CARPENTIER-EDWARDS PHYSIO TRICUSPID ANNULOPLASTY RING
Applicant
Edwards Lifesciences, LLC
Product Code
KRH · Cardiovascular
Decision Date
Jun 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model 6200, is intended for use in patients with tricuspid valvular insufficiency. It is intended to correct annular dilatation, increase leaflet coaptation, reinforce annular suture lines, and prevent further dilatation of the annulus.
Device Story
Carpentier-Edwards Physio Tricuspid annuloplasty ring (model 6200) is a prosthetic implant used during cardiac surgery to treat tricuspid valvular insufficiency. The device functions by providing structural support to the tricuspid valve annulus; it corrects annular dilatation, improves leaflet coaptation, and reinforces suture lines to prevent future dilation. It is implanted by a cardiac surgeon in an operating room setting. The device consists of a titanium core encased in a silicone sewing ring margin, covered with polyester cloth. It is a passive implantable device; it does not involve sensors, software, or active electronic components. Clinical benefit is derived from the mechanical stabilization of the valve annulus, restoring proper valve function and preventing disease progression.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included computational stress analysis, finite element analysis, tensile strength, suture retention, flexibility, ring/holder removal force, corrosion resistance, MR compatibility (ASTM F2182, F2052, F2119, F2213), comprehensive biocompatibility testing (ISO 10993-1), bioburden, pyrogen testing, packaging qualification (ISO 11607-1, ASTM D4169), and design validation by surgeons in a simulated use environment.
Technological Characteristics
Titanium core, silicone sewing ring margin, polyester cloth cover. Passive implantable ring. MR compatible per ASTM F2182, F2052, F2119, F2213. Biocompatibility per ISO 10993-1. Packaging per ISO 11607-1. Sterilization method not explicitly stated but implied by terminal sterilization requirements.
Indications for Use
Indicated for patients with tricuspid valvular insufficiency requiring correction of annular dilatation, increased leaflet coaptation, reinforcement of annular suture lines, and prevention of further annular dilatation.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
Edwards MC3® Tricuspid annuloplasty ring (K020864)
Carpentier-Edwards® Classic® Tricuspid annuloplasty ring (K912554)
Submission Summary (Full Text)
{1}------------------------------------------------
K103520
Image /page/0/Picture/1 description: The image shows a logo with the word "Edwards" written below a square graphic. The graphic contains a large letter "E" in white, overlaid on a grid pattern. Circular designs are visible behind the letter "E" and within the grid, adding a complex, layered effect to the logo. The overall design appears to be for a company or brand named Edwards.
510(k) Summary
:
JUN 2 4 2011
| Submitter: | Edwards Lifesciences LLC |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Daryl Richardson<br>Regulatory Affairs Associate<br>One Edwards Way<br>Irvine, CA 92614<br>daryl richardson@edwards.com<br>(ph) 949-250-5036 / (fax) 949-250-3630 |
| Date Prepared: | June 15, 2011 |
| Trade Name: | Carpentier-Edwards® Physio Tricuspid™ annuloplasty ring,<br>model 6200 |
| Classification Name: | Class II, CFR 870.3800 Annuloplasty Ring, KRH |
| Predicate Device(s): | Edwards MC3® Tricuspid annuloplasty ring (K020864)<br>Cosgrove-Edwards® annuloplasty system (K923367)<br>Carpentier-Edwards® Classic® Tricuspid annuloplasty ring (K912554) |
| Device Description: | The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model<br>6200, is constructed of a titanium core with a silicone sewing ring<br>margin covered with a polyester cloth. |
| ·Indications for Use: | The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model<br>6200, is intended for use in patients with tricuspid valvular<br>insufficiency. It is intended to correct annular dilatation, increase<br>leaflet coaptation, reinforce annular suture lines, and prevent further<br>dilatation of the annulus. |
| Comparative Analysis: | It has been demonstrated that the Carpentier-Edwards Physio<br>Tricuspid annuloplasty ring, model 6200, is comparable to its predicate<br>devices in design, intended use, materials, fundamental technology<br>and principal of operation. |
| Functional/Safety Testing: | The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model<br>6200, has successfully completed design verification and validation<br>testing. The following studies were conducted:<br>Computational stress analysis |
| | |
- Finite element analysis .
