial resection of internal organs, tumours, lesions -
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for starmedtec, which is written in lowercase letters. Below the company name is the phrase "innovative solutions". The number K103654 is written in a large font below the logo.
- Mastectomy i
- Hepatectomy
- Pancreatectomy -
- Splenectomy -
- Thyroidectomy —
- Parathyroidectomy -
- Herniorrhaphy -
- Tonsillectomy -
- Lymphadenectomy -
- -Partial nephrectomy
- Pilonidal Cystectomy -
- Resection of lipoma -
- Debridement of decubitus ulcer -
- Hemorrhoids …
- Debridement of statis ulcer -
#### Arthroscopy
Arthroscopic/orthopaedic surgery (excision, ablation, and coagulation of soft and cartilaginous (hard) tissue):
Ablation of soft and cartilaginous (hard) tissue in minimal invasive spinal surgery including
- Percutaneous laser disc decompression/disectomy —
- Foraminoplasty -
- Ablation and coagulation of soft vascular and non vascular tissue in minimally --invasive spinal surgery
- 6. Performance testing:
The vela® XL is tested according to following standards:
IEC 60601-1:1988 + A1:1991 + A2:1995 IEC 60601-1-2:2005 IEC 60601-1-4: 1996 (First Ed.) + Am.1: 1999 (Consolidated 1.1 Ed.) for use with IEC 60601-1 (1988), Amts 1 (1991) and 2 (1995)60601-1-6 IEC 60601-2-22:2005 IEC 60825-1:2007 IEC 62366:2007 IEC 62304:2007 The device also complies with European Medical Device 93/42/EEC + Amendment 2007/47/EC
- 7. Performance data
Laboratory testing was conducted to verify and validate that the vela® XL met all design specifications and is substantially equivalent to the predicate device. Clinical data: No clinical information is required.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "starmedtec" with the tagline "innovative solutions." The logo consists of a black square with a white target symbol on the left, and the company name in black text on the right. Below the company name is the tagline in a smaller font.
K103654
- 8. Conclusion:
The vela® XL is as effective and save as the predicate device. The vela® XL is substantially equivalent to the cited legally marked predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing segments.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
StarMedTec GmbH % Mr. Gregor Weidemann Kreuzstraß22 82319 Starnberg Germany
APR - 6 2011
Re: K103654 Trade/Device Name: vela XL Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 7, 2011 Received: March 9, 2011
Dear Mr. Weidemann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
Page 2 - Mr. Gregor Weidemann
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Pan 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Aing B. Rete
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol inside, followed by the word "starmedtec" in a sans-serif font. Below the word is a line, and below the line is the phrase "innovative solutions" in a smaller font.
K103654
pg 1 of 6
### Indications for Use
510(k) Number (if known): n.a.
Device Name: vela XL
Indications For Use:
The vela XL laser system including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft and hard tissue in medical specialities including:
Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CD Device Evaluation (ODE)
> (Division Sign-Of Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103654
{8}------------------------------------------------
### 510(k) vela XL Indication for Use Statement (section 4)
Image /page/8/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black letters on the right. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font.
