K103385 · Hologic, Inc. · LLZ · Feb 2, 2011 · Radiology
Device Facts
Record ID
K103385
Device Name
SECURVIEW DX DIAGNOSTIC WORKSTATION
Applicant
Hologic, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 2, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Hologic SecurView DX is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards. The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image. The SecurView DX is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
Device Story
SecurView DX is a multi-modality review workstation software for digital radiology images; focuses on mammography. Software accepts standard/multi-frame DICOM images (MG, MR, US, CT, SC, CR, PET, DR) via network or media. Operates on Windows-based PCs with high-resolution FDA-cleared displays, mouse, keyboard, and dedicated workflow keypad. Provides tools for window/level, magnification, pan, rotation, mirroring, inversion, and CAD result display. Supports breast MR CAD analysis and motion correction. Allows display of information/images calculated by separate FDA-cleared medical devices (e.g., AIE Region of Interest Image Enhancement). Used by radiologists and clinicians in clinical settings for image interpretation; facilitates clinical decision-making by providing optimized visualization of diagnostic images. Enables PACS connectivity for query/retrieve operations and DICOM printing.
Clinical Evidence
Bench testing only. No clinical data presented. Software modifications were validated through system verification and validation protocols, including risk analysis and design reviews, to ensure performance equivalent to the predicate.
Technological Characteristics
Software-based workstation; runs on Windows XP/7+ OS. Hardware requirements: Intel CPU (2.0 GHz+), 2.0 GB RAM+, 140 GB HDD+. Connectivity: DICOM standard (PS 3) for image exchange, PACS query/retrieve, and printing. Supports high-resolution mammography-cleared displays. Image processing is external; software displays 2D/3D images and CAD results. Quality system: ISO 13485 and 21 CFR Part 820.
Indications for Use
Indicated for use by trained professionals (physicians, radiologists, nurses, technicians) for the display, manipulation, and interpretation of multi-modality DICOM images, including screening and diagnostic mammography (lossless compressed or non-compressed).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
AIE Region of Interest Image Enhancement for Digital Mammography software (K062059)
Submission Summary (Full Text)
{1}------------------------------------------------
### 510(k) Summary SecurView DX
FEB - 2 2011
Product Name: SecurView DX
Product Classification Name: Picture archiving and communication system
Product Classification Code: LLZ CFR Section: 892.2050
Classification Panel: Radiology · Class II
.. . Manufacturer: Hologic, Inc. *** 35 Crosby Drive Bedford, MA 01730 USA
Contact Person: Gail Yaeker-Daunis Telephone Number: (203) 731-8337 Fax Number: (203) 731-8440
Date Prepared: January 5, 2011
Predicate Device: K062107 SecurView DX, Hologic Inc.
#### Predicate Device Description: ----
.. The Hologic SecurView DX is mainly a software product. It is used for visualization and manipulation of digital radiology images. The SecurView DX is a multi-modality review workstation software focused on mammographic applications. The software can drive high-resolution displays (FDA cleared for Mammography), a PC mouse, a keyboard, optional lower resolution monitors and dedicated workflow keypad.
The software accepts standard or multi-frame mammography images that have been created according to the "FOR PRESENTATION" specification of the DICOM Standard with no compression or using lossless compression. Image processing is external to the SecurView DX software.
The software accepts standard or multi-frame multimodality images of US, MR, DR, CR, SC, CT, PET, and other DICOM formats for display and manipulation on high resolution displays (FDA cleared for mammography) or on other lower resolution color displays in 2-D or 3-D view.
The SecurView DX software can be used in a single or in a multi-workstation configuration.
The minimum computer requirements to run the Secur View DX software are:
{2}------------------------------------------------
Hologic, Inc.
- ♥ Windows Operating System
- . Intel CPU with a clock rate of 2.0 GHz or greater
- 2.0 GB RAM or greater .
- 140 GB Hard drive or greater, operating at 160 Mb/s or greater .
- CD-ROM/R
- 10/100 Base -T Network Interface .
#### Intended Use:
The Hologic SecurView DX is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images using FDA cleared displays for screening and diagnostic mammography, as well as any other DICOM multi-modality image.
The SecurView DX software is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
#### Comparison with Predicate Device:
The modified SecurView DX is substantially equivalent to the SecurView DX previously cleared as K062107 as described above and has the same indications and intended use.
The comparison table below shows that the modified Secur View DX is substantially equivalent to the predicate with the exception of the following features:
- Ability to run on both Windows XP or Windows 7 or higher Operating System .
- Ability to display information or images calculated by separate medical devices . cleared by FDA, such as: AIE Region of Interest Image Enhancement for Digital Mammography software, cleared as K062059.
Image /page/1/Picture/18 description: The image is a black and white image that appears to be a scan of a document. The document contains some text, but it is difficult to read due to the low resolution and poor quality of the scan. The text appears to be arranged in a vertical format. There are also some other marks and smudges on the document.
