← Product Code [JJY](/productcode/JJY) · K103601

# AUDIT MICROCV IMMUNOASSAY LINEARITY SET (K103601)

_Aalto Scientific, Ltd. · JJY · Jun 10, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K103601

## Device Facts

- **Applicant:** Aalto Scientific, Ltd.
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Jun 10, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Audit™ MicroCV™ Free T4 / Free T3 Linearity Set consists of five levels of human and bovine albumin based matrix. Each level contains the following analytes: Free T4 and Free T3. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents, and instruments. This product may also be used as a quality control material for these analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The product is intended for use with quantitative assays on automatic analyzers. The Audit™ Free T4 / Free T3 Linearity Set is “For In Vitro Diagnostic Use Only”.

## Device Story

Audit™ MicroCV™ Free T4 / Free T3 Linearity Set is a 5-level quality control solution; human and bovine albumin-based matrix; contains Free T4 and Free T3 analytes. Used in clinical laboratories to verify calibration, linear operating range, and reportable range of quantitative assays on automatic analyzers; also serves as quality control material. Levels A-E represent concentrations across the analytical range; Levels B-D are linear dilutions of A and E. Healthcare providers use the output to monitor instrument performance and ensure accuracy of patient test results. Benefits include improved diagnostic reliability through validated instrument linearity.

## Clinical Evidence

No clinical data. Bench testing only. Linearity verified using CLSI EP6-A guidelines; regression analysis performed on Biomerieux Mini-Vidas and Siemens Centaur analyzers. Stability (shelf life 24 months, open vial 5 days) confirmed with recovery within ±15%.

## Technological Characteristics

Matrix: Human and bovine serum albumin. Form: Lyophilized. Levels: 5. Storage: 2–8 °C. Analyte assignment: Repeat analysis on automated immunoassay analyzers. Linearity verification: CLSI EP6-A statistical approach. Connectivity: None (standalone control material).

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Audit™ MicroCV™ Immunoassay Linearity Set (k062668)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k103601

B. Purpose for Submission:
New device

C. Measurand:
Quality control materials for Free T3 and Free T4

D. Type of Test:
Not applicable

E. Applicant:
Aalto Scientific, Ltd.

F. Proprietary and Established Names:
Audit MicroCV™ FreeT4/FreeT3 Linearity Set

G. Regulatory Information:
1. Regulation section:
21CFR 862.1660, Quality control material
2. Classification:
Class I, reserved
3. Product code:
JJY, multi-analyte controls, all kinds
4. Panel:
Clinical Chemistry (75)

{1}

H. Intended Use:

1. Intended use(s):

See indications for use below

2. Indication(s) for use:

The Audit™ MicroCV™ Free T4 / Free T3 Linearity Set consists of five levels of human and bovine albumin based matrix. Each level contains the following analytes: Free T4 and Free T3. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents, and instruments.

This product may also be used as a quality control material for these analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The product is intended for use with quantitative assays on automatic analyzers. The Audit™ Free T4 / Free T3 Linearity Set is “For In Vitro Diagnostic Use Only”.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

Performance testing was conducted on the Biomerieux V12 Compact Mini-Vidas Immunoassay Analyzer (for Free T4) and the Siemens Centaur XP (for Free T3)

I. Device Description:

The Audit™ MicroCV™ Free T4 / Free T3 Linearity Set is a 5 level quality control solution set containing Free T4 / Free T3 analytes as the messurand. The base matrix is a human and bovine based serum albumin and the free T4 and free T3 in each control are the endogenous levels found in the human and bovine serum used to make each control. The linearity set is used to confirm the proper calibration, linear operating range, and reportable range of Free T4 / Free T3 analytes. Level A is near the lower limit level and Level E has concentrations near the upper limit of instruments. Levels B – D are related by linear dilution of Level A and Level E.

All human source materials used to produce this product have been tested for HbsAg, anti-HCV, HIV-1 and HIV-2 and found to be non-reactive by FDA cleared/approved tests.

J. Substantial Equivalence Information:

{2}

1. Predicate device name(s):

Audit™ MicroCV™ Immunoassay Linearity Set

2. Predicate K number(s):

k062668

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Indications for Use | The Audit™ MicroCV™ Free T4 / Free T3 Linearity Set is assayed quality control material consisting of human based serum. It is intended to simulate human patient serum samples to detect systematic analytical deviations of laboratory testing procedures | Same  |
|  Number of levels per set | 5 | Same  |
|  Kit Contents | 5 x 5 mL | Same  |
|  Matrix | Human and Bovine Albumin Serum | Same  |
|  Form | Lyophilized | Same  |
|  Storage | 2 – 8 °C for 24 months | Same  |
|  Open Bottle Stability | 2 – 8 °C for 5 days | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Analytes | Free T4 and Free T3 | Cortisol. Digoxin, Estradiol, Ferritin, Folate, Free T4, FSH, hCG, LH, Progesterone, Prolactin, Testosterone, Total PSA, Total T3, Total T4, TSH, and Vitamin B12  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI EP6-A: evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline

L. Test Principle:

Not applicable

{3}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

The Audit™ Micro CV™ Free T4 / Free T3 Linearity Set is prepared such that a linear relationship exists among all five levels (Levels A through E). Level A has a concentration near the lower limit level and Level E has concentrations near the upper limit level of instruments. Levels B-D are related by linear dilution of Level A and Level E using dilution schemes based on guidelines provided by CLSI EP6-A.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability and Value Assignment

All components of the Audit™ Micro CV™ Free T4 / Free T3 Linearity Set are obtained from a commercial vendor and inspected in-house.

Value assignment for each analyte was conducted by repeat analyses on the Biomerieux Mini-Vidas analyzer (Free T4) or the Siemens Centaur analyzer (Free T3). All control material levels for both analytes were measured 4 times (1 lot) and the mean values were used to assign the target concentration values for each of the five linearity set levels. The target ranges were calculated as ± 15% of the target value. The mean concentration values of each level were plotted (concentration value vs. assigned level) and a linear regression value obtained. If the five-point linear regression R² value was greater than 0.95 and if the plots are linear, the product passed the linearity test. Values may change depending on lot, instrumentation, methodology, and assay temperature. Lot-specific expected value ranges are provided in the package insert. Expected value ranges for each level of the linearity set are summarized in the table below.

Target Values:

|  Analyte | Units | A | B | C | D | E  |
| --- | --- | --- | --- | --- | --- | --- |
|  Free T4 | pmol/L | ≤ 10 | 30 | 50 | 70 | 90  |
|  Free T3 | pg/mL | ≤ 5 | 13.75 | 17.5 | 21.25 | 25  |

{4}

5

# Stability

Stability studies have been performed to determine the open vial stability and shelf life for the Audit™ MicroCV™ Free T4 / Free T3 Linearity Set. Real time studies are ongoing to support the shelf life of this product. Stability claims meet the stated acceptance criteria of recovery from unstressed material values of ± 15%, and are as follows:

Open Vial Stability: Once a vial has been reconstituted, all analytes will be stable for 5 days when stored tightly capped at 2-8 C.

Shelf Life: 24 months at 2 - 8° C.

d. Detection limit:

Not applicable

e. Analytical specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Not applicable

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

{5}

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

The expected values are provided in the labeling for each specific lot.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda.innolitics.com/device/K103601](https://fda.innolitics.com/device/K103601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