# K 10365
Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Strictures of Urethra and Ureter
- Bladder Neck Incisions (BNI) ー
- Ablation and resecetion of Bladder Tumors, Uretheral Tumors and -Ureteral Tumors
- Ablation of Benign Prostatic Hyperplasia (BPH) -
- Transurethral incision of prostate (TUIP) -
- Holmium Laser Resection of the prostate (HoLRP) -
- Holmium Laser Enucleation of the prostate (HoLEP)
- Holmium Laser Ablation of the prostate (HoLap) -
- Bladder/Renal Calculi (Lithotripsy) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- Condylomas -
- Lesions or external genitalia
#### Gastroenterology
Open and endoscopic gastroenterological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy
- Polyps
- Biopsy -
- Gall Bladder Calculi (Lithotripsy) -
- Biliary/Bile duct Calculi (Lithotripsy ﺴﯩ
- Ulcers
- Gastric ulcers
- Duodenal ulcers
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-O Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103654
{9}------------------------------------------------
### 510(k) vela XL Indication for Use Statement (section 4)
starmedtec
K103654 pg
- Non bleeding ulcers -
- Pancreatitis -
- Hemorrhoids -
- Cholecystectomy ー
- Benign and malignant neoplasm -
- Angiodysplasia -
- Colorectal cancer -
- Telangiectasias -
- Telangiectasias of Osler-Weber-Renu disease -
- Vascular malformation ー
- Gastritis -
- Esophagitis -
- Esophageal ulcers -
- Varices -
- Colitis --
- Mallory-Weiss tear ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- Gastric Erosions
#### Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
- Laryngeal lesions l
- Airway obstructions including carcinoma
- Polyps and granuloma
- Palliation of obstructing carcinoma of the tracheobroncial tree
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NehRP Dyke for nxm
Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103654
{10}------------------------------------------------
### 510(k) vela XL Indication for Use Statement (section 4)
Image /page/10/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol inside, followed by the word "starmedtec" in a simple, sans-serif font. Below the word "starmedtec" is the phrase "innovative solutions" in a smaller font.
K103654
pg 4 of 6
#### Gynaecology
Open and endoscopic gynaecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis):
- Intra-uterine treatment of sub mucous fibroids -
- Benign endometrial polyps and uterine septum by incision, excision and or vessel coagulation
- Soft tissue excision procedures such as excisional conisation of the | cervix
#### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Endonasal / sinus surgery
- Partial turbinectomy -
- Polypectomy -
- Dacryocystorhinostomy -
- Frontal sinustomy -
- Ethmoidectomy -
- Maxillary antrostomy -
- Functional endoscopic sinus surgery -
- Lesions or tumours (oral, nasal, glossal, pharyngeal or laryngeal) ー
- Tonsillectomy -
- Adenoidectomy -
#### Dermatology and Plastic Surgery
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Dable for
vision of Surgical, Orthopedic. Restorative Devi
510(k) Number K103654
{11}------------------------------------------------
### 510(k) vela® XL Indication for Use Statement (section 4)
Image /page/11/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in a sans-serif font on the right. Below the word "starmedtec" is the phrase "innovative solutions" with a line above it.
K103654 pg 5 of 6
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue, muscosal tissue, fatty tissue and cartilaginous tissue (hard tissue) including:
- Basal Cell Carcinomas -
- Lesion of skin and subcutaneous tissue -
- Skin tags -
- Plantar warts -
#### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectomy ー
- Lysis of adhesion |
- -Appendectomy
- Biopsy -
- Skin incision -
- -Tissue dissection
- Excision of external tumours and lesions -
- Complete or partial resection of internal organs, tumours, lesions -
- Mastectomy -
- = Hepatectomy
- -Pancreatectomy
- Splenectomy -
- -Thyroidectomy
- ﺳﮯ Parathyroidectomy
- Herniorrhaphy -
- Tonsillectomy -
- Lymphadenectomy
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP. Ooster for maxim
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103654
{12}------------------------------------------------
### 510(k) vela® XL Indication for Use Statement (section 4)
Image /page/12/Picture/1 description: The image shows the logo for starmedtec. The logo consists of a black square with a white target symbol on the left, and the word "starmedtec" in black font on the right. Below the word "starmedtec" is the phrase "innovative solutions".
- Partial nephrectomy -
- Pilonidal Cystectomy --
- Resection of lipoma -
- Debridement of decubitus ulcer -
- Hemorrhoids -
- Debridement of statis ulcer -
#### Arthroscopy
Arthroscopic/orthopaedic surgery (excision, ablation, and coagulation of soft and cartilaginous (hard) tissue):
Ablation of soft and cartilaginous (hard) tissue in minimal invasive spinal surgery including
- Percutaneous laser disc decompression/disectomy -
- -Foraminoplasty
- Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive spinal surgery
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K103654
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.