" :
04
{3}------------------------------------------------
.
| | COMPARISON TABLE | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hardware | SecurView DX K062107 | Modified Device |
| Processor | Dual Processor Intel CPU<br>with clock rate of 2.0 GHz<br>or higher | Same |
| Memory | 2.0 GB RAM | Same or greater |
| Hard Disks | 140 GB total | Same or greater |
| Display | Multiple display systems<br>(display with graphics<br>controller) which has been<br>FDA cleared for<br>mammography specific<br>medical diagnostic<br>applications & Optional<br>Color High Contrast<br>Display for multimodality<br>image review other than<br>mammography | Same<br>Multiple display systems<br>(displays and graphics<br>controllers) which have been<br>FDA cleared for<br>mammography specific<br>medical diagnostic<br>applications and optional<br>additional displays for<br>multimodality image review<br>other than mammography |
| Networking | Network Interface Card<br>(NIC) 10/100 Base-T | Same or greater |
| Removable Media | CD-ROM/R | Same or greater |
| Keypad | Custom keypad single click<br>access to commonly used<br>features | Same |
| Operating System | Windows XP | Same (Windows XP) or<br>higher, Windows 7 |
| Memory | 2.0 GB RAM | Same or greater |
| Hard Disks | 140 GB total | Same or greater |
| Software | SecurView DX K062107 | Modified SecurView DX |
| Administrative<br>functions | A menu exists for administrator use<br>which includes access to functions<br>such as:<br>User account administration Setting system wide defaults Patient List administration | Same |
| Service<br>functions | A menu exists for a service user<br>which includes access to functions<br>such as:<br>DICOM settings License administration | Same |
| User Preference<br>selection | A menu exists for radiologist users<br>which includes access to functions<br>such as:<br>Setting individual preferences Setting personal workflow and<br>Reportflows (series of<br>configurable hanging snapshots<br>created as user preferences and<br>associated with different types<br>of studies | Same |
| Receiving<br>Exams | Images are transmitted to the<br>SecurView DX software according<br>to the DICOM standard.<br>Modalities supported include:<br>MG, MR, US, CT, SC, CR, PET,<br>DR, and multiframe mammography<br>images<br>After an exam has been received by<br>the software, an entry is added to<br>the patient list. | Same |
| Viewing Exams | All exams which are stored on the<br>system appear in a patient list. The<br>user selects exams from this list for<br>viewing. | Same |
| Printing | Images can be printed on DICOM<br>printers | Same |
| PACS Archive<br>Connectivity | The system can perform a<br>Query/Retrieve operation on a<br>DICOM archive to retrieve studies<br>of interest | Same |
| Image viewing<br>tools | The software provides tools for<br>image viewing and manipulation<br>such as:<br>Window / Level Magnification Window Display at variable<br>resolutions Image Pan Rotation Mirroring Image inversion Display of CAD results Display of patient<br>demographic information Multimodality Viewer has cine<br>mode tools for multiframe images,<br>breast MR CAD analysis tools and<br>breast MR motion correction | Same with addition of ability<br>to display information or<br>images calculated by separate<br>medical devices cleared by<br>FDA, such as:<br>AIE Region of Interest<br>Image Enhancement for<br>Digital Mammography<br>software, cleared as<br>K062059 |
| Annotation and<br>Measurement<br>tools | The software provides tools for:<br>Free hand drawing Ellipse drawing Ability to add textual<br>annotations Ability to measure features | Same |
| Send GSPS or<br>SC images to<br>other DICOM<br>nodes | The system can send grayscale<br>softcopy presentation state (GSPS)<br>objects or secondary capture<br>DICOM images to other DICOM<br>nodes | Same |
| Export Image | Capable of export to local storage<br>in TIFF format de-identified such<br>that image file content contains no<br>patient identifying information<br>Capable of exporting DICOM<br>format images | Same |
| Application<br>Synchronization<br>and Data<br>Exchange | Provides synchronization interface<br>to synchronize patient and patient<br>data with external applications such<br>as RIS, data tracking systems,<br>dictation systems, etc. | Same |
DISON TA r
.
05
{4}------------------------------------------------
. . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
{5}------------------------------------------------
. ' . *. 1.6 ! ;;
ト ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
.
{6}------------------------------------------------
# Safety and Effectiveness - Performance Standards:
The SecurView DX software is specified, validated and tested by Hologic, Inc., under a registered ISO 13485 and 21 CFR Part 820 Quality System.
The SecurView DX software conforms to the ACR/NEMA Digital Imaging Communications in Medicine PS 3 (DICOM) Standard. Hologic publishes a DICOM Conformance Statement.
A comprehensive Operator's Manual is provided with each system, to facilitate intended operation.
#### Conclusion:
The modified SecurView DX software is substantially equivalent to the predicate software in terms of intended use, indications of use and conformance to the DICOM standard. The Secur View DX software changes were specified, validated and tested by Hologic, Inc., under an ISO 13485 and 21 CFR Part 820 Quality System. The following quality assurance measures were applied to the device software modification:
- Risk Analysis .
- Requirements Reviews �
- Design Reviews .
- System Verification ●
- System Validation �
The modified software has successfully undergone bench testing designed to simulate clinical factors.
Based on the information supplied in this 510(k), Hologic concludes that the modified device is substantially equivalent to the predicate device and is safe and effective.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002
Ms. Gail Yaeker-Daunis, MBA, RAC Sr. Regulatory Affairs Specialist Hologic, Inc. 36 Apple Ridge Road DANBURY CT 06810
FEB - 2 2011
Re: K103385
Trade/Device Name: SecurView DX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 10, 2011 Received: January 12, 2011
Dear Ms. Yaeker-Daunis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean · · · that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
{8}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S Pastel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{9}------------------------------------------------
## Indications For Use
#### Premarket Notification: SecurView DX
510(k) No.
Device Name: SecurView D
#### Indications For Use
The Hologic SecurView DX is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image.
The SecurView DX is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Division Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K Number | |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| | :<br>ﻣﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍ |
| | बहु |
| | . . |
Page 1 of
006
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.