Page 1 of 3
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letter 'E' inside a square. The letter 'E' is white and bold, and it is surrounded by a circle. The background of the square is black and white. Below the square is the word 'Edwards' in a serif font.
- Tensile strength .
- Suture retention .
- . Flexibility
- Ring / Ring holder removal force .
- Corrosion resistance .
- Magnetic resonance compatibility (RF heating, image artifact, . displacement force, induced torque)
- Biocompatibility (percent inhibition of cell growth, medium . eluate method - cytotoxicity, blood compatibility, complement activation, gene mutation assay, chromosomal aberration effect assay, agar overlay - cytotoxicity, DNA damage / effects assay, Mouse systemic injection - systemic toxicity, rabbit pvrogen - systemic toxicity, rabbit intracutaneous reactivity, guinea pig maximization, rabbit intramuscular implantation sub-chronic and chronic evaluations)
- Relative resistance evaluation and comparison for sterilization .
- . Bioburden
- Limulus amebocyte lysate pyrogen testing .
- Packaging qualification (visual inspection, sterile barrier . integrity, dye leak, burst)
- Shelf-life assessments of packaging and product components .
- Design validation by surgeons in simulated use environment .
Standards used in support of the application
See the attached table
Conclusion:
The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model 6200, is substantially equivalent to its predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the word "Edwards" written below a square graphic. The graphic contains a white letter "E" superimposed over a grid pattern and circular shapes. The letter "E" is stylized and appears to be partially obscured by the circular elements within the square.
# Utilization of Standards
| Standard No. | Standard Title |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO<br>10993-1:2009 | Biological evaluation of medical device -<br>Part 1: Biological Testing |
| ASTM<br>F2182:2009 | Standard Test Method for Measurement<br>of Radio Frequency Induced Heating<br>Near Passive Implants During Magnetic<br>Resonance Imaging |
| ASTM<br>F2052:2006 | Standard Test Method for Measurement<br>of Magnetically Induced Displacement<br>Force on Medical Devices in the<br>Magnetic Resonance Environment |
| ASTM<br>F2119:2007 | Standard Test Method for Evaluation of<br>MR Image Artifacts from Passive<br>Implants |
| ASTM<br>F2213:2006 | Standard Test Method for Measurement<br>of Magnetically Induced Torque on<br>Medical Devices in the Magnetic<br>Resonance Environment |
| BS EN ISO<br>11607-1:2009 | Packaging for Terminally Sterilized<br>Medical Devices, Part 1: Requirements<br>for Materials, Sterile Barrier Systems<br>and Packaging Systems (ISO 11607-<br>1:2006) |
| ASTM<br>D4169:2005 | Standard Practice for Performance<br>Testing of Shipping Containers and<br>Systems |
| ISO<br>14971:2007 | Medical Devices - Application of risk<br>management to medical devices |
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white circular logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the top and left side of the circle. The eagle is composed of three curved lines that form the body and wings, and a smaller curved line that forms the head and beak.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Edwards Lifesciences, LLC c/o Mr. Daryl Richardson Regulatory Affairs Associate One Edwards Way Irvine, CA 92614
JUN 2 4 2011
Re: K103520
Carpentier-Edwards™ Physio Tricuspid™ Annuloplasty Ring, Model 6200 Regulation Number: 21 CFR 870.3800 Regulation Name: Annuloplasty ring Regulatory Class: Class II Product Code: KRH Dated: June 22, 2011 Received: June 23, 2011
Dear Mr. Richardson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{5}------------------------------------------------
### Page 2 - Mr. Daryl Richardson
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
R. D. Hinderaker, M.D.
Bram D. Auckerman, M.D. Director) Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known): 《103520
Device Name: Carpentier-Edwards Physio Tricuspid annuloplasty ring, model 6200
Indications for Use:
The Carpentier-Edwards Physio Tricuspid annuloplasty ring, model 6200, is intended for use in patients with tricuspid valvular insufficiency. It is intended to correct annular dilatation, increase leaflet coaptation, reinforce annular suture lines, and prevent further dilatation of the annulus.
Prescription Use: 冈 (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: [] (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
yascular Devices 510(k)
Page 1of